SOP: Rapid Detection of Listeria spp. by the VIDAS Listeria Method (MFHPB-29)
GIFSQ ID: SOP-0953
Culture confirmation for Listeria takes days. An automated enzyme-linked fluorescent immunoassay (ELFA) shortens the screening stage: after a two-step enrichment, a boiled portion of broth goes into a VIDAS reagent strip and the instrument returns a fluorescence answer in about 45 minutes. Presumptive positives then move to cultural confirmation, while negatives can be reported far sooner than a full cultural work-up allows.
Inside the instrument, sample cycles in and out of a disposable Solid Phase Receptacle (SPR), a pipette-like device whose inner surface carries anti-Listeria antibodies. Antigen in the sample binds there, unbound material is washed away, an alkaline phosphatase conjugate binds to the captured antigen, and the substrate 4-methyl-umbelliferyl-phosphate is converted to fluorescent 4-methyl-umbelliferone, read at 450 nm.
Purpose
Screen foods and environmental samples rapidly for Listeria spp. so compliance with Sections 4 and 7 of the Food and Drugs Act can be assessed without waiting for full cultural isolation on every sample. A relative fluorescence value (RFV) at or above the cut-off marks a sample as presumptively positive and sends it forward for cultural confirmation and speciation.
Scope
Validation covers environmental samples and all foods except ready-to-eat meat and poultry, smoked fish, and yogurt, which sit outside the approved application. Other commodities may still be run, but fitness for purpose has to be established first by proper validation against the criteria in the Compendium of Analytical Methods, and those validation data should be forwarded to the Microbiological Methods Committee so the application can be widened. Findings are genus-level: a Listeria spp. detection cannot be used to claim that L. monocytogenes is absent, and every Listeria species identified must be reported.
Equipment and Materials
- VIDAS Listeria assay kit, Ref. 30 700 (bioMérieux), sufficient for 60 tests, including standards and controls, stored at 2 to 8°C
- Large-capacity VIDAS 30 or miniVIDAS instrument
- Palcam broth, pre-warmed to approximately 35°C before use
- UVM 2 broth for secondary enrichment
- Palcam agar and Oxford agar (or equivalent agars) for confirmation plating
- Optional chromogenic media: RAPID’L.mono plates (Bio-Rad) or Agar Listeria according to Ottaviani and Agosti (ALOA)
- Pipettor calibrated to dispense 500 µL, with disposable tips
- Stomacher and stomacher bags, blender or equivalent, and vortex mixer
- Water bath at 95 to 100°C (or equivalent heating system)
- Incubators holding 30°C and 35°C
- Refrigerated storage at 2 to 8°C for kits and for holding enrichment broths for confirmation
- Suitable surface disinfectant
Commercial media are prepared and sterilised as their manufacturers direct. Any variation from the listed media, commercial or in-house, carries an equivalency obligation for the analyst or laboratory supervisor.
Procedure
- Hold sample units refrigerated or frozen before analysis, according to the product; shelf-stable foods are excepted. Thaw frozen units in a refrigerator or under conditions that prevent growth or death, and analyse as soon as possible after receipt.
- Have sterile Palcam broth ready, pre-warmed to approximately 35°C, and disinfect the work surface.
- Build a representative analytical unit: mix liquids or free-flowing materials until homogeneous, and for solids take portions from several locations within the sample unit.
- Start the primary enrichment for foods: add 25 g or mL of food to 225 mL of Palcam broth in a blender jar or stomacher bag. For composites, keep the ratio at 1 part sample material to 9 parts Palcam broth, with analytical unit sizes set by the Listeria policy.
- For environmental samples: place a sponge or large swabs in 100 mL of Palcam broth; up to 10 sponges may be composited with 100 mL of broth per sponge. Put smaller swabs, such as cotton tips, in 10 mL portions of broth; up to 10 swabs may be composited with 10 mL of broth per swab.
- Blend, stomach or vortex as needed for thorough mixing, then incubate the primary enrichment at 35°C for 26 to 28 hours.
- Move to secondary enrichment: mix the Palcam broth enough to resuspend contents while letting large food particles settle, then transfer 1 mL into 9 mL of UVM 2. Incubate at 30°C for a minimum of 26 hours and up to 48 hours.
- Optional holding point: UVM 2 cultures may be refrigerated, which is how Friday cultures from Wednesday or Thursday samples are carried over the weekend. On the next working day, resuspend the refrigerated culture contents and continue with the next step.
- After incubation, transfer 2 mL of the UVM 2 broth to a tube and heat it for 15 ± 2 minutes in a water bath at 95 to 100°C. Let the tube cool. Keep the remaining enrichment broth at 2 to 8°C; it is the material used later to confirm any presumptive positive.
