Login Register

Access the GIFSQ Portal

Select your user type to log in or register a new account.

Student Portal

Access your food safety courses and exams.

Instructor Portal

Manage courses, view student submissions, and grade quizzes.

Company Portal

Manage corporate setup, view employee logs, and access QA services.

SOP: Detecting Listeria monocytogenes in Foods by the VIDAS LMO 2 Method (MFLP-33)

GIFSQ ID: SOP-1035

Automated immunoassay shortens Listeria monocytogenes screening without giving up enrichment. Broth cultures grow first, then a VIDAS instrument runs an enzyme-linked fluorescent assay, ELFA, in which sample cycles in and out of an antibody-coated Solid Phase Receptacle, SPR, and fluorescence betrays captured antigen.

Reading is a number, not a colour judgement. Relative fluorescence at 450 nm comes from 4-methyl-umbelliferone, produced when alkaline phosphatase conjugate acts on 4-methyl-umbelliferyl-phosphate substrate, and a fixed threshold separates presumptive positives from negatives.

Purpose

Purpose is rapid qualitative detection of L. monocytogenes in foods for compliance under Sections 4 and 7 of the Food and Drugs Act. Presumptive positives must be confirmed with a cultural method.

Scope

Scope covers all foods except environmental samples, with named exclusions inside food categories: within ready-to-eat meat and poultry, fermented, other, and cured/smoked food types are excluded; smoked fish, yogurt, kefir and fermented dairy drinks are also excluded. Everything else validated here follows one of two enrichment routes below, chosen by food type: Half-Fraser/Fraser for heat processed ready-to-eat meat and poultry, Palcam broth/UVM 2 for all other foods.

Equipment and Materials

  • VIDAS LMO 2 assay kit, Ref. 30704, holding reagents, standards and controls for 60 test samples; components from different lots must never be mixed or used past expiry
  • Large capacity VIDAS 30 or reduced capacity miniVIDAS instrument
  • Palcam broth; UVM 2; Half-Fraser broth; Fraser broth
  • Palcam agar (or equivalent) and Oxford agar (or equivalent) for confirmation plating; optional chromogenic media: Rapid L. mono plates or Aloa plates
  • Pipettor calibrated to dispense 500 μL, with disposable tips
  • Stomacher and stomacher bags, blender or equivalent
  • Waterbath at 95–100°C or equivalent system
  • Incubators capable of maintaining 30°C and 35°C

Procedure

  1. Keep sample units refrigerated or frozen by product nature, except shelf-stable foods. Thaw frozen units in a refrigerator, or under time and temperature conditions preventing microbial growth or death. Analyse as soon as possible after receipt.
  2. Prepare media ahead: sterile Palcam broth pre-warmed to 35 ± 2°C for general route; for heat processed ready-to-eat meat and poultry, sterile Half-Fraser broth pre-warmed to room temperature. Clean work surface with disinfectant.
  3. Build a representative analytical unit: agitate liquids or free-flowing materials until homogeneous; for solids, take portions from several locations.
  4. Heat processed ready-to-eat meat and poultry, primary enrichment: add 25 g or mL to 225 mL Half-Fraser broth in a blender jar or stomacher bag. Composites keep 1 part sample to 9 parts broth, up to a maximum of 125 g. Blend, stomach or vortex for thorough mixing, then incubate at 30 ± 1°C for 25 ± 1 hours.
  5. Heat processed ready-to-eat meat and poultry, secondary enrichment: agitate Half-Fraser broth enough to mix and let large particles settle. Transfer 1 mL into 10 mL Fraser broth and incubate at 30 ± 1°C for 25 ± 1 hours. Secondary broth feeds VIDAS step below.
  6. All other foods, primary enrichment: add 25 g or mL to 225 mL Palcam broth in a blender jar or stomacher bag. Composites keep 1 part sample to 9 parts Palcam broth, up to a maximum of 125 g. Blend, stomach or vortex for thorough mixing and incubate at 35°C for 26–28 h.
  7. All other foods, secondary enrichment: agitate Palcam broth enough to mix and let large particles settle. Transfer 1 mL into 9 mL UVM 2 and incubate at 30°C for 26 h minimum to 48 h.
  8. Optional holding step for UVM 2 only: refrigerate secondary enrichment for up to 72 hours when laboratory scheduling needs flexibility. Afterwards resuspend broth and plate onto agar media as confirmation requires; VIDAS testing still uses enrichment as below.
  9. After incubation, vortex UVM 2, allow particles to settle, and proceed to analysis. Heating enrichment broths before VIDAS testing is not necessary. Store remaining enrichment broth at 2–8°C for later confirmation of any positive assay result.
  10. Take VIDAS LMO 2 kit from refrigerator, remove components needed, and let them reach room temperature for a minimum of 30 min. Return unused components to storage at 2–8°C. Label LMO 2 reagent strips with appropriate sample identification numbers in space provided.
  11. Enter master lot, sample, standard and control data following operating instructions for instrument in use, VIDAS 30 or miniVIDAS. With a new kit lot, enter factory master calibration curve specifications first, using master lot entry card included with kit.
  12. Calibrate standard at least every 14 days and whenever a new kit lot starts. When a standard must be tested, enter AS1@ as its sample identification; place standard at beginning of Work List, followed by controls C1 and C2. A standard stored in memory must be tested in duplicate, both portions under same S1 designation.
  13. Vortex standard, controls and samples. Pipette 500 ± 50 μL of standard, control or sample into centre of sample well of an LMO reagent strip.
  14. Load LMO 2 reagent strips and LMO 2 SPRs into positions matching VIDAS section shown on Work List. Check colour labels carrying three-letter assay code on SPRs and reagent strips match each other.
  15. Start assay processing as instrument manual directs. Assay completes in approximately 70 minutes, with instrument performing all steps automatically and sequentially: antigen binding to antibody-coated SPR interior, washing away unbound material, alkaline phosphatase conjugate binding, washing away unbound conjugate, then substrate cycling and fluorescence measurement at 450 nm.

