ICP-MS Quiz
GIFSQ ID: ART-1450
Test your knowledge with this free 20-question ICP-MS quiz from Food Analysis & Analytical Chemistry, part of…
What this quiz covers
ICP-MS is the instrument food laboratories reach for when the question is how much of a toxic element is present at levels older techniques cannot see. This quiz starts with what makes it different: compared with ICP-OES it generally offers lower detection limits, down in the parts per trillion range, and isotopic information, because ions are counted directly rather than measured as emitted light. The argon plasma runs at roughly 6,000 to 10,000 K, hot enough to desolvate droplets, break molecules into atoms, and strip an electron from most elements, and the mass analyser then separates those ions by their mass to charge ratio. A quadrupole hops between masses in milliseconds, so one run quantifies dozens of elements in about a minute, and a nine orders of magnitude range lets major minerals and ultra trace contaminants be measured together.
Argon chloride overlaps arsenic at m/z 75, argon dimers overlap selenium, and doubly charged ions appear at half their mass, so tuning checks sensitivity, oxide formation and doubly charged ion levels before any sample runs. Collision and reaction cells remove polyatomic interferences with gases like helium or hydrogen, and internal standards such as rhodium or indium correct for drift and matrix effects. Sample preparation matters just as much: solid foods are digested by microwave assisted acid digestion into a clear solution, and dissolved solids are kept low because they clog the nebulizer and salt up the cones.
At parts per trillion, the blank is the analysis
At these detection limits the laboratory itself becomes part of the measurement. Reagents, vessels and even laboratory air can contribute signals comparable to the analyte, which is why method blanks run with every batch and acid washed vessels and high purity reagents are standard. Mercury needs special care because it sticks to the sample introduction system and slowly releases into later samples, and clean room practice exists because zinc, iron and lead contaminate samples at the very levels being measured. The reward is real: lead, cadmium, arsenic and mercury, the four toxic elements with maximum levels across food categories, can be monitored together, and liquid chromatography adds speciation, distinguishing toxic inorganic arsenic from non toxic arsenobetaine in seafood.
How to use this quiz
Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the methods and standards a food analyst is expected to know by heart.
Choose how you want to play
A short note: ICP-MS
ICP-MS uses an argon plasma, around 6,000 to 10,000 K, to ionize sample atoms. The analyzer separates ions by mass-to-charge ratio. Against ICP-OES it offers lower detection limits and isotopic information. It determines many elements at once, and a 9-order range covers major minerals and ultra-trace contaminants in one run.
Solid foods need microwave-assisted acid digestion for a clear solution. The nebulizer and spray chamber make a fine aerosol for the plasma. Rhodium or indium internal standards correct drift and matrix effects. Standard addition is used when matrix effects are severe and no matrix-matched standards exist.
Argon chloride overlaps arsenic at m/z 75, a classic polyatomic interference. Collision and reaction cells clear polyatomic interferences with helium or hydrogen gas. Doubly charged ions form when elements with low second ionization energies make M2+ ions at half their mass. Tuning checks sensitivity, oxide formation and doubly charged levels.
Method blanks matter: reagents, vessels and laboratory air can add detectable contamination at ppt levels. Acid-washed vessels and high-purity reagents matter: everyday elements like zinc, iron and lead can spoil samples at the low levels measured. High dissolved solids can clog the nebulizer and coat the cones, suppressing signals. Lead, cadmium, arsenic and mercury are the toxic elements most routinely monitored. Mercury clings to the introduction system, a memory effect. Selenium work must account for argon dimers and variable ionization. LC-ICP-MS enables speciation, such as telling inorganic arsenic from arsenobetaine apart.
Practical takeaway: digest clean, run blanks, correct with internal standards.
Questions and answers
Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.
Question 1: Compared with ICP-OES (optical emission), ICP-MS generally offers:
Correct answer: Lower detection limits and isotopic information
ICP-MS typically reaches detection limits in the parts-per-trillion range, roughly two to three orders of magnitude lower than ICP-OES, because ions are counted directly by the mass spectrometer rather than measured as emitted light. The mass analyzer also resolves individual isotopes (for example 206Pb, 207Pb and 208Pb), which optical emission cannot do. These capabilities are documented in standard ICP-MS references such as Thomas, Practical Guide to ICP-MS (CRC Press).
Question 2: A classic polyatomic interference in food ICP-MS is argon chloride (⁴⁰Ar³⁵Cl⁺) overlapping:
Correct answer: Arsenic at m/z 75
In the argon plasma, chloride from the sample or reagents combines with argon to form the polyatomic ion 40Ar35Cl+, which has the same nominal mass (m/z 75) as the only stable arsenic isotope, 75As. This overlap falsely elevates the arsenic signal in high-chloride foods such as seafood unless it is corrected. The interference is a textbook example in ICP-MS literature, covered in Thomas, Practical Guide to ICP-MS, and is handled with collision/reaction cells or mathematical correction per EPA Method 200.8.
