SOP: Counting Mould Filaments in Tomato and Vegetable Juice Products by the Howard Method (MFO-5)

GIFSQ ID: SOP-0960

Mould in a tomato product is not counted colony by colony. Instead, a small, standardised portion is spread in a Howard mould counting chamber and examined field by field under a microscope, and each field is scored simply as positive or negative for mould filaments. Done carefully, that percentage of positive fields gives a reproducible picture of how much mouldy raw material went into a lot.

Consistency carries the whole method. Mixing counts, spatula size, chamber cleanliness, field diameter and the scanning pattern are all fixed, because any drift in one of them changes the percentage even when the product itself has not changed. Analysts also need a practised eye: mould filaments must be told apart from plant tissue, lint and other look-alikes before a field can be called positive.

Purpose

Purpose here is to determine mould filaments in the tomato products within scope, expressed as a percentage of microscopic fields containing mould, and to judge each lot against the tolerance that applies to its product group.

Scope

Scope covers canned tomatoes, tomato juice and vegetable juice under Section B.11.016 of the Food and Drug Regulations, and tomato puree, tomato paste, tomato pulp and tomato catsup under Section B.11.017. Sampling language matches the other official methods: a lot is a batch or production unit identified by the same code, or, where no code exists, product made under essentially the same conditions at one establishment representing no more than one day’s production, or a quantity of the same kind of product from one manufacturer available for sampling at a fixed location. A sample unit is usually a consumer-size container of at least 100 g (ml), and an analytical unit is the amount withdrawn from it for examination.

One related supplement was reviewed for this draft: a March 2009 supplement covering official Salmonella methods MFO-6, MFO-10, MFO-11 and MFO-12, directing users of those methods to MFHPB-20 because of safety and environmental concerns with Selenite Cystine. Its subject is Salmonella detection, not mould counting, so none of its content applies to this procedure.

Equipment and Materials

  • Compound microscope, binocular or monocular, with mechanical stage, condenser with iris diaphragm and a source of illumination
  • Objectives: 10× (16 mm) for counting and 20× (8 mm) for confirmation; oculars of 8× to 12.5×
  • Ocular micrometer disk, cross-ruled in sixths of the ocular diaphragm opening, prepared as described in the procedure
  • Howard mould counting chamber or cell, with cover glass
  • Spatula with a 5.0 mm flat blade, marked 10.0 mm from the tip so the working area is 50 sq. mm; grind a blade down to that width and flat surface if needed
  • Dissecting needle, Bunsen burner, lint-free clean towel or cloth
  • U.S. standard sieve No. 2, for canned tomatoes
  • Wide-mouth bottles with screw caps or other suitable containers, and a spoon or other suitable utensil for puree, pulp, paste and catsup
  • Refractometer, for puree, pulp and paste; coarse filter paper or celluwipe
  • Distilled water
  • Stage micrometer, or a Howard cell scribed with two parallel lines or a circle measuring 1.382 mm, for calibration

