Accuracy Quiz

GIFSQ ID: ART-1441

Test your knowledge with this free 20-question Accuracy quiz from Food Analysis & Analytical Chemistry, part of…

What this quiz covers

Accuracy is closeness of a measured value to the true value, and this quiz covers how a food laboratory proves it has it. The proof is never a feeling about the instrument: it is evidence, and the quiz walks through each kind. Certified reference materials provide values certified with uncertainty by an authoritative body, and a CRM that reads consistently low reveals a negative bias needing investigation. A recovery experiment tests accuracy directly by adding a known amount of analyte and measuring how much is found, with acceptable spike recoveries in food analysis typically 80 to 120 percent, tighter at high concentrations. Accuracy across the working range is checked with CRMs or spikes at low, mid and high levels, and method validation must demonstrate accuracy among the other performance characteristics before a method is trusted for routine work.

The quiz also covers the systems that keep accuracy honest over time. Traceability of calibration to SI units supports accuracy by linking measurements through an unbroken chain to international standards. Standardizing a titrant against a primary standard confirms the titrant’s true concentration, the basis of every accurate titration that follows. Correcting for blank values improves accuracy by removing background signal, and accuracy can be lost through expired standards, matrix interference or calibration errors. Proficiency testing, the interlaboratory comparison, assesses a laboratory’s accuracy against peers and assigned values, with a z-score calculated as the laboratory result minus the assigned value divided by the standard deviation for proficiency: a |z| score above 3 means unsatisfactory performance, investigate and correct. Documenting all of this evidence is needed for audits, accreditation and defending results when a customer or regulator asks how the laboratory knows its numbers are right.

Accuracy is checked, not assumed

Trueness and precision together determine accuracy: trueness is lack of bias, precision is lack of scatter, and an accurate method performed inconsistently shows poor precision despite good trueness, which points to training and SOP issues rather than a faulty method. That split matters most in food safety decisions, where a precise but biased result can confidently declare unsafe food safe. Bias itself is the systematic difference between the expected result and the true value, and naming it is the first step to removing it.

How to use this quiz

Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the definitions and limits a food analyst is expected to know by heart.

Accuracy Quiz
Uncategorized

Choose how you want to play

Sound effects
🔈 Volume
Show explanations
🔊 Auto-read every question
Reading speed

A short note: accuracy in food analysis

Accuracy in food analysis is how near a measured value sits to the true value. ISO 5725-1 splits accuracy into trueness and precision: lack of bias, and lack of scatter. A result can be precise yet wrong, with readings clustering tightly around a value far from the truth. For food safety decisions that gap is dangerous: a precise but biased result can call unsafe food safe.

Bias is a steady offset: the gap between what you expect to get and the true value. Analysts test for it with certified reference materials, CRMs, which carry certified values with uncertainty from an authoritative body, and with recovery experiments, where a known amount is added and the fraction found is measured. Spike recoveries of 80 to 120 percent are accepted. A CRM that reads consistently low shows a negative bias to investigate. Blank correction removes background signal. Accuracy is checked across the range with CRMs or spikes at the low, middle and high ends.

Under ISO 13528, a lab in proficiency testing measures the same sample as its peers. The z-score divides the lab result minus the assigned value by the standard deviation for proficiency; above 3 it signals unsatisfactory performance and triggers investigation. SI-traceable calibration links every measurement through an unbroken chain to international standards, as ISO/IEC 17025 requires. CRM results, recoveries and calibration records are the evidence audits ask for.

Practical takeaway: before trusting a result, check the CRM reading, the spike recovery and the latest z-score.

Questions and answers

Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.

Question 1: Acceptable spike recoveries in food analysis are typically:

  • A. 80, 120% (tighter for high concentrations)
  • B. Spike recoveries of 10, 20% are acceptable
  • C. Spike recoveries of 200, 300% are acceptable
  • D. Spike recoveries of 0% are acceptable

Correct answer: 80, 120% (tighter for high concentrations)

Food-analysis method guidelines typically accept spike recoveries of 80-120%, with tighter windows expected at high concentrations where the method is most capable. The AOAC Guidelines for Standard Method Performance Requirements (Appendix F) publish recovery ranges by concentration level. Recoveries far outside this band signal bias or matrix interference.

Question 2: Method validation must demonstrate:

  • A. Method validation must demonstrate speed
  • B. Method validation must demonstrate cheapness
  • C. Method validation must demonstrate popularity
  • D. Accuracy (bias/recovery) among other performance characteristics

Correct answer: Accuracy (bias/recovery) among other performance characteristics

Validation must show that a method measures what it claims: accuracy : how close results sit to the true value through bias or recovery studies : plus precision, selectivity, and the other performance characteristics. ISO/IEC 17025 requires this documented evidence before results are reported. A method proven only fast or cheap is not a validated method.

