SOP: Rapid PCR Screening for Listeria Species on Stainless Steel Surfaces (MFLP-51)

GIFSQ ID: SOP-1052

Speed is the point of this screen. A sponge or swab wiped across stainless steel goes into a proprietary enrichment broth for roughly two days, and real-time PCR then answers within about 3 hours of enrichment finishing. Primers and molecular beacon probes built for Listeria species do the recognising: a beacon only lights up once it has bound its target sequence, so fluorescence rises in step with amplified Listeria DNA and stays absent when no target organism was on the sampled surface.

Nothing here stands alone as a final answer. Any presumptive positive still needs cultural confirmation before it is reported as confirmed, and approval covers stainless steel only.

Purpose

Purpose is rapid detection of Listeria species on stainless steel surfaces, supporting compliance with Sections 4 and 7 of the Food and Drugs Act. Positive screening results must be confirmed by a cultural method.

Scope

Validation covers stainless steel surfaces. Other environmental surfaces sit outside the approved application: before such a surface is tested routinely, validate fitness for purpose against Compendium criteria and forward the validation data to the Microbiological Methods Committee, so the application can be widened if the data meet committee requirements. Work is performed under ISO/IEC 17025:1999 (or latest version). Where a laboratory substitutes a variation of any listed medium or kit component, commercially made or prepared in-house, the analyst or laboratory supervisor must demonstrate equivalency.

Equipment and Materials

  • Listerboost enrichment broth (AEB 910623/4, AES Chemunex Canada)
  • ADIAFOOD kit (DFS 6102A, AES Chemunex Canada): extraction buffer EX-1; lyophilised extraction reagent EX-2; detection buffer DT-1; lyophilised detection reagent DT-2; detection microplates or strip tubes holding pre-dispensed PCR reagents; control strip tubes (with detection strip kits only); PCR optical-grade caps
  • ADIAFOOD extraction kit (EXT9997A): extraction plates, domed caps and seals
  • Sponges pre-moistened with 10 mL of D/E Neutralizing buffer, with stomacher bags; swabs pre-moistened with D/E Neutralizing buffer, with culture tubes
  • AES-validated real-time PCR thermocycler: 96-well (or 48-well) microplate (low and high profile) or 96 (48) × 0.2 mL tube capacity; excitation of fluorophores peaking at 485 to 520 nm; detection of fluorophores with peak emission at 500 to 600 nm; Sentinel software
  • Conventional thermocycler for DNA extraction (optional); centrifuge with rotor and microplate carriers; vortex fitted for microplates; stomacher or equivalent; multi-channel and single-channel pipettors; PCR enclosure (optional); PCR capping tool; strip holders
  • Aerosol barrier pipettor tips; disposable multi-channel pipettor basins; powder-free gloves; denatured alcohol and bleach; incubator holding 37 ± 1°C
  • Sterile, DNase-free water, pipettes, pipette tips and all other PCR-contact materials

Procedure

  1. Collect and handle surface samples with sponges pre-moistened with 10 mL of D/E Neutralizing buffer, held in stomacher bags, or with swabs pre-moistened with D/E Neutralizing buffer, held in culture tubes, following standard environmental surface sampling practice for Listeria.
  2. Set up each sponge sample by transferring the sponge into a sterile stomacher bag with 90 mL of Listerboost enrichment broth and mixing thoroughly in a stomacher. Set up each swab sample by transferring the swab into a sterile culture tube with 10 mL of Listerboost enrichment broth and vortexing.
  3. Composite environmental samples where compositing applies, adding broth per item composited: 10 mL of Listerboost enrichment broth for each swab, or 90 mL for each sponge. Stomach for thorough mixing.
  4. Enrich every sample in Listerboost broth for 42 to 48 hours at 37 ± 1°C.
  5. Reconstitute extraction reagent just before extraction: dispense the contents of EX-1 (squeeze bottle) into the EX-2 vial, recap, and mix by inversion until the desiccated reagent has completely dissolved. Pour reconstituted EX-2 into a multi-channel pipettor basin.
  6. Aliquot 90 µL of reconstituted EX-2 into each well of the extraction microplate. Transfer 10 µL of enriched cell suspension into the plate, following the sample assignment layout. Seal wells with the domed cap strips provided.
  7. Place the extraction microplate in the thermocycler and run the DNA extraction program. That automated cycle lyses the bacterial cells; once it finishes, remove the plate, all organisms in it are now lysed and inactivated.
  8. Centrifuge the extraction microplate for 5 minutes at 1,800 × g. Supernatant from this spin is the extracted DNA used for amplification.
  9. Reconstitute detection reagent: dispense the contents of DT-1 (squeeze bottle) into the DT-2 vial, recap, and mix by inversion until fully dissolved. Pour reconstituted DT-2 into a multi-channel pipettor basin.
  10. Remove a detection microplate or strips from the pouch, setting strips on the strip holder. Aliquot 15 µL of reconstituted DT-2 into each well of the detection microplate or strips, then transfer 10 µL of extracted DNA from the extraction microplate into the matching wells.
  11. Seal the detection microplate or strips with the optical-grade caps provided. Seal the extraction microplate with the microplate seals provided and store it at 4°C until analysis is complete.
  12. Vigorously invert the detection microplate or strips 4 times so well contents mix thoroughly, then gently tap to remove air bubbles from the bottom of the wells. Vortex for 1 minute at maximum speed, then centrifuge for 1 minute at 1,800 × g so the whole PCR mixture, sample DNA extract included, sits at the bottom of the wells.
  13. Secure the detection microplate or strips tightly in the PCR instrument, well A1 in the upper left corner, and click “Start Detection” to initialise the PCR protocol. Automated cycling amplifies target DNA and reads beacon fluorescence cycle by cycle.
  14. Remove and discard the detection microplate or strips once the detection program has completed.
  15. Confirm every presumptive positive culturally: work from the Listerboost enrichment broth into the plating and confirmation steps of a cultural Listeria method, such as MFHPB-07 or MFHPB-30, as appropriate.

