SOP: Concentrating Salmonella by Immunomagnetic Separation with Dynabeads™ anti-Salmonella (MFLP-84)

GIFSQ ID: SOP-1070

Antibody-coated magnetic beads do a fishing job here. Dynabeads™ anti-Salmonella are magnetically charged polystyrene beads carrying Salmonella-specific polyclonal and monoclonal antibodies. Mixed into an enrichment culture, beads bind Salmonella cells through those antibodies; an external magnet then pulls a bead–bacteria complex out of suspension, washing removes unbound components, and a cleaned complex can be plated on indicator media or carried into further enrichment.

Two routes perform a separation: an automated BeadRetriever™ run, preferred for avoiding cross-contamination and for safety, and a manual Dynal® MPC®-S procedure for laboratories without that instrument. Either way, this SOP is a concentration step inside a wider Salmonella method, not a stand-alone identification. Primary enrichment, and secondary enrichment plus confirmation after separation, sit in companion methods named under Procedure; this source gives no enrichment broth or incubation of its own.

Purpose

Purpose is concentration of Salmonella spp. from enrichment cultures by immunomagnetic separation (IMS), so a bead–bacteria complex can be selectively plated on indicator media for identification, or carried into secondary enrichment and confirmation. No reading in this procedure declares a sample positive: a presumptive finding is established downstream.

Scope

Validated scope is animal feed. Handling, preparation for analysis and primary enrichment are performed as in MFLP-75, whose enrichment supplies a test sample used here. Secondary enrichment on semisolid agar likewise follows MFLP-75, and presumptive positives proceed to biochemical screening and serological confirmation under MFHPB-20. Dynabeads™, Dynal® MPC® and BeadRetriever™ are trademarks of Thermo Fisher Scientific Inc. and its subsidiaries; equivalents are allowed where a materials list says “or equivalent.”

Equipment and Materials

  • Dynabeads™ anti-Salmonella, Cat. No. 710.02 (or equivalent immunomagnetic beads or concentration steps)
  • Dynal® MPC®-S (Magnetic Particle Concentrator) unit (or equivalent), with removable magnetic plate and tube opener; 1.5 mL Eppendorf tubes, one per sample for a manual route
  • BeadRetriever™ instrument, with sample racks, sample tube strips and sterile protective sample tip combs, for an automated route
  • Sample mixer capable of tilt, swirl or rotational mixing, for example a Dynal® Sample Mixer (MX4 named for manual incubation)
  • Wash buffer, as used in tube strips and manual washing; vortex mixer; pipettes, changed for each new sample as directed under Procedure
  • Pre-enriched sample from a primary enrichment performed under MFLP-75; for a manual route, a filtered pre-enriched sample aliquot
  • Biohazard waste container for used tip combs and tube strips

Procedure

  1. Handle sample units following MFLP-75, carry out preparation for analysis following MFLP-75, and perform a primary enrichment of MFLP-75. Use resulting enrichment culture as a test sample for immunomagnetic separation below.
  2. Choose a route. Use a BeadRetriever™ where one is available: source text strongly recommends IMS be performed using a BeadRetriever™, to avoid cross-contamination and for safety reasons. Use a manual route only in absence of a BeadRetriever™, with strict adherence to good laboratory practice as a prerequisite to obtaining valid results.
  3. For an automated run, load one BeadRetriever™ sample tube strip for each sample into a sample rack.
  4. Resuspend Dynabeads™ anti-Salmonella by vortexing until a pellet in a bottom of a tube disappears. Aseptically add 10 µL of properly mixed Dynabeads™ anti-Salmonella into sample tubes 1 and 2 of each strip.
  5. Aseptically add 500 µL of wash buffer to sample tubes 1 and 2. Aseptically add 1 mL of wash buffer to tubes 3 and 4 within a strip, and aseptically add 100 µL of wash buffer to tube 5.
  6. For each sample, remove a labelled sample tube strip from a sample rack and place in a second sample rack one metre away. Add 500 µL of a test sample to tubes 1 and 2 and return an inoculated tube strip to a first sample rack. Repeat for remaining samples.
  7. Aseptically insert sterile protective sample tip combs into an instrument. Insert a rack with filled tubes into an instrument to lock it in place, check that everything is properly aligned, and close an instrument door.
  8. Select a SALMONELLA program sequence by scrolling with an arrow key and press a START button. Keep a door closed while an instrument is in operation; each processing step and total time remaining can be followed on an LC display.
  9. At end of a program run, remove a sample rack from an instrument and process bead–bacteria complexes for each sample by resuspension in step 14. Remove sample tip combs and discard into a biohazard waste container together with tube strips.
  10. For a manual run, remove a magnetic plate and load one 1.5 mL Eppendorf tube for each sample into a Dynal® MPC®-S. Resuspend Dynabeads™ anti-Salmonella by vortexing until a pellet in a bottom of a tube disappears, and dispense 20 µL into each tube using a pipette.
  11. Add 1 mL of a pre-enriched filtered sample aliquot to each manual tube and close a tube, changing to a new pipette for each new sample. Invert a Dynal® MPC®-S rack five times to mix sample and beads, then incubate at room temperature for 10 minutes with gentle continuous agitation to prevent beads from settling, for example in a Dynal® MX4 sample mixer.
  12. Insert a magnetic plate into a Dynal® MPC®-S and allow 3 minutes for proper bead recovery, inverting a rack several times to concentrate beads into a pellet on a side of a tube. Open a tube cap using a tube opener provided and carefully aspirate and discard supernatant and liquid in a cap, without disturbing a pellet on a side wall.
  13. Remove a magnetic plate from a Dynal® MPC®-S. Add 1 mL of wash buffer, changing to a new pipette for each sample and not touching a tube with a pipette, since this can cross-contaminate samples as well as wash buffer. Close a cap and invert a Dynal® MPC®-S a few times to resuspend beads.
  14. Repeat a manual wash cycle once: plate in, 3 minutes for bead recovery with rack inversions, aspirate supernatant and cap liquid without disturbing a pellet, plate out, add 1 mL of wash buffer with a fresh pipette per sample, close a cap and invert to resuspend beads. Then repeat capture once more (plate in, 3 minutes with inversions, aspirate as above), remove a magnetic plate, and proceed to step 14.
  15. Resuspend a Dynabeads™–bacteria complex in 100 µL of wash buffer, whichever route was used. Mix briefly by using a vortex mixer.
  16. Follow secondary enrichment steps of MFLP-75 for secondary enrichment on semisolid agar. For presumptive positive Salmonella spp., proceed to biochemical screening and serological confirmation steps of MFHPB-20.

