HPLC Quiz
GIFSQ ID: ART-1449
Test your knowledge with this free 20-question HPLC quiz from Food Analysis & Analytical Chemistry, part of…
What this quiz covers
HPLC is the workhorse of food analysis, and this quiz walks through the whole system the way an analyst meets it: the column, the mobile phase, the detector, and the checks that prove the system is fit before any sample is trusted. Separation happens as analytes partition between a pressurized liquid mobile phase and a stationary phase, most commonly in reversed phase mode: a non polar C18 column with polar water organic mobile phases. Gradient elution changes the mobile phase composition during the run, usually ramping up the strong solvent, so early and late peaks both elute in reasonable time with good shape. Mobile phase pH matters because it decides whether an organic acid is ionized or neutral, and only the neutral form is retained strongly on a reversed phase column.
The middle of the set covers the parts that protect and extend the separation. Smaller particles give higher efficiency and resolution at the cost of higher backpressure, a trade UHPLC pushes further with sub 2 micrometre particles at very high pressure. A guard column sacrifices itself, trapping particulates and strongly retained contaminants before they can foul the expensive main column. Size exclusion separates by molecular size with large molecules eluting first, ion pair chromatography retains charged analytes on reversed phase columns, and normal phase excels at isomers and non polar compounds such as fat soluble vitamins and carotenoids. On the detection side, a diode array detector records full UV Vis spectra for identification and peak purity checks, fluorescence suits aflatoxins and some vitamins, and the refractive index detector remains the classic choice for sugars, which have no useful UV chromophore.
Retention is a choice, not an accident
Good chromatography is designed, and the quiz keeps returning to the design rules: keep the retention factor between 1 and 10, so peaks stay clear of the void without running so late they broaden; check peak purity with diode array spectra or MS when a peak may hide a co eluting impurity; and remember that tailing usually comes from unreacted silanol groups or overloading. Filter and degas samples and mobile phases, because particles clog columns and bubbles disturb detectors, and run system suitability before the batch.
How to use this quiz
Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the methods and standards a food analyst is expected to know by heart.
Choose how you want to play
A short note: HPLC
HPLC separates analytes by partitioning a pressurized liquid mobile phase against a solid stationary phase. Reversed-phase HPLC, the usual mode for food analysis, pairs a C18 column, non-polar, with polar water-organic mobile phases. In gradient elution the mobile phase changes during the run, raising the strong solvent. Mobile-phase pH sets the ionized or neutral form of an organic acid, changing retention.
A diode-array detector captures full UV-Vis spectra of each peak, helping identification and peak-purity checks. A retention factor from 1 to 10 keeps peaks clear of the void, avoiding late, broad peaks. Smaller particles raise efficiency and resolution but also backpressure; UHPLC runs on sub-2-µm particles for faster separations. Size-exclusion chromatography separates by molecular size, big molecules first. Normal-phase HPLC suits isomers and non-polar compounds like fat-soluble vitamins. Ion-pair chromatography retains charged analytes with a pairing reagent.
Sugars lack a useful UV chromophore, so analysis uses the refractive index detector. Ascorbic acid and B-vitamins are measured with reversed-phase HPLC, UV or fluorescence detection. Fluorescence detection suits compounds that fluoresce natively or as derivatives. Caffeine is measured by reversed-phase HPLC, UV detection near 272 nm. A guard column ahead of the analytical column traps particulates and retained contaminants. Peak tailing usually stems from extra interactions with silanol groups left unreacted, or column overloading. Samples and mobile phases are filtered and degassed, so particles cannot clog it. System suitability testing confirms resolution, precision and peak-shape criteria.
Practical takeaway: keep the retention factor between 1 and 10, and check suitability first.
Questions and answers
Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.
Question 1: In gradient elution HPLC, the mobile phase composition is:
Correct answer: Changed during the run, usually increasing the strong solvent
Gradient elution programs the mobile phase to change composition during the run : typically ramping up the strong solvent : so early and late peaks both elute in reasonable time with good shape. Gradient theory is covered in Snyder, Kirkland & Glajch, Practical HPLC Method Development (2nd ed., 1997).
