SOP: Calculation of the Most Probable Number of Growth Units for Hydrophobic Grid-Membrane Filter (HGMF) Methods (Appendix C)
GIFSQ ID: SOP-0963
A hydrophobic grid-membrane filter turns one filtration into a large multiple-tube test. Hydrophobic printed lines divide a membrane into 1600 discrete grid-cells, each cell acting like a miniature tube: growth either occupies it or does not. Because colonies stay confined, one dilution can span contamination levels that would overgrow an ordinary plate, up to about 5,000 organisms per mL in a suspension.
Scoring and conversion need discipline. A cell counts once however many colonies sit inside it, duplicate filters are converted separately before averaging, and partial counts follow fixed geometry. Procedure below covers scoring, conversion by formula or table, dilution correction, and reporting exactly as HGMF methods require.
Purpose
Aim is to convert an HGMF grid-cell score into a Most Probable Number of Growth Units (MPNGU) and then into organisms per g (mL), with counting rules that correct for overlapping growth units sharing one cell. Growth in one grid-cell is equivalent to one colony forming unit in concept, but MPNGU, not a raw score, is reportable quantity.
Scope
Apply to any HGMF method where a sample suspension is filtered through a hydrophobic grid membrane and growth is scored by grid-cell. Conversion table and standard formula assume a filter carrying 1600 grid-cells; do not apply that table to a filter with a different cell total. Membrane transfer between media for resuscitation or biochemical steps, removal of soluble inhibitors by filtration, and concentration benefit for detection all belong to wider HGMF methods, while scoring and calculation here govern every such method’s counting stage.
Equipment and Materials
- Inoculated and incubated HGMF membranes, normally in duplicate, with center indicator visible for orientation
- Conversion Table of HGMF Score to MPNGU for a 1600-cell filter (Section 5 of source)
- Calculator capable of natural logarithms (loge)
- Counting aid and bench record showing volume filtered (V), dilution concentration (C), and whether whole or one-fifth of a filter was counted
- Hand tally counter for positive or negative cell counts
Procedure
- Examine each membrane and form a rough estimate of occupancy before detailed counting. Decide counting strategy from density, using steps below.
- Score each grid-cell showing typical growth as one positive cell. Never count individual colonies inside a cell: a cell holding several typical colonies still scores one.
- Count whole filter when rough estimate suggests fewer than 200 occupied cells. Tally every positive cell across entire membrane to obtain a score.
- Count one-fifth of a filter at higher densities, up to 50% of cells showing typical growth. Rotate membrane so center indicator lies to left or right. Count positive cells in four rows immediately below center plus four rows immediately above center (eight rows, equal to one-fifth of filter). Multiply partial count by 5 to obtain score.
- Count negative cells where that route is more practical. Negative means no growth or non-typical growth. If fewer than 200 negative cells are estimated, count all of them across whole filter. If estimate exceeds 200, count negatives in same eight-row band described above. Derive score by subtraction: subtract negative count from 1600 for a whole-filter count, or from 320 for an eight-row count, then multiply by 5 if only one-fifth was counted.
- Respect countable range. Treat only filters holding 20 to 1520 occupied grid-cells as fully countable. Regard counts outside that range as estimated because uncertainty is greater.
- Convert each duplicate score separately. Either look up MPNGU in conversion table or calculate it from formula in step 8. Never average duplicate scores first and then convert averaged score.
- Apply formula where calculation is chosen: MPNGU = N loge (N / (N − x)), where N is total grid-cells on a filter and x is score. With usual N = 1600, formula becomes MPNGU = 1600 loge (1600 / (1600 − x)). Use natural logarithm throughout.
- Average MPNGU values from duplicates after individual conversion. Averaged MPNGU, not averaged score, carries forward to dilution correction.
- Correct for dilution and volume. If 1.0 mL of a 1 in 10 suspension inoculated a filter, multiply average MPNGU by 10 to obtain organisms per g (mL). If V mL of a 1 in 10 suspension was used, multiply average MPNGU by 10 / V. If V mL of a 1 in C dilution was used, multiply average MPNGU by C / V.
