SOP: Measuring Water Activity in Foods with the Novasina aw Center (MFLP-63)

GIFSQ ID: SOP-1058

Water activity decides how much of a food’s water is available to microbes, so it sits at the heart of shelf-life and pathogen control. A Novasina aw Center answers that question without drying or titration: a sample equilibrates inside a temperature-controlled measuring box fitted with three sensors, and each sensor reads equilibrium relative humidity, which converts directly to water activity. Careful calibration against salt humidity standards, checked on control charts before any food is measured, is what makes those readings defensible.

Rigour around the instrument matters as much as the reading itself. Standards are validated, sensors are protected by pre-filters, and a sample result is corrected with the most recent calibration regression line before it is reported.

Purpose

Purpose is to determine water activity (aw) of foods, including for judging compliance with Sections 4 and 7 of the Food and Drug Act. Water activity is the ratio of water vapour pressure in a food system to vapour pressure of pure water at the same temperature. An instrument expresses its measurement as equilibrium relative humidity in % rh, and water activity follows from aw = ERH/100. Equilibration, for this method, means an aw reading that changes by no more than ± 0.001 aw over a 10 minute period. A humidity standard is a saturated solution of a salt with a known % relative humidity at a designated temperature.

Scope

Scope covers foods generally, measured on a Novasina aw Center with three sensors in a temperature-controlled aw-box. Published studies cited by the source method found this instrument satisfactory for measuring water activity in foods. A numerical result can only be reported for samples falling between the upper and lower calibration points chosen for the run; results outside that bracket are reported as greater than the upper point or less than the lower point. Sample units must stay sealed during transport so no moisture is lost or gained before measurement. Work under this method is carried out in accordance with ISO/IEC 17025:2005 (or latest version), General Requirements for the Competence of Testing and Calibration Laboratories, as the laboratory supervisor must ensure.

Equipment and Materials

  • Novasina aw Center, comprising a measuring station and a temperature-controlled aw-box with 3 sensors (Novasina AG, a division of Axair Ltd., Pfäffikon, Switzerland, or JEB Controls Inc., Kirkland, QC)
  • Novasina humidity control standards, including SAL-53, SAL-75, SAL-84, SAL-90 and SAL-98 (same suppliers as above)
  • Disposable aw sample bowls, plastic, with stainless steel measuring bowls and tongs for installing them
  • Pre-filters for the sensors; chemically-neutral, fast-flowing paper filters may be used except where special pre-filters are required
  • Printer for the instrument (optional channel recording), graph paper or software for control charts, plastic bags for sealing filled sample bowls, distilled water and absorbent tissue for standard maintenance
  • Reference salt slushes at 25°C, for selecting calibration salts: magnesium chloride 0.328; potassium carbonate 0.432; magnesium nitrate 0.529; sodium bromide 0.576; cobalt chloride 0.649; sodium chloride 0.753; potassium bromide 0.809; ammonium sulfate 0.81; potassium chloride 0.843; strontium nitrate 0.851; barium chloride 0.902; potassium nitrate 0.936; potassium sulfate 0.973

