SOP: Checking Canned Foods for Commercial Sterility and Living Microbes (MFHPB-01)
GIFSQ ID: SOP-0934
Commercial sterility is a pass-or-fail promise: a sealed food should carry no living organism, spore included, that could grow at temperatures the product will meet in storage and distribution. Work in this SOP tests that promise directly. Aseptically taken portions go into enrichments and onto plates, and any growth is then characterised by oxygen need, temperature preference, population mix and cell shape.
Cans, jars, retort pouches and similar hermetically sealed packs all fall within scope. Approach every swollen unit as a botulism risk until evidence says otherwise, and treat cleanliness of air, hands and container surfaces as part of the test itself, because a contaminant introduced at opening can look exactly like a process failure.
Purpose
Detect viable microorganisms in commercially sterile foods in hermetically sealed containers, decide whether commercial sterility has been achieved, and, where growth occurs, describe isolates as mixed or pure, aerobic, anaerobic or facultative, mesophilic or thermophilic, and coccal, bacillary or spiral in form.
Scope
Use for shelf-stable canned and similarly sealed foods. A low-acid food has any component at pH above 4.6 with water activity above 0.85, water activity being food vapour pressure divided by pure-water vapour pressure at matching temperature and pressure. Acid and acidified foods sit at pH 4.6 or below. Refrigeration means 4°C or less without freezing. Commercial sterility wording here does not cover low-acid foods labelled for refrigeration or freezing, nor tomato products at pH 4.7 or less after processing. Where normal holding temperature exceeds 40°C, test for thermophiles; otherwise test for mesophiles only. Lots linked to under-processing, illness or visible swelling get both mesophile and thermophile work.
Equipment and Materials
- Laminar-flow or Class II biological safety cabinet in a clean environment: no open windows and no direct draught across the work area; October 2001 supplement condition, applied in full here. Laminar-flow particle count must stay at or below 100 particles of 0.5 µm or larger per cubic foot (Class 100). A Class 10,000 clean room is acceptable, and a Class 100,000 room has also proved workable only with added air-quality checks during every operation.
- Clean smock with protective sleeves and tight cuffs, surgical gloves, hair cap, and mask covering mouth, nose and facial hair; disinfect gloved hands as needed. Sterile gowns, caps, gloves and masks, or equivalent, may be worn.
- Sanitary can openers such as Bacti-Disc cutters, sterile screw-cap tubes 20 × 150 mm, sterile cotton-plugged transfer pipettes, sterile spoons, spatulas, probes, pliers and cork borers, sterile loops, anaerobe jars, sterile examination trays and sealable sterile containers for retained food.
- Chlorine sanitising solution and a 2–10 L soaking vessel, chlorine-indicating tape or test papers, optional 70% alcohol, pH meter or paper resolving 0.3–0.5 pH unit across pH 6.0–7.5, Gram stain reagents, light microscope, diamond-point pen, sterile plastic bags for swollen units.
- Incubators at 30°C, 35°C and 55°C; water baths at 80°C and 100°C. Tolerance is ±1.0°C at 35°C, 30°C and 25°C; hold 43°C or 45.5°C settings within 0.5°C because excess heat can kill target organisms.
- Low-acid foods: PE-2 and Cooked Meat Medium (CMM) for enrichment; Plate Count Agar (PCA), Tryptic Soy Agar (TSA) or Blood Agar (BA) for plating, plus Reinforced Clostridial Medium (RCM) or Liver Veal Agar (LVA) for anaerobes; Baird-Parker or MacConkey agar optionally where a specific organism is suspected.
- Acid and acidified foods: Acid Broth (AB) for enrichment; Potato Dextrose Agar (PDA), Sabouraud Dextrose Agar (SDA), Thermoacidurans Agar (TA) and Orange Serum Agar (OSA, with filter aid) for plating. Adjust broth and agar pH to match food pH where needed; measure at least one container first when food pH is unknown.
- Control strains or equivalents: Clostridium sporogenes ATCC 7955 (NFPA 3679) or C. histolyticum as anaerobic mesophilic sporeformer; Bacillus coagulans ATCC 8038 (NFPA 43P) for aerobic growth at 30°C; Bacillus stearothermophilus ATCC 12016 (NFPA 2184) as obligate aerobic thermophile; Clostridium thermosaccharolyticum ATCC 7956 as obligate anaerobic thermophile at 55°C.
