UV-Vis Quiz

GIFSQ ID: ART-1453

Test your knowledge with this free 20-question UV-Vis quiz from Food Analysis & Analytical Chemistry, part of…

What this quiz covers

UV-Vis spectrophotometry is one of the oldest quantitative tools in the food laboratory, and it still rewards the analyst who understands its simple physics. The Beer-Lambert law states that absorbance is directly proportional to concentration and path length, A equals epsilon b c, so doubling the concentration or the cuvette path length doubles the absorbance within the linear range. That is why samples are diluted so their absorbance falls roughly between 0.1 and 1.0: below 0.1 noise dominates, and above 1.0 stray light and detector limits cause curvature. The wavelength of maximum absorbance is chosen because it gives maximum sensitivity and the smallest error from wavelength drift, since the peak is flat there. A calibration curve relates measured absorbance to concentration using standards, and a reagent blank zeroes the instrument against everything except the analyte.

A deuterium lamp covers the ultraviolet region and a tungsten or halogen lamp covers the visible, the monochromator selects a narrow wavelength band from the source, and a double beam design compares the sample beam with a reference beam to cancel lamp drift. Quartz cuvettes are required below about 350 nm because glass and most plastics absorb ultraviolet light there, and fingerprints on cuvette windows must be avoided because they scatter and absorb the beam, adding an error no blanking can correct. Turbidity is the same problem in the sample itself: suspended particles scatter light away from the detector, and the instrument reads the lost light as absorbance.

The blank sets the zero

Phosphate is measured as a yellow vanadomolybdate complex or molybdenum blue, iron with 1,10-phenanthroline as the orange red ferroin complex, and nitrite by the Griess reaction as a pink azo dye near 540 nm. NADH is followed at 340 nm because NADH absorbs there while NAD+ does not, and carotenoids are estimated in the visible near 450 nm because their long conjugated systems absorb strongly there. When two analytes absorb at the same wavelength, simultaneous equations at two wavelengths or derivative analysis still separate them, and derivative spectroscopy sharpens overlapping bands while suppressing baseline shifts. At high concentrations the law itself bends, because solute molecules interact, the refractive index shifts, and stray light flattens the response.

How to use this quiz

Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the principles and methods a food analyst is expected to know by heart.

UV-Vis Quiz
Uncategorized

Choose how you want to play

Sound effects
🔈 Volume
Show explanations
🔊 Auto-read every question
Reading speed

A short note: UV-Vis spectrophotometry

Ultraviolet visible spectrophotometry, UV-Vis, turns colour into concentration: the Beer-Lambert law ties absorbance to concentration times path length, so doubling either doubles the absorbance. Samples are diluted so absorbance falls roughly between 0.1 and 1.0, where noise stays low.

Two lamps share the work: deuterium for ultraviolet, tungsten-halogen for visible. The monochromator picks a narrow band, and lambda max is chosen for maximum sensitivity with the smallest error from drift. Quartz cuvettes are required below about 350 nanometres, since glass and plastic absorb ultraviolet.

Practice decides whether the number is trusted. A reagent blank, everything except the analyte, zeroes the background, and standards build a calibration curve the sample is read off. Smudged windows and turbid samples scatter light and give falsely high readings, so samples must be clear, filtered or centrifuged. Stray light makes absorbance read falsely low at high absorbance, and molecular interactions bend the law at high concentrations too.

Food labs use UV-Vis through colour forming reactions: phosphate as vanadomolybdate yellow or molybdenum blue, iron as orange red phenanthroline near 510, nitrite by the Griess reaction as a pink azo dye near 540. NADH is followed at 340, where NAD plus does not absorb, and carotenoids absorb strongly near 450. Two analytes sharing one wavelength are resolved by simultaneous equations or derivative analysis, and a double beam instrument cancels lamp drift by comparing sample and reference beams.

Practical takeaway: keep the sample clear, the cuvettes clean, and absorbance between 0.1 and 1.0.

Questions and answers

Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.

Question 1: Phosphorus in foods is commonly determined spectrophotometrically as:

  • A. A red iron complex
  • B. Phosphorus in foods is determined spectrophotometrically
  • C. A yellow vanadomolybdate complex or molybdenum blue
  • D. A green copper complex

Correct answer: A yellow vanadomolybdate complex or molybdenum blue

Phosphate reacts with ammonium molybdate to form a yellow vanadomolybdophosphoric acid complex or, with a reducing agent, the intensely blue molybdenum-blue complex. Both colored complexes absorb in the visible range in proportion to phosphorus concentration, enabling Beer-Lambert quantification. These molybdate-based methods are the classical spectrophotometric approaches for phosphorus in foods described in standard quantitative analysis texts such as Harris, Quantitative Chemical Analysis.

