SOP: Detecting E. coli O157:H7 in Raw Beef with the Atlas EG2 Assay (MFLP-03)

GIFSQ ID: SOP-0996

Raw beef carries a zero-tolerance problem, and enrichment plus molecular screening answers it faster than culture alone. Enriched broth feeds a fully automated instrument that captures target messenger RNA, amplifies it isothermally and reads a light signal, so a walk-away run screens many samples while the analyst keeps the matching enrichments warm for culture confirmation of anything flagged.

Sample size changes almost everything downstream here. Small and large beef portions enrich at different ratios and for different windows, and ground beef alone earns an extra heat step, so sorting samples correctly at the bench matters more than any later adjustment.

Purpose

Purpose is the rapid qualitative detection of Escherichia coli O157:H7, including non-motile strains that are genetically H7, in raw ground beef and raw beef trim. Every positive screen requires cultural confirmation.

Scope

Scope is limited to raw ground beef and raw beef trim, the matrices the assay was validated on. Kit design detects all E. coli O157:H7, including all H7 and non-motile strains, while excluding other O157 strains. Application to other foods needs prior validation against Compendium criteria before use. Keep sample units refrigerated or frozen by product type, thaw frozen units in a refrigerator or under conditions preventing growth or death, and analyse as soon as possible after receipt.

Equipment and Materials

  • Atlas E. coli O157:H7 EG2 Detection Assay Kit (PRT-01252) and Atlas System (PRT-00641)
  • Atlas EG2 Calibrator Kit (PRT-01253); Roka G2 Sample Transfer Tube Kit (PRT-01291)
  • Atlas Auto Detection Reagents (PRT-00261), comprising Auto Detect 1 and Auto Detect 2; Atlas Assay Fluids (PRT-01397); Wash Solution; Buffer for Deactivation Fluid; Oil Reagent
  • Modified tryptic soy broth with 20 mg/L novobiocin (mTSB-n), prepared and sterilised to manufacturer instructions
  • Incubator capable of maintaining 42°C; stomacher and stomacher bags with mesh filter
  • Pipettes capable of dispensing 0.22 to 1.5 mL; heating block with 13 mm diameter wells
  • Powder-free gloves; sterile cotton-tipped applicators for bubble removal

Procedure

  1. Disinfect the work surface and have sterile mTSB-n ready, pre-warmed to 35 ± 2°C. Build each analytical unit from portions taken at several locations within the sample unit.
  2. Weigh the sample into a stomacher bag with mesh filter by size band: 25–65 g of raw ground beef, 25–65 g of raw beef trim, or 66–375 g of raw beef trim.
  3. Homogenise in pre-warmed mTSB-n for 2 minutes in the stomacher, holding the ratio each band demands. Portions of 25–65 g take 1 portion sample to 9 portions broth, for example 25 g in 225 mL. Portions of 66–375 g take 1 portion sample to 4 portions broth, for example 375 g in 1500 mL.
  4. Incubate every enrichment at 42 ± 0.5°C, with the window set by matrix and size: 25–65 g raw ground beef for 10–22 h; 25–65 g raw beef trim for 8–22 h; 66–375 g raw beef trim for 8–18 h.
  5. Sample the enrichment for the instrument: transfer 1.2 mL of broth, avoiding food particles, into a Roka G2 Sample Transfer Tube containing lysis reagent.
  6. Apply the heat step to ground beef without exception: place its transfer tube in a heating block pre-warmed to 95°C for 10 minutes. Trim samples get the same treatment whenever homogenised trim resembles homogenised ground beef in consistency during sampling or preparation; other trim tubes skip heating and move on.
  7. Load all transfer tubes onto the fully automated Atlas instrument. Return the enrichment bags to a 42°C incubator and keep incubating to the maximum time for their band, so broth remains available for confirmation. Broth can support confirmation once it has incubated at least 8 h for raw beef trim or 10 h for raw ground beef. After the maximum time is reached, enrichments may be held refrigerated for a maximum of 48 h for confirmation use.
  8. Reconstitute reagents in a clean area before any work on the Atlas System, in this order: Probe Reagent first, then Amplification and Enzyme Reagents. Put on a fresh pair of powder-free gloves. Pair each reconstitution solution with its lyophilised reagent by matching label colours, and check lot numbers against the Master Lot Barcode Sheet.
  9. Perform each reconstitution the same way: open the lyophilised vial and firmly seat the notched end of the reconstitution collar in the vial opening; open the matching solution bottle and rest its cap on a clean, covered surface; hold the bottle on the bench and firmly seat the collar’s other end in the bottle opening; invert slowly and let solution drain into the vial; swirl gently until the lyophilised reagent has completely dissolved; invert again at a 45° angle to minimise foaming and drain everything back into the bottle; run two further inversion cycles; then remove and discard collar and vial into non-hazardous waste and recap the bottle.
  10. Mix Amplification, Enzyme and Probe Reagents by gentle inversion just before loading, avoiding foam, because foam compromises level sensing on the Atlas System. Where bubbles form in a reagent centre, remove them with a sterile cotton-tipped applicator.
  11. Load reagents and transfer tubes on the Atlas instrument and start the program. Inside the run, target capture concentrates and purifies target mRNA on magnetic particles, transcription-mediated amplification copies the target isothermally using RNA polymerase and reverse transcriptase, and the hybridisation protection assay reads chemiluminescence from an acridinium ester-labelled probe. A built-in Internal Control guards against false negatives from reaction failure.