Recording Results

A relative fluorescence value, RFV, of ≥ 0.05 indicates presumptive presence of L. monocytogenes in test sample. Values below that threshold read negative for screening. Record RFV, strip and SPR lot, standard calibration date, control results C1 and C2, enrichment route, and incubation times for each sample. Report every presumptive positive as unconfirmed until cultural confirmation finishes.

Quality Control

Standards and controls anchor each run: standard calibration at least every 14 days and at each lot change, controls C1 and C2 placed immediately after standard on Work List. Lot discipline is absolute: never mix components across lots and never use expired components. Temperature control follows house pattern for these methods: 35°C may run at 35 ± 1.0°C, lower temperatures such as 30 or 25°C also carry ± 1.0°C, while higher temperatures such as 43 or 45.5°C must hold within 0.5°C. Any media variation, commercial or made in-house, must be proven equivalent by laboratory using it. Work should meet ISO/IEC 17025 requirements.

Troubleshooting

  • Standard or control out of range: check calibration age, lot entry, and whether kit components stood at room temperature for full minimum 30 min before loading.
  • Colour codes mismatched: stop before starting; SPR and strip three-letter codes must match, or results cannot be trusted to LMO 2.
  • Cloudy sample well or poor cycling: let particles settle after vortexing enrichment, and pipette carefully into centre of sample well at 500 ± 50 μL.
  • Borderline RFV near 0.05: treat threshold literally, confirm culturally from stored secondary enrichment, and do not average repeat RFVs into a verdict.

Interferences and Limitations

Excluded matrices named under Scope can behave differently in this assay, so results there carry no validation. Assay configuration is built to prevent non-specific SPR reactions, yet antigen detection remains presumptive: dead cells and cross-reacting material cannot be separated by fluorescence alone. Confirmation depends on retaining secondary enrichment at 2–8°C; discarding broth early removes that safety net.

Safety Notes

  • Treat enrichments as potentially carrying L. monocytogenes; use gloves, careful pipetting, and biohazard disposal for strips, SPRs and broths.
  • Handle hot waterbath work at 95–100°C with thermal protection where that system is used in laboratory.
  • Keep VIDAS reagents refrigerated at 2–8°C when stored, and keep food samples and assay chemicals in separate areas as general laboratory practice.

Source Method

Adapted in original wording from Health Canada method MFLP-33, Detection of Listeria monocytogenes in foods by the VIDAS LMO 2 method, August 2014, Compendium of Analytical Methods, Volume 3.