SOP: Oxygen Transmission Rate of Plastic Films by Coulometric Sensor (ASTM D3985)

GIFSQ ID: SOP-0990

Oxygen creeping through a film decides how fast a packed food oxidises, loses colour or goes stale, so a laboratory needs a number for that movement rather than a guess from a resin datasheet. Coulometric measurement supplies that number with unusual sensitivity: every oxygen molecule reaching the detector generates electrical current, and current becomes the transmission rate once flow through the film has settled into a steady state.

Running the test is an exercise in patience and gas purity. Film must first saturate with oxygen before flow becomes constant, and traces of oxygen hiding in the carrier gas would be counted as if they had crossed the specimen.

Purpose

Purpose is to measure the rate at which oxygen passes through a flat plastic film or sheeting specimen under controlled temperature and humidity, expressed per unit area and per unit time, using a coulometric sensor as specified in ASTM D3985.

Scope

Scope covers flat films and sheeting tested as a sealed barrier between two chambers of a permeation cell. One source describes a 4 in. by 4 in. flat film specimen. Design intent was low-permeating, high-barrier material, with transmission rates typically in the range 0.0003 to 50 cc/100 in.2/day/atm, equivalently 0.005 to 775 cc/m2/day/atm; newer instrumentation described in the same source extends measurement up to 10,000 cc/100 in.2/day/atm for highly permeable films.

Note on neighbouring methods: a pressure-rise gas transmission method already published in this library measures gas movement through an evacuated chamber and works for gases generally. That technique is different. Material here is oxygen-specific: oxygen alone is detected, by a coulometric sensor, while pressure stays essentially equal on both faces of the film. Results from the two methods should not be interchanged without checking what each instrument actually measured.

Equipment and Materials

  • Isostatic permeation cell in which the specimen forms a sealed semi-barrier between two chambers, with a fixed exposed test area
  • Coulometric oxygen sensor with its load resistor and readout or computer acquisition; sensor construction described in the source is a cadmium anode with a graphite cathode
  • Oxygen test gas, used at 100% on the test-gas side in the arrangement described in Book 2
  • Oxygen-free nitrogen carrier gas; a nitrogen and hydrogen mixture is required where trace oxygen in process gas must be removed catalytically before the sensor
  • Temperature control for the cell, and humidity control by humidifier, with input-pressure adjustment of test gas and carrier gas
  • Film specimens cut flat and free of visible damage, with thickness known where a permeability coefficient is to be calculated
  • Reference or calibration arrangements supplied with the instrument, used as general laboratory practice for checking sensor response before a series

Procedure

  1. Mount the specimen. Seal the flat film into the permeation cell so it separates the two chambers across the fixed test area. Any path around the edge would bypass the film entirely, so check the seal before gases are started.
  2. Start test gas on one face. Pass oxygen slowly over one side of the specimen. Book 2 describes that side held under a slow flow of 100% oxygen.
  3. Start carrier gas on the other face. Pass nitrogen, at 100% in the Book 2 arrangement, slowly over the opposite face at an absolute pressure similar to the test-gas side. Equal absolute pressure is what makes the method isostatic; a constant partial-pressure difference across the film is what makes it steady-state in character.
  4. Protect the sensor from stray oxygen. Where process gases carry trace oxygen, feed the nitrogen and hydrogen mixture described in the source so oxygen contamination is removed catalytically. Keep the sensor bathed in a stream of oxygen-free gas between and during tests, because sensor life depends on total oxygen exposure.
  5. Set and hold test conditions. Hold the film at a precise temperature and humidity, since both affect oxygen permeation through many polymers. Humidity is controlled with a humidifier and the input pressures of test and carrier gas; as input pressures rise, relative humidity falls. Record the conditions actually used: examples in the sources include 23°C at 0% relative humidity on a film specification sheet, 37.8°C (100°F) at 90% relative humidity given as typical conditions in Book 2, and a produce film compared at 6°C and 23°C, where transmission at 6°C came out at less than half the room-temperature value. Test at the temperature of intended use wherever shelf life is the question.
  6. Wait for steady state. Allow oxygen to saturate the polymer and the emerging flow to level off. Early readings climb as the film fills; only when the amount permeating per unit time reaches its equilibrium, constant level has the transmission rate been established.
  7. Read the sensor. Carrier gas sweeps oxygen that has crossed the film into the coulometric sensor. Inside the sensor, oxygen is reduced at the cathode, written in the source as: ½ O2 + H2O + 2e– → 2OH–, while cadmium is oxidised at the anode: Cd + 2OH– → Cd(OH)2 + 2e–. Book 2 states the same event in whole-molecule terms: four electrons are released for each oxygen molecule the sensor sees. Electrons flowing across the scaling load resistor produce a voltage proportional to oxygen arriving at the sensor, and that signal gives the amount transmitted.
  8. Finish and preserve the sensor. At the end of a run, return the sensor to its oxygen-free gas stream rather than leaving it exposed to oxygen, and log total exposure where the instrument maker requires it, as general laboratory practice.