Procedure

  1. Collect the sample. Take six sample units at random from each lot, each containing at least 100 g or 100 ml. Use aseptic technique when sampling from bulk, placing each unit into a separate sterile container, and analyse each sample unit separately.
  2. Calibrate the counting field. Using the 10× objective and oculars in the 8× to 12.5× range, measure field diameter with a stage micrometer or with the lines or circle scribed on a Howard cell, adjusting until field diameter is 1.382 mm. Where diameter is less than 1.382 mm use lower-power oculars; where it is greater, raise ocular height until diameter coincides with 1.382 mm, or fit an accessory drop-in ocular diaphragm cut accurately to the needed aperture size.
  3. Fit the micrometer disk. Use or make a disk of suitable diameter to fit the ocular (approximately 21 mm) and 1 mm thick, marked with a centre grid of 36 small squares, six to a side, sized so the length of six squares equals the diameter of the ocular diaphragm as adjusted in step 2. Calculate the grid width (10–14 mm) that coincides with 1.382 mm on the stage, mark it, place the disk in the ocular and check coincidence, adjusting lines as needed; once width is correct, etch the grid with very fine lines, centred on the disk.
  4. Set illumination. Locate and focus a mould filament, focus the light source into the condenser, and adjust condenser height, iris diaphragm diameter and light intensity to give clear, uniform illumination: enough light to see all particles, but not so intense that filament characteristics are masked. Use a coloured filter if needed to increase contrast.
  5. Prepare juice. For tomato juice and vegetable juice, shake the unopened container 60 times in 30 sec through a 30 cm arc. Open it, and where considerable foam forms pass a Bunsen flame lightly over the surface to disperse foam, then go to chamber preparation in step 9.
  6. Prepare canned tomatoes. Shake the unopened container 60 times in 30 sec through a 30 cm arc, open it, drain the liquid through a No. 2 sieve into a suitable clean receptacle, transfer the liquid to a wide-mouth bottle, secure the lid, and then handle the liquid as for juice.
  7. Prepare puree, pulp and paste. Open the container and mix the product 60 times in 30 sec with a spoon or other suitable utensil. Transfer a small portion onto coarse filter paper or celluwipe and measure the refractive index of the filtrate; pulp removal does not affect refractive index, which rests on soluble solids alone, while leaving pulp in gives a hazy image that is hard to centre and read. From the dilution table in Recording Results, find the distilled water to add to 100 ml of sample to reach a final refractive index of 1.3448–1.3454 at 20°C or 1.3442–1.3448 at 25°C. Mix again as above, transfer 100 ml to a wide-mouth bottle, add the required water, secure the lid and repeat mixing, then re-measure refractive index and correct if necessary.
  8. Prepare catsup. Open the container and mix 60 times in 30 sec with a spoon or other suitable instrument. Transfer a measured, well-mixed representative portion to a wide-mouth bottle, dilute with an equal volume of distilled water, secure the lid and shake 60 times in 30 sec through a 30 cm arc.
  9. Clean the Howard cell. Clean cell and cover glass, making certain the central area is clean, rinse with distilled water, dry with a lint-free cloth and pass lightly over a Bunsen flame. Check cleanliness by placing the cover glass in position and pressing firmly against the shoulders: Newton’s rings appearing between each shoulder and the cover glass, and remaining after pressure is released, show the slide is sufficiently clean; viewed with light reflected from the cover glass, properly formed rings look like broken arcs of concentric circles in rainbow colours. Re-wash slide and cover glass if rings do not form, because their absence means dirt is preventing proper seating and the chamber would hold an incorrect sample volume. Clean spatula and dissecting needle, rinse in distilled water, flame and cool.
  10. Make the mount, inclined technique. Dip the spatula into well-mixed sample up to the 10 mm line and transfer a portion to an area on the central disk halfway between centre and far edge, using a dissecting needle to help the transfer; neither spatula nor needle may touch the central disk, only the sample. Rest one edge of the cover glass in a slanting position on the shoulders nearest the portion, lower it slightly until it almost touches the material, then lower rapidly but gently into place so material spreads evenly over the whole disk. Lowering too rapidly can splash sample onto a shoulder and ruin the mount; lowering too gently leaves material unevenly spread.
  11. Or make the mount, parallel technique. Transfer a portion as in step 10, but onto the approximate centre of the disk. Hold the cover glass parallel to the disk surface and lower slowly until it just touches the sample, then, maintaining contact, raise and lower it very slightly 2 or 3 times; without stopping, lower rapidly but gently until it just touches the shoulders, so the portion spreads evenly.
  12. Accept or discard the mount. Accept only a mount with sufficient material to fill the counting area, Newton’s rings visible, even distribution of material, and absence of air bubbles. Discard any mount showing uneven distribution, absence of Newton’s rings, liquid drawn across the moat and between cover glass and shoulder, or numerous air bubbles. Examine each sample unit immediately after preparation; where delay is unavoidable, shake thoroughly again before examination. Repeat preparation for the remaining five sample units.
  13. Examine the mount. Place the cell on the stage and examine at 90–125× magnification with field diameter 1.382 mm (1.5 sq. mm); use higher magnification (180–250×) only to confirm mould. From each of 2 or more mounts, examine at least 25 fields representative of all sections: examine alternate fields horizontally across the preparation until 5 fields are done, move the mechanical stage vertically to the next alternate row and examine 5 more alternate fields in the reverse horizontal direction, repeating until 25 fields are examined. Where a field holds an air bubble, move to another field unless mould is seen at first glance, because such a field contains insufficient sample; otherwise never move the slide purposely to include or exclude mould filaments.
  14. Score each field. Count a field as positive when the aggregate length of not more than three of the longest filaments present exceeds 1/6 of the diameter of the field; those filaments may be separate or attached to each other, and a clump or mass of mould carries the same value as a single filament.

