SOP: Rapid VIDAS UP Screening for Listeria spp. in Foods and Surface Samples (MFLP-59)

GIFSQ ID: SOP-1056

Phage protein does the catching here. VIDAS® UP Listeria, sold as the LPT assay, uses recombinant phage proteins coated inside a Solid Phase Receptacle (SPR®) to bind Listeria receptors from an enrichment broth, then an enzyme-linked fluorescent assay reads the result automatically. Foods go through two successive enrichments in LPT broth; environmental swabs and sponges need only one. A heated aliquot feeds the instrument, while the unheated broth stays refrigerated so any presumptive positive can still be confirmed culturally.

Foods are tested as 125 g units only, and a genus-level result carries a specific reporting limit: a negative finding also covers L. monocytogenes, but a positive genus finding can never be used to claim that species is absent.

Purpose

Purpose is rapid qualitative detection of Listeria spp. in foods and on environmental surfaces, supporting compliance with Sections 4 and 7 of the Food and Drugs Act and/or other relevant federal regulations. Presumptive positives move on to cultural confirmation, and every Listeria species identified along the way gets reported.

Scope

Validation covers raw and raw processed fruit and vegetable–based products, processed raw meat products, heat processed ready-to-eat meat and poultry, heat processed fish and seafood, spices within the Miscellaneous food category, and a variety of environmental surfaces. Other commodities may well work, but fitness for purpose must first be shown by validation following Compendium of Analytical Methods criteria, with the data forwarded to the Microbiological Methods Committee so the approved scope can be widened. Work is carried out under ISO/IEC 17025 (latest version), and a laboratory choosing a variation on a listed medium formulation carries responsibility for demonstrating equivalency, forwarding that data to the Editor of the Compendium of Analytical Methods. Only 125 g food samples are run, under the enrichment and incubation conditions given below.

Equipment and Materials

  • VIDAS® family instrument: miniVIDAS® or VIDAS®
  • VIDAS® UP Listeria (LPT) kit, ref. 30126, holding reagents plus the standards and controls needed for 60 test samples
  • VIDAS® Heat and Go, ref. 93554, or a water bath at 95–100°C, or an equivalent heating system
  • LPT broth, sterile
  • Stomacher and stomacher bags, blender or equivalent; pipette dispensing 500 µL with disposable tips
  • Incubators capable of holding 30°C
  • Environmental collection devices: cellulose sponges or swabs, moistened with 10 mL Neutralizing buffer or D/E neutralizing broth

Procedure

  1. Keep sample units refrigerated or frozen before analysis according to the nature of the product, shelf-stable foods excepted. Thaw frozen units in a refrigerator, or under time and temperature conditions that prevent microbial growth or death, and analyse units as soon as possible after receipt. Incubator and water bath temperatures must actually be held as recommended: where the method states a growth temperature of ≤ 37°C, a tolerance of ± 1.0°C applies, whereas higher recommended temperatures must be held within 0.5°C because of the potential lethality of excess heat for the organism being isolated.
  2. Have sterile LPT broth ready and equilibrated to room temperature (18–25°C) before use. Clean the working area surface with a suitable disinfectant.
  3. Build each food analytical unit by taking portions from several locations within the sample unit, so the unit is representative and at least 125 g. Compositing is allowed: up to 5 analytical units of 25 g each may be combined to reach the 125 g sample size.
  4. Prepare the food dilution at 1 in 4 (1:3). Aseptically place 125 g of food in a sterile container and add 375 mL of LPT broth. Stomach or blend according to the type of food, and stomach, vortex or homogenise as required for thorough mixing.
  5. Incubate the food enrichment for 24–30 h at 30°C ± 1°C. Afterwards, mix the contents manually, then transfer 1 mL of the suspension into 10 mL of LPT broth and incubate this second enrichment for 22–26 h at 30°C ± 1°C.
  6. Handle environmental samples in a single enrichment. Combine 1 to 10 swabs with 10 mL of LPT broth per swab, or 1 to 10 sponges with 100 mL of LPT broth per sponge. Stomach, vortex or homogenise as required for thorough mixing, then incubate for 22–30 h at 30°C ± 1°C.
  7. Take only the required reagents out of the refrigerator from the VIDAS® UP Listeria kit and let them equilibrate to room temperature for at least 30 min. Use one strip and one SPR® for each sample, control or standard to be tested, and reseal the storage pouch carefully once the needed SPR®s are out. Components from different lots cannot be mixed and cannot be used past the expiry date. For a new kit lot, enter the factory master calibration curve data first by scanning the MLE (Master Lot Entry) label from the kit, and calibrate with the standard provided in the kit for each new lot of reagent upon opening, and every 28 days thereafter, following the package insert for PTC protocol data entry, master lot data entry and calibration.
  8. Mix the incubated enrichment broth manually after incubation and let the sample contents settle before transfer to the VIDAS® strip. Heat the aliquot destined for the assay by one of the two routes in the next steps. Never heat the standard and controls supplied with the kit, and for poultry samples always use a water bath rather than the VIDAS® Heat and Go.
  9. Heat and Go route: transfer 0.5 mL of enrichment broth into the sample well on the strip, heat for 5 ± 1 min following the VIDAS® Heat and Go Operator’s Manual, then remove the strip and leave it to cool for 10 min.
  10. Water bath route: transfer 2–3 mL of enrichment broth into a tube, seal the tube, and heat for 5 ± 1 min at 95–100°C. Cool the tube, mix the boiled broth on a vortex type mixer, and transfer 0.5 mL into the sample well on the VIDAS® strip. Keep the remaining non-boiled enrichment broth refrigerated, for up to 72 h, for confirming presumptive positive results, and mix it thoroughly after refrigeration before confirmation begins.
  11. Insert the SPR® and strip into the instrument, checking that the colour label carrying the assay code on the SPR® matches the reagent strip. Start the assay as directed in the Operator’s Manual. Inside the instrument, the reaction medium cycles in and out of the SPR®; Listeria receptors bind the phage proteins coating its interior, washing removes unbound material, alkaline phosphatase–conjugated protein binds any captured receptors, and a final wash clears unbound conjugate. Substrate (4-methyl-umbelliferyl phosphate) is cycled in last, and the conjugate enzyme hydrolyses it to a fluorescent product (4-methyl-umbelliferone) measured at 450 nm.
  12. Remove the SPR® and strip once the run finishes. Dispose of used SPR®s and strips in an appropriate biohazard receptacle, in accordance with applicable local regulations.
  13. Confirm presumptive positives culturally. Using the non-boiled enrichment broth held back in the refrigerator, carry the sample through the plating and confirmation steps of a cultural method (i.e. MFHPB-30), as appropriate, and report all species of Listeria identified.