Question 3: Tuning the ICP-MS with a multi-element solution before analysis checks:
Correct answer: Sensitivity, oxide formation and doubly charged ion levels across the mass range
A multi-element tuning solution (typically containing Mg, In, Ce, Ba, Pb and U) is aspirated to verify sensitivity across the mass range before any samples run. The same tune measures oxide formation (CeO+/Ce+) and doubly charged ion formation (Ba2+/Ba+), which must be kept low, commonly below about 3%, to control spectral interferences. These daily performance criteria follow the tuning requirements described in EPA Method 200.8 and standard instrument protocols.
Question 4: Doubly charged ion interferences in ICP-MS arise when:
Correct answer: Elements with low second ionization energies form M²⁺ ions at half their mass
Elements with low second ionization energies, such as barium and the rare earths, can lose two electrons in the plasma to form M2+ ions. A quadrupole separates ions by mass-to-charge ratio, so an M2+ ion appears at half its atomic mass (for example 138Ba2+ at m/z 69, overlapping 69Ga). EPA Method 200.8 addresses these doubly charged interferences through tune criteria and mathematical correction equations.
Question 5: Method blanks are critical in ultra-trace ICP-MS food analysis because:
Correct answer: Reagents, vessels and laboratory air can contribute detectable contamination at ppt levels
At parts-per-trillion detection limits, trace metals leached from reagents, digestion vessels or even laboratory air produce signals comparable to the analyte itself. A method blank carried through the entire digestion and analysis reveals this background so it can be subtracted or investigated. EPA Method 200.8 therefore requires laboratory reagent blanks with every batch as a core quality control element.
Question 6: The wide dynamic range of ICP-MS (often 9 orders of magnitude) means the analyst can:
Correct answer: Measure major minerals and ultra-trace contaminants in the same run
Modern ICP-MS detectors combine pulse-counting and analog modes, giving a linear working range spanning up to about nine orders of magnitude. This lets one run quantify major minerals at mg/kg levels and ultra-trace contaminants at µg/kg levels without re-dilution or re-analysis. The wide dynamic range is a documented performance characteristic in standard ICP-MS references such as Thomas, Practical Guide to ICP-MS.
Question 7: Before ICP-MS analysis, solid foods are usually prepared by:
Correct answer: Microwave-assisted acid digestion to give a clear solution
Solid foods must be converted to a clear aqueous solution before nebulization, since particles would clog the nebulizer and incompletely vaporize in the plasma. Microwave-assisted digestion with nitric acid (often with hydrogen peroxide) destroys the organic matrix quickly in sealed vessels, minimizing contamination and volatile-element losses. AOAC Official Method 2015.01 prescribes microwave digestion as the sample preparation step for heavy metals in foods by ICP-MS.
Question 8: Collision/reaction cells in ICP-MS are used to:
Correct answer: Remove polyatomic interferences by colliding or reacting them away with gases like helium or hydrogen
A collision/reaction cell placed before the mass analyzer is pressurized with a gas such as helium, hydrogen or ammonia. Polyatomic interfering ions are larger than analyte ions, so they collide more often and lose energy (kinetic energy discrimination), or they react chemically with the cell gas and are converted to a different mass. This interference-reduction approach is described in standard ICP-MS references including Thomas, Practical Guide to ICP-MS.
Question 9: Clean-room practices (acid-washed vessels, high-purity reagents) matter in food ICP-MS because:
Correct answer: Ubiquitous elements like zinc, iron and lead contaminate samples at the low levels being measured
Zinc, iron, lead and copper are present in dust, tap water, glassware and even on skin, and at ppt-level targets these sources overwhelm the analyte signal. Acid-washed vessels, high-purity reagents and clean handling keep procedural blanks low enough for ultra-trace work. EPA clean-sampling guidance (EPA Method 1669, the clean hands/dirty hands technique) was developed specifically because ambient contamination dominates at these levels.
Question 10: The sample introduction system (nebulizer plus spray chamber) in ICP-MS is designed to:
Correct answer: Create a fine aerosol of the sample solution for the plasma
The nebulizer uses the argon gas flow to shear the liquid sample into a fine mist, and the spray chamber removes the larger droplets so only a fine, uniform aerosol reaches the plasma. A stable, fine aerosol is essential because droplet size controls how completely the sample is desolvated, vaporized and ionized. This sample-introduction design is standard across ICP-MS instruments as described in Thomas, Practical Guide to ICP-MS.
Question 11: Internal standards (such as rhodium or indium) are added to ICP-MS samples to:
Correct answer: Correct for instrument drift and matrix effects on signal intensity
Internal standard elements such as rhodium or indium are added at the same concentration to every blank, standard and sample; they are chosen to be absent from the samples and similar in mass to the analytes. Because the internal standard experiences the same plasma drift and matrix suppression as the analytes, ratioing the signals cancels those effects. EPA Method 200.8 requires internal standardization to compensate for physical interferences and instrument drift.