Recording Results

Report oxygen transmission rate as amount of oxygen per unit film area per unit time, in cc/m2/day/atm or cc/100 in.2/day/atm, together with test temperature, relative humidity, oxygen concentration used and specimen thickness. A transmission rate belongs to the particular film tested, at its actual thickness, unlike a permeability coefficient, which is a material property independent of thickness.

Where a permeability coefficient is needed, use the relationship given in the source: permeability P is the product of diffusion coefficient D and solubility coefficient S, and transmission rate follows GTR = (P / l) × Δp, where l is film thickness and Δp is the partial-pressure difference across the film. Rearranged from the amount Q crossing area A in time t, P = Q × l / (A × t × Δp). Common permeability units in the literature are cm3 (STP) µm/(m2 day atm) and cm3 mil/(100 in.2 day atm).

Quality Control

Sensor health dominates quality here. Coulometric sensors are extremely sensitive to oxygen, comparatively expensive to acquire and maintain, and have a limited life span governed by overall oxygen exposure, so a drifting response should be checked against the instrument maker’s reference arrangements before specimens are blamed. Gas purity deserves equal attention: oxygen-contaminated carrier gas inflates the apparent rate, which is why the catalytic nitrogen and hydrogen mixture exists. Duplicate testing is worthwhile on new materials; Book 2 reports duplicate runs on a 3.1 mil coextruded celery-packaging film as its worked example.

Troubleshooting

  • Reading never levels off: check the specimen seal for an edge leak, then allow more time; thick or slowly saturating films take longer to reach steady state.
  • Baseline sits above zero with no specimen transmission expected: suspect oxygen contamination in the carrier or process gas; renew the catalytically cleaned nitrogen and hydrogen supply and re-check.
  • Results shift between runs at different settings: compare temperature and humidity first. Permeability follows an Arrhenius-type temperature relationship over small ranges, and humidity changes the diffusion behaviour of water-sensitive polymers such as ethylene vinyl alcohol and polyamide.
  • High-transmission film overloads the detector: older practice masked the specimen to a smaller area or lowered the oxygen concentration to reduce driving force, but the source warns those workarounds suffered poor repeatability and poor agreement between laboratories; prefer instrumentation built for the high range.
  • Testing below 10°C proves difficult: condensation can damage the sensor, a known limitation for chilled modified-atmosphere work; protect the sensor and treat low-temperature results with caution.

Interferences and Limitations

Humidity is both a test condition and an interferent: for polar polymers such as ethylene vinyl alcohol and polyamide, diffusion depends on water content, so a dry-gas result may not describe a film in contact with a moist food. Temperature carries the same warning, since a rate at room temperature can be more than double a rate at chill temperature. Prediction from early data is not part of this method; transmission is read at steady state. Sensor cost, gas cost (the cleaned mixture tends to cost about ten times unmixed nitrogen) and sensor replacement limit how casually the method can be run.

Safety Notes

  • Handle hydrogen-containing carrier mixtures as flammable gas and oxygen supplies as oxidising gas, following their safety data sheets and cylinder rules, as general laboratory practice.
  • Dispose of spent coulometric sensors through the route for cadmium-containing waste required locally, as general laboratory practice; cadmium is part of the sensor anode described in the source.
  • Vent exhaust gases safely and check pressurised gas lines for leaks before a run, as general laboratory practice.

Source and Standard

Drafted from Chapter 5 of Food Packaging Materials: Testing & Quality Assurance (Singh, Wani and Langowski, eds., CRC Press, 2017), especially the isostatic steady-state section, the ASTM D3985 discussion and the sensing-mechanism section, and from Chapters 11 and 13 of Food Packaging: Testing Methods and Applications (Risch, ed., ACS Symposium Series 753, 2000). Governing standard: ASTM D3985, Standard Test Method for Oxygen Gas Transmission Rate Through Plastic Film and Sheeting Using a Coulometric Sensor.