SOP: Screening Environmental Surfaces for Listeria Species with the BAX System PCR Assay (MFLP-15)

GIFSQ ID: SOP-1019

Environmental swabs answer a sanitation question, and speed matters before a line restarts. BAX screening enriches a sponge or swab for two days, lyses a small portion, then amplifies a DNA fragment found only in Listeria species, reading fluorescence inside a closed tube so amplified product never opens to the room.

Two instruments share one workflow. Q7 and X5 kits use identical reagents apart from tube optics, because X5 reads from the side through transparent tubes while Q7 reads from above through opaque ones, plus an extra lysing agent in the X5 kit that improves lysis.

Purpose

Purpose is detection of Listeria spp. on environmental surfaces, supporting compliance work under Sections 4 and 7 of the Food and Drugs Act and other relevant federal regulations.

Scope

Scope is environmental surfaces only, validated across a variety of surface types, run on either a BAX Q7 or a BAX X5 instrument. Every positive needs confirmation by a cultural method before it counts as confirmed. Collection devices are moistened with D/E Neutralizing broth, Buffered Peptone Water, or Hi-Cap Neutralizing Broth, and samples are collected and transported following MFLP-41 practice: in practical terms, keep units refrigerated on arrival and analyse them as soon as possible. Any variation in a listed medium formulation, bought or made in-house, must be shown equivalent by the laboratory using it.

Equipment and Materials

  • BAX System Q7 Start-Up Package (ASY2018): cycler/detector with verification plates; computer workstation with Windows operating system, BAX amplification software and printer; Automated Thermal Block (DMCH2023) or dry block heaters with thermometer inserts; capping/decapping tools; pipettes (5 µL, 10 µL, 10–50 µL 8-channel, repeat pipettor); cooling block with inserts for lysis and PCR tubes; PCR tube holders; lysis tubes with caps and rack; barrier or filter tips; powder-free nitrile gloves; Q7 User Guide
  • BAX System X5 Start-Up Package (MCH2007): cycler/detector; computer workstation with Windows operating system, BAX amplification software and printer; Automated Thermal Block (DMCH2023); capping/decapping tools; same pipette set as Q7; cooling block with inserts; PCR tube holders; lysis tubes with caps and rack; barrier or filter tips; powder-free nitrile gloves; X5 User Guide
  • BAX System PCR Assay Genus Listeria for Q7 (KIT2016, 96 tests): PCR tubes with tablets, flat optical caps, 2 lysis buffer bottles at 12 mL each, protease solution 400 µL per vial for working lysis reagent
  • BAX System X5 PCR Assay for Genus Listeria (KIT2024, 64 tests): PCR tubes with tablets, optical caps, 2 lysis buffer bottles at 12 mL each with Lysis Agent 2 at 1.1 mL per bottle, protease solution 400 µL per vial
  • BAX System Enrichment Media for Listeria (MED2002), prepared by manufacturer directions
  • Wetting agents: D/E Neutralizing broth or equivalent, Hi-Cap Neutralizing Broth (World Bioproducts BLU-10HC), or Buffered Peptone Water
  • Incubator holding 35 ± 1°C; sterile stomacher bags

