SOP: Determining Water Activity in Foods with the Aqualab Chilled-Mirror Instrument (MFLP-66)

GIFSQ ID: SOP-1060

Principle

An Aqualab meter finds water activity by watching for dew. Its chilled-mirror dew point technique cools a clean mirror until condensation forms, and that dew point temperature relates directly to the vapour pressure of the air in the chamber, and so to how much of a food’s water is available for chemical and biological reactions, including microbial growth. Daily verification against salt standards of known aw, capped by a distilled-water check at 1.000, is what anchors every sample reading that follows.

Sample handling decides accuracy just as much as the mirror does. Packages stay intact until measurement, cups are filled and wiped with care, and awkward products, butter, foods packed in oil, multi-ingredient items, each get their own preparation so the number reflects the food as eaten.

Purpose

Purpose is to measure water activity (aw) in foods and food ingredients, including for judging compliance with Sections 4 and 7 of the Food and Drugs Act and any other relevant federal regulations. Water activity is the ratio of the water vapour pressure of a food to the vapour pressure of pure water at the same temperature and pressure; thermodynamically it rests on chemical potential and its relationship to the osmotic pressure of an aqueous solution. Equilibration, for this instrument, means any two aw readings less than 0.001 apart, at which point the measurement process stops. An Aqualab verification standard is an unsaturated salt solution of a specific molality representing a known aw, while a verification standard in saturated form is a saturated solution of a salt of known % relative humidity at a designated temperature.

Scope

Scope covers foods and food ingredients, with each sample unit analysed individually or analytical units combined. One named exclusion governs applicability: this method may not be applicable to foods containing volatile components, because volatiles such as alcohols interfere with chilled-mirror dew point technology, and the concentration that causes interference varies with the food matrix. Model-specific operating detail for different Aqualab instruments sits in the operator’s manuals supplied by the manufacturer; this procedure covers sample preparation and water activity measurement in general. Work is done in accordance with ISO/IEC 17025 (latest version), General Requirements for the Competence of Testing and Calibration Laboratories, as the laboratory supervisor must ensure. Samples must be kept sealed during transport to prevent loss or gain of moisture.

Equipment and Reagents

  • Aqualab instrument (chilled-mirror dew point), sited on a flat, levelled surface in an area of stable ambient temperature
  • Aqualab verification standards (recommended standard solutions), or in-house saturated saline verification standards prepared from the reference salts below
  • Reference salts, ACS reagent grade, fine crystal — water activity values (± 0.003) at 20°C and 25°C: LiCl 0.113 / 0.113; 13.41M LiCl 0.245 / 0.250; MgCl2 0.331 / 0.328; K2CO3 0.432 / 0.432; 8.57M LiCl 0.496 / 0.500; Mg(NO3)2 0.544 / 0.529; NaBr 0.591 / 0.576; 6.0M NaCl 0.760 / 0.760; KBr 0.817 / 0.809; KCl 0.851 / 0.843; KNO3 0.946 / 0.936; K2SO4 0.976 / 0.973; 0.5M KCl 0.984 / 0.984; distilled H2O 1.000 / 1.000
  • Aqualab disposable sample cups or equivalent, always clean and new for verification standards; lids, tape or Parafilm for sealing filled cups
  • Aqualab cleaning kit (optional); sieves or strainers of appropriate mesh; absorbent material such as Kimwipes, cheesecloth or paper towels; spatula
  • Distilled water (steam distilled water where a water reading will not come right), sealed containers for stored cups, and for long-term storage glass jars or mylar or plastic bags
  • Optional ambient temperature incubator capable of maintaining 20–25°C, for bringing samples to room temperature

