SOP: PCR Detection of Verotoxin Genes in E. coli Isolates (MFLP-62)
GIFSQ ID: SOP-1057
Colonies do the travelling in this method; foods never enter the PCR itself. Presumptive verotoxigenic Escherichia coli (VTEC) colonies, isolated beforehand by a cultural method, are boiled in Triton X-100 to release their DNA, and a multiplex PCR then asks a narrow question: are the verotoxin 1 and/or verotoxin 2 genes (VT1 and/or VT2) present? Amplicons are sized by rapid agarose gel electrophoresis on a FlashGel® system, alongside an internal amplification control (IAC) built from a template-independent primer pair with complementary 3′ ends, which flags any reaction where inhibition has silenced the PCR.
Controls decide whether a run counts at all. A VTEC positive control, a generic E. coli negative control, a lysis buffer blank and two no-template reactions accompany every set, and bands in the wrong lane, or a missing IAC in a gene-negative sample, invalidate the run outright.
Purpose
Purpose is to determine the presence of the verotoxin genes in verotoxigenic E. coli colonies isolated from foods, supporting compliance with Sections 4 and 7 of the Food and Drugs Act and/or other relevant federal regulations. Individual colonies positive by this VT-PCR are confirmed as possessing the genes encoding the virulence factors verotoxin 1 and/or verotoxin 2.
Scope
Scope begins where isolation ends. Foods are enriched, and plated on a validated selective medium, under the applicable Compendium method for verotoxigenic E. coli; this procedure characterises the colonies that come off those plates. A colony from Brain Heart Infusion (BHI) agar or Nutrient Agar (NA) may also be characterised, provided it was purified from a selective agar listed in the cultural method and/or had previously been confirmed as E. coli. Work follows ISO/IEC 17025:2005 (or latest version); equivalency is the laboratory’s responsibility wherever a listed medium formulation is varied, with data forwarded to the Editor of the Compendium of Analytical Methods. One reagent allows no substitution at all: HotStarTaq® must be used as specified.
Equipment and Materials
- Biological safety cabinet (BSC) and/or PCR workstation; chiller block for 0.2 mL or 0.5 mL PCR tubes; freezer capable of maintaining −20°C; refrigerator capable of maintaining 4°C
- Heat block accommodating 1.5 mL microcentrifuge tubes and holding 100°C; microcentrifuge for 1.5 mL tubes; vortex mixer; micropipettors adjustable across 1.0 to 1000 µL
- Thermal cycler equipped with a heated lid; FlashGel® DNA System (Lonza Group Ltd.) or equivalent; photodocumentation system (optional)
- Disposable gloves; sterile nuclease-free 1.5 mL microtubes, screw-cap and regular; microtube racks; cold racks for 0.2 mL, 0.5 mL and 1.5 mL tubes; sterile nuclease-free PCR tubes (0.2 mL or 0.5 mL); sterile filtered nuclease-free pipettor tips
- FlashGel® DNA Cassette, 2.2% agarose; FlashGel Loading Dye; FlashGel DNA Marker 50 bp–1.5 kb
- Rainbow® Agar dehydrated media (Biolog, Inc., Cat. no. 80102); Brain Heart Infusion (BHI) agar or Nutrient Agar (NA); Triton X-100
- Bovine serum albumin (BSA), non-acetylated, 20 mg/mL; dNTP mix, PCR grade, 10 mM; HotStarTaq® (Qiagen, Cat. no. 203205), no substitutions permitted; DNase and RNase free water (Ultrapure or molecular grade)
- Primers, forward and reverse, for VT1, VT2 and IAC: VT1-F, GTG GCA AGA GCG ATG TTA CGG TTT G; VT1-R, ATG ATA GTC AGG CAG GAC GCT ACT C (182 bp amplicon from the vt1 gene, designed from NCBI #M19473); VT2-F, ACG AGG GCT TGA TGT CTA TCA GGC G; VT2-R, GCG ACA CGT TGC AGA GTG GTA TAA C (200 bp amplicon from the vt2 gene, designed from NCBI # X07865); IAC-1, CAT AAT ATC ACT CGC GTC CGT TGA AGC TTA; IAC-2, GAC GAA ATC GTA AGC TTC AA (40 bp amplicon)
- Control cultures: positive control, E. coli O157:H7 containing vt1 and vt2 (e.g., ATCC 35150); negative control, generic E. coli negative for both verotoxin genes
Procedure
- Process controls with every set of samples: a positive control isolate of the VTEC laboratory strain and a negative control isolate of the generic E. coli (non-VTEC) strain go through the full procedure alongside the test colonies. To avoid delays, well isolated typical colonies may be streaked concurrently to BHI, NA or Rainbow agar when typical isolates are picked for confirmation, so suspect E. coli O157 isolates can afterwards be assayed by VT-PCR from colonies grown on the Rainbow, BHI or NA plates.