Recording Results

Calculate for each sample unit: percentage of positive fields = (number of positive fields × 100) / number of fields examined. Average for the whole sample = (percentage for unit 1 + unit 2 + unit 3 + unit 4 + unit 5 + unit 6) / 6. Report the percentage individually for each sample unit as well as the sample average.

Dilution table for puree (pulp) at 20°C, giving water to add to 100 ml of sample unit: refractive index 1.3462 — dilution factor 1.145, water 14.5 ml, total volume 114.5 ml; 1.3478 — 1.292, 29.2 ml, 129.2 ml; 1.3494 — 1.440, 44.0 ml, 144.0 ml; 1.3511 — 1.585, 58.5 ml, 158.5 ml; 1.3527 — 1.730, 73.0 ml, 173.0 ml; 1.3544 — 1.876, 87.6 ml, 187.6 ml; 1.3560 — 2.024, 102.4 ml, 202.4 ml; 1.3577 — 2.171, 117.2 ml, 217.2 ml; 1.3593 — 2.322, 132.2 ml, 232.2 ml; 1.3610 — 2.474, 147.4 ml, 247.4 ml.

Apply tolerances by product group. For canned tomatoes, tomato juice and vegetable juice, the percentage of positive fields must not exceed 25% in any sample unit in the sample taken from the lot. For tomato puree, tomato paste, tomato pulp and tomato catsup, the percentage must not exceed 50% in any sample unit.

Quality Control

Calibration is the first control: counting is valid only at a verified field diameter of 1.382 mm with the micrometer disk grid coinciding with that diameter. Chamber cleanliness is verified by Newton’s rings on every mount, and mounts failing the acceptance criteria in step 12 are discarded and remade rather than counted. Analyst competence is built deliberately: prepare known material by removing mouldy areas from fresh tomatoes infected with various types of mould, boiling them in low-count tomato juice to simulate actual conditions, and examining them microscopically; practise distinguishing mould filaments from plant remnants such as tracheal tube thickenings, pieces of cell wall, lint or fabric segments. Classification of mould types is not required, positive identification of mould filaments is.

Troubleshooting

  • No Newton’s rings: dirt is keeping the cover glass off the shoulders and chamber volume is wrong; re-wash cell and cover glass and remake the mount.
  • Uneven spread or material concentrated at the centre: the sample was not thoroughly mixed, or the cover glass was lowered too gently; remix, and remake the mount using the rapid-but-gentle lowering motion.
  • Sample splashed onto a shoulder: the cover glass came down too rapidly; discard the mount and remake it.
  • Numerous air bubbles, or liquid across the moat: discard the mount; both defects are listed grounds for rejection.
  • Hazy refractometer reading on puree, pulp or paste: pulp is still present; filter a small portion through coarse filter paper or celluwipe and read the filtrate.
  • Field diameter will not reach 1.382 mm: change ocular power if the field is too small, or raise the oculars or fit a cut accessory diaphragm if the field is too large.

Interferences and Limitations

Plant remnants, tracheal tube thickenings, pieces of cell wall, plus lint and fabric segments can be mistaken for mould filaments by an untrained eye, and mould filaments themselves take diverse forms. Uneven distribution biases a count because insoluble material, and with it mould, can concentrate at the centre of a poorly made mount. Air bubbles reduce the sample actually present in a field, and a chamber seated without Newton’s rings holds the wrong volume. Results are percentages of fields, not counts of organisms, and they stand only for the six units examined, judged unit by unit against the tolerance.

Safety Notes

  • Use the Bunsen flame only for its stated tasks, dispersing foam, flaming the cell, spatula and needle, and keep flammable materials clear of it, as general laboratory practice.
  • Handle the Howard cell and cover glass with care: thin glass breaks easily and broken edges cut; dispose of damaged glass in a sharps container, as general laboratory practice.
  • Follow aseptic technique when sampling from bulk, both to protect the sample and to avoid contaminating product remaining in the bulk container.

Source Method

Source method is Health Canada Official Method MFO-5, Examination of Canned Tomatoes, Tomato Juice and Vegetable Juice, Tomato Puree, Tomato Paste, Tomato Pulp and Tomato Catsup for Mould Filaments, issued November 30, 1981, in Volume 1 of the Compendium of Analytical Methods.