SOP: Oxygen Transmission Rate of Whole Packages (ASTM F1307)

GIFSQ ID: SOP-0992

A film result, however careful, still leaves one question unanswered: how much oxygen enters the finished pack, seals, closures and all? Whole-package measurement answers that question directly. One completed, empty package is mounted on a fixture, its inside is washed with nitrogen, and whatever oxygen works its way through the package walls is carried to the same kind of coulometric sensor used for flat film.

Because the specimen is a three-dimensional object rather than a flat coupon, mounting and equilibration carry the method. Given a sound mount, the measurement itself is a steady-state one: wait until the package and its surroundings have settled, then read the oxygen arriving in the carrier stream.

Purpose

Purpose is to determine the oxygen transmission rate of a complete dry package, in volume of oxygen per unit time for the package as a whole, using a coulometric sensor under ASTM F1307.

Scope

Scope covers dry packages tested whole: bottles, pouches and finished packs mounted on a test fixture, as distinct from flat film cut from the same material. Reporting is per package, not per unit area, because a finished pack has no single meaningful test area in the way a film coupon does, and oxygen may enter through every part of its wall. Packages are tested in the dry state named in the standard’s title; a pack holding a moist product is outside the arrangement described in the source. An unsteady-state variant, the ambient oxygen ingress rate method, is included as a second branch because the source presents it on the same mounting.

Equipment and Materials

  • Test fixture that holds a complete package and connects its interior to a nitrogen purge line and to the sensor line
  • Coulometric oxygen sensor and associated equipment in an arrangement similar to ASTM D3985, with a cadmium anode and graphite cathode as described for that sensor
  • Nitrogen carrier gas at a known, constant flow rate; a nitrogen and hydrogen mixture where trace oxygen in process gas must be removed catalytically, as in the film method
  • Known oxygen environment for the outside of the package: ambient room air at 20.9% oxygen, or an atmosphere of 100% oxygen enclosing the package
  • Finished, empty, dry packages representative of the packs to be judged, undamaged and with their normal closures fitted
  • Oxygen measurement for the ingress branch: a means of measuring oxygen concentration inside the package periodically, of the kind described in the source study

Procedure

  1. Select the package. Take a finished package in its normal closed form. Inspect it first; a creased, cracked or badly sealed pack measures its damage, not its design.
  2. Mount the package on the fixture. Fix the package so its interior can be purged while its outside stays exposed to the test atmosphere. Sealing at the fixture connection must not add a second, easier path for oxygen, and must not block part of the package wall in a way that changes what is being measured, as general laboratory practice for fixture-mounted specimens.
  3. Equilibrate with the test environment. Leave the mounted package until it has reached equilibrium with the surrounding test conditions. Transmission figures taken before equilibrium describe the approach to equilibrium, not the package.
  4. Purge the inside with nitrogen. Run a slow stream of nitrogen through the package interior at a known, constant flow rate, and keep that flow running for the whole test. Choose a flow high enough that the oxygen concentration inside stays below the ambient concentration yet not zero.
  5. Expose the outside to known oxygen. Either leave the package in ambient room air containing 20.9% oxygen, or immerse it in an atmosphere of 100% oxygen. Record which atmosphere was used; the two are different driving forces and results belong with the condition that produced them.
  6. Carry permeated oxygen to the sensor. Oxygen crossing the package walls enters the nitrogen stream and is transported to the coulometric sensor, where it generates current in proportion to the amount arriving, as in the film method.
  7. Wait for the equilibrium reading. Continue until the oxygen concentration in the carrier stream reaches its equilibrium level under the steady purge. That plateau, not an early rising value, carries the result.
  8. Branch B: ambient oxygen ingress rate. Where the unsteady-state variant is wanted, mount the package in the same way, then purge it free of oxygen at the start. Stop the through-flow arrangement used above and let oxygen accumulate inside the package over time, measuring the internal oxygen concentration periodically. Transmission is worked out from how fast oxygen accumulates.

Recording Results

For the steady-state branch, determine oxygen transmission rate in volume per unit time from the equilibrium oxygen concentration in the nitrogen stream, whose flow rate is known and constant: the fraction of oxygen in that stream, multiplied by the stream flow, gives the volume of oxygen entering the package per unit time. Report the value for the package as a whole, together with the external oxygen atmosphere used (ambient air at 20.9% oxygen or 100% oxygen), test temperature and humidity where controlled, and package identification. For the ingress branch, report the rate of oxygen accumulation inside the initially oxygen-free package and the interval over which concentration was followed. Never convert a whole-package result into a per-area film value and present it as a film property; geometry, seals and closures are built into the package figure.

Quality Control

Mounting quality decides whether a run means anything, so repeat runs on comparable packages deserve comparison before any single value is quoted. Sensor care follows the film method, because the same coulometric sensor is used: sensors are sensitive, costly and finite in life, with life governed by total oxygen exposure, and they should stay bathed in oxygen-free gas when not measuring. Gas purity matters just as much at package scale: oxygen hiding in the nitrogen supply would be read as oxygen that had crossed the package wall. A package that gives a wildly different result from its duplicates should be examined for damage or a fixture leak before its value is accepted, as general laboratory practice.

Troubleshooting

  • Internal oxygen concentration falls to zero: the purge flow is too strong for the ingress rate; the source requires an inside concentration below ambient but not zero, so reduce the flow toward a level that keeps a measurable concentration, without inventing a set value the source does not give.
  • Reading never reaches a plateau: allow longer for the package and test environment to equilibrate, and check the fixture connection for leakage.
  • Result far higher than the film data suggest: inspect seals, closures and the fixture mount; whole-package testing captures every entry path, which is precisely its value, but a damaged pack should be replaced and the run repeated.
  • Low-temperature testing is planned: coulometric sensors are difficult to use below 10°C because condensation can damage them, a limitation carried over from the film method’s sensor.

Interferences and Limitations

Only dry packages sit inside the described method, so moisture-driven changes in oxygen barrier, familiar in water-sensitive polymers such as ethylene vinyl alcohol and polyamide whose diffusion behaviour depends on water content, are not reproduced by a dry, empty pack. Temperature shifts results in the same Arrhenius fashion described for permeation generally, so a value belongs to its test temperature. A whole-package figure also cannot say which part of the pack let the oxygen in; separating wall, seal and closure contributions requires film testing or further investigation, not this measurement alone. Finally, the source gives no fixture design or flow-rate value, so laboratories must take those from the standard itself and from the instrument maker, and record what was used.

Safety Notes

  • Handle oxygen-enriched atmospheres, nitrogen supplies and any hydrogen-containing cleaned carrier gas under their safety data sheets and cylinder rules, as general laboratory practice.
  • Guard against pressure build-up when a closed package is purged; keep flows slow, as the method requires, and vent the exhaust safely, as general laboratory practice.
  • Route spent cadmium-anode sensors to cadmium-containing waste, as general laboratory practice.

Source and Standard

Drafted from Chapter 5 of Food Packaging Materials: Testing & Quality Assurance (Singh, Wani and Langowski, eds., CRC Press, 2017), section 5.2.4 on measuring the oxygen transmission rate of whole packages, including its ambient oxygen ingress rate variant, with sensor detail drawn from the same chapter’s ASTM D3985 sections. Governing standard: ASTM F1307, Standard Test Method for Oxygen Transmission Rate Through Dry Packages Using a Coulometric Sensor.