- Key media notes: AB contains proteose peptone 5 g, yeast extract 5 g, dextrose 5 g and K2HPO4 4 g per litre, pH adjusted to food pH with 1N HCl, sterilised at 121°C for 20 minutes. CMM tubes hold 2.5 g dehydrated medium plus 20 mL cold distilled water, sterilised 15 minutes at 121°C, final pH 7.2. PE-2 basal medium holds peptone 20 g, yeast extract 3 g and 2 mL of 2% alcoholic bromocresol purple per litre, dispensed 19 mL over 8–10 untreated Alaska seed peas per tube, sterilised 30 minutes at 121°C. CMM and PE-2 keep refrigerated at 0–4°C for up to 4 weeks; deaerate refrigerated medium at 100°C for 10 minutes and cool before use.
Procedure
- Keep swollen, leaking or illness-linked units contained during storage and transport, ideally in sealable plastic bags inside watertight metal or plastic bins, refrigerated. Transfer contents of already-open containers aseptically to sterile containers whenever possible.
- Pre-incubate aseptically processed and packaged units collected less than 30 days after packaging for 7 days at 30–35°C before taking an analytical unit, unless an illness investigation or obvious loss of sterility makes waiting inappropriate.
- Mark each unit directly on its container with a diamond pen or equivalent so identity survives handling, and record its full code. Place coincident marks on label and container in indelible ink, carborundum or diamond point, remove the label carefully and intact where a printed pack allows it, and mark sample and unit numbers on that label.
- Scrub dirty containers with detergent and water. Disinfect by complete immersion in buffered chlorine solution, or by flooding the end to be opened, for at least 20 minutes. Buffered chlorine is made by diluting a 1.0 M KH2PO4 stock buffer (136 g per litre, adjusted to pH 6.2 with 1N NaOH) twenty-fold, then adding 10 mL commercial chlorine solution at 5.25% available chlorine per litre, giving 500–525 ppm available chlorine at final pH 6.5. Check free chlorine periodically; remake solution once it falls below 100 ppm. An alternative is flooding or spreading with 2% peracetic acid in a wetting agent such as 0.1% polysorbate 80 for 5 minutes. Lift units out with sterile tongs or gloves.
- Set disinfected units, end to be opened uppermost (usually the uncoded end), on a disinfected surface under a running laminar-flow hood. Confirm the surface is fully dry before piercing; air-dry under flow or use sterile disposable tissues.
- Open normal units aseptically. For liquids and semi-liquids, mix first by turning the unopened container end over end at least 20–30 times; semi-liquids pour with difficulty but mix by shaking, while semi-solids are highly viscous and cannot be mixed that way. Place coded end down and, with a sterile or disinfected opener, cut a hole ending ¼ inch (6.35 mm) from the double seam, large enough for sampling and later micro-leak vacuum testing. Between units, strip adhering product from an opener, immerse its tip in 70% aqueous ethanol and flame piercing and cutting edges, then cool under the hood; or soak it in disinfecting solution for at least 20 minutes and air-dry or flame-dry.
- Open swollen units inside a sterile plastic bag held firmly around the opener shaft, and keep the unit inside until venting finishes completely. Pouch products suit this route especially well. Then cut the sampling hole as for normal units.
- Sample promptly. Aseptically move at least 50 g of food into a sealable sterile container, label it, and refrigerate it protected from contamination as a retained portion for possible toxin, chemical or repeat testing, remembering refrigeration can shift bacterial numbers. Take 1–2 mL or g per liquid medium. For direct plating, streak each plate with at least one loopful, about 0.01 mL; pour plates may also be used. Sample solids from both centre, by sterile cork-borer core where under-processing is a concern, and outer surface near seams or closure, where leakage is a concern, scraping with a sterile spatula after removing contents to a sterile tray if needed. Finish any planned vacuum leak test before removing a whole can end.
- Run mesophile work for low-acid foods: inoculate 2 tubes PE-2 and 2 tubes CMM, with positive and negative media controls, and incubate aerobically at 30–35°C for at least 14 days, examining daily; a single set temperature may be chosen inside that range provided it holds within ±1.0°C. From every positive or suspect tube, streak duplicate PCA, TSA or BA plates, incubating one aerobically and one anaerobically at 35°C for 2–5 days with daily checks; RCM or LVA may carry the anaerobic side. Direct-plate two PCA plates (RCM or LVA for anaerobes), one aerobic and one anaerobic at 35°C for 2–5 days. Record Gram reaction and cell morphology for every colony type.
- Run mesophile work for acid and acidified foods: inoculate 2 tubes AB; re-sterilised normal product itself may serve as medium, alone or in addition, for fluid or semi-fluid foods, adding recently boiled distilled water to solids to reach a workable consistency. Incubate one set aerobically and one anaerobically at 30–35°C for at least 14 days, checking at 3, 7 and 14 days, and use smears plus streak plates because growth can be hard to see in food-based medium. Prepare duplicate streak PDA plates and duplicate pour TA plates, incubated aerobically and anaerobically at 30°C for 2–5 days with daily checks. Direct-plate two PDA and two TA plates, one of each atmosphere, at 35°C for 2–5 days. Record Gram reaction and morphology per colony type.
