Instrument Calibration Quiz
GIFSQ ID: ART-1394
Test your knowledge with this free 20-question Instrument Calibration quiz from Food Analysis & Analytical Chemistry, part…
What this quiz covers
Calibration is the comparison of an instrument’s readings against known standards, with adjustment or correction as needed, and this quiz covers the practice end to end. It starts with the calibration curve itself, instrument response plotted against known standard concentrations, and the blank that anchors the bottom of the curve by measuring the response with zero analyte. Standards should bracket the expected sample range and be matrix-matched where food components change the response, internal standards correct for injection and instrument variation, and a check standard analysed with the samples verifies the calibration still holds during the run. If a check standard fails, the run stops: investigate, recalibrate and re-analyse the affected samples.
The rest of the quiz is about the traps that a good-looking number can hide. A correlation coefficient of at least 0.99 is typical for quantitative methods, but residual plots reveal the non-linearity and outliers that a good r value can hide, and even software-fitted curves need analyst review of r, residuals and range before acceptance. Extrapolating beyond the calibrated range is risky because the response may stop being linear, single-point calibration is acceptable only when the response is proven linear through zero over a narrow range, and temperature matters because volumes and instrument responses both shift with it. Balances are calibrated with certified weights traceable to national standards, pH meters with certified buffers such as pH 4.00, 7.00 and 10.00, and volumetric glassware is calibrated for critical work because manufacturing tolerances have limits. Calibration intervals follow a schedule based on stability, use and risk, plus recalibration after maintenance, and a due-date system stops anyone using an instrument whose calibration has expired.
One idea runs through the set: a number is only as good as the curve behind it
Every quantitative result in a food laboratory traces back to a calibration, and the records, standards used, dates, results, acceptance criteria and actions taken, are the audit trail that proves it. The curve, the residuals and the check standard are the three things to look at before trusting any result.
How to use this quiz
Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the standards, ranges and record-keeping rules a food analysis student is expected to know by heart.
Choose how you want to play
A short note: instrument calibration
Calibration means checking an instrument's readings against known standards and correcting as needed. A calibration curve plots response against known standard concentrations. A blank measures response with zero analyte, anchoring the baseline. Standards should bracket the expected sample range, matrix-matched where needed.
The correlation coefficient, r, should be 0.99 or higher for most quantitative methods. Residual plots help to show non-linearity or outliers that a good-looking r value hides. Software curves still need analyst review of r, residuals and range. Extrapolating beyond the range is risky: the response may not stay linear. Single-point calibration works only when the response runs linear through zero over a narrow range.
Instruments are calibrated on a schedule set by stability, use and risk, plus after maintenance. A due-date system stops instruments with expired calibration being used. Records must show standards used, dates, results and actions taken. Temperature affects calibration because volumes and instrument responses change with temperature.
A check standard is a known standard run with samples to verify calibration holds. If it fails, stop, investigate, recalibrate and repeat the affected samples. Balances use certified test weights traceable to national standards. A pH meter is calibrated with certified buffers: pH 4, 7 and 10. Volumetric glassware needs calibration for critical work: manufacturing tolerances have limits. Use matrix-matched calibration when food components alter response versus pure-solvent standards. Internal standards correct for injection and instrument variation, lifting accuracy and precision.
Practical takeaway: calibrate on schedule, check the standard mid-run, never extrapolate.
Questions and answers
Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.
Question 1: Residual plots of a calibration help to:
Correct answer: Reveal non-linearity or outliers hidden by a good-looking r value
Residuals - the gaps between measured points and the fitted line - reveal curvature or outliers that a good-looking r value can hide. A random scatter of residuals is the real check on linearity.
Question 2: Calibration records must show:
Correct answer: Standards used, dates, results, acceptance and actions taken
Good records show which standards were used, the dates, the results, the acceptance criteria, and what actions were taken - the audit trail ISO/IEC 17025:2017 expects. The analyst's mood is not a record.
Question 3: How often should instruments be calibrated?
Correct answer: Per schedule based on stability, use and risk, plus after maintenance
Calibration intervals are set by schedule from the instrument's stability, how heavily it is used, and the risk of bad data - plus recalibration after any maintenance. 'Once ever' and 'yearly only' ignore all three factors (ISO/IEC 17025:2017 §6.4).
Question 4: A calibration due-date system prevents:
Correct answer: Use of instruments with expired calibration
A due-date system stops analysts from using instruments whose calibration has expired. No result is then ever produced on an unqualified instrument.