- Prepare the cell lysate. Pick a portion of a single suspect isolated colony from Rainbow agar, BHI agar or nutrient agar with a loop and suspend it in 200 µL of 1% (v/v) Triton X-100 in a sterile 1.5 mL microfuge tube with a secured cap (e.g. screw-cap). Vortex to mix. Set up a lysis buffer negative control tube containing 200 µL of 1% (v/v) Triton X-100 with no colony, and carry it, plus the positive and negative control cultures, through the remainder of the procedure.
- Make the PCR template by heating the tubes at 100°C in a heat block for 10 min. Cool the resulting lysates for 5 min at room temperature and pulse-spin for a few seconds to bring down condensation accumulated under the tube cap. Use 5 µL of the supernatant as the PCR template.
- Label reaction tubes. Prepare the VT-PCR master mix (composition in step 5) and dispense 45 µL into 0.2 or 0.5 mL PCR tubes. Add 5 µL of colony lysate to the corresponding tubes, including the positive and negative control cultures. Include two further controls: one reaction of PCR master mix with 5 µL of water and no template, and one reaction of PCR master mix with 5 µL of the 1% Triton X-100 devoid of template.
- Composition of the master mix, per 50 µL total reaction volume: molecular grade water 32.25 µL; 10 × HotStar Taq PCR Buffer (included with the Qiagen HotStarTaq polymerase kit) 5 µL, 1× final; dNTP mix (10 mM) 1 µL, 0.2 mM final; BSA, non-acetylated (20 mg/mL) 1.25 µL, 0.5 µg/µL final; VT1-F (10 µM) 1 µL, 0.2 µM final; VT1-R (10 µM) 1 µL, 0.2 µM final; VT2-F (10 µM) 1 µL, 0.2 µM final; VT2-R (10 µM) 1 µL, 0.2 µM final; IAC-1 (10 µM) 0.5 µL, 0.1 µM final; IAC-2 (10 µM) 0.5 µL, 0.1 µM final; Qiagen HotStar Taq® (5 U/µL) 0.5 µL, 2.5 U; template (lysate) 5 µL. Master mix may be prepared in larger quantities, aliquoted in portions of >150 µL into screw-cap 1.5 mL-capacity tubes, and stored at −20°C for up to 2 months.
- Place the tubes in the thermal cycler and run the VT-PCR cycling program: HotStarTaq® polymerase activation at 94°C for 15 min; then 35 cycles of denaturation at 94°C for 30 sec, annealing at 55°C for 30 sec and extension at 72°C for 1 min; then a final elongation at 72°C for 2 min; then hold at 4°C indefinitely. Where products cannot be analysed at once, store them at 4°C for up to 24 h or at −20°C for up to one week.
- Analyse the PCR products by FlashGel® electrophoresis, following the manufacturer’s instructions. Load 5 µL of FlashGel® 50 bp–1.5 kb DNA marker into the first well of a FlashGel® 2.2% DNA cassette. Prepare each PCR product by mixing 2 µL of PCR product with 2 µL of ultrapure or molecular grade water and 1 µL of 5X Flash Gel® loading dye, and load the prepared samples into the remaining wells.
- Run the FlashGel® for approximately 6.5 min at 275 V. Under these conditions both VT bands and the IAC resolve as follows: VT1, 182 bp; VT2, 200 bp; IAC, 40 bp. Expect the IAC to appear faint for VT positive samples.