Question 3: A |z|-score above 3 in proficiency testing means:

  • A. Unsatisfactory performance, investigate and correct
  • B. A |z|-score above 3 means excellent
  • C. A |z|-score above 3 means average
  • D. A |z|-score above 3 means lucky

Correct answer: Unsatisfactory performance, investigate and correct

In proficiency testing under ISO 13528, a |z|-score of 3 or more marks unsatisfactory performance: the laboratory's result lies too far from the assigned value to be chance. The standard sets satisfactory at |z| <= 2 and questionable between 2 and 3. A score above 3 triggers investigation and corrective action.

Question 4: Traceability of calibration to SI units supports accuracy by:

  • A. Traceability of calibration to SI units supports accuracy by convention
  • B. Linking measurements through an unbroken chain to international standards
  • C. Traceability of calibration to SI units supports accuracy by cost
  • D. Traceability of calibration to SI units supports accuracy by tradition

Correct answer: Linking measurements through an unbroken chain to international standards

Calibration traceability means every measurement links back through an unbroken chain of calibrations to international (SI) standards, each step carrying a stated uncertainty. The JCGM's International Vocabulary of Metrology (VIM) defines traceability exactly this way. That chain is what lets a result in one lab be meaningfully compared with one in another.

Question 5: Standardising a titrant against a primary standard confirms:

  • A. Standardising a titrant against a primary standard results in faster titrations
  • B. The titrant's true concentration, the basis of accurate results
  • C. Standardising a titrant against a primary standard produces pretty colours
  • D. Standardising a titrant against a primary standard means less washing

Correct answer: The titrant's true concentration, the basis of accurate results

A titrant's nominal concentration is only a starting guess; standardizing it against a primary standard : a pure, stable substance of known stoichiometry such as potassium hydrogen phthalate : fixes its true concentration. Every titration calculation downstream rests on that value. Without standardization, titrant strength is an assumption, not a measurement.

Question 6: Accuracy across the working range is checked by:

  • A. Accuracy across the working range is checked by one test
  • B. Accuracy across the working range is checked by estimating
  • C. CRMs or spikes at low, mid and high levels
  • D. Accuracy across the working range is checked by averaging

Correct answer: CRMs or spikes at low, mid and high levels

Accuracy is not uniform by assumption: certified reference materials or spikes must be run at low, mid, and high levels across the working range. The Eurachem Guide on method validation calls for exactly this multi-level check. A method accurate at one concentration can still fail at the extremes.

Question 7: Certified reference materials (CRMs) provide:

  • A. Certified reference materials (CRMs) provide cheap samples
  • B. Certified reference materials (CRMs) provide practice material
  • C. Values certified with uncertainty by an authoritative body
  • D. Certified reference materials (CRMs) provide waste

Correct answer: Values certified with uncertainty by an authoritative body

A certified reference material ships with a certified value and an associated uncertainty, assigned by an authoritative body (often through interlaboratory certification under ISO Guide 35). That certificate is the anchor the lab's result is judged against. Cheap or practice materials carry no such certified values.

Question 8: If a CRM reads consistently low, the method has:

  • A. The method has good accuracy
  • B. A negative bias needing investigation
  • C. The method has great precision
  • D. The method has no problem

Correct answer: A negative bias needing investigation

A CRM that reads consistently low means the method carries a systematic error : a negative bias : rather than random scatter. Under ISO 5725, bias is assessed against reference values precisely this way. The finding calls for investigation of the cause, not acceptance of the results.

Question 9: Correcting for blank values improves:

  • A. Correcting for blank values improves precision
  • B. Correcting for blank values improves speed
  • C. Correcting for blank values improves morale
  • D. Accuracy, by removing background signal

Correct answer: Accuracy, by removing background signal

The blank captures signal from reagents, glassware, and the instrument itself : background that is not the analyte. Subtracting it removes that systematic contribution from the result. The correction improves trueness, hence accuracy, and it is standard practice in trace analysis.

Question 10: Accuracy can be lost through:

  • A. Accuracy can be lost through good training
  • B. Accuracy can be lost through clean glassware
  • C. Expired standards, matrix interference, or calibration errors
  • D. Accuracy can be lost through new equipment

Correct answer: Expired standards, matrix interference, or calibration errors

Accuracy dies quietly: standards degrade past their expiry, matrix components suppress or enhance the analyte signal, and calibration mistakes shift every result. Each is a classic source of bias documented in method-validation literature. Good training and new equipment do not prevent these losses : checks do.

Question 11: An accurate method that analysts perform inconsistently shows:

  • A. An accurate method that analysts perform inconsistently shows good precision
  • B. Poor precision despite good trueness, training/SOP issues
  • C. An accurate method that analysts perform inconsistently shows bias
  • D. An accurate method that analysts perform inconsistently shows fraud

Correct answer: Poor precision despite good trueness, training/SOP issues

When an accurate method is run inconsistently by different analysts, the lab sees good trueness but poor precision : results scatter around the right value. That pattern points to technique variation and training or SOP weaknesses, not method bias. ISO 5725 separates the two components precisely so the fix targets the right one.

Question 12: Documenting accuracy evidence (CRMs, recoveries) is needed for:

  • A. Accuracy evidence is needed for presentation
  • B. Accuracy evidence is needed for fun
  • C. Audits, accreditation and defending results
  • D. Accuracy evidence is needed for filing practice

Correct answer: Audits, accreditation and defending results

Audits and accreditation bodies demand objective evidence : CRM results, recovery data, calibration records : that reported numbers are trustworthy. ISO/IEC 17025 makes such records a requirement, not a courtesy. Filing them is how a lab defends its results when challenged.