Sample Preparation

  1. Keep sample units refrigerated or frozen before analysis according to the nature of the product, except for shelf-stable foods, and maintain package integrity until the point of measurement to prevent water loss that could alter water activity. Analyse units as soon as possible after receipt.
  2. Bring samples to room temperature (~ 21°C) and let them equilibrate there; an ambient temperature incubator may be used for this. Have verification standards at ambient temperature as well.
  3. Prepare the sample so it represents the entire product, leaving it intact where possible so its aw is not altered; most materials need no preparation beyond cutting to fit the sample cup. Where an investigation requires it, only part of a sample may be tested, for example separating garlic and herbs in oil. Discard non-edible parts where appropriate, such as cheese wax, packing liquid or olive pits.
  4. Scrape any loose salt crystals or similar material off the sample surface with a spatula, then wipe off remaining crystals with a paper towel.
  5. Fill plastic sample cups half full, in a manner that covers most of the bottom of the cup. Clean the exterior surfaces, sides and rim, of each cup to avoid contaminating the reading chamber.
  6. For a multi-ingredient food, choose the approach circumstances require: prepare the product as a whole by the applicable preparation below, and/or analyse each component individually. Component testing is at the laboratory’s discretion for foods where aw is critical for control of pathogens such as Listeria monocytogenes or Clostridium botulinum. To test components, drain the container contents into a suitable strainer or sieve of appropriate mesh to retain the solids, retain any liquid portion, soak excess liquid off the solids with absorbent paper without drying them, place each solid ingredient in its own sample cup, and measure each ingredient, including the liquid if applicable.
  7. For a product packed in a medium such as oil, brine or vinegar that the consumer normally discards, drain it through sieves of appropriate mesh, draining sufficiently to remove as much packing medium as possible, then soak off excess remaining medium with absorbent paper, do not dry, and measure aw on the solid portion. For a product packed in a medium the consumer normally uses, shake well before opening, and take a portion of the medium for aw measurement before draining, for other analyses such as pH determination.
  8. For butter or other water-in-oil emulsions such as pesto, expect dramatically prolonged equilibration times because lipids act as a barrier to vapour diffusion. Run these samples in a chamber that has a lower aw than the sample or that has been dried with desiccant before measurement, and take 3 readings if needed to fall within the ± 0.003 range.
  9. For coated foods, dried foods, highly viscous water-in-oil and high fat samples, apply the special preparation for the commodity and take special care to keep preparation consistent from sample to sample.
  10. Hold prepared samples correctly when measurement cannot happen at once: a prepared unit should be analysed the same day. If not, tape the cup/lid junction of filled cups with tape or parafilm, place them in a sealed container with excess air removed, and store at a temperature appropriate to the sample: shelf-stable samples at room temperature, degradable samples refrigerated. For long-term storage use glass jars or mylar/plastic bags, minimize headspace to reduce water vapour and condensation effects, and store the containers in a freezer. Pre-warm refrigerated or frozen samples to ambient temperature before measurement to reduce equilibration time and give a more accurate reading, taking care to avoid loss of water.

Procedure

  1. Prepare in-house verification standards if used: place the selected reference salt in a labelled test container to a depth of about 4 cm, then add distilled water in gradual increments with constant stirring until no more water is absorbed and some free liquid sits on top. Such standards are then ready to use and keep indefinitely in a sealed container that prevents evaporation losses. When measuring with them, use the liquid portion of the mixture only.
  2. Verify calibration daily, before sample measurement, with two different verification standards, always starting with the standard of lowest aw value. Use clean, new sample cups for the standards.
  3. For routine food analysis, either use 6M NaCl, which has aw = 0.760, or choose a verification standard with an aw value lower than the anticipated sample range, remembering that only samples within the range of that standard and the aw of distilled water can be reported as a numerical value.
  4. Measure the chosen standard in duplicate. Compare its readings and temperatures with the theoretical values in the reference-salt table in Equipment and Materials.
  5. If the standard readings fall within the stated ranges, take duplicate readings of distilled water. Verification is complete when the distilled water readings fall within 1.000 ± 0.003. If the water readings are not correct, clean the sensor as the operating manual describes and repeat the verification from the duplicate standard measurement; if the aw of water still will not read correctly, use steam distilled water, which will always read 1.000.
  6. If standard readings fall outside the stated ranges even after re-checking, adjust the meter reading to its correct value by carefully adjusting the potentiometer or by performing an instrument calibration, following the operator’s manual, then repeat the whole verification from the duplicate standard measurement.
  7. Optionally, check linearity across the range: measure the aw of 4 verification standards, the upper and lower standards plus 2 others with aw values between them. Once the upper and lower standards have been adjusted as above, measure the other 2 standards in duplicate.
  8. Place the prepared sample cup in the drawer or sample chamber and slide the drawer closed, taking care not to spill or splash liquid samples. Move the device to the “Read” position to seal the cup in the chamber and start the read cycle; the reading is complete when the instrument beeps and/or the decimal points or LCD flash, and the display shows a final aw and temperature.
  9. Measure each sample in duplicate, taking 2 readings of the same sample cup. Record the aw reading and the temperature. To minimize reading time, work through samples from the driest food to the most humid, and check calibration stability by measuring distilled water after each food type and/or at the end of the day, repeating the verification steps if values are out of range.

