HPLC Quiz

GIFSQ ID: ART-1449

Test your knowledge with this free 20-question HPLC quiz from Food Analysis & Analytical Chemistry, part of…

What this quiz covers

HPLC is the workhorse of food analysis, and this quiz walks through the whole system the way an analyst meets it: the column, the mobile phase, the detector, and the checks that prove the system is fit before any sample is trusted. Separation happens as analytes partition between a pressurized liquid mobile phase and a stationary phase, most commonly in reversed phase mode: a non polar C18 column with polar water organic mobile phases. Gradient elution changes the mobile phase composition during the run, usually ramping up the strong solvent, so early and late peaks both elute in reasonable time with good shape. Mobile phase pH matters because it decides whether an organic acid is ionized or neutral, and only the neutral form is retained strongly on a reversed phase column.

The middle of the set covers the parts that protect and extend the separation. Smaller particles give higher efficiency and resolution at the cost of higher backpressure, a trade UHPLC pushes further with sub 2 micrometre particles at very high pressure. A guard column sacrifices itself, trapping particulates and strongly retained contaminants before they can foul the expensive main column. Size exclusion separates by molecular size with large molecules eluting first, ion pair chromatography retains charged analytes on reversed phase columns, and normal phase excels at isomers and non polar compounds such as fat soluble vitamins and carotenoids. On the detection side, a diode array detector records full UV Vis spectra for identification and peak purity checks, fluorescence suits aflatoxins and some vitamins, and the refractive index detector remains the classic choice for sugars, which have no useful UV chromophore.

Retention is a choice, not an accident

Good chromatography is designed, and the quiz keeps returning to the design rules: keep the retention factor between 1 and 10, so peaks stay clear of the void without running so late they broaden; check peak purity with diode array spectra or MS when a peak may hide a co eluting impurity; and remember that tailing usually comes from unreacted silanol groups or overloading. Filter and degas samples and mobile phases, because particles clog columns and bubbles disturb detectors, and run system suitability before the batch.

How to use this quiz

Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the methods and standards a food analyst is expected to know by heart.

HPLC Quiz
Uncategorized

Choose how you want to play

Sound effects
🔈 Volume
Show explanations
🔊 Auto-read every question
Reading speed

A short note: HPLC

HPLC separates analytes by partitioning a pressurized liquid mobile phase against a solid stationary phase. Reversed-phase HPLC, the usual mode for food analysis, pairs a C18 column, non-polar, with polar water-organic mobile phases. In gradient elution the mobile phase changes during the run, raising the strong solvent. Mobile-phase pH sets the ionized or neutral form of an organic acid, changing retention.

A diode-array detector captures full UV-Vis spectra of each peak, helping identification and peak-purity checks. A retention factor from 1 to 10 keeps peaks clear of the void, avoiding late, broad peaks. Smaller particles raise efficiency and resolution but also backpressure; UHPLC runs on sub-2-µm particles for faster separations. Size-exclusion chromatography separates by molecular size, big molecules first. Normal-phase HPLC suits isomers and non-polar compounds like fat-soluble vitamins. Ion-pair chromatography retains charged analytes with a pairing reagent.

Sugars lack a useful UV chromophore, so analysis uses the refractive index detector. Ascorbic acid and B-vitamins are measured with reversed-phase HPLC, UV or fluorescence detection. Fluorescence detection suits compounds that fluoresce natively or as derivatives. Caffeine is measured by reversed-phase HPLC, UV detection near 272 nm. A guard column ahead of the analytical column traps particulates and retained contaminants. Peak tailing usually stems from extra interactions with silanol groups left unreacted, or column overloading. Samples and mobile phases are filtered and degassed, so particles cannot clog it. System suitability testing confirms resolution, precision and peak-shape criteria.

Practical takeaway: keep the retention factor between 1 and 10, and check suitability first.

Questions and answers

Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.

Question 1: In gradient elution HPLC, the mobile phase composition is:

  • A. Changed during the run, usually increasing the strong solvent
  • B. In gradient elution, it is kept constant throughout the run
  • C. In gradient elution, it is changed only between runs
  • D. In gradient elution HPLC, the mobile phase composition is replaced with gas

Correct answer: Changed during the run, usually increasing the strong solvent

Gradient elution programs the mobile phase to change composition during the run : typically ramping up the strong solvent : so early and late peaks both elute in reasonable time with good shape. Gradient theory is covered in Snyder, Kirkland & Glajch, Practical HPLC Method Development (2nd ed., 1997).