Question 13: Proficiency testing (interlaboratory comparison) assesses:

  • A. Proficiency testing (interlaboratory comparison) assesses staff popularity
  • B. Proficiency testing (interlaboratory comparison) assesses equipment brands
  • C. Proficiency testing (interlaboratory comparison) assesses coffee quality
  • D. A lab's accuracy against peers and assigned values

Correct answer: A lab's accuracy against peers and assigned values

Proficiency testing (interlaboratory comparison) scores a laboratory's result against an assigned value and against peer laboratories under ISO 13528. The exercise assesses accuracy in the real world, including sample handling and calibration. It tests the whole measurement process, not staff or equipment brands.

Question 14: Trueness and precision together determine:

  • A. Accuracy, trueness is lack of bias, precision is lack of scatter
  • B. Trueness and precision together determine the overall monetary cost
  • C. Trueness and precision together determine the overall operational speed
  • D. Trueness and precision together determine the overall staffing levels

Correct answer: Accuracy, trueness is lack of bias, precision is lack of scatter

ISO 5725-1 defines accuracy as the combination of trueness (closeness of the mean to the true value : freedom from bias) and precision (closeness of repeat results : freedom from scatter). Both components are needed: a result can be precise yet biased, or unbiased yet scattered. Cost, speed, and staffing are operational concerns, not metrological ones.

Question 15: A recovery experiment tests accuracy by:

  • A. Adding a known amount of analyte and measuring how much is found
  • B. A recovery experiment tests accuracy by recovering from illness
  • C. A recovery experiment tests accuracy by recycling all waste
  • D. A recovery experiment tests accuracy by always resting more

Correct answer: Adding a known amount of analyte and measuring how much is found

A recovery experiment spikes a known amount of analyte into the sample matrix and measures how much is found, expressed as a percentage. The result directly tests the method's accuracy in the presence of matrix effects. Recovery near 100% means the method finds what is truly there.

Question 16: Accuracy is assessed using:

  • A. Accuracy is assessed using replicates alone
  • B. Accuracy is assessed by estimating
  • C. Certified reference materials or spiked recoveries
  • D. Accuracy is assessed using averages

Correct answer: Certified reference materials or spiked recoveries

Accuracy is assessed against the truth, and the truth comes from certified reference materials or from spiked recoveries of known additions. Replicates alone measure precision : how tightly results cluster : not closeness to the true value. ISO 5725 keeps these assessments distinct for this reason.

Question 17: Bias is:

  • A. In measurement science, bias is mere human prejudice
  • B. In measurement science, bias is just random noise
  • C. In measurement science, bias is just an estimate
  • D. The systematic difference between the expected result and the true value

Correct answer: The systematic difference between the expected result and the true value

In measurement science, bias is the systematic difference between the expected value of measurement results and the true value : a consistent offset, not human prejudice or random noise. ISO 3534-2 carries this definition. Random scatter around the truth is precision's problem; bias is a directional error.

Question 18: Why does accuracy matter more than precision for food safety decisions?

  • A. Accuracy does matter more than precision for food safety decisions because it doesn't
  • B. A precise but biased result can confidently declare unsafe food safe
  • C. Accuracy does matter more than precision for food safety decisions because precision is expensive
  • D. Accuracy does matter more than precision for food safety decisions because auditors prefer it

Correct answer: A precise but biased result can confidently declare unsafe food safe

For food safety decisions, trueness matters more than tight clustering: a precisely repeatable but biased result can confidently declare contaminated food safe. The 2008 melamine adulteration crisis showed how systematically wrong numbers carry real public-health consequences (FDA/EFSA investigations). Precision without trueness is confidence without truth.

Question 19: A z-score in proficiency testing is calculated as:

  • A. A z-score in proficiency testing is calculated as (result − median)/range
  • B. A z-score in proficiency testing is calculated as result × 2
  • C. A z-score in proficiency testing is calculated as intuition
  • D. (lab result − assigned value) / standard deviation for proficiency

Correct answer: (lab result − assigned value) / standard deviation for proficiency

ISO 13528 defines the proficiency-testing z-score as the laboratory's result minus the assigned value, divided by the standard deviation for proficiency assessment: z = (x - X)/sigma_pt. It standardizes how far a result sits from the consensus. Median-over-range and doubling-the-result are not recognized formulas.

Question 20: Accuracy in analysis means:

  • A. Closeness of the measured value to the true value
  • B. Accuracy in analysis means tight clustering of data
  • C. Accuracy in analysis means fast results only
  • D. Accuracy in analysis means cheap tests only

Correct answer: Closeness of the measured value to the true value

ISO 5725-1 defines accuracy as the closeness of a measured value to the true value : trueness plus precision combined. Tight clustering alone is precision; speed and cost are logistics. Accuracy is the property that tells a laboratory its number is right, not merely repeatable.