SOP: Screening Select Foods for E. coli O157:H7 with the BAX System MP Assay (MFLP-30)
GIFSQ ID: SOP-1032
Two enrichment media divide this screen, and picking the wrong one invalidates the run for that food. Raw ground meat, meat trim and leafy greens grow fastest in proprietary BAX MP medium, with ground meat ready in as little as 8 h, while most other validated foods use modified tryptic soy broth carrying novobiocin. Lysis volume follows the medium, not the food, doubling as a built-in check that the pairing was right.
One exclusion deserves attention before any sample is logged: seasoned raw processed meat and leafy greens must not be enriched in mTSB-n for this assay, because that pairing failed the method performance criteria behind the source.
Purpose
Purpose is presence/absence screening for Escherichia coli O157:H7 in validated food categories on BAX System Q7 and X5 instruments, supporting compliance with Sections 4 and 7 of the Food and Drugs Act and other relevant federal regulations, with positives confirmed by cultural method MFHPB-10.
Scope
Validated scope is dairy products, raw meat products, ready-to-eat meat and poultry products, fruit- and vegetable-based products, and dry cereal and dry animal feed products in the miscellaneous category. Within that scope, BAX MP medium covers raw ground meat and raw meat trim in 8–24 h and leafy greens in 18–24 h, while mTSB-n covers dairy, raw meat including ground meat and trim (excluding seasoned raw ground meat), ready-to-eat meat and poultry, and fruit- and vegetable products excluding leafy greens, in 18–24 h. Other commodities may be attempted only after proper validation against Compendium criteria, with data forwarded to the Microbiological Methods Committee so the application can be expanded.
Equipment and Materials
- BAX System Q7 Start-Up Package (Cat. No. ASY2018) or X5 Start-Up Package (Cat. No. MCH2007): cycler/detector (verification plates on Q7), workstation with Windows, BAX software and printer, Automated Thermal Block (Cat. No. DMCH2023) or dry block heaters with thermometer inserts, capping/decapping tools, pipettes (5 µL, 10 µL, 10–50 µL 8-channel, repeat pipettor), cooling blocks with inserts, PCR tube holders, lysis tubes with caps and rack, barrier or filter tips, powder-free nitrile gloves, and the matching User Guide
- BAX System PCR Assay for E. coli O157:H7 MP kit (KIT2004, 96 tests) or X5 PCR Assay for E. coli O157:H7 kit (KIT2022, 64 tests): PCR tubes with tablets, flat optical caps, 2 lysis buffer bottles at 12 mL, protease solution 400 µL vial. X5 tubes are optically transparent for side reading; Q7 tubes are opaque for top reading, with all other reagents the same
- Enrichment broths: modified tryptic soy broth with 20 mg/L novobiocin (mTSB-n) prepared as directed in the Procedure; BAX System MP media (Cat. No. MED2016 Stat-Media soluble packets, or MED2003 bulk powder 2.5 kg) prepared to container-label instructions; buffered peptone water (BPW) for the dry-foods pre-treatment
- Stomacher, blender or equivalent; incubator holding 42°C
Procedure
- Hold units refrigerated or frozen by product nature before analysis, except shelf-stable foods, thawing frozen units in a refrigerator or under conditions preventing growth or death. Analyse as soon as possible after receipt. Build representative units by mixing liquids and free-flowing materials until homogeneous and, for solids, sampling several locations.
- Prepare media ahead. For mTSB-n basal medium, dissolve 30.0 g tryptic soy broth or tryptone soya broth, 1.5 g bile salts No. 3 and 1.5 g dipotassium phosphate (K2HPO4) in 1.0 L distilled water to pH 7.4 ± 0.2 at ambient temperature; heat to boiling, dispense, autoclave 15 min at 121°C, cool and store at 4°C. For the novobiocin supplement, dissolve 100 mg novobiocin sodium salt in 1.0 mL distilled water, filter-sterilise through a 0.2 µm filter, and store for several months in a dark bottle at 4°C; add 0.2 mL of that solution per litre before use, giving the 20 mg/L working strength. For BPW, dissolve 10.0 g peptone, 5.0 g sodium chloride, 3.5 g disodium phosphate and 1.5 g potassium dihydrogen phosphate in 1.0 L distilled water to a final pH of 7.2 ± 0.2, dispense and autoclave at 121°C for 15 min. Prepare BAX MP media to its container label.
- Have broths pre-warmed before samples arrive: BAX MP media and mTSB-n to 40 ± 2°C, and BPW, where needed, to 35 ± 2°C. Make a 1 in 10 dilution as the standard unit, 25 g or mL into 225 mL of the specified enrichment broth; other units and composites keep 1 part sample to 9 parts broth. Stomach or blend as thorough mixing requires.
- Enrich raw ground meat and raw meat trim in BAX MP media with no pre-treatment, at 42°C for 8–24 h. Lysis volume for this group is 20 µL.
- Enrich leafy greens in BAX MP media with no pre-treatment, at 42°C for 18–24 h. Lysis volume is 20 µL.
- Enrich raw meat products other than seasoned raw ground meat, together with ready-to-eat meat and poultry products, in mTSB-n with no pre-treatment, at 42°C for 18–24 h. Continue incubation so total time reaches 22–24 h whenever confirmation may be required. Lysis volume is 5 µL.