Recording Results

Results are calculated automatically at the end of the assay: the instrument derives a test value for each sample, compares it against internal references (thresholds) and interprets each result as positive or negative. A test value of ≥ 0.05 indicates the presumptive presence of Listeria spp. in the test sample, and presumptive positives require cultural confirmation before reporting. Record the reporting logic exactly: where Listeria spp. have not been detected, L. monocytogenes is likewise understood as not detected, because no species of the genus was found; where Listeria spp. are detected, no amount of colony screening or additional confirmation can support a statement that L. monocytogenes is absent. Every identified Listeria species is named in the report.

Quality Control

Standards and controls supplied in the kit run as part of the assay, one strip and one SPR® each, and are never heated before loading. Calibration with the kit standard is performed for each new reagent lot upon opening and every 28 days thereafter, after the MLE label for that lot has been scanned into the system. Kit components from different lots are never mixed, and nothing is used past its expiry date. Incubator and water bath temperatures are verified against the tolerances in step 1, and media variations are used only where the laboratory has demonstrated equivalency.

Troubleshooting

  • Weak or failed calibration on a new lot: confirm the MLE label was scanned before testing and that calibration used the standard provided in that kit; recalibrate rather than reporting results against the previous lot’s curve.
  • Poultry sample heated in the Heat and Go: discard that preparation and repeat through a water bath, since poultry samples must not use the Heat and Go device.
  • Heated standard or control: replace it with an unheated one from the kit; kit standards and controls are never heated.
  • Presumptive positive with no broth left to confirm: prevent this by always holding back the non-boiled enrichment broth under refrigeration (up to 72 h) before heating any aliquot.
  • Reagents loaded straight from the refrigerator: allow at least 30 min at room temperature first; cold reagents compromise the automated assay.
  • SPR® and strip codes disagree: check the colour label assay code on both before starting, and never combine components from different kit lots.

Interferences and Limitations

Approved use is limited to the validated commodities and environmental surfaces listed under Scope; other foods require validation data forwarded to the Microbiological Methods Committee before results carry the same standing. Food testing is defined for 125 g samples only, with compositing capped at 5 units of 25 g. Genus-level detection has an inherent interpretive limit: detection of Listeria spp. says nothing about the absence of L. monocytogenes, so speciation through cultural confirmation decides any species-level statement. Confirmation capacity also has a time limit, since the retained non-boiled broth keeps for up to 72 h refrigerated only.

Safety Notes

  • Treat all enrichments as potentially carrying Listeria monocytogenes, a human pathogen; work with standard precautions for pathogenic enteric and environmental organisms, and contain aerosols during stomaching, blending and vortexing.
  • Dispose of used SPR®s and strips in an appropriate biohazard receptacle in accordance with applicable local regulations, and disinfect the working area surface with a suitable disinfectant before analysis.
  • Take care around the 95–100°C water bath and the heated strips and tubes; allow the prescribed cooling periods before handling.
  • As generic laboratory practice, wear gloves and eye protection when handling enrichment broths and kit reagents, and decontaminate spills of enrichment culture promptly.

Source Method