SOP: Counting Escherichia coli in Foods with the TEMPO EC System (MFLP-107)

GIFSQ ID: SOP-1007

Specificity in this count comes from a single enzyme. Escherichia coli in a filled card well acts on the medium substrate through its β-glucuronidase activity, and fluorescence appears as that reaction proceeds. Sample identity, well sizes and fluorescence pattern feed an automated most probable number calculation, giving an E. coli count in a little over a day without colony picking.

Handling before the card matters as much as reading after it. Homogenisation, dilution choice for coloured or inhibitory foods, and a firm 45-minute window to get suspension into the card protect the final number.

Purpose

Purpose is to enumerate Escherichia coli in a variety of foods, supporting compliance work under Sections 4 and 7 of the Food and Drugs Act and other relevant federal requirements.

Scope

Scope covers all foods apart from strongly coloured foods, with enumeration complete in 22–27 hours. Use the TEMPO® EC test exclusively with the TEMPO® system; its medium vial and card are specific to this method. Any laboratory varying a listed medium formulation must ensure equivalency itself. Listed media are commercially available and are prepared and sterilised by manufacturer instructions, and work is done under ISO/IEC 17025 competence requirements overseen by the laboratory supervisor.

Equipment and Materials

  • TEMPO® system complete for its intended use, including TEMPO® Filler and TEMPO® Reader
  • TEMPO® EC assay kit, reference 80004, holding cards and culture medium vials for the analysis of 48 test samples
  • TEMPO® Filter Bags, reference 80015
  • Pipette able to deliver 0.10 mL or 1.0 mL, with disposable tips; vortex mixer; stomacher, blender or equivalent
  • Incubator able to hold 35°C
  • 0.1% peptone water; sterile deionised or distilled water
  • TEMPO® Control Kit, reference 80000

Procedure

  1. Store units before analysis refrigerated or frozen according to product nature, shelf-stable foods excepted. Thaw frozen samples in a refrigerator, or under time and temperature conditions that prevent microbial growth or death. Analyse units as soon as possible after receipt in the laboratory.
  2. Have primary and secondary sterile diluent prepared beforehand and pre-warmed to room temperature. Clean the working area surface with a suitable disinfectant.
  3. Secure a representative analytical unit: agitate liquids or free-flowing materials until homogeneous, and for a solid sample unit take a portion from several locations within it.
  4. Prepare a 1 in 10 primary dilution in a TEMPO® filter bag by aseptically adding 25 g or 25 mL, the analytical unit, to 225 mL of 0.1% peptone water. Shake, stomach or blend according to food type. A smaller sample size is acceptable when the 1 in 10 dilution is maintained, for example 11 g in 99 mL or 10 g in 90 mL.
  5. Blend or stomach for the minimum time needed for a homogeneous suspension; blending must not exceed 2 min, to avoid overheating. Foods prone to foaming need low blender speed, with the aliquot taken from below the liquid and foam interface.
  6. Assess colour at this stage: a strongly coloured primary dilution, for example fruit purées and cocoa, can affect the fluorescent signal, so a dilution of at least 1 in 400 is recommended for the TEMPO® EC test with such matrices.
  7. Do not let more than 45 minutes elapse between homogenisation of the primary dilution and its transfer into the TEMPO® card.
  8. Set the dispenser or dispensers containing secondary diluent to the appropriate volumes and prime the pump by eliminating the first two volumes dispensed. Log on to the TEMPO® preparation station.
  9. Reconstitute the culture medium in its TEMPO® vial by dispensing the appropriate volume of secondary diluent using the dispenser, then set up the dilution. For a 1 in 40 dilution, enabling enumeration between 10 and 4.9 × 104 CFU/g, aseptically transfer 1 mL of the primary dilution into one TEMPO® EC vial containing 3.0 mL of sterile deionised or distilled water. For a 1 in 400 dilution, enabling enumeration between 100 and 4.9 × 105 CFU/g, aseptically transfer 0.1 mL of the primary dilution into one vial containing 3.9 mL of sterile deionised or distilled water.
  10. Apply the inhibition branch when the food calls for it: aromatic herbs, spices, teas and herbal teas may have an inhibitory effect, so a minimum dilution of 1 in 400 should be used for them.
  11. Cover higher expected counts with serial dilutions of the primary dilution, reaching 1 in 4,000, 1 in 40,000, 1 in 400,000 and beyond, prepared according to manufacturer instructions.
  12. Agitate each dilution to resuspend material that may have settled. With a sterile pipette, transfer the appropriate volume from the TEMPO® bag containing the sample to its corresponding vial holding reconstituted medium. Vortex or agitate manually for approximately 5 seconds.
  13. Identify the sample at the preparation station interface, following its instructions. Remove one TEMPO® EC card and scan the card information into the system as prompted. Scan the TEMPO® EC vial containing the sample, insert the TEMPO® EC fill tube into its corresponding vial, and place the combined card and vial into a filler rack. Repeat for each prepared vial.
  14. Fill the TEMPO® cards using the TEMPO® Filler. The filler moves inoculated medium into 48 wells arranged as three sets of 16 wells, small, medium and large, with a one log difference in volume between sets, mimicking a most probable number layout; the card is then hermetically sealed so later handling carries no contamination risk. After the cycle, visually inspect that the vials are empty and dispose of the used vials in an appropriate receptacle. Remove filled TEMPO® EC cards from the filler rack and insert them into the TEMPO® incubation rack.
  15. Incubate the cards at 35 ± 1°C for 22–27 hours.
  16. Follow the software timing rule. Incubation time is managed by TEMPO® Read software, which integrates a theoretical interval of 15 minutes between reading the card barcode and the start of incubation. If the real interval is greater than 15 minutes, without exceeding 2 hours, that extra time must be added to the remaining incubation time displayed by the software. Reading must always be performed within the time limit authorised by the software.
  17. Remove the incubation rack from the incubator, log on to the TEMPO® reading station, and introduce the rack containing the cards into the TEMPO® Reader. The reader scans the barcode of each card and interprets fluorescence in the wells, automatically associating the sample identifier with the type of test, dilution and enumeration result.
  18. At the end of analysis, remove the cards from the rack and dispose of them into an appropriate receptacle.