Recording Results

Record a separation, not a detection. For each sample, note its primary enrichment under MFLP-75, route used (automated BeadRetriever™ or manual MPC®-S), bead and wash volumes per tube position, and final resuspension of a bead–bacteria complex in 100 µL of wash buffer. Carry sample identity forward into secondary enrichment under MFLP-75 and, for presumptive positives, into biochemical and serological confirmation under MFHPB-20, recording those outcomes against same identity. A completed IMS run alone is recorded as “complex recovered and forwarded,” never as Salmonella detected or not detected.

Quality Control

Fully resuspend beads before dispensing: vortex Dynabeads™ anti-Salmonella until a pellet in a bottom of a tube disappears, since a settled suspension cannot deliver a consistent 10 µL automated or 20 µL manual dose. Change to a new pipette for each new sample wherever Procedure directs, and keep pipettes from touching tubes during wash additions. Check automated alignment before starting: rack locked, tip combs inserted, door closed, SALMONELLA program selected. As general laboratory practice, label strips and tubes before adding a sample; an automated layout itself moves a strip being inoculated to a second rack one metre away before returning it.

Troubleshooting

  • Beads settling during manual incubation: maintain gentle continuous agitation for a full 10 minutes at room temperature; a settled tube has not had even bead–sample contact.
  • Poor bead recovery on a magnet: allow a full 3 minutes after inserting a magnetic plate and invert a rack several times to concentrate beads into a pellet on a tube wall before aspirating.
  • Pellet disturbed during aspiration: aspirate carefully, including liquid held in a cap, without touching a pellet on a side wall; lost beads carry bound target cells away with a supernatant.
  • Suspected cross-contamination in a manual run: switch to a BeadRetriever™ route where available; in manual work, change pipettes for each sample and never let a pipette touch a tube during wash addition.
  • Uneven bead dose: re-vortex until no pellet remains, then re-dispense; never pipette from a partially settled suspension.

Interferences and Limitations

Validation is for animal feed only; other matrices are not covered as written. Separation depends on enrichment supplied by another method: without primary enrichment under MFLP-75 there is no defined test sample, and without secondary enrichment under MFLP-75 and confirmation under MFHPB-20 there is no result to report. Manual separation carries higher contamination and safety risk, which is why a BeadRetriever™ is strongly recommended and strict good laboratory practice is a prerequisite for valid manual results. Extra or skipped washes are not neutral changes: washing removes unbound sample components, while each aspiration risks a bead pellet holding bound target cells.

Safety Notes

  • Treat enrichments and bead–bacteria complexes as potentially carrying Salmonella; work aseptically, contain spills, and dispose of used tip combs and tube strips in a biohazard waste container.
  • Prefer an automated BeadRetriever™ route where available, for safety as well as cross-contamination control.
  • Use a tube opener provided with an MPC®-S unit for tightly capped tubes rather than forcing caps by hand.
  • Keep pipettes and hands clear of tube openings during additions, protecting both operator and sample.

Source Method

Adapted in original wording from Health Canada method MFLP-84, Method for the ImmunoMagnetic Separation of Salmonella species by Dynabeads™ anti-Salmonella, March 2016, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-84 dated October 2001.