SOP: Screening Foods for E. coli O157 with the Merck Singlepath® Kit (MFLP-82)
GIFSQ ID: SOP-1069
A lateral-flow style device does a screening job in this procedure, and culture still decides a final answer. Singlepath® E. coli O157 is an immunological screening test based on an immune flow principle, designed to avoid time-consuming and personnel-intensive working steps. Enrichment culture, optionally boiled first, is applied to a small test device, and red lines in test and control zones are read 20 minutes later.
Read this SOP for what it is: a presumptive, qualitative screen for E. coli O157, including H7, applied to enrichment broths produced under a companion culture method, MFLP-80. Enrichment itself, selective isolation and cultural confirmation all sit inside MFLP-80, and the source for this kit procedure gives no enrichment broth, sample weight or enrichment incubation of its own. No such values are invented here; a laboratory running this screen must run its enrichment, handling and confirmation under MFLP-80 as written there.
Purpose
Purpose is presumptive qualitative detection of Escherichia coli O157, including H7, in food products, to determine compliance with Sections 4 and 7 of the Food and Drugs Act. Positive results must be confirmed with a cultural method. A negative screen is a screen result on an enrichment, reported as described under Recording Results, and a positive screen is never reported as a confirmed finding on its own.
Scope
Validated scope is the food category Raw Meat Products and the food type Fermented Dairy Products. Method text notes that use with other foods is assumed possible, but a commodity outside that application is fit for purpose only after proper validation following criteria in the Compendium of Analytical Methods; validation data are requested to be forwarded to the Microbiological Methods Committee (MMC) so an Application section can be expanded where data comply with MMC requirements. Work is carried out in accordance with ISO/IEC 17025:2005 (or latest version). Media listed are commercially available and are prepared and sterilised according to manufacturer instructions. Where a laboratory varies a listed medium formulation, whether a commercial product or one made from scratch, showing equivalency is that laboratory’s responsibility, and equivalency data go to the Editor of the Compendium of Analytical Methods for consideration of a method modification.
Equipment and Materials
- Singlepath® E. coli O157 GLISA test devices (Merck, Cat. No. 1.04141.0001)
- Media (including broths and agars), materials and control cultures specified in MFLP-80, covering enrichment, selective isolation and cultural confirmation
- Water bath for boiling samples (optional)
- Disposable polypropylene tubes for boiling samples (optional)
- Disposable plastic transfer pipettes and/or appropriate micropipetters and disposable tips, for dispensing 1–2 mL (sample for boiling) and 150 µL (application of boiled sample into test devices)
- Flat surface for holding test devices during incubation and reading; sample identification labels
Procedure
- Handle sample units following MFLP-80, and carry out preparation for analysis and preparation of a sample, including enrichment, following MFLP-80. After incubation of enrichment broths, proceed with steps below.
- Decide whether to use an optional boiling step. Boiling is not mandatory for test performance, but it can be done in order to reduce a potential risk of contamination when handling live bacteria.
- Where boiling is used, transfer approximately 1–2 mL of enrichment culture to an appropriate polypropylene tube and cover with a loose-fitting cap. Place tubes in a boiling water bath for 15 min, then remove and allow to cool to room temperature prior to use.
- Remove foil pouches from a required number of Singlepath® E. coli O157 devices. Place each test device on a flat surface and label with appropriate sample identification. Perform tests within a period of 2 hours after opening.
- Apply a sample. Using a micropipetter and disposable pipette tip, draw up 150 µL from a boiled and cooled, or a non-boiled, enrichment and dispense 150 µL into a circular sample port on a test device. Alternatively, use a disposable transfer pipette and dispense five (5) drops, about 150–160 µL, into a circular sample port.
- Incubate a test device at room temperature and read a test result 20 minutes after applying a sample to a device.
- Confirm every positive result culturally as described in MFLP-80. For selective isolation and confirmation under that method, re-suspend a Dynabeads® (immunomagnetic separation) pellet and inoculate onto selective agars.
Recording Results
Judge validity from a control zone first. A test is working correctly if a distinct red line appears in a control zone (C) within a read time of 10–20 minutes; if a control zone (C) line does not appear, discard a test and repeat with a new test kit, recording no result for that device. Consider a sample POSITIVE if, at or prior to 20 minutes, red lines appear in both a test (T) zone and a control (C) zone. Consider a sample NEGATIVE if no red line appears in a test (T) zone but a red line does appear distinctly in a control (C) zone 20 minutes after application of a sample to a device. Record, for every device, a sample identification, whether a boiling step was used, a time of application and reading, presence or absence of lines in both zones, and, for positives, a cultural confirmation outcome obtained under MFLP-80. Positive entries are recorded as presumptive until that confirmation is complete.
Quality Control
Run media, materials and control cultures specified in MFLP-80 alongside enrichment and confirmation work, as that method directs. Use a control zone on every device as an in-test validity check: no distinct red control line within 10–20 minutes invalidates that device, regardless of what a test zone shows. Respect two time disciplines built into a kit: devices are used within 2 hours after opening a foil pouch, and results are read at 20 minutes after application, not at a laboratory’s convenience later. As general laboratory practice, label each device before applying a sample and use a fresh disposable tip or transfer pipette for each enrichment, so a result cannot be assigned to a wrong sample or carried over from a previous one.
Troubleshooting
- No red line in a control zone (C): discard a test and repeat with a new test kit; do not report a test zone reading from an invalid device as either positive or negative.
- A line appearing only after 20 minutes: a positive requires red lines in both zones at or prior to 20 minutes, and a negative is judged at 20 minutes, so a late change falls outside a reading rule and a test should be repeated rather than reinterpreted.
- Volume uncertainty with a transfer pipette: count five drops, about 150–160 µL, or switch to a micropipetter delivering an exact 150 µL into a circular sample port.
- Boiled sample applied while still hot: allow a boiled enrichment to cool to room temperature before use, as a procedure specifies, and repeat with a cooled aliquot if a hot sample was applied.
- Device left in an opened pouch: discard a device held beyond 2 hours after opening and open a fresh pouch for a repeat.
Interferences and Limitations
Approval is narrow. Only Raw Meat Products and Fermented Dairy Products carry validation in this method, and any other food needs validation first, as set out under Scope. Detection here is presumptive and qualitative: an immunological flow test on an enrichment cannot confirm identity, so positives must be confirmed culturally, by re-suspending an immunomagnetic separation pellet and inoculating selective agars under MFLP-80. An optional boiling step changes handling risk, not a result’s status; boiled and non-boiled enrichments are both screened and read under identical line rules. Finally, this kit procedure is not self-sufficient: enrichment conditions, sample handling and confirmation media come from MFLP-80, and without that companion method performed as written, a line on a device has no defined meaning.
Safety Notes
- Treat enrichment cultures as potentially carrying E. coli O157, a human pathogen; where risk reduction is wanted, use an optional 15-minute boiling step before handling an aliquot for a device, and dispose of cultures and used devices as biohazard waste.
- Take care with a boiling water bath and with hot polypropylene tubes; use loose-fitting caps during boiling rather than sealed caps, and allow tubes to cool to room temperature before handling a sample further.
- Avoid cross-contamination between enrichments by changing disposable tips and transfer pipettes for every sample, and decontaminate work surfaces after spills of enrichment culture.
Source Method
Adapted in original wording from Health Canada method MFLP-82, Detection of E. coli O157 by the Merck Singlepath® E. coli O157 Kit, July 2014, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-82 dated July 2007.