- Check work against standard examples, all with 1.0 mL of a 1 in 10 dilution unless stated. Full-filter positives 157 and 164 convert to MPNGU 165 and 173, average 169, giving 169 × 10 = 1690, reported as 1700 per mL. Eight-row positives 157 and 164 give scores 785 and 820, MPNGU 1079 and 1150, average 1114, reported after correction and rounding as 1100 per mL in source worked example presentation. Eight-row negatives 57 and 49 give scores 1600 − (5 × 57) = 1315 and 1600 − (5 × 49) = 1355, MPNGU 2760 and 3002, average 2881, reported as 2900 per mL.
- Use conversion table correctly. Table maps score directly to MPNGU for a 1600-cell filter only; illustrative pairs include score 100 → 103, score 157 → 165, score 200 → 214, score 500 → 600, score 785 → 1079. At low scores MPNGU tracks score closely; at high scores MPNGU rises far above score because multiple growth units increasingly share cells, which formula corrects.
Recording Results
Report average MPNGU after dilution correction, rounded to two significant figures (for example, record 2881 as 2900). Where lowest dilution shows no positive cells (no growth or only non-typical growth), report lowest average obtainable for volume plated onto replicate filters, preceded by a less-than sign; for duplicate filters at 1 mL per filter that value is <0.5 before multiplication by dilution factor of inoculum, and factor must then be applied. Where every grid-cell holds typical growth, report as ≥11804 (before any further dilution correction presentation required by a method). Bench records must show, per filter: counting mode (whole, one-fifth positive, or negative count), raw tally, derived score, individual MPNGU, volume V and dilution C, so averaging and correction can be audited.
Quality Control
Recount a filter independently when a score sits near 20 or 1520 boundaries, as generic good counting practice, and resolve discrepancies before conversion. Verify orientation and row selection for one-fifth counts: eight rows straddling center, four above and four below, with center indicator to left or right. Confirm N = 1600 before using either standard formula simplification or conversion table. Check that duplicate MPNGU values were averaged, not duplicate scores, on every worksheet; this single error is a common source of understated results at high occupancy.
Troubleshooting
- Filter is almost fully occupied and positive counting is slow: switch to negative counting. Count negatives (all of them if under 200, otherwise in eight-row band) and subtract from 1600 or 320 as procedure directs.
- Score exceeds 1520 or falls below 20: still calculate, but flag result as estimated due to greater uncertainty, and consider a different dilution if a method requires a countable-range result.
- Duplicate MPNGU values differ sharply: re-examine both membranes for uneven filtration, non-typical growth mis-scored as typical, or row-selection error before averaging.
- Every cell shows typical growth: no finite score-based conversion applies in normal way; report as ≥11804 and repeat at higher dilution if quantification is needed.
- No positive cells at lowest dilution: report less-than value for replicate set and volume used, corrected by dilution factor, rather than reporting zero.
Interferences and Limitations
Lateral spread is restrained by hydrophobic grid, yet overlapping growth units within one cell remain unavoidable at high inoculum; MPNGU formula corrects that overlap statistically, it does not reveal how many colonies shared a cell. Filtration removes many soluble inhibitors and concentrates organisms, improving detection limits, but particulate load can still distribute unevenly across a membrane. Visual scoring still depends on recognising typical growth correctly: counting non-typical cells as positive inflates a score, while missing faint typical growth deflates it, although precision is generally better than conventional plates or plain membrane filters because grid confinement reduces effect of individual visual acuity. Single-dilution breadth (up to about 5,000 organisms per mL) does not remove need for dilution correction or for a further dilution when a membrane saturates.
Safety Notes
- Handle incubated membranes as potentially pathogenic cultures; use gloves and avoid aerosol generation when handling dried filters, as generic laboratory practice.
- Dispose of used membranes and filtration waste through sterilisation and approved biological waste routes.
- Keep filtration apparatus decontaminated between samples to protect both operator and next sample from carryover.
Source Method
Based on Health Canada Appendix C, Calculating the Most Probable Number of Growth Units for Hydrophobic Grid-Membrane Filter (HGMF) Methods, Volume 1 of Compendium of Analytical Methods, issued May 2014, including its Table 1 formula, worked examples, reporting rules and score-to-MPNGU conversion table. Presentation and phrasing here are original; counting limits, formula, factors and reporting thresholds reproduce source exactly.