Question 2: The Beer-Lambert law states that absorbance is:

  • A. Directly proportional to concentration and path length (A = εbc)
  • B. The Beer-Lambert law states that absorbance is inversely proportional to concentration
  • C. The Beer-Lambert law states that absorbance is independent of wavelength
  • D. The Beer-Lambert law states that absorbance is equal to transmittance

Correct answer: Directly proportional to concentration and path length (A = εbc)

The Beer-Lambert law states that absorbance equals the molar absorptivity times concentration times path length (A = εbc). Doubling the concentration or the cuvette path length therefore doubles the absorbance, within the linear range. The law is derived and applied in standard instrumental analysis texts such as Skoog, Holler and Crouch, Principles of Instrumental Analysis.

Question 3: The monochromator in a UV-Vis spectrophotometer functions to:

  • A. The monochromator in a UV-Vis spectrophotometer functions to produce white light
  • B. The monochromator in a UV-Vis spectrophotometer functions to hold the cuvette
  • C. Select a narrow wavelength band from the source
  • D. The monochromator in a UV-Vis spectrophotometer functions to amplify the signal

Correct answer: Select a narrow wavelength band from the source

The light source emits a broad spectrum, but quantitative analysis needs nearly monochromatic light at the chosen wavelength. The monochromator, a grating or prism plus slits, isolates a narrow band, typically a few nanometers wide, from that spectrum. Its function and the effect of spectral bandwidth are covered in Skoog, Holler and Crouch, Principles of Instrumental Analysis.

Question 4: Enzyme activity assays often follow NADH formation or consumption at 340 nm because NADH:

  • A. Is colored blue
  • B. Precipitates proteins
  • C. Absorbs at 340 nm while NAD⁺ does not
  • D. Emits fluorescence only

Correct answer: Absorbs at 340 nm while NAD⁺ does not

Reduced NADH has an absorption maximum at 340 nm due to its dihydronicotinamide ring, while the oxidized form NAD+ does not absorb there. Enzyme assays therefore track reaction progress by the rise or fall of absorbance at 340 nm. This 340 nm NADH assay principle is standard biochemistry, described in texts such as Berg, Tymoczko and Stryer, Biochemistry.

Question 5: Fingerprints or smudges on cuvette windows should be avoided because they:

  • A. They change the wavelength
  • B. Scatter and absorb light, causing noisy, inaccurate readings
  • C. They improve sensitivity
  • D. They calibrate the instrument

Correct answer: Scatter and absorb light, causing noisy, inaccurate readings

Fingerprints and smudges on the optical windows scatter and absorb the beam, adding a variable, wavelength-dependent error to every reading. The result is noisy, drifting absorbance values that cannot be corrected by blanking. Good spectrophotometric practice, as described in Harris, Quantitative Chemical Analysis, requires clean, scratch-free cuvette windows.

Question 6: When two analytes absorb at the same wavelength, UV-Vis quantification can still work by:

  • A. Using simultaneous equations at two wavelengths or derivative/multicomponent analysis
  • B. Measuring at only one wavelength anyway regardless of overlap
  • C. Diluting until one of the analytes disappears completely
  • D. Changing the cuvette path length instead of diluting further

Correct answer: Using simultaneous equations at two wavelengths or derivative/multicomponent analysis

When two analytes absorb at the same wavelength, measuring at two different wavelengths gives two equations in two unknowns, solvable for both concentrations if their spectra differ. Alternatively, derivative spectroscopy or full-spectrum multicomponent analysis separates overlapping bands mathematically. These approaches are standard in Skoog, Holler and Crouch, Principles of Instrumental Analysis.

Question 7: A blank (reagent blank) is measured in UV-Vis analysis to:

  • A. To increase the absorbance
  • B. To calibrate the wavelength scale
  • C. To dilute the sample
  • D. Zero the instrument against everything except the analyte

Correct answer: Zero the instrument against everything except the analyte

The reagent blank contains every reagent and the solvent but no analyte, so measuring it captures the background absorbance of the chemistry itself. Subtracting or zeroing against the blank leaves only the absorbance due to the analyte. Blank correction is fundamental quantitative practice described in Harris, Quantitative Chemical Analysis.