Recording Results

Report the enumeration as colony forming units per gram or per millilitre of the initial sample, exactly as the reader outputs it. Calculation rests on the most probable number approach: the number and type, meaning size class, of positive wells determine the estimate for E. coli in the original sample. Record sample identification, dilution branch, incubation period, and any added time for a barcode-to-incubation gap over 15 minutes, so the reported figure can be traced to its card.

Quality Control

Use the TEMPO® Control Kit, reference 80000, for routine system control following manufacturer directions; no additional numeric acceptance limit is invented here. Incubator control is a named quality point: for growth steps at 37°C or below tolerance is ± 1.0°C, so 35 ± 1.0°C here, while higher temperatures must be maintained within 0.5°C because of potential lethality to the microorganism sought. Dispenser priming, disinfected surfaces, room-temperature diluents and verified-empty vials complete the run checks, framed as standard laboratory quality practice.

Troubleshooting

  • Fluorescence doubtful on a deeply coloured food: repeat using a dilution of at least 1 in 400; fruit purées and cocoa are named problem matrices.
  • Herb, spice or tea results depressed: confirm the minimum 1 in 400 dilution was used to offset inhibitory effects.
  • Count saturates the card: prepare serial dilutions at 1 in 4,000, 1 in 40,000 or higher and retest; do not extend the stated range by calculation.
  • Homogenate waited too long: any primary dilution older than 45 minutes before card transfer falls outside the method; prepare a fresh dilution.
  • Delayed incubation start: add the real extra interval, up to 2 hours total gap, to the remaining time shown by TEMPO® Read, and read only within the software-authorised limit.

Interferences and Limitations

Strongly coloured foods sit outside the validated scope, and colour in the primary dilution can distort the fluorescent signal that detection depends on. Detection itself depends on β-glucuronidase activity: E. coli present must express that activity and generate fluorescence under the stated incubation before the reader can count a well as positive. Raising dilution to manage colour or inhibition raises the reporting floor too, from 10 CFU/g at 1 in 40 to 100 CFU/g at 1 in 400.

Safety Notes

  • Handle foods, suspensions and incubated cards as potentially containing pathogenic E. coli and other enteric organisms; use aseptic technique, gloves and biohazard disposal throughout.
  • Keep blending to the stated maximum, operate blenders at low speed for foaming foods, and guard against aerosols and splashes when opening bags and jars.
  • Place used vials and finished cards in appropriate receptacles and disinfect the working area once reading is complete.

Source Method

Adapted in original wording from Health Canada method MFLP-107, Enumeration of Escherichia coli in a Variety of Foods using the TEMPO EC (E. coli Count) method, October 2018, Compendium of Analytical Methods, Volume 3.