SOP: Counting Yeasts and Moulds in Foods with the TEMPO YM System (MFLP-108)
GIFSQ ID: SOP-1008
Yeasts and moulds spoil quietly and grow slowly, so a count that finishes in days rather than a week changes how a lab can act. TEMPO YM replaces a manual most probable number series with a sealed card of miniature wells: fungi growing in those wells alter a substrate, fluorescence appears, and an automated reader converts a pattern of positive wells back into a count for the original food.
Patience still matters here. Incubation runs a full three days at a cool temperature, diluent choice changes for high-sugar foods, and strongly coloured matrices need extra dilution before fluorescence can be read reliably.
Purpose
Purpose is to enumerate viable yeasts and moulds in foods, for checking compliance with Sections 4 and 7 of the Food and Drugs Act and other relevant federal requirements, with results available in 72–76 hours.
Scope
Scope covers foods generally, validated for all foods except strongly coloured foods. Strong colour does not exclude a food outright in practice, but such matrices need the higher dilution described in the procedure so colour does not mask the fluorescent signal. Work under ISO/IEC 17025 competence requirements, and prove equivalency before substituting any medium formulation, whether bought ready-made or prepared in-house.
Equipment and Materials
- TEMPO system: TEMPO Filler, TEMPO Reader and all components needed for its intended use
- TEMPO YM assay kit, reference 80001: cards and culture medium in vials, sufficient for 48 test samples
- TEMPO Filter Bags, reference 80015; TEMPO Control Kit, reference 80000
- Pipettes dispensing 0.10 mL or 1.0 mL with disposable tips; vortex mixer; stomacher, blender or equivalent
- Incubator capable of maintaining 25°C
- 0.1% peptone water; for high-sugar foods with aw 0.2–0.8, 0.1% peptone water with 20% sucrose
- Sterile deionized or distilled water for reconstituting and diluting in the vial
Procedure
- Hold sample units refrigerated or frozen according to product nature until analysis, except for shelf-stable foods. Thaw frozen units in a refrigerator, or under conditions that prevent both growth and death, and analyse as soon as possible after receipt.
- Have primary and secondary sterile diluent ready and pre-warmed to room temperature. Disinfect the working surface before starting.
- Make a representative unit: mix liquids and free-flowing materials until homogeneous; for solids, take portions from several locations within the unit.
- Prepare a 1 in 10 primary dilution in a TEMPO Bag: aseptically add 25 g or mL to 225 mL of 0.1% peptone water, using 0.1% peptone water with 20% sucrose for high-sugar foods at aw 0.2–0.8. Shake, stomach or blend as the food requires. A smaller unit is acceptable provided the 1 in 10 ratio is kept, for example 11 g in 99 mL or 10 g in 90 mL.
- Blend or stomach only long enough to give a homogeneous suspension, never longer than 2 min, to avoid overheating. With foaming foods, run a blender at low speed and draw the aliquot from below the liquid/foam interface. Transfer into the TEMPO card within 45 minutes of homogenizing the primary dilution; that interval must not be exceeded.
- Set dispensers holding secondary diluent to the correct volumes and prime the pump by discarding the first two volumes dispensed. Log on to the TEMPO preparation station and reconstitute each TEMPO vial of culture medium with the appropriate volume of secondary diluent from the dispenser.
- Choose the vial dilution by the count expected. For a 1 in 40 dilution, enumerating between 10 and 4.9 × 104 CFU/g, transfer 1 mL of primary dilution into one TEMPO YM vial containing 3.0 mL sterile deionized or distilled water. For a 1 in 400 dilution, enumerating between 100 and 4.9 × 105 CFU/g, transfer 0.1 mL of primary dilution into one vial containing 3.9 mL sterile deionized or distilled water.
- Apply two minimum-dilution rules in place of the standard choice where they apply. Strongly coloured primary dilutions, such as fruit purées and cocoa, need at least 1 in 400 because colour can affect the fluorescent signal. Aromatic herbs, spices, teas and herbal teas, which may inhibit growth, likewise need a minimum dilution of 1 in 400.
- Where a higher range is needed, prepare serial dilutions from the primary dilution to reach 1 in 4,000, 1 in 40,000, 1 in 400,000 and so on, following manufacturer instructions.