Question 8: Stray light in a spectrophotometer is best described as:

  • A. Stray light in a spectrophotometer is best described as the desired monochromatic beam
  • B. Stray light in a spectrophotometer is best described as the blank signal
  • C. Stray light in a spectrophotometer is best described as fluorescence from the sample
  • D. Unwanted light of other wavelengths reaching the detector, causing errors at high absorbance

Correct answer: Unwanted light of other wavelengths reaching the detector, causing errors at high absorbance

Stray light is radiation reaching the detector that is outside the selected wavelength band, from scattering inside the monochromator or leaks around it. At high absorbance the true transmitted light is tiny, so even a little stray light dominates the detector signal and makes absorbance read falsely low. Skoog, Holler and Crouch, Principles of Instrumental Analysis, identify stray light as the main cause of negative Beer-Lambert deviations at high absorbance.

Question 9: The Griess reaction is a classic spectrophotometric method for determining:

  • A. The Griess reaction is a classic spectrophotometric method for determining protein
  • B. The Griess reaction is a classic spectrophotometric method for determining fat
  • C. The Griess reaction is a classic spectrophotometric method for determining starch
  • D. Nitrite (and nitrate after reduction) via a pink azo dye

Correct answer: Nitrite (and nitrate after reduction) via a pink azo dye

In the Griess reaction, nitrite diazotizes sulfanilamide, which then couples with N-(1-naphthyl)ethylenediamine to form a pink azo dye absorbing near 540 nm. Nitrate is first reduced to nitrite so that total nitrite plus nitrate can be determined. The Griess assay is the classical spectrophotometric method for nitrite described in standard food analysis references such as Nielsen, Food Analysis (Springer).

Question 10: For accurate UV-Vis quantification, samples are diluted so their absorbance falls roughly between:

  • A. 0.1 and 1.0
  • B. 2.0 and 3.0
  • C. 0.001 and 0.01
  • D. 5.0 and 10.0

Correct answer: 0.1 and 1.0

The Beer-Lambert relationship stays linear roughly between 0.1 and 1.0 absorbance units on most instruments; below 0.1, noise dominates, and above 1.0, stray light and detector limits cause curvature. Diluting samples into this window keeps quantification accurate and precise. Skoog, Holler and Crouch, Principles of Instrumental Analysis, discuss this optimum absorbance range for photometric accuracy.

Question 11: Iron in foods can be quantified spectrophotometrically with 1,10-phenanthroline because it forms:

  • A. A colorless complex
  • B. An orange-red complex with iron(II)
  • C. A blue precipitate
  • D. A fluorescent derivative

Correct answer: An orange-red complex with iron(II)

1,10-phenanthroline chelates iron(II) to form the intensely orange-red tris(phenanthroline)iron(II) complex, ferroin, absorbing near 510 nm. Iron(III) in the sample is first reduced to iron(II) so that total iron is measured. This phenanthroline method is the classical spectrophotometric determination of iron in foods, described in standard references such as Nielsen, Food Analysis (Springer).

Question 12: Deviations from Beer's law at high concentrations commonly occur because of:

  • A. Too little analyte present in the cuvette
  • B. Stray light, molecular interactions and refractive-index changes
  • C. Using quartz cuvettes for the measurement
  • D. Measuring absorbance exactly at the λmax

Correct answer: Stray light, molecular interactions and refractive-index changes

At high concentrations, solute molecules interact with each other so the absorptivity itself changes, while the solution refractive index shifts; both break the law assumptions. Stray light additionally flattens the response because the detector sees unabsorbed light. These chemical and instrumental causes of Beer-Lambert deviation are detailed in Skoog, Holler and Crouch, Principles of Instrumental Analysis.

Question 13: In UV-Vis spectrophotometry, a calibration curve is prepared to:

  • A. Relate measured absorbance to analyte concentration using standards
  • B. Clean the cuvettes thoroughly before use
  • C. Set the instrument wavelength correctly each time
  • D. Warm up the instrument lamp beforehand

Correct answer: Relate measured absorbance to analyte concentration using standards

Standards of known concentration are measured to build the calibration curve of absorbance versus concentration, which should be linear through the working range. The sample absorbance is then read off this curve to find its concentration. Calibration with standards is the basis of quantitative analysis in Harris, Quantitative Chemical Analysis.