Question 12: Standard addition calibration is sometimes used in food ICP-MS when:
Correct answer: Matrix effects are severe and matrix-matched standards are unavailable
In the method of standard additions, known amounts of analyte are spiked directly into aliquots of the sample itself, so calibration occurs in the exact sample matrix. This cancels severe matrix suppression or enhancement that external calibration cannot correct. EPA Method 200.8 permits standard additions as an alternative calibration approach when matrix effects cannot otherwise be controlled.
Question 13: In inductively coupled plasma mass spectrometry (ICP-MS), the argon plasma is used to:
Correct answer: Ionize sample atoms at very high temperature so the mass spectrometer can detect them
The argon plasma operates at roughly 6,000 to 10,000 K, hot enough to desolvate droplets, break molecules into atoms, and strip an electron from most elements. These singly charged positive ions are then extracted into the mass spectrometer for separation and detection. The high-temperature ionization step is the defining principle of ICP-MS as described in standard references such as Thomas, Practical Guide to ICP-MS.
Question 14: Coupling liquid chromatography to ICP-MS (LC-ICP-MS) enables:
Correct answer: Elemental speciation, such as distinguishing inorganic arsenic from arsenobetaine
Liquid chromatography separates the chemical forms of an element before they enter the plasma, so each species is quantified independently by its element signal. This distinguishes toxic inorganic arsenic from essentially non-toxic arsenobetaine, the dominant arsenic form in seafood, a distinction total-element analysis cannot make. LC-ICP-MS speciation is the established approach used in food safety work, for example in FDA Elemental Analysis Manual methods for arsenic species.
Question 15: A major strength of ICP-MS for food elemental analysis is its ability to:
Correct answer: Determine many elements simultaneously at parts-per-trillion detection limits
A quadrupole or time-of-flight mass analyzer can hop between masses in milliseconds, so a single run quantifies dozens of elements in about a minute. Detection limits routinely reach the parts-per-trillion range for most metals. AOAC Official Method 2015.01 exploits exactly this capability, determining arsenic, cadmium, mercury and lead together in foods.
Question 16: High dissolved-solids content in a digested food sample can cause problems in ICP-MS by:
Correct answer: Clogging the nebulizer and depositing salts on the cones, suppressing signals
High dissolved solids deposit salts on the sampler and skimmer cones, narrowing their orifices and reducing ion transmission into the mass spectrometer. They also clog the nebulizer and change aerosol formation, suppressing analyte signals. EPA Method 200.8 therefore recommends keeping dissolved solids at or below 0.2% (w/v) and using internal standards to compensate for residual physical effects.
Question 17: Which toxic elements in foods are most routinely monitored by ICP-MS for regulatory compliance?
Correct answer: Lead, cadmium, arsenic and mercury
Lead, cadmium, mercury and inorganic arsenic are the toxic elements with established maximum levels across food categories. In the EU these limits are laid down in Commission Regulation (EU) 2023/915 on maximum levels for certain contaminants in food, which replaced Regulation (EC) No 1881/2006. AOAC Official Method 2015.01 was validated specifically for these four elements in foods by ICP-MS.
Question 18: Mercury analysis by ICP-MS needs special care because mercury:
Correct answer: Shows memory effects, sticking to the sample introduction system
Mercury adsorbs onto the surfaces of the spray chamber, tubing and nebulizer, then slowly releases into subsequent samples, causing carryover (memory) and high blanks. EPA Method 200.8 explicitly addresses mercury memory interferences and notes that adding gold to samples and standards alike eliminates the effect by keeping mercury in solution. Extended rinse times are also standard practice.
Question 19: Selenium determination in foods by ICP-MS must account for:
Correct answer: Polyatomic interferences (e.g., argon dimers) on its main isotopes and its variable ionization
The most abundant selenium isotope, 80Se, is overlapped by the argon dimer 40Ar40Ar+, and the other isotopes suffer similar polyatomic overlaps, so analysts must use collision/reaction cells or less-abundant isotopes such as 77Se or 82Se. Selenium also has a relatively high first ionization energy, which makes its ionization efficiency variable. These interference challenges for selenium are documented in standard ICP-MS references such as Thomas, Practical Guide to ICP-MS.
Question 20: ICP-MS distinguishes between isotopes of an element because the mass analyzer separates ions by:
Correct answer: Their mass-to-charge ratio (m/z)
The mass analyzer, typically a quadrupole, applies oscillating electric fields that only let ions of one mass-to-charge ratio pass through at a time. Isotopes of the same element have different masses, so they are transmitted and counted separately. This m/z-based separation is the fundamental principle of mass spectrometry described in standard texts such as Gross, Mass Spectrometry: A Textbook (Springer).