Procedure

  1. Keep sample units sealed in transport. On receipt, hold units refrigerated or frozen according to the nature of the product, except for shelf-stable foods. Thaw frozen units in a refrigerator, or under time and temperature conditions that prevent moisture loss, and analyse units as soon as possible after receipt.
  2. Prepare the instrument before calibrating: switch the power on and leave it for a minimum of 2 hours; set the temperature dial to 190 (25°C) and confirm temperature is holding; set the RTD-24 switches to “S”; remove the pre-filter from each sensor.
  3. Choose calibration points that bracket the anticipated sample range, because only samples inside that range earn a numerical result. For routine food analysis, use SAL-75 as the lower point and SAL-98 as the upper point. Where more than one humidity standard shares the same water activity, dedicate one standard to one sensor and label it.
  4. Calibrate each sensor either with the built-in dummy values, following the Novasina Operation Manual, or with the D-Pot procedure set out in the steps below. With only one D-Pot cable, expect some connecting and disconnecting to calibrate all three sensors at the same time.
  5. For D-Pot calibration, insert the D-Pot into the junction of each wire between the conditioning chamber and the control apparatus. Install SAL-75 and leave it about 2 hours to reach equilibrium.
  6. Set the D-Pot selector button to “S” or “sensor”, depending on the unit, and the channel selector to the channel being calibrated. At equilibrium, if the digital readout differs from the SAL-75 value of 0.753 at 25°C, adjust by turning the screw of trimmer potentiometer “M”.
  7. Set the D-Pot selector to “D” or “POT” and adjust the thumb wheels until the digital readout shows 0.753. Record that D-Pot value on the calibration record.
  8. Return the D-Pot selector to “S” or “sensor”. Record the time the final readings were taken, the temperature, the channel number and the humidity source on the calibration record.
  9. Replace SAL-75 with SAL-98, noting the time the salt was placed under the sensor, and wait about 2 hours for equilibrium. With the selector on the channel being calibrated, if the readout differs from the SAL-98 value of 0.980 at 25°C, adjust by turning trimmer potentiometer “H”.
  10. Set the D-Pot selector to “D” or “POT” and adjust the thumb wheels until the readout shows 0.980; record that D-Pot value. Then set the thumb wheels back to the value recorded during the “M” adjustment and turn the “M” trimmer screw until the readout shows 0.753; set the thumb wheels to the value recorded during the “H” adjustment and turn the “H” screw until the readout shows 0.980. Repeat these paired adjustments until both 0.753 and 0.980 appear on the readout, since moving either trimmer changes the other value reciprocally and the pair converges with each round.
  11. Adjust the low range only if desired, and only after the medium and high range adjustments are done, using SAL-53.
  12. Check calibration linearity: measure the aw of 4 humidity standards with each sensor, the upper and lower points plus two in between, and calculate a regression line. Values used for the regression salts are SAL-53 0.529, SAL-75 0.753, SAL-84 0.843, SAL-90 0.901 and SAL-98 0.980. Record calibration and regression data on the calibration record form.
  13. Calculate the regression line from the standards’ known values (y) and the readings taken (x): slope b = [Σxy − (Σx Σy)/n] / [Σx² − (Σx)²/n], and intercept a = (Σy − bΣx)/n, giving y = a + bx, where a is the y-intercept, b the slope and r the correlation coefficient. Correct later aw values with y = a + bx, where x is the reading.
  14. Install fresh pre-filters in front of each sensor once calibration is complete.
  15. Before sample readings on any day, prepare the instrument again: power on for a minimum of 2 hours; temperature dial at 190 (25°C); RTD-24 switches to “S”; printer on if used, selecting “1” to record only one channel or “6” to record all three channels; confirm readiness as described under Quality Control; install fresh, appropriate pre-filters; and confirm the aw calibration range includes the anticipated sample readings.
  16. Analyse samples in triplicate. Prewarm refrigerated and frozen samples to 25°C before opening, to prevent condensation on the sample and shorten equilibration time, while taking care to avoid evaporation of water from the sample.
  17. Crush or cut up the sample material to homogenize it. Fill plastic sample bowls to the rim, with material not protruding beyond the rim, and avoid handling the sample material directly. Samples not analysed right away may be held by sealing the filled bowls in plastic bags at an appropriate temperature, prewarming refrigerated or frozen material to 25°C again before measurement.
  18. Load each plastic bowl into a stainless steel measuring bowl: pull up on the appropriate sensor knob and use tongs to install the measuring bowls under the sensors, making sure the fit is proper and the seal is snug. Close the chamber door securely.
  19. When equilibrium has been reached, note the water activity value to three places after the decimal, together with the temperature reading, for the appropriate channel.
  20. After every reading, dry the sensors by removing the bowls and letting the sensors return to ambient humidity. Clean and dry the stainless steel bowls and filter as required, and discard the pre-filter.

Recording Results

Correct each aw value with the latest calibration regression line for its channel (y = a + bx, x being the raw reading). Report a sample result by first taking the mean of the multiple readings, samples are run in triplicate, and then reporting that mean to 2 places after the decimal, rounded as necessary, even though individual equilibrium readings were noted to three decimal places along with the channel temperature. Record the sample aw on the appropriate lab sheet. A numerical value may only be reported when the sample falls inside the calibration bracket; a sample above the upper point is reported as greater than that point and one below the lower point as less than that point. Keep calibration and regression-line data on the calibration record for the run.

Quality Control

Validate humidity standards before trusting them. To establish reproducibility or variation, take 20 daily aw measurements of each humidity standard for each sensor, using the standard dedicated to that sensor, and record them on a Novasina Control Chart; for Novasina-supplied standards, upper and lower limits should not exceed the labelled value ± 0.005, the manufacturer’s stated claim. Judge accuracy of uncertified standard material by an independent route: send blind samples to another CFIA facility for checking on the same type of aw apparatus, test blind samples on a different type of apparatus, or send blind samples to an external calibration laboratory for certification; certified humidity standards are best where available.