- Agitate each dilution to resuspend material that settled during preparation. With a sterile pipette, transfer the appropriate volume from the TEMPO Bag to its corresponding vial with reconstituted medium, then vortex or agitate manually for about 5 seconds.
- Identify each sample through the preparation station interface. Take one TEMPO YM card, scan its information into the system as prompted, scan the corresponding vial, insert the vial fill tube into that vial, and place card and vial together in a filler rack. Repeat scanning and racking for every prepared vial.
- Fill the cards with the TEMPO Filler. After the cycle, visually confirm vials are empty and discard used vials in an appropriate receptacle. Move filled cards from the filler rack into a TEMPO incubation rack.
- Incubate cards at 25 ± 1°C for 72–76 hours, holding that ±1°C tolerance strictly. Incubation timing is managed by the TEMPO Read software, which assumes a theoretical 15 minutes between reading the card barcode and the start of incubation. If the real interval is longer than 15 minutes, without exceeding 2 hours, add the extra time to the remaining incubation time shown by the software, and always read within the time limit the software authorizes.
- Remove the incubation rack, log on to the reading station, and place the rack in the TEMPO Reader. After reading, remove cards and discard them in an appropriate receptacle.
Recording Results
Reader software scans each card barcode, interprets fluorescence across the wells, and links sample identity, test type and dilution to a result expressed as colony forming units (CFU) per gram or millilitre of the original sample. Card design explains how that conversion works: each card holds 48 wells in 3 sets of 16, small, medium and large, with a one-log volume difference between sets, simulating a most probable number (MPN) determination. Record the dilution actually used, especially a 1 in 400 or higher dilution for coloured or inhibitory matrices, alongside the reported CFU/g or CFU/mL, the incubation period and any timing adjustment made for a delayed start.
Quality Control
Run the TEMPO Control Kit, reference 80000, as the system directs and keep control results with the batch record. Temperature control is part of validity: for growth steps at or below 37°C tolerance is ±1.0°C, while higher recommended temperatures must be held within 0.5°C. Media prepared or substituted in-house need documented equivalency, with data forwarded to the Editor of the Compendium for consideration of a method modification. As generic lab practice, verify dispenser volumes after priming and reject any card that did not fill completely.
Troubleshooting
- Weak or unreadable fluorescence in a coloured food: repeat from a dilution of at least 1 in 400; fruit purée and cocoa colour can suppress the signal at 1 in 10 or 1 in 40.
- Counts unexpectedly low for herbs, spices or teas: check whether the minimum 1 in 400 dilution for inhibitory matrices was applied.
- Vial not empty after filling: do not report that card; inspect the fill tube and filler cycle, then prepare a fresh vial and card within the 45-minute window or remake the primary dilution.
- Reading window missed: follow the software-authorized time limit; a delayed start of up to 2 hours is corrected by adding the extra time, beyond that a repeat is safer than a late read.
- Foam carried into the aliquot: resample from below the foam interface at low blender speed, as foam skews the volume transferred.
Interferences and Limitations
Validation excludes strongly coloured foods, and colour remains the chief interference even with extra dilution. Inhibitory compounds in aromatic herbs, spices, teas and herbal teas can depress recovery at low dilutions. High-sugar matrices need the sucrose-containing diluent to protect cells osmotically; using plain peptone water there risks undercounting. Results depend on the dilution chosen: a 1 in 40 preparation cannot report below 10 CFU/g, and a 1 in 400 preparation cannot report below 100 CFU/g.
Safety Notes
- Treat all food homogenates as potentially carrying spoilage and pathogenic fungi; use gloves, eye protection and biohazard disposal for cards and vials.
- Handle blender jars and stomacher bags as aerosol sources, opening them only after contents settle.
- Disinfect benches before and after work, and decontaminate spills of primary dilution promptly as generic lab practice.
Source Method
Adapted in original wording from Health Canada method MFLP-108, Enumeration of Yeasts and Molds in a Variety of Foods using the TEMPO YM (Yeasts/Molds Count) method, October 2018, Compendium of Analytical Methods, Volume 3.