Recording Results

Read results against the baseline set by the negative controls, which the system itself determines and sets as the cut-off fluorescence value. A sample counts as a positive PCR test when its fluorescence response rises higher than that baseline within 40 cycles of amplification. A negative PCR test normally produces no visible fluorescence. Record each surface sample as presumptive positive or negative on that basis, and record the cultural confirmation outcome alongside every presumptive positive before any confirmed result is reported.

Quality Control

Negative controls anchor every run: fluorescence above the baseline cut-off in a negative control invalidates the whole test, and analysis must be repeated with precautions taken to eliminate possible sources of error. Store ADIAFOOD detection kits refrigerated in original packaging until use; reconstituted EX-2 and DT-2 keep a 14 day shelf life refrigerated in their original containers, sealed with Parafilm. Hold the enrichment incubator at 37 ± 1°C, and keep all water, pipettes, tips and other materials sterile and DNase free, since stray DNA or nuclease activity corrupts a PCR screen.

Troubleshooting

  • Fluorescence in the negative controls: treat the run as invalid, discard the results, decontaminate work areas and consumables (alcohol and bleach are on hand for this), and repeat the analysis from retained material with contamination sources eliminated.
  • Weak or absent signal from a suspect sample: check that EX-2 and DT-2 were fully dissolved on reconstitution, that both reagents are inside their 14 day refrigerated life, and that the detection plate was inverted, vortexed and centrifuged as specified so DNA and reagents actually meet at the well bottom.
  • Air bubbles in wells: tap the sealed plate gently before vortexing; bubbles left at the well bottom distort optical readings.
  • Plate seated wrongly: confirm well A1 sits in the upper left corner of the instrument before starting, otherwise sample identities will not match the assignment layout.

Interferences and Limitations

Primer and beacon specificity carries the method: under the stated reaction conditions the oligonucleotides amplify Listeria species DNA and do not amplify DNA from other organisms, so signal means target DNA was present in the enrichment, not, by itself, that viable Listeria was recovered, which is why cultural confirmation is mandatory. Approval extends only to stainless steel; other surfaces need prior validation. Enrichment is fixed at 42 to 48 hours, and the post-enrichment identification itself takes about 3 hours, so the screen cannot be shortened at the enrichment stage. Detection depends on an AES-validated thermocycler meeting the stated excitation and emission ranges; an unvalidated instrument is outside the method.

Safety Notes

  • Treat enrichments as potentially carrying Listeria monocytogenes, a human pathogen: wear powder-free gloves, contain aerosols during stomaching, vortexing and centrifugation, and dispose of cultures and plates as biohazard waste.
  • Handle denatured alcohol and bleach as corrosive/flammable laboratory chemicals, keeping them away from PCR reagents and optical surfaces.
  • Extracted DNA plates hold lysed, inactivated organisms, but unextracted enrichment broth remains fully infectious until that extraction step is complete.

Source Method

Adapted in original wording from Health Canada method MFLP-51, Detection of Listeria species in Stainless steel Surfaces by the ADIAFOOD Rapid Pathogen Detection System, issued July 2012, Compendium of Analytical Methods, Volume 3.