- Run thermophile work where required. For low-acid foods, inoculate 2 tubes PE-2 and 2 tubes CMM, heat-shock one tube of each at 80°C for 10 minutes timed only after the medium itself reaches 80°C, then incubate all tubes at 55°C for up to 7 days with daily checks; handle positive tubes as in mesophile work except plates go at 55°C. Direct-plate two PCA plates (RCM or LVA for anaerobes), one aerobic and one anaerobic, at 55°C for 2–5 days. For acid foods, use 2 tubes AB or re-sterilised product medium in the same heat-shock pattern, and direct-plate two PDA aerobically and two TA anaerobically at 55°C for 2–5 days. For shelf-stable milk and milk products, add 1 mL re-sterilised product per litre of tempered medium before pouring plates, to aid recovery.
- After microbiological sampling, prepare smears for direct microscopic examination, measure pH as soon as practical because growth often shifts pH, and record odour and appearance such as frothing, curdling or discolouration without tasting anything.
- Identify isolates in detail only where circumstances require it, for example compliance action or recall; in such cases identify organisms from positive tubes and consider sending cultures to a reference laboratory.
- Autoclave container and contents before disposal whenever non-sterility is suspected, and always where C. botulinum is suspected.
Recording Results
Log each sample unit separately: full container code, media inoculated, atmosphere, temperature and duration, daily growth observations, and, for every colony type, Gram reaction and cell morphology. State whether growth was aerobic, anaerobic or both, mesophilic or thermophilic, and mixed or pure. A unit supports commercial sterility only when its required enrichments and direct plates show no viable growth under the conditions run; any confirmed growth is reported as presence of viable microorganisms, with identification added where performed.
Quality Control
Challenge-test laminar-flow hoods with a dioctyl phthalate (DOP) test or equivalent and measure air velocity at least yearly, recording results; certify biological safety cabinets at least yearly. Check clean-room air with a slit-type sampler or one-hour exposed plates: at least once per operation, or weekly in continuous use, for Class 10,000 rooms, and continuously during every operation for Class 100,000 rooms. Prepare a negative food control, when interpretation may be unclear or earlier results conflicted, by autoclaving a unit or aliquot at 121°C for 15 minutes and inoculating 2 tubes PE-2 and 2 tubes CMM with the sterile food. Run media negative controls by pre-incubating medium for at least two days, or one unit per batch per sample, and positive controls at one unit per batch: C. sporogenes or C. histolyticum in PE-2, CMM and LVA at 35°C; B. coagulans in Thermoacidurans Agar, Acid Broth and PDA at 30°C; B. stearothermophilus in PE-2, CMM and PCA at 55°C; C. thermosaccharolyticum in PE-2, CMM and LVA at 55°C. Record all control outcomes.
Troubleshooting
- Growth in controls: a positive negative-control or sterile-food control points to contamination during opening or inoculation; discard that run’s sterility conclusion and repeat from a fresh unit.
- No growth from a swollen unit: check pH, smear findings and odour before concluding; gas producers may have died off, so retain the portion for toxin or chemical follow-up.
- Cloudy food-based enrichment: judge growth by smears and subculture plates rather than turbidity alone, as directed for acid foods.
- Weak recovery from milk products: confirm 1 mL re-sterilised product per litre was added to tempered plating medium.
- Opener carry-over suspected: re-disinfect by the 70% ethanol-and-flame route or a full 20-minute disinfectant soak and cool fully before reuse.
Interferences and Limitations
Localised contamination in solid packs can be missed if only one zone is sampled, hence centre-plus-surface sampling. Media pH mismatched to an acid food can suppress recovery, hence pH adjustment. Refrigerated retained portions change with time, so repeat results from them need cautious interpretation. Routine work establishes presence or absence and broad character; species-level identification is an added step, not an automatic output.
Safety Notes
- Handle all swollen containers as if they held Clostridium botulinum or its toxin until absence is established.
- Never taste food from a container under investigation, and always use mechanical pipetting devices.
- Expect hard swells to spray on puncture; work in a Class II biological safety cabinet and use bag containment while venting.
- Take chemical-disinfectant precautions for chlorine and peracetic acid, and autoclave suspect material before disposal.
Source Method
Adapted in original wording from Health Canada Health Products and Food Branch method MFHPB-01, Determination of Commercial Sterility and the Presence of Viable Microorganisms in Canned Foods, issued March 2001, incorporating its Supplement issued October 2001.