SOP: VIDAS Immunoassay Screening for Salmonella in Foods (MFLP-40)
GIFSQ ID: SOP-1041
Antigen capture replaces plating at the screening stage here. Two enrichments build up Salmonella first, a heated aliquot of the second broth goes into an automated enzyme immunoassay, and fluorescence decides whether the sample screens presumptive positive. Hands-on time is short once the instrument takes over, because every wash, binding and reading step inside the assay runs automatically.
Only the screen is automated. A positive fluorescence value never stands alone: unheated selective broth is held back precisely so cultural isolation can follow, and calibration discipline decides whether any run can be trusted at all.
Purpose
Purpose is rapid detection of Salmonella spp. in food products for compliance with Sections 4 and 7 of the Food and Drugs Act. Positive results must be confirmed with a cultural method.
Scope
Validated scope covers dairy products, chocolate and bakery products, and fish and seafood as food categories, plus heat processed eggs and derivatives, raw processed meat, raw unprocessed meat, and processed cooked multi-ingredient composite foods as food types. Other foods sit outside the approved application; use on another commodity requires proper validation under Compendium criteria first, with data forwarded to the Microbiological Methods Committee for consideration. Sample units may be analysed individually, or analytical units combined as composites where the procedure allows. Work follows ISO/IEC 17025, and any variation in listed media needs equivalency shown by the laboratory.
Equipment and Materials
- SX2 broth (bioMérieux ref. 42121, 20 tubes), mandatory supply from bioMérieux Canada
- VIDAS family instrument, mini-VIDAS or VIDAS 30; VIDAS Salmonella (SLM) kit (ref. 30702) holding reagents, standards and controls for 60 tests, including one solid phase receptacle (SPR) and one reagent strip per test
- VIDAS Heat and Go device, or a water bath capable of 95–100°C, for heating the assay aliquot
- Buffered peptone water (BPW)
- Stomacher with filter bags, blender or equivalent; pipette calibrated to dispense 500 µL with disposable tips
- Incubators capable of maintaining 35°C and 42°C
Procedure
- Hold sample units refrigerated or frozen according to product nature, except for shelf-stable foods. Thaw frozen units in a refrigerator or under time and temperature conditions preventing microbial growth or death, and analyse units as soon as possible after receipt.
- Have sterile BPW ready, pre-warmed to 35 ± 1°C, and clean the working surface with a suitable disinfectant before starting.
- Make the analytical unit representative: agitate liquids and free-flowing materials until homogeneous, and for solids take portions from several locations within the unit.
- Run non-selective enrichment: aseptically place 25 g or 25 mL of sample in a stomacher bag and dilute 1:10 with BPW, meaning 225 mL BPW for a 25 g unit. Composites may combine units up to 125 g or mL, for example 125 g or mL in 1125 mL BPW, and any alternate unit keeps the ratio of 1 part sample to 9 parts BPW. Homogenise with a stomacher and incubate at 35 ± 1°C for 16–22 hours.
- Run selective enrichment: let SX2 broth reach room temperature before use, mix the incubated BPW culture, and transfer 0.1 mL of suspension into 10 mL SX2 broth. Incubate at 42 ± 1°C for 22–26 hours.
- Use the optional holding branch when scheduling requires it: incubated SX2 broth may be refrigerated, provided total refrigeration across this hold and any later hold before confirmation never exceeds 72 hours in total.
- Set up the assay: take the VIDAS Salmonella (SLM) kit from the refrigerator, remove only the reagents needed, and let them reach room temperature for at least 30 minutes. Allow one SLM strip and one SLM SPR per sample, control or standard, and reseal the storage pouch carefully after removing SPRs. Never mix components from different lots and never use components past expiry. When a new kit lot is opened, enter the factory master calibration curve data from the master lot entry card into the system before testing, calibrate with the standard supplied in the kit at that point, and repeat calibration every 14 days thereafter, following the package insert for entry and calibration detail.