Recording Results
Interpretation follows four fixed rules. A sample is positive where it is positive for VT1 and/or VT2. A sample negative for both VT1 and VT2 genes and positive for the IAC is negative. A sample negative for both VT1 and VT2 genes and negative for the IAC shows evidence of PCR inhibition by sample matrix contaminants, failure of one of the detector reagents, or omission of a critical component: the results of the run are invalidated and the VT-PCR must be repeated. VT bands observed in the negative and/or non-template or sterility controls likewise invalidate the run, which must be repeated with precautions taken to eliminate possible sources of contamination. Record, for each colony, the agar it was picked from, the lysate and control outcomes, band sizes observed (VT1, 182 bp; VT2, 200 bp; IAC, 40 bp) and the final call, noting any repeat runs and why they were required.
Quality Control
Every set carries the VTEC positive control (E. coli O157:H7 containing vt1 and vt2, e.g. ATCC 35150), the generic E. coli negative control, a lysis buffer negative control, a water no-template reaction and a Triton X-100 no-template reaction; a run whose controls misbehave is invalid rather than interpretable. HotStarTaq® (Qiagen, Cat. no. 203205) is used exactly as specified, with no substitutions permitted, and stored master mix aliquots keep at −20°C for no more than 2 months. Temperatures must be held as recommended: a stated 35°C allows 35 ± 1.0°C, and 30°C or 25°C likewise allow ± 1.0°C, while higher temperatures such as 43°C or 45.5°C must be held within 0.5°C because of potential lethality to the organism being isolated, unless otherwise specified. As generic laboratory practice, keep pre- and post-amplification work separated and use filtered, nuclease-free tips throughout against amplicon carryover.
Troubleshooting
- No VT band and no IAC band in a test lane: read this as PCR inhibition, a failed detector reagent or an omitted component. The run is invalid; repeat the VT-PCR rather than reporting a negative.
- VT bands in negative, non-template or sterility controls: contamination is present. Invalidate the run and repeat it, taking precautions to eliminate possible sources of contamination.
- Faint IAC beside a clear VT band: expected behaviour in VT positive samples, not a fault; the sample still reads positive for VT1 and/or VT2.
- Weak product after storage: check holding conditions were respected: PCR products keep at 4°C for up to 24 h or at −20°C for up to one week only.
- Master mix older than its limit: discard aliquots stored at −20°C beyond 2 months and prepare fresh mix; never substitute another polymerase for HotStarTaq®.
- Condensation under caps after boiling: pulse-spin lysates for a few seconds after the 5 min room-temperature cool, so the full 5 µL template volume is recovered from the supernatant.
Interferences and Limitations
Characterisation depends on starting material: only colonies from Rainbow agar, BHI agar or nutrient agar are assayed, and colonies from BHI or NA qualify only where they were purified from a selective agar listed in the cultural method and/or previously confirmed as E. coli. A gene result is not a toxin result: the method determines presence of the vt1 and vt2 genes in an isolate, and inhibition by sample matrix contaminants can silence a reaction entirely, which is why the IAC outcome governs whether a gene-negative result may be reported at all. Band spacing is tight by design (VT1 at 182 bp against VT2 at 200 bp, with the IAC at 40 bp), so electrophoresis must be run as specified, approximately 6.5 min at 275 V on the 2.2% cassette, for the three bands to resolve.
Safety Notes
- Handle VTEC control cultures and suspect colonies as human pathogens: work in a biological safety cabinet, wear disposable gloves, and dispose of cultures, lysates and gels as biohazard waste under local rules.
- Take care with the 100°C heat block and freshly boiled microfuge tubes; use screw-cap tubes with secured caps for the lysis boil so caps cannot pop and release hot liquid or aerosol.
- Treat Triton X-100 and loading dye as laboratory chemicals, avoid skin and eye contact, and clean spills promptly.
- As generic laboratory practice, never open post-amplification tubes in the template preparation area, and decontaminate the PCR workstation after any run invalidated for contamination.
Source Method
Adapted in original wording from Health Canada method MFLP-62, Determination of Verotoxin genes in Verotoxigenic Escherichia coli isolates, November 2015, Compendium of Analytical Methods, Volume 3.