Laboratory Calibration Management Quiz
GIFSQ ID: ART-1463
Test your knowledge with this free 20-question Laboratory Calibration Management quiz from Food Laboratory Management, part of…
What this quiz covers
Calibration is the comparison of an instrument’s readings against a known standard, with adjustment or correction as needed, and this quiz works through everything a food laboratory must get right to keep its instruments honest. It opens with the reason balances are calibrated regularly: weighing errors feed directly into every quantitative result calculated from sample mass. Calibration must cover the instrument’s full working range, because samples across the whole range must be measured accurately, not just at one point, and multiple calibration points are better than a single point because they reveal non-linearity and cover the working range. A calibration curve in chemical analysis is a plot of instrument response against standard concentrations, used to find sample concentrations, and a calibration standard is simply a material of known value used as the reference point.
The middle of the set covers drift, traceability and the paperwork. Calibration drift is the gradual change in an instrument’s readings over time away from true values, and the as-found reading, taken before any adjustment, shows how far the instrument had drifted. Traceability of calibration standards means their values are linked through an unbroken chain to national or international standards, and the measurement uncertainty stated on a calibration certificate quantifies the doubt about the calibration result, which the laboratory needs for its own uncertainty budgets. A certificate must contain the instrument identification, the standards used, the results, the uncertainty, the date and an authorised signature. Reference weights check balances because their certified mass provides a known true value.
A failed calibration stops the instrument
The discipline the quiz stresses is what happens when things go wrong. If an instrument fails calibration, it is taken out of service, labelled, investigated, and the results it produced are assessed. Calibrations are performed by trained, authorised personnel following documented procedures, at intervals set by risk, usage, manufacturer guidance and accreditation requirements, and records are retained because they prove instruments were fit for use when results were generated. A calibration within tolerance ends simply: the instrument is approved for use and the result is recorded. And the distinction to keep straight is that calibration is the comparison against a standard, while adjustment changes the instrument to correct it.
How to use this quiz
Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the definitions and the habits a laboratory professional is expected to know by heart.
Choose how you want to play
A short note: laboratory calibration management
Calibration compares readings against a known standard, adjusting or correcting as needed. Adjustment differs: it changes the instrument to correct it. A calibration standard is a substance of known value, the reference for calibration. Traceability links values in an unbroken chain to national and international standards.
A calibration curve plots response against standard concentrations to find samples. Multiple points beat one: they expose non-linearity and cover the working range. Calibration must cover the full working range, since samples across it must be measured accurately. Drift is the gradual change in readings over time away from true values. The as-found reading comes before adjustment, showing how far it had drifted.
Balances need regular calibration: weighing errors feed into every result from sample mass. Reference weights check balances with certified mass, a known true value. Calibrations need trained, authorized personnel following documented procedures. Intervals follow risk, usage, manufacturer guidance and accreditation requirements. An intermediate check is a quick periodic check between full calibrations. Recalibrate a pipette after repair or mishandling, at scheduled intervals, and whenever accuracy is doubted.
Stated uncertainty quantifies doubt about the result, feeding the lab's uncertainty budgets. A certificate lists instrument, standards used, results, uncertainty, date and signature. A failed calibration means: out of service, labelled, investigated, affected results assessed. Within tolerance, the instrument passes for use and the result is recorded. Records show instruments were fit for use at the time results were generated.
Practical takeaway: calibrate across the full range, and remove failed instruments from service.
Questions and answers
Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.
Question 1: Why must analytical balances be calibrated regularly?
Correct answer: Weighing errors directly affect every quantitative result calculated from sample mass
Every quantitative result that starts from a weighed mass inherits the balance's error, so an uncalibrated balance corrupts concentrations, recoveries, and reported values across the whole laboratory. Regular calibration against reference weights verifies the balance still reads true, as ISO/IEC 17025:2017 (clause 6.4) requires for equipment influencing results. Warranties, dust, and colour are not the point.
Question 2: Why should calibration cover the instrument's full working range?
Correct answer: Samples across the whole range must be measured accurately, not just at one point
An instrument can be accurate at one point and wrong elsewhere, so calibration must cover the full working range the laboratory actually uses : otherwise results at the extremes are unverified. ISO/IEC 17025:2017 requires equipment to be capable of the accuracy needed for its intended use (6.4.5). Auditor preferences and long reports are not the reason.
Question 3: Why is the measurement uncertainty stated on a calibration certificate?
Correct answer: It quantifies the doubt about the calibration result, needed for the lab's own uncertainty budgets
The stated measurement uncertainty quantifies the doubt about the calibration result itself, and the laboratory needs it to build its own uncertainty budgets for test results, as required by ISO/IEC 17025:2017 (clauses 7.6 and 7.8.4). Without it, the laboratory cannot state how reliable its measurements are. It is metrological information, not decoration or a fee driver.
Question 4: What is calibration of laboratory equipment?
Correct answer: Comparing an instrument's readings against a known standard and adjusting or correcting as needed
The International Vocabulary of Metrology (JCGM 200:2012, the VIM) defines calibration as the operation that establishes the relation between quantity values provided by measurement standards and the corresponding indications of the instrument, with adjustment or correction applied as needed. Cleaning, moving, and painting the instrument are maintenance, not calibration. Comparison against a known standard is the essential act.
Question 5: Who should perform calibrations in an accredited lab?
Correct answer: Trained, authorized personnel following documented procedures
ISO/IEC 17025:2017 (clause 6.2) requires that personnel performing laboratory activities : including calibration : be competent, trained, and authorised, and that calibrations follow documented procedures. Visitors, passers-by, and cleaning staff meet none of these conditions. Competence is demonstrated, not assumed.
Question 6: What is calibration drift?