Calculation and Reporting

Report sample aw values using three or four decimal digits, depending on the accuracy of the device. From multiple readings, each sample is measured in duplicate, determine the mean without rounding, then report the final mean value rounded to two decimal places. Only samples within the aw range of the tested verification standards and the aw of distilled water earn a numerical value. Report a sample below that range as less than the value of the lower humidity standard used — for example < 0.760 if 6M NaCl is used — and a sample above it as greater than the higher humidity standard used — for example > 0.984 if 0.5M KCl is used. Where components of a multi-ingredient food were measured separately, report the aw for each ingredient as required. Record the temperature alongside every aw reading.

Quality Control

Daily verification before use, with two different verification standards starting from the lowest aw, is the core control, and it is only complete when the standard readings sit inside their stated ranges and duplicate distilled water reads 1.000 ± 0.003. Replace verification standards, commercial or in-house, whenever they become contaminated, are found to be inaccurate, or are expired. Judge the accuracy of uncertified verification standards by an independent route: test blind samples at another facility on the same type of aw apparatus, test blind samples on a different type of instrument, or send blind samples to an external calibration laboratory to certify the standards. Recheck stability during the day by measuring distilled water after each food type and/or at day’s end, returning to full verification whenever values fall out of range. Always use clean, new cups for standards, and keep the instrument on a flat, levelled surface in a stable ambient temperature so verification reflects the meter rather than its surroundings.

Troubleshooting

  • Distilled water does not read 1.000 ± 0.003: clean the sensor as the operating manual describes and repeat verification from the duplicate standard measurement; if the reading still fails, switch to steam distilled water, which will always read 1.000.
  • Verification standard outside its stated range after re-checking: adjust the meter to the correct value by careful potentiometer adjustment or a full instrument calibration per the operator’s manual, then repeat the whole verification.
  • Standards cannot be made linear even after chamber cleaning, calibration verification or potentiometer adjustment: consult the Troubleshooting section of the operator’s manual; if non-linearity continues, contact the manufacturer.
  • Very slow readings on butter, pesto or other water-in-oil emulsions: lipids block vapour diffusion; run the sample in a chamber of lower aw than the sample or dried with desiccant beforehand, and take 3 readings if needed to land within ± 0.003.
  • Suspected volatile interference: compare dew point readings against capacitance readings from a meter fitted with a capacitance sensor. A difference greater than 0.018 aw units, as specified by the manufacturer, is significant; without such a meter, samples with volatile components cannot be tested by this method.
  • Spilled or splashed liquid in the chamber: avoid it at the loading step by closing the drawer carefully, and clean the chamber and sensor as the operating manual directs before verifying again.

Interferences and Limitations

Volatiles are the standing limitation of chilled-mirror measurement: alcohols and similar components interfere with dew point readings at concentrations that vary with the food matrix, and foods containing volatile components may not be applicable to this method at all. Significance of the interference is decided only by comparison with a capacitance-sensor meter, with a difference greater than 0.018 aw units counting as significant; a laboratory without that comparison meter cannot test volatile samples under this method. High-fat and water-in-oil products equilibrate slowly because lipids bar vapour diffusion, so their readings carry extra handling conditions. Numerical reporting is bounded by the verified range, from the lower verification standard up to distilled water; anything outside is a less-than or greater-than result. Moisture loss before measurement, through broken packaging, delayed analysis or warm storage, changes the very quantity being measured, and special commodities such as coated, dried and highly viscous samples demand commodity-specific preparation applied consistently.

Safety and Waste

  • As generic laboratory practice, wear gloves and eye protection when preparing samples and salt standards, wash hands after handling foods and chemicals, and keep foods for testing separate from anything intended for consumption.
  • Reference salts are laboratory chemicals: prepare saturated solutions with care, label every container, and dispose of spent standards and cups as laboratory waste under local practice.
  • Handle knives, spatulas and sieve edges used in cutting samples to fit cups as sharps, and clean spills of oils or brines promptly to keep bench and floor surfaces safe.
  • Follow the manufacturer’s operator manual for cleaning the chilled mirror and chamber; never introduce cleaning materials or procedures the manual does not specify.

References and Control

Adapted in original wording from Health Canada method MFLP-66, Determination of Water Activity Using the Aqualab Instrument, January 2020, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-66 dated August 2014. Published 23 October 2026. No validation record is given on this page.