- Enrich dairy products in mTSB-n with no pre-treatment, at 42°C for 18–24 h. Lysis volume is 5 µL.
- Enrich fruit- and vegetable-based products excluding leafy greens in mTSB-n with no pre-treatment, at 42°C for 18–24 h. Lysis volume is 5 µL.
- Enrich dry cereal and dry animal feed (miscellaneous foods) in mTSB-n after a pre-treatment: moisten the sample with pre-warmed BPW without exceeding a 1:2 BPW to sample ratio, equilibrate 1 h at 35°C, then add pre-warmed mTSB-n and incubate at 42°C for 18–24 h. Lysis volume is 5 µL.
- Hold enrichments correctly around the assay. Remove aliquots for BAX analysis once incubation finishes, but keep samples incubating while the test proceeds so the full 22–24 h is available if MFHPB-10 confirmation becomes necessary. Do not shake enriched samples before removing lysis aliquots; when a sample has been agitated after enrichment, let it sit at least 10 min before transferring aliquots to lysis tubes.
- Ready the workstation: PCR cooling blocks stored at −20°C until just before use and all other cooling blocks stored refrigerated until just before use; heating blocks warmed and verified at 37°C and 95°C; cycler/detector initialised with verification when prompted; rack file created; RUN FULL PROCESS selected to warm the instrument.
- Lyse samples. Arrange lysis tubes, one per sample plus one blank, to match the rack file. Make lysis reagent with 150 µL protease in one 12 mL bottle of lysis buffer. Add 200 µL lysis reagent to each tube, then transfer the volume set by the enrichment medium: 20 µL for BAX MP enrichments, 5 µL for mTSB-n enrichments.
- Secure caps. On a manual block, heat at 37°C for 20 min, then at 95°C for 10 min; on the Automated Thermal Block, use preset parameters including cooling. Place tubes in a cooling block for at least 5 min.
- Hydrate PCR tablets. Take the PCR cooling block from the freezer immediately before this step and set the PCR tube holder into that frozen block. Arrange one PCR tube per sample, discard lids one strip at a time, transfer 50 µL of each lysed sample into its PCR tube with a multi-channel pipettor, and cover with fresh optical cap strips secured tightly. Keep tubes in the cooling block until loading, for no more than 30 min after tablet hydration.
- Amplify and detect through PCR Wizard prompts as the BAX System User Guide specifies.
- Confirm presumptive positives from the presumptive enrichment broth by the immunomagnetic separation, plating and confirmation steps of MFHPB-10.
Recording Results
Record food category, enrichment medium, incubation duration actually completed, lysis volume used and the BAX outcome for each sample. Indeterminate results repeat starting from the lysis transfer step with a fresh aliquot at the correct medium-specific volume. Negatives report as E. coli O157:H7 Not Detected. Repeated indeterminate results (more than one) or error results count as presumptive positive and go to MFHPB-10 confirmation like any positive.
Quality Control
Blank tubes accompany every rack, and controls follow laboratory requirements for the assay. Growth temperature tolerance at 42°C is tight: incubation runs at 42 ± 0.5°C, while the general rule allows ± 1.0°C only at or below 37°C. Media variations, commercial or made from scratch, require laboratory equivalency, with data forwarded to the Compendium editor as the source requests. PCR cooling blocks held at −20°C until use, rather than merely refrigerated, form part of run validity for this assay.
Troubleshooting
- Single indeterminate result: repeat from the lysis transfer, checking that 20 µL was used for BAX MP samples and 5 µL for mTSB-n samples.
- Ground meat reported before 8 h: the enrichment window had not opened; results before 8 h in BAX MP media fall outside the validated range.
- Dry feed or cereal under-recovered: verify the BPW moistening stayed within the 1:2 ratio and the 1 h equilibration at 35°C ran before mTSB-n was added.
- Confirmation broth too young: confirm total incubation reached 22–24 h, using the continued incubation that runs while the BAX test proceeds.
- Seasoned meat or leafy greens logged into mTSB-n: stop and reroute; that pairing is expressly excluded, with leafy greens belonging in BAX MP media.
Interferences and Limitations
Approved scope is deliberately narrow, and commodities outside it need validation before reporting. The mTSB-n exclusion for raw processed seasoned (with spice) meat and for leafy greens reflects failed performance criteria, not a paperwork preference. Shaking enrichments before aliquoting disturbs the sample and forces a 10 min settling pause. Tablet hydration allows at most 30 min before loading, and the serotype-specific DNA target means a positive remains presumptive until MFHPB-10 culture, including immunomagnetic separation, confirms it.
Safety Notes
- Treat enrichments as potentially carrying a Shiga toxin-producing pathogen with a low infectious dose; use strict enteric precautions, sealed waste handling and biohazard disposal.
- Handle novobiocin supplement preparation and filtration with gloves and eye protection, as generic laboratory practice.
- Use thermal protection for 95°C blocks and 121°C autoclaving, and keep frozen cooling blocks handled briefly to avoid cold burns.
Source Method
Adapted in original wording from Health Canada method MFLP-30, Detection of Escherichia coli O157:H7 in Select Foods Using the BAX System PCR Assay for E. coli O157:H7 MP, January 2025, Compendium of Analytical Methods, Volume 3. This revision replaces the November 2021 edition named in the source.