Correct answer: The gradual change in an instrument's readings over time away from true values
Calibration drift is the gradual change in an instrument's readings over time away from the true values, caused by wear, ageing, or environmental effects. Recognising drift is why calibration intervals exist at all. It is not an error message, a method, or a weighing technique.
Question 7: What does traceability of calibration standards mean?
Correct answer: Their values are linked through an unbroken chain to national or international standards
Traceability means the values of the calibration standards are linked through a documented, unbroken chain of calibrations to national or international standards, each step contributing to the uncertainty : the VIM (JCGM 200:2012) definition applied via ISO/IEC 17025:2017 (6.4.6) and ILAC P10:2013. Barcodes, paper trails, and price tags do not create traceability; the calibration chain does.
Question 8: What is a calibration curve in chemical analysis?
Correct answer: A plot of instrument response versus standard concentrations to find sample concentrations
A calibration curve is a plot of instrument response against the known concentrations of standards, used to convert a sample's response into its concentration. It is the working tool of quantitative chemical analysis. Wall maps, graph paper, and pencil sketches are not it.
Question 9: How often should calibration be performed?
Correct answer: At intervals defined by risk, usage, manufacturer guidance, and accreditation requirements
ILAC G24:2022, the guidance on calibration intervals, bases the interval on risk, usage, drift history, manufacturer recommendations, and the requirements of the accreditation body : not on fixed calendar dogma. Intervals are therefore set, reviewed, and justified, not felt. 'Once ever' and 'every century' are not intervals.
Question 10: Why are calibration records retained?
Correct answer: They prove instruments were fit for use when results were generated, as auditors require
Calibration records are the evidence that instruments were fit for purpose at the time results were generated, and auditors under ISO/IEC 17025:2017 (clauses 7.5 and 8.4) expect to see them. Without the record, the result's traceability cannot be demonstrated. Shelves, cheap paper, and art projects are not the reason.
Question 11: Why are reference weights used to check balances?
Correct answer: Their certified mass provides a known true value to verify balance accuracy
Reference weights are themselves calibrated with certified mass values and traceability, so they provide a known true value against which the balance's accuracy can be verified. The check compares the balance's reading with that known value. Shininess, heaviness, and tradition are not metrological properties.
Question 12: What must a calibration certificate contain?
Correct answer: Instrument identification, standards used, results, uncertainty, date, authorized signature
ISO/IEC 17025:2017 (clause 7.8.4) specifies what a calibration certificate must contain: identification of the instrument, the standards used, the calibration results, the measurement uncertainty, the date, and an authorised signature. These elements make the certificate a traceable, accountable document. Poems, prices, and photographs are not required content.
Question 13: What should be done if an instrument fails calibration?
Correct answer: Take it out of service, label it, investigate, and assess affected results
ISO/IEC 17025:2017 (clause 6.4.9) requires equipment that is defective or outside specified requirements to be taken out of service, clearly labelled, and not used until repaired and shown to perform correctly : with an investigation into the validity of results already produced. Hiding it, using it anyway, or recalibrating until it passes all violate this. Affected results must be assessed.
Question 14: What action follows a calibration that is within tolerance?
Correct answer: The instrument is approved for use and the result is recorded
A calibration result within tolerance means the instrument meets its specified requirements, so it is approved for continued use and the result is recorded as evidence. That is the normal, documented outcome of a successful calibration. Celebration, immediate recalibration, and doing nothing are not procedures.
Question 15: What is the difference between calibration and adjustment?
Correct answer: Calibration is the comparison against a standard; adjustment changes the instrument to correct it
The VIM (JCGM 200:2012) distinguishes the two operations: calibration is the comparison of the instrument's indications against a measurement standard, while adjustment is the set of operations that bring the instrument into proper working condition. They are related but not the same : and neither is cleaning or painting. Knowing which was done matters for the record.
Question 16: Why are multiple calibration points better than a single point?
Correct answer: They reveal non-linearity and cover the working range, improving accuracy
Multiple calibration points reveal non-linearity that a single point cannot, and they cover the working range so interpolation is valid across it : the foundation of a reliable calibration curve. Longer procedures, auditor impressions, and vague 'better' claims are not the reason. More points mean better-characterised accuracy.
Question 17: What is intermediate check (between formal calibrations)?
Correct answer: A simplified periodic check that the instrument still performs acceptably between full calibrations
An intermediate check is a simplified, periodic verification : such as weighing a check weight : that the instrument still performs acceptably between full calibrations, required by ISO/IEC 17025:2017 (clause 6.4.10) where needed to maintain confidence. It is not a full recalibration, and it is certainly not a coffee break, a fire drill, or a staff meeting.
Question 18: When must a pipette be recalibrated?
Correct answer: After repair, mishandling or scheduled intervals, and whenever accuracy is doubted
A pipette must be recalibrated after repair, after mishandling or dropping, at scheduled intervals, and whenever its accuracy is doubted : the conditions recognised in the ISO 8655 series for piston-operated volumetric apparatus. Volume delivery drifts with wear and damage, so 'never' and 'once a decade' are not options. A brand-new pipette still needs its initial verification.
Question 19: What is a calibration standard?
Correct answer: A material of known value used as the reference point for calibration
A calibration standard is a material of known value : such as a certified reference weight or a standard solution : used as the reference point against which the instrument is calibrated. Its known value is what transfers accuracy to the instrument. Staff members, flags, and textbook chapters are not calibration standards.
Question 20: What is an 'as-found' calibration reading?
Correct answer: The instrument's reading before any adjustment, showing how far it had drifted
The 'as-found' reading is the instrument's reading taken before any adjustment during calibration, and it shows how far the instrument had drifted since its last calibration. Comparing as-found with as-left data is how laboratories assess whether past results were affected. It is not a guess, a lost reading, or an outdoor measurement.