Recording Results

Instrument software reports each sample directly as Positive, Negative or Invalid, where Invalid means the Internal Control failed or a mechanical issue occurred. Retest an invalid sample from a fresh aliquot of enrichment broth; repeated invalid results are treated as presumptive positive and go forward to confirmation. Confirm every presumptive positive culturally by the MFHPB-10 route: immunomagnetic separation from enrichment broth incubated at least 8 h for beef trim or 10 h for ground beef, followed by inoculation onto at least two selective agars as that cultural method specifies. Where confirmation fails from the early broth, repeat it with the 18 h or 22 h enrichment broth retained under the holding rules above. Report screening and confirmation outcomes separately, with sample size band, ratio and incubation window recorded for each unit.

Quality Control

Calibrators from the Atlas EG2 Calibrator Kit and the per-reaction Internal Control form the run’s control backbone; an Internal Control failure invalidates that sample rather than letting a silent negative stand. Lot pairing is verified against the Master Lot Barcode Sheet before reconstitution, and reagent handling, glove change and clean-area reconstitution protect against carryover. Incubation at 42°C is a higher-temperature step and must be held within 0.5°C; more generally, steps at or below 37°C follow the 35 ± 1.0°C style tolerance pattern stated in the method notes. Work proceeds under ISO/IEC 17025 competence requirements, and media variations require demonstrated equivalency.

Troubleshooting

  • Invalid result on first run: take a new aliquot of enrichment broth and retest; only repeated invalids escalate to presumptive positive.
  • Foam in a reagent bottle: stop and clear bubbles with a sterile cotton-tipped applicator before loading, since foam defeats level sensing.
  • Screen positive, culture silent at the early time point: repeat confirmation from the 18 h or 22 h broth before closing the sample out.
  • Trim sample looked like ground beef: apply the 95°C, 10 minute heat step; consistency, not the label alone, decides that branch.
  • Reagent pairing in doubt: recheck label colours and the Master Lot Barcode Sheet before reconstitution, never after loading.

Interferences and Limitations

Validation covers raw ground beef and raw beef trim only, so results on other matrices carry no standing until validated. Food particles carried into the transfer tube can compromise the run, which is why the aliquot step avoids them deliberately. A screen result, positive or negative, never substitutes for culture where confirmation is required, and detection excludes non-H7 O157 strains by design.

Safety Notes

  • Treat enrichments as potentially carrying E. coli O157:H7, a pathogen with a low infective dose; use strict enteric precautions and biohazard disposal throughout.
  • Take care at the 95°C heating block and with hot tubes; allow tubes to be handled safely after the heat step.
  • Follow kit and instrument safety guidance for assay fluids, deactivation buffer and oil reagent, and dispose of used transfer tubes as biohazardous waste.

Source Method

Adapted in original wording from Health Canada method MFLP-03, The Atlas® E. coli O157:H7 EG2 Detection Assay for detection of Escherichia coli O157:H7 in raw ground beef and raw beef trim, February 2018, Compendium of Analytical Methods, Volume 3. This revision replaces MFLP-03 dated June 2016.