Question 2: The pH of the mobile phase matters in reversed-phase HPLC of organic acids because it controls:
Correct answer: Whether the acid is ionized or neutral, which strongly affects retention
Mobile-phase pH decides whether an organic acid is ionized or neutral, and only the neutral form is retained strongly on a reversed-phase column : so pH directly controls its retention. pH effects on retention are detailed in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 3: A diode-array detector (DAD) in HPLC is advantageous because it:
Correct answer: Records full UV-Vis spectra of each peak, aiding identification and peak-purity checks
A diode-array detector records a complete UV, visible spectrum for every point of every peak, enabling compound identification and peak-purity checks that a single-wavelength detector cannot do. DAD capabilities are described in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 4: Size-exclusion chromatography (SEC) separates food analytes on the basis of:
Correct answer: Molecular size, with large molecules eluting first
Size-exclusion chromatography separates by molecular size: large molecules are excluded from the pores and sweep through first, while small ones wander into the pores and elute later. SEC principles are given in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 5: In high-performance liquid chromatography, separation occurs as analytes:
Correct answer: Partition between a pressurized liquid mobile phase and a stationary phase
HPLC separates analytes as they partition between a pressurized liquid mobile phase and a stationary phase packed in the column : no evaporation, electric fields, or precipitation involved. The partition basis of HPLC is the starting point of Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 6: Reversed-phase HPLC, the most common mode for food analysis, typically uses:
Correct answer: A non-polar C18 column with polar water-organic mobile phases
Reversed-phase HPLC : the workhorse mode for food analysis : pairs a non-polar C18 stationary phase with polar water, organic mobile phases. This standard configuration is described in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 7: When an HPLC peak is suspected to hide a co-eluting impurity, the analyst can check peak purity using:
Correct answer: DAD spectral comparison across the peak or MS detection
Peak purity is checked by comparing diode-array spectra across the peak, or by MS detection: a hidden co-eluting impurity distorts the spectrum or adds unexpected ions. Peak-purity assessment methods are covered in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 8: In HPLC method development, the retention factor (k) ideally falls between 1 and 10 because:
Correct answer: Peaks eluting near the void are poorly resolved while very late peaks waste time and broaden
A retention factor between 1 and 10 keeps peaks clear of the unretained void (where nothing resolves) without letting them run so late that they broaden, tail, and waste time. The k = 1, 10 guideline is given in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 9: UHPLC (ultra-high-performance LC) differs from conventional HPLC mainly in:
Correct answer: Using sub-2-µm particles at very high pressure for faster, higher-resolution separations
UHPLC uses sub-2-µm particles at very high pressure, which delivers faster, higher-resolution separations than conventional HPLC with larger particles. Sub-2-µm ultra-performance LC was introduced by Swartz in the Journal of Liquid Chromatography & Related Technologies (2005).
Question 10: Ion-pair chromatography is used in HPLC to:
Correct answer: Retain charged analytes on reversed-phase columns via pairing reagents
Ion-pair chromatography adds a pairing reagent to the mobile phase that associates with charged analytes, giving them the neutral character needed to be retained on reversed-phase columns. The technique is explained in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 11: For HPLC analysis of sugars, which detector is most commonly used?
Correct answer: Refractive index (RI) detector
Sugars have no useful UV chromophore, so the refractive index detector : which responds to any change in the eluent's refractive index : became the classic HPLC detector for them. HPLC-RI determination of sugars follows AOAC 982.14.
Question 12: A guard column is installed before the analytical HPLC column to:
Correct answer: Trap particulates and strongly retained contaminants, protecting the main column
The guard column sits ahead of the analytical column and sacrifices itself: it traps particulates and strongly retained contaminants before they can foul the expensive main column. Guard-column protection is recommended in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 13: Normal-phase HPLC (polar stationary phase, non-polar mobile phase) is best suited to:
Correct answer: Separation of isomers and non-polar compounds such as fat-soluble vitamins and carotenoids
Normal-phase HPLC, with its polar stationary phase and non-polar mobile phase, excels at separating isomers and non-polar compounds such as fat-soluble vitamins and carotenoids. Mode selection guidance is given in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 14: Smaller stationary-phase particle sizes in HPLC columns generally give:
Correct answer: Higher efficiency and resolution, but higher backpressure
Smaller stationary-phase particles shorten diffusion paths and raise efficiency and resolution : at the cost of substantially higher backpressure. The particle-size/efficiency relationship follows van Deemter theory, summarized in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 15: Water-soluble vitamins such as ascorbic acid and B-vitamins in foods are commonly analyzed by:
Correct answer: Reversed-phase HPLC with UV or fluorescence detection
Water-soluble vitamins such as ascorbic acid and the B-vitamins are routinely measured by reversed-phase HPLC with UV or fluorescence detection, which pairs the right separation mode with sensitive detection. Vitamin HPLC methods are compiled in Nielsen, Food Analysis (5th ed., 2017).
Question 16: Fluorescence detection in HPLC is especially valuable for:
Correct answer: Compounds with native fluorescence or fluorescent derivatives, such as aflatoxins and some vitamins
Fluorescence detection is prized for compounds that fluoresce natively or can be derivatized to do so : aflatoxins and several vitamins among them : because it is both sensitive and selective. Aflatoxin determination by HPLC with fluorescence detection follows AOAC 999.07 (1999).
Question 17: Caffeine in beverages is routinely quantified by reversed-phase HPLC with:
Correct answer: UV detection near 272 nm
Caffeine absorbs strongly in the UV, so reversed-phase HPLC with UV detection near 272 nm is the standard way to quantify it in beverages. Caffeine determination by HPLC is specified in ISO 20481:2008 (coffee and coffee products).
Question 18: Peak tailing in HPLC chromatograms is commonly caused by:
Correct answer: Secondary interactions with unreacted silanol groups or column overloading
Peak tailing usually comes from secondary interactions between analytes and unreacted silanol groups on the silica, or from simply overloading the column. These causes are catalogued in Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 19: Why are HPLC samples and mobile phases filtered (typically 0.45 µm) and degassed?
Correct answer: To prevent particles from clogging the column and bubbles from disturbing the detector
Filtering (typically 0.45 µm) removes particles that would clog the column and fittings, while degassing keeps dissolved air from forming bubbles that disturb the detector. Both are standard system-care steps per Snyder, Kirkland & Glajch (2nd ed., 1997).
Question 20: System suitability testing before an HPLC run verifies that:
Correct answer: The chromatographic system meets resolution, precision and peak-shape criteria for the method
System suitability testing confirms the chromatographic system still meets the method's criteria : resolution, precision, peak shape : before any sample results are trusted. Suitability requirements are defined in USP <621> (Chromatography).