Question 14: A UV-Vis spectrophotometer uses a deuterium lamp for the UV region and:

  • A. A second deuterium lamp for visible
  • B. No lamp for visible
  • C. An infrared source
  • D. A tungsten (or halogen) lamp for the visible region

Correct answer: A tungsten (or halogen) lamp for the visible region

A deuterium lamp emits strongly in the ultraviolet (about 160-375 nm) but little in the visible, so a tungsten or tungsten-halogen lamp covers the visible region (about 350-900 nm). The instrument switches between lamps near the UV-visible boundary. This dual-source design is standard in UV-Vis instruments as described in Skoog, Holler and Crouch, Principles of Instrumental Analysis.

Question 15: Quartz cuvettes are required (rather than glass or plastic) when measuring in the:

  • A. Visible region only
  • B. Infrared region
  • C. Ultraviolet region below about 350 nm
  • D. Entire spectrum always

Correct answer: Ultraviolet region below about 350 nm

Glass and most plastics absorb ultraviolet light below about 350 nm, which would mask the sample own absorbance. Quartz (fused silica) transmits down into the UV, so quartz cuvettes are required for measurements in that region. This material requirement is standard practice noted in Harris, Quantitative Chemical Analysis.

Question 16: Carotenoid pigments in foods are conveniently estimated by UV-Vis because they:

  • A. Carotenoid pigments are estimated by UV-Vis
  • B. They absorb only in the far UV
  • C. They cannot be extracted
  • D. Absorb strongly in the visible region (around 450 nm)

Correct answer: Absorb strongly in the visible region (around 450 nm)

Carotenoids have long conjugated double-bond systems that absorb visible light strongly, with maxima typically around 430-480 nm, the absorption that gives them their yellow-to-red colors. This strong visible absorption makes simple UV-Vis measurement a convenient estimation method. The visible spectra of carotenoids are covered in standard food analysis references such as Nielsen, Food Analysis (Springer).

Question 17: A double-beam UV-Vis spectrophotometer improves measurement stability by:

  • A. A double-beam UV-Vis spectrophotometer improves measurement stability by doubling the sample volume
  • B. Continuously comparing the sample beam with a reference beam to cancel lamp drift
  • C. A double-beam UV-Vis spectrophotometer improves measurement stability by using two detectors for color
  • D. A double-beam UV-Vis spectrophotometer improves measurement stability by heating two cuvettes

Correct answer: Continuously comparing the sample beam with a reference beam to cancel lamp drift

A double-beam instrument splits the source light into a sample beam and a reference beam, and the detector continuously measures their ratio. Lamp intensity drift and electronic drift affect both beams equally, so ratioing cancels them out. This drift-cancellation design is described in Skoog, Holler and Crouch, Principles of Instrumental Analysis.

Question 18: The wavelength of maximum absorbance (λmax) is chosen for quantitative UV-Vis analysis because it gives:

  • A. Maximum sensitivity and smallest error from wavelength drift
  • B. The lowest sensitivity
  • C. Faster measurements
  • D. No need for a blank

Correct answer: Maximum sensitivity and smallest error from wavelength drift

At the absorption maximum the absorptivity is highest, giving the greatest signal per unit concentration, which is maximum sensitivity. The peak is also flat, so a small wavelength-setting error changes the measured absorbance very little. Choosing λmax for these two reasons is recommended practice in Harris, Quantitative Chemical Analysis.

Question 19: Turbidity or suspended particles in a sample cause problems in UV-Vis because they:

  • A. They increase true absorption
  • B. Scatter light, falsely increasing apparent absorbance
  • C. They decrease the path length
  • D. They change the wavelength

Correct answer: Scatter light, falsely increasing apparent absorbance

Suspended particles scatter the beam away from the detector, and the instrument interprets the lost light as absorbance, inflating the reading. This apparent absorbance has nothing to do with the analyte true absorption. Samples for UV-Vis must therefore be clear, filtered or centrifuged, standard practice in Harris, Quantitative Chemical Analysis.

Question 20: Derivative spectroscopy (plotting dA/dλ) is useful in food UV-Vis analysis for:

  • A. Increasing the sample volume in the cuvette
  • B. Changing the solvent used for the analysis
  • C. Resolving overlapping absorption bands and suppressing baseline shifts
  • D. Heating the sample before the measurement

Correct answer: Resolving overlapping absorption bands and suppressing baseline shifts

Taking the derivative of absorbance with respect to wavelength sharpens overlapping bands into better-resolved peaks and turns constant or sloping baselines into zero or constant offsets. This makes it possible to quantify components whose ordinary spectra overlap. Derivative techniques are covered in Skoog, Holler and Crouch, Principles of Instrumental Analysis.