Question 5: Temperature affects calibration because:
Correct answer: Volumes and instrument responses change with temperature
Temperature changes liquid volumes and shifts instrument responses, so calibrations are temperature-sensitive. Standards and instruments are therefore specified at reference temperatures.
Question 6: A blank in calibration:
Correct answer: Measures the response with zero analyte, anchoring the curve's baseline
The blank measures the instrument's response with zero analyte, anchoring the bottom of the calibration curve. Without it, the curve's baseline is a guess.
Question 7: Automatic (software) calibration curves still need:
Correct answer: Analyst review, checking r, residuals and range before accepting
Software can fit the line, but the analyst must still review r, the residuals, and whether samples fall inside the calibrated range before accepting it. Blind trust in the fit is how bad curves get used.
Question 8: A calibration curve is:
Correct answer: A plot of instrument response versus known standard concentrations
A calibration curve plots instrument response against known standard concentrations, giving the conversion from signal to concentration. It is the working reference for every quantitative result.
Question 9: What is calibration?
Correct answer: Comparing an instrument's readings to known standards and adjusting/correcting as needed
Calibration compares an instrument's readings against known standards and adjusts or corrects the readings as needed. It is neither estimating, cleaning, nor repairing (VIM definition, via ISO/IEC 17025:2017).
Question 10: Calibrating volumetric glassware is:
Correct answer: Needed for critical work, manufacturing tolerances have limits
For critical work, volumetric glassware must be calibrated because manufacturing tolerances have limits - a nominal 100 mL flask is not exactly 100.00 mL. Routine work can rely on the class tolerance; critical work cannot.
Question 11: Extrapolating beyond the calibration range is:
Correct answer: Risky, response may not stay linear outside measured standards
Beyond the highest and lowest standards, the response may stop being linear, so extrapolated results are unreliable. Samples should be diluted or concentrated into the calibrated range instead.
Question 12: Balance calibration uses:
Correct answer: Certified test weights traceable to national standards
Balances are calibrated with certified test weights traceable to national mass standards, giving an unbroken traceability chain. Feathers, coins, and stones are not standards.
Question 13: Matrix-matched calibration is needed when:
Correct answer: Food components alter instrument response versus pure-solvent standards
Matrix-matched calibration is needed when food components change the instrument's response compared with pure-solvent standards - the matrix effect. Standards prepared in blank matrix cancel it out (EU SANTE/11312/2021).
Question 14: Internal standards in chromatography:
Correct answer: Correct for injection and instrument variation, improving accuracy and precision
An internal standard added to every sample experiences the same injection and instrument variation as the analyte. Ratioing the analyte signal to it corrects that variation and improves accuracy and precision.
Question 15: What is a 'check standard'?
Correct answer: A known standard analysed with samples to verify calibration holds during the run
A check standard is a known standard analyzed alongside samples to verify the calibration still holds during the run. It is the routine guard against drift.
Question 16: If a check standard fails, you should:
Correct answer: Stop, investigate, recalibrate and re-analyse affected samples
A failed check standard means the calibration may have shifted, so the run stops: investigate, recalibrate, and re-analyze the affected samples. Ignoring, deleting, or averaging it in hides the problem.
Question 17: Single-point calibration is acceptable when:
Correct answer: Response is proven linear through zero and the range is narrow
Single-point calibration is acceptable only when the response is proven linear through zero and the working range is narrow - then one point plus the origin defines the line. Otherwise, multiple points are required.
Question 18: A pH meter is calibrated with:
Correct answer: Certified buffer solutions (e.g. pH 4.00, 7.00, 10.00)
pH meters are calibrated with certified buffer solutions - typically pH 4.00, 7.00, and 10.00 - that bracket the measurement range. Water and juice are not standards; their pH is unknown.
Question 19: Calibration standards should:
Correct answer: Bracket the expected sample range and be matrix-matched where needed
Calibration standards should bracket the expected sample concentrations and be matrix-matched where the matrix affects response. Standards that sit off to one side of the samples invite extrapolation error.
Question 20: The correlation coefficient (r) of a calibration curve should typically be:
Correct answer: ≥0.99 for most quantitative methods
For most quantitative methods, the calibration curve's correlation coefficient should be at least 0.99 - evidence of the proportional response the method assumes. Values near 0.5, 0, or negative mean the 'calibration' is not one.