- Bring the VIDAS LIS kit out of the refrigerator and let the components needed reach room temperature, at least 30 minutes. Return unused components to 2 to 8°C storage. Never mix components from different lots and never use a kit past its expiry date.
- Label LIS reagent strips with the sample identification numbers in the space provided. Enter master lot, sample, standard and control data following the operating instructions for the instrument in use, whether a VIDAS 30 or a miniVIDAS.
- For a new kit lot, enter the factory master calibration curve first, using the master lot entry (MLE) card supplied with the kit.
- Calibrate the standard at least every 14 days and whenever a new lot enters service. Enter AS1@ as the sample identification for the standard, place it at the start of the work list, followed by controls C1 and C2. To store the calibration in memory, run the standard in duplicate, both positions carrying the same S1 designation.
- Vortex the standard, controls and boiled samples. Pipette 500 ± 50 µL of standard, control or boiled sample into the centre of the sample well of the labelled LIS reagent strip. Every enrichment broth must have had its full 15 minutes at 95 to 100°C and cooled to room temperature before it goes on the instrument; unheated broth is never tested.
- Load the LIS reagent strips and LIS SPRs into the positions the work list indicates, and check that the colour labels carrying the three-letter assay code match between each SPR and its strip.
- Start the assay as the VIDAS operator’s manual directs. Processing finishes in approximately 45 minutes.
- Read the result against the cut-off in the Recording Results section, and send every presumptive positive to cultural confirmation from the refrigerated UVM 2 broth held in step 9.
Recording Results
An RFV of 0.10 or higher indicates the presumptive presence of Listeria spp. in the test sample; below that value the sample is reported as not detected for Listeria spp. Because detection is at genus level, a not-detected result also means L. monocytogenes was not detected, since no member of the genus was found. The reverse logic does not hold: once Listeria spp. are detected, no amount of colony screening or extra confirmation can support a statement that L. monocytogenes is absent. Report every Listeria species identified during confirmation.
Quality Control
Controls C1 and C2 travel with each calibration and work list, positioned directly after the standard, so a reagent or instrument problem shows up before sample results are trusted. Calibration discipline is fixed: standard calibration at least every 14 days and at every lot change, with the master calibration curve loaded from the MLE card for each new lot. Components from different lots are never mixed. Kit storage is 2 to 8°C, with at least 30 minutes at room temperature before use. Incubator temperatures follow the standing pattern, 35 ± 1°C and 30 ± 1°C for the two enrichments. Analysts follow the kit package insert for assay detail beyond this SOP.
Troubleshooting
- Unexpected positives in clean product: verify the broth was fully boiled and cooled before loading, confirm strip and SPR assay codes matched, and resolve the result by cultural confirmation from the refrigerated UVM 2 rather than repeating the immunoassay alone.
- Standard or controls out of range: check kit expiry, lot mixing, storage at 2 to 8°C, and whether the 14-day calibration interval or a lot change was missed; re-enter the master lot data from the MLE card for a new lot.
- Weak enrichment growth suspected: confirm the Palcam broth was pre-warmed to about 35°C, the full 26 to 28 hour primary incubation was given, and the 1 mL into 9 mL transfer went into UVM 2, not back into Palcam.
- Weekend timing pressure: use the refrigerated hold on UVM 2 cultures rather than shortening either enrichment; resuspend the culture contents before boiling the test portion.
- Dispensing errors: keep the aliquot at 500 ± 50 µL, vortex boiled samples first, and place the volume in the centre of the sample well.
Interferences and Limitations
Approval is limited to the validated scope: ready-to-eat meat and poultry, smoked fish and yogurt are excluded, and any other unvalidated commodity needs its own validation before results are relied on. Findings are presumptive until cultured; the immunoassay detects genus-level antigen, including material from cells killed by the boiling step, and cannot speciate. A positive therefore always needs the plating and confirmation steps of a cultural method before species are reported.
Safety Notes
- Handle the 95 to 100°C water bath and freshly boiled broth tubes as burn hazards; allow cooling before pipetting.
- Treat all enrichment cultures as potentially containing L. monocytogenes, a serious pathogen: work with gloves, contain aerosols from blending and stomaching, and decontaminate spills promptly.
- Dispose of used SPRs, strips and cultures as infectious waste.
- Keep disinfectant practice routine on benches before and after each run.
Source Method
This SOP is an original-wording adaptation of Health Canada method MFHPB-29, Detection of Listeria spp. in Foods and Environmental Samples by the VIDAS Listeria Method, issued March 2012 in Volume 2 of the Compendium of Analytical Methods. Confirmation referred to in this SOP follows the cultural isolation approach of MFHPB-30, summarised in its own SOP.