Procedure

  1. Pre-warm prepared BAX System Listeria enrichment broth to 35 ± 1°C before testing. Place each sponge or swab in a sterile stomacher bag.
  2. Add 190 mL pre-warmed enrichment broth per sponge, ensuring complete coverage; composites may combine up to 10 sponges at 190 mL per sponge. Homogenize sponges by hand for 1–2 min. Place smaller swabs in 10 mL enrichment broth each, compositing up to 10 swabs at 10 mL per swab, and rotate swabs by hand for 1 min.
  3. Incubate enrichments for 48 ± 2 h at 35 ± 1°C. Hands-on preparation afterwards takes about 1.5 hours, with automated results about 4 hours later.
  4. Ready the equipment while enrichments finish: refrigerate cooling blocks overnight, warm heating blocks and confirm settings of 55°C and 95°C, initialize the cycler/detector and run verification if prompted, create a rack file, then select RUN FULL PROCESS from the menu bar so the cycler/detector warms up before samples are loaded.
  5. Arrange lysis tubes in the rack to match the rack file, preparing blank, positive and negative controls as the laboratory’s own requirements specify. Make working lysis reagent for Q7 by pipetting 150 µL protease into one 12 mL bottle of lysis buffer; for X5, pipette 150 µL protease and 200 µL Lysing Agent 2 into one 12 mL bottle of lysis buffer.
  6. Add 200 µL lysis reagent to each lysis tube, then transfer 5 µL enriched sample into its corresponding tube and secure caps. Heat manually at 55°C for 60 minutes, then at 95°C for 10 minutes; an Automated Thermal Block instead runs its preset parameters, which include the cooling step that follows. Move tubes into a cooling block for at least 5 minutes.
  7. Set the PCR tube holder onto a PCR cooling block and arrange one PCR tube per sample. Remove and discard the lid from one strip at a time, transfer 50 µL of each lysed sample into its corresponding PCR tube with a multi-channel pipette, then cover tubes with a fresh optical cap strip secured tightly. Repeat strip by strip for all samples.
  8. Carry the whole cooling block to the cycler/detector. Hydrated tablets wait in the block only until the instrument is ready, and never longer than 30 minutes after hydration. Follow PCR Wizard screen prompts to load samples, run the program and unload, exactly as the Q7 or X5 User Guide specifies; automated fluorescent detection then calls each sample positive or negative.
  9. Read the run outcome for every sample and control before reporting anything, and hold enrichment broth for any sample needing repeat testing or confirmation.

Recording Results

Report a valid negative as Listeria spp. Not Detected. Record an indeterminate outcome and repeat analysis starting from the transfer of enriched sample into lysis tubes (5 µL into prepared lysis reagent). Treat any sample that comes back indeterminate or in error more than once as presumptive positive, and send it to confirmation rather than reporting it negative. Confirm presumptive positives from the enrichment broth without delay, plating and confirming by a cultural method such as MFHPB-07 or MFHPB-30; isolating L. monocytogenes or other Listeria spp. from that broth requires at least one chromogenic agar listed in MFHPB-30 alongside Oxford agar.

Quality Control

Blank, positive and negative controls form part of each run, set up under the laboratory’s own control requirements. Cycler/detector verification runs whenever the system prompts during initialization. Temperature discipline matters throughout: growth steps at or below 37°C allow ± 1.0°C tolerance, while higher-temperature steps must hold within 0.5°C because excess heat can kill the organisms being recovered.

Troubleshooting

  • Indeterminate result on first run: repeat from the lysis transfer step using fresh enriched sample; a second indeterminate or error flips the sample to presumptive positive.
  • Hydrated PCR tubes waiting too long: discard any tube held beyond 30 minutes after tablet hydration and rehydrate a fresh tube from retained lysate or enrichment.
  • Weak or failed lysis suspected on X5: confirm Lysing Agent 2 was added at 200 µL per bottle alongside protease; Q7 reagent omits it by design, X5 reagent does not.
  • Composites giving inconsistent coverage: check broth volume scales per unit, 190 mL per sponge and 10 mL per swab, so every device stays covered during enrichment.

Interferences and Limitations

Screening stops at genus level and at presumptive status: fluorescence shows Listeria DNA was amplified, not that viable L. monocytogenes was isolated, so cultural confirmation decides final reporting. Closed-tube detection limits carry-over contamination from amplified DNA, yet poor enrichment, uncovered sponges or delayed analysis still suppress recovery. Surfaces or collection devices outside the validated set carry unproven performance.

Safety Notes

  • Wear powder-free nitrile gloves and treat enrichments as potentially carrying Listeria; dispose of tubes and bags as biohazardous waste, as generic laboratory practice.
  • Respect hot blocks at 55°C and 95°C during lysis, and keep caps secured so tubes cannot pop open while heated.
  • Handle neutralizing broths and enrichment media by their own safety data sheets.

Source Method

Adapted in original wording from Health Canada method MFLP-15, Detection of Listeria Species from Environmental Surfaces Using the BAX® System Genus Listeria Assay, April 2020, Compendium of Analytical Methods, Volume 3.