- Homogenise the incubated SX2 broth. Heat the assay portion by one of two routes, never heating standards or controls. Water bath route: transfer an aliquot of SX2 broth to a tube, seal it, heat for 15 ± 1 minutes at 95–100°C, cool the tube, mix the boiled broth, and transfer 0.5 mL into the sample well of the VIDAS strip, labelling the strip with the sample identification. Heat and Go route: transfer 0.5 mL SX2 broth directly into the strip sample well, label the strip, heat for 15 ± 1 minutes following the device manual, remove the strip, and let it cool for 10 minutes.
- Start the instrument run as directed in the VIDAS operator’s manual: load each SPR and strip into the positions matching the section shown on the work list, and check that the colour labels carrying the assay code on each SPR and its strip match before starting. Inside the instrument, sample cycles in and out of the antibody-coated SPR so Salmonella antigens bind, washes remove unbound material, alkaline phosphatase conjugate binds to captured antigen and is washed again, and substrate 4-methyl-umbelliferyl-phosphate is hydrolysed to fluorescent 4-methyl-umbelliferone, read at 450 nm.
- Confirm every presumptive positive culturally: from the non-boiled SX2 broth, or from refrigerated non-boiled SX2 broth held within the 72-hour total limit, proceed with the plating and confirmation steps of cultural method MFHPB-20.
Recording Results
Instrument software analyses results automatically once the assay finishes, comparing each sample’s relative fluorescence value (RFV) against internal thresholds and reporting positive or negative. An RFV of 0.23 indicates the presumptive presence of Salmonella in the test sample. Record the RFV outcome, kit lot, calibration status (new-lot calibration plus 14-day recalibration), enrichment times and temperatures, heating route used, and the cultural confirmation outcome for every presumptive positive. A screening negative is reported as the screen result for that unit; no screening positive is reported as final until MFHPB-20 confirmation is complete.
Quality Control
Kit standards and controls run as part of each assay series, and calibration state is itself a quality gate: an unentered master lot or an overdue 14-day calibration invalidates results regardless of how clean they look. Temperature tolerance is ± 1.0°C at 35°C and at lower temperatures such as 30 or 25°C, tightening to within 0.5°C at higher temperatures such as 43 or 45.5°C because excess heat can kill the organisms being isolated; the 42°C SX2 incubation and the 95–100°C heating step are therefore checked, not assumed. Keep lot integrity strict: components from different lots are never combined.
Troubleshooting
- Assay code mismatch flagged at loading: stop and match each SPR to a strip carrying the same colour label and assay code before starting.
- Borderline fluorescence near the decision point: verify calibration is current for the lot in use and that reagents stood at least 30 minutes at room temperature; confirm presumptive positives culturally rather than re-screening repeatedly.
- No unheated broth left for confirmation: the screen cannot be closed out; prevent recurrence by always reserving non-boiled SX2 before the heating step.
- Refrigerated SX2 held too long: count total refrigeration across both holds; beyond 72 hours combined, the broth no longer supports a valid confirmation.
- Mixed-lot reagents suspected: discard the run setup and rebuild it from a single lot, including its own standards and controls.
Interferences and Limitations
Detection rests on antigen binding after enrichment, so very low initial numbers, uneven homogenisation, or a shortened enrichment can leave antigen below the fluorescence threshold. Approval covers only the categories and types listed under Scope; performance in other matrices is unproven until validated. Heating destroys the assay aliquot for culture, which is why confirmation depends entirely on the separate non-boiled broth, and any break in the 72-hour refrigeration accounting weakens that confirmation.
Safety Notes
- Treat BPW and SX2 enrichments as potentially carrying Salmonella; use standard enteric precautions and biohazard disposal.
- Handle sealed tubes in a 95–100°C water bath and hot strips from the Heat and Go device with thermal protection, and let heated material cool for the stated time before transfer.
- Disinfect the working surface before analysis as the method directs, and keep disinfectant away from open enrichment bags as general laboratory practice.
Source Method
Adapted in original wording from Health Canada method MFLP-40, Detection of Salmonella in food products by the VIDAS Easy Salmonella (SLM) method, November 2011, Compendium of Analytical Methods, Volume 3.