GIFSQ Tools · Module 2 — Food Chemistry & Physicochemical Testing · Tool 21 of 500
Log a pH reading with its temperature and calibration buffers. The tool checks the reading against the calibration range and checks the electrode slope against the theoretical Nernst slope.
Try sample values — fills the form with the worked example below.
Reading log
Saved on this device only. Use "Add current reading to log" after a calculation.
| Date / time | pH | Temp (°C) | Slope % | Range | Slope OK |
|---|
Method: AOAC 981.12 (pH of acidified foods) and AOAC 943.02 — glass electrode, two-point calibration with standard buffers that bracket the sample, reading taken at the stated temperature.
Assumptions: the electrode follows the Nernst response; the buffers are fresh and at their stated pH; the sample and the buffers are at the stated temperature.
Limitations: a passing slope does not prove accuracy — a drifting electrode can pass the slope check and still read wrong, so check the reading against a known buffer as well. Junction error, sodium error at high pH, and protein or fat fouling are not detected by this check.
Rounding: pH to 2 decimal places; slope to 2 decimal places. The accepted slope range is 95–105 % of theoretical.
This is a calculation aid, not a laboratory report — confirm against your accredited method before citing a value.
Enter values above to see results.
Typical food pH: milk 6.5–6.7, yogurt 4.0–4.5, orange juice 3.3–4.2, pickles 3.0–3.5, meat 5.4–6.2. Calibrate daily, and re-check with a buffer after every 10 readings. Tool 21 of 500 — GIFSQ