SOP: Measuring Salt (Sodium Chloride) in Foods with Quantab Chloride Titrators (MFLP-45)
GIFSQ ID: SOP-1046
Salt measurement here needs no burette and no bench titration. A pre-calibrated Quantab Chloride Titrator strip carries its own reagents, draws up a filtered hot-water extract of the food by capillary action, and converts chloride into a white column whose height reads directly as salt. Convenience is the point: a result in percent sodium chloride, or in ppm chloride ion, can be obtained from most foods with little more than a blender, boiling water and filter paper.
Simplicity does not excuse sloppy sampling. Everything depends on a representative, finely blended portion and on accurate weights and volumes, because the strip can only measure the extract it is given.
Purpose
Purpose is to determine chloride content in most foods, reported as % sodium chloride or as ppm Cl− ions on a wet-weight basis. With a separate moisture result, the same measurement also gives % salt in the water phase, also called % brine. Quantab titrators measure chloride in aqueous solutions or in dilute aqueous extracts of solids, and each strip already contains the reagents needed for a standard chloride titration.
Scope
Scope covers most foods. One boundary governs regulated work: where a regulation names an Official Method, follow that method instead: for example, FO 38 for % salt in smoked fish products. Moisture content must be measured separately whenever a water-phase or brine result is required; choose an acceptable moisture method for the food from the Official Methods of the Association of Official Analytical Chemists (AOAC) or an equivalent source. Quantab is a registered trademark of Environmental Test Systems, Inc., Ames Co., a division of Miles Laboratories Inc., Elkhart, IN.
Equipment and Materials
- Quantab Chloride Titrator kit, with titrator strips and the information summary sheet supplied with the kit; kits were listed from Fisher Scientific (catalogue no. AM 1176) or VWR (catalogue no. 66130-502)
- Test tubes, 16 × 100 mm; 50 mL beakers may be substituted
- Filter paper, Whatman #1, with glass (conical) funnels
- Blender with a mason jar for sample blending; balance and volumetric measures accurate enough for the stated weights and volumes
- Boiling water supply; means of cooling extracts to room temperature
Procedure
- Read the information summary sheet supplied with the kit before starting. Kit instructions govern strip handling, reading times, range selection and calculation details; this procedure covers sample preparation and the steps around the strip itself.
- Take a representative portion from the sample unit. Blend that portion in a mason jar for two minutes so the portion taken for extraction represents the whole unit fairly.
- Weigh 10 g of the finely blended sample accurately. Add 90 mL of boiling water and blend for 30 seconds. Let the mixture stand for one minute, then blend again. Cool the extract to room temperature.
- Filter at least 5 mL of the cooled extract through Whatman #1 filter paper folded into a cone, as specified in the kit directions, and supported in a conical funnel. Collect the filtrate in a test tube (16 × 100 mm), or in a 50 mL beaker used in place of a test tube.
- Place a Quantab Chloride Titrator into the filtrate in the tube and carry out the strip test exactly as the kit procedure directs.
- Read the strip within the time the kit specifies, which falls between 2 and 30 minutes depending on the strip. Judge the height of the white column in Quantab units. Dilute samples as needed so readings fall inside the readable range of the strip selected, and record every dilution applied.
- Calculate % sodium chloride by following the directions specified in the kit, applying the dilution used. Where a water-phase result is also needed, determine % moisture for the same food by an acceptable AOAC or equivalent method before calculating, as set out under Recording Results.
Recording Results
Report the primary result as % sodium chloride or as ppm Cl− ions on a wet-weight basis, calculated from the white-column height by the kit directions and corrected for any dilution made to bring the reading into range. When no sodium chloride is detected, report the result as less than the detection limit for the strip used and the amount the sample was diluted. For the water-phase result, use: % salt in the water phase (WPS) or % brine = [% sodium chloride] × 100 / ([% sodium chloride] + [% moisture]). Equivalently, in masses: % salt in the water phase or % brine = [grams sodium chloride] × 100 / ([grams sodium chloride] + [grams moisture]). Report both inputs alongside the derived value: the % sodium chloride read from the Quantab strip and the % moisture value used. Record sample weight, water volume, strip type or range, dilution, reading time and white-column height in Quantab units, so the result can be retraced.
Quality Control
Accuracy of weights and volumes is imperative throughout, because the calculation assumes the stated 10 g in 90 mL extraction. As generic laboratory practice, run a duplicate extraction when sample quantity allows and investigate any wide disagreement before reporting, check each new kit lot against a salt solution of known concentration prepared in the laboratory, and confirm strips are in date and stored as the kit directs. Keep the kit summary sheet with the bench record for the lot in use, since reading windows and conversion details are kit-specific.
Troubleshooting
- Reading off scale or above the readable range: dilute the sample or extract as the kit allows and repeat, so the white column falls inside the range of the strip selected; apply that dilution in the calculation.
- No colour change or no detectable salt: confirm enough filtrate was collected (at least 5 mL), that the strip was fully inserted as directed, and that reading happened inside the specified 2–30 minute window; if sodium chloride is genuinely absent, report a less-than detection-limit result rather than zero.
- Cloudy or slow-filtering extract: re-blend to a finer consistency and refilter through fresh Whatman #1 paper; a turbid filtrate makes the column end-point hard to judge.
- Duplicate results disagree: suspect an unrepresentative portion first; take a fresh portion, blend a full two minutes in the mason jar, and repeat with freshly weighed sample and measured boiling water.
Interferences and Limitations
Strips respond to chloride, and results are expressed as sodium chloride or chloride ion; other chloride sources in a food therefore count toward the result. Coarse or uneven blending biases the extract, because only dissolved chloride reaches the strip. Foods needing a regulation-specified Official Method sit outside this kit route for compliance purposes. A brine or water-phase figure is only as good as the separate moisture determination combined with it, and no water-phase result can be given without one.
Safety Notes
- Handle boiling water and hot blended extracts with heat-resistant gloves and eye protection; secure the mason jar lid before blending hot liquids.
- Treat blender blades as a sharps hazard during washing, and dispose of used titrator strips and filter papers as laboratory waste following local practice.
- As generic laboratory practice, never pipette by mouth and keep food samples for chemical testing separate from samples intended for microbiological examination.
Source Method
Adapted in original wording from Health Canada method MFLP-45, DETERMINATION OF SODIUM CHLORIDE (SALT) IN FOODS AND OTHER SAMPLES BY QUANTABTM CHLORIDE TITRATORS, November 2007, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-45 dated April 1997.