Build control charts on graph paper or with software. For the standards used to judge readiness, usually SAL-90 and SAL-98, take the 20 readings above, calculate mean and standard deviation for each sensor and each standard, and set upper and lower warning limits at 1.5 standard deviations and upper and lower control limits at 2 standard deviations.

Standards used to judge readiness must not be the same physical standards used to calibrate the unit. Before sample readings, read the humidity standard nearest the anticipated sample water activity, usually SAL-90 or SAL-98 for most foodstuffs, correct that reading with the regression line or formula from the last valid calibration, and plot it on the control chart for that sensor and standard. Proceed to samples only if the value falls between the upper and lower limits. If a readiness reading falls outside the limits, repeat it 3 times, letting the sensor return to ambient humidity between readings: if all 3 fall within the upper and lower control limits, no adjustment is needed and sample readings may continue; if the 3 readings are still abnormal, recalibrate. Should the problem persist, run a 4 or 5 point calibration check: results in a different range than earlier checks but still linear point to an equipment malfunction, while non-linear results point to one or more humidity standards that need replacing.

Maintain standards so they stay valid. Buy standards with a labelled water activity from a reputable source. For standards above room humidity that show signs of drying out, rejuvenate them with distilled water until water is just visible on the surface, then seal them in their containers with tape and store them in an airtight container. For standards below room humidity that have filled up with water, remove water with absorbent tissue until the normal level returns, and store them sealed in a dry area such as a desiccator. Replace any standard that becomes contaminated or in which all of the salt has dissolved, and replace standards occasionally as routine practice.

Troubleshooting

  • Readiness standard outside its limits: repeat the reading 3 times, allowing the sensor to return to ambient humidity between readings; continue without adjustment if all 3 fall within the control limits, otherwise recalibrate.
  • Problem persists after recalibration: run a 4 or 5 point calibration check. A linear result shifted into a different range indicates an equipment malfunction; a non-linear result indicates one or more humidity standards at fault, which must be replaced.
  • Long equilibration or drifting readings: confirm refrigerated or frozen samples were prewarmed to 25°C before opening, that the chamber door is closed securely, and that the measuring bowl fit and seal are snug; dry sensors back to ambient humidity between readings.
  • Standard drying out or taking up water: rejuvenate a drying high-humidity standard with distilled water until water is just visible on the surface, or blot a flooded low-humidity standard with absorbent tissue back to its normal level; replace the standard if contaminated or fully dissolved.
  • Sensor response degrading over time: fatty, acidic and other foods damage unprotected sensors, so fit a fresh pre-filter for every sample measurement and clean the stainless steel filter frequently.

Interferences and Limitations

Numerical reporting is bounded by the calibration bracket: nothing outside the upper and lower points can be given a number. Moisture exchange before analysis distorts results, which is why units stay sealed in transport, filled bowls are bagged if held, and samples are not handled directly or left open to evaporate. Condensation is a direct error source: cold samples must reach 25°C before opening. Sensor exposure to fatty or acidic foods can cause sensor failure over time, so pre-filters are a measurement requirement, not an accessory. Calibration by the D-Pot route is slow, about 2 hours of equilibration at each of SAL-75 and SAL-98, and low-range adjustment with SAL-53 is optional, done only after the medium and high ranges.

Safety Notes

  • Never touch the button sensor when changing pre-filters: lift the gold-coloured retaining ring while lifting the sensor with its knob, grasp the sensor and pull down firmly to disengage it, turn it upward, remove the C-clip, then turn it downward to drop the metal screen out before removing or replacing the filter; afterwards replace the screen and C-clip, re-engage the sensor and replace the retaining ring.
  • Do not submerge sensors in water or any other liquid, spray them with liquid or blow on them, and do not leave them in a draft.
  • Do not subject sensors to temperature changes that allow condensation to form, do not cool or heat them beyond specified limits, and do not place them in a humid environment; store them in a sealed container for long-term storage.
  • Protect sensors from strong substances such as acid or alkaline vapours, from dust and particles, and from physical shocks; avoid direct contact between sample and sensor, use the proper filters, and let sensors return to ambient humidity after each use.
  • As generic laboratory practice, wear gloves when handling samples and standards, and keep refrigerated or frozen foods under temperature control until preparation so sample condition is not compromised.

Source Method

Adapted in original wording from Health Canada method MFLP-63, Measuring Water Activity (aw) Using the Novasina aw Center, January 2006, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-63 dated April 1999.