Question 2: The pH of the mobile phase matters in reversed-phase HPLC of organic acids because it controls:

  • A. In reversed-phase HPLC, the pH controls the pump pressure
  • B. Whether the acid is ionized or neutral, which strongly affects retention
  • C. In reversed-phase HPLC, the pH controls the detector wavelength
  • D. In reversed-phase HPLC, the pH controls the column length

Correct answer: Whether the acid is ionized or neutral, which strongly affects retention

Mobile-phase pH decides whether an organic acid is ionized or neutral, and only the neutral form is retained strongly on a reversed-phase column : so pH directly controls its retention. pH effects on retention are detailed in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 3: A diode-array detector (DAD) in HPLC is advantageous because it:

  • A. A diode-array detector (DAD) in HPLC is advantageous because it measures only one wavelength
  • B. A diode-array detector (DAD) in HPLC is advantageous because it detects only fluorescent compounds
  • C. A diode-array detector (DAD) in HPLC is advantageous because it requires no light source
  • D. Records full UV-Vis spectra of each peak, aiding identification and peak-purity checks

Correct answer: Records full UV-Vis spectra of each peak, aiding identification and peak-purity checks

A diode-array detector records a complete UV, visible spectrum for every point of every peak, enabling compound identification and peak-purity checks that a single-wavelength detector cannot do. DAD capabilities are described in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 4: Size-exclusion chromatography (SEC) separates food analytes on the basis of:

  • A. SEC separates food analytes on the basis of charge
  • B. Molecular size, with large molecules eluting first
  • C. SEC separates food analytes on the basis of polarity
  • D. SEC separates food analytes on the basis of volatility

Correct answer: Molecular size, with large molecules eluting first

Size-exclusion chromatography separates by molecular size: large molecules are excluded from the pores and sweep through first, while small ones wander into the pores and elute later. SEC principles are given in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 5: In high-performance liquid chromatography, separation occurs as analytes:

  • A. In high-performance liquid chromatography, separation occurs as analytes evaporate in a heated column
  • B. In high-performance liquid chromatography, separation occurs as analytes migrate in an electric field
  • C. In high-performance liquid chromatography, separation occurs as analytes precipitate from solution
  • D. Partition between a pressurized liquid mobile phase and a stationary phase

Correct answer: Partition between a pressurized liquid mobile phase and a stationary phase

HPLC separates analytes as they partition between a pressurized liquid mobile phase and a stationary phase packed in the column : no evaporation, electric fields, or precipitation involved. The partition basis of HPLC is the starting point of Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 6: Reversed-phase HPLC, the most common mode for food analysis, typically uses:

  • A. A polar silica column with hexane mobile phase
  • B. An ion-exchange resin with salt gradients only
  • C. A non-polar C18 column with polar water-organic mobile phases
  • D. A chiral column for all separations

Correct answer: A non-polar C18 column with polar water-organic mobile phases

Reversed-phase HPLC : the workhorse mode for food analysis : pairs a non-polar C18 stationary phase with polar water, organic mobile phases. This standard configuration is described in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 7: When an HPLC peak is suspected to hide a co-eluting impurity, the analyst can check peak purity using:

  • A. When impurity is suspected, use a higher flow rate
  • B. DAD spectral comparison across the peak or MS detection
  • C. When impurity is suspected, use a longer column only
  • D. When a peak hides impurity, use visual inspection

Correct answer: DAD spectral comparison across the peak or MS detection

Peak purity is checked by comparing diode-array spectra across the peak, or by MS detection: a hidden co-eluting impurity distorts the spectrum or adds unexpected ions. Peak-purity assessment methods are covered in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 8: In HPLC method development, the retention factor (k) ideally falls between 1 and 10 because:

  • A. In HPLC method development, the retention factor (k) ideally falls between 1 and 10 because it is a regulatory requirement
  • B. In HPLC method development, the retention factor (k) ideally falls between 1 and 10 because detectors only work in that range
  • C. Peaks eluting near the void are poorly resolved while very late peaks waste time and broaden
  • D. In HPLC method development, the retention factor (k) ideally falls between 1 and 10 because pumps cannot exceed k = 10

Correct answer: Peaks eluting near the void are poorly resolved while very late peaks waste time and broaden

A retention factor between 1 and 10 keeps peaks clear of the unretained void (where nothing resolves) without letting them run so late that they broaden, tail, and waste time. The k = 1, 10 guideline is given in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 9: UHPLC (ultra-high-performance LC) differs from conventional HPLC mainly in:

  • A. UHPLC differs by using gas as the mobile phase
  • B. UHPLC (ultra-high-performance LC) differs from conventional HPLC mainly in requiring no detector
  • C. Using sub-2-µm particles at very high pressure for faster, higher-resolution separations
  • D. UHPLC (ultra-high-performance LC) differs from conventional HPLC mainly in operating without columns

Correct answer: Using sub-2-µm particles at very high pressure for faster, higher-resolution separations

UHPLC uses sub-2-µm particles at very high pressure, which delivers faster, higher-resolution separations than conventional HPLC with larger particles. Sub-2-µm ultra-performance LC was introduced by Swartz in the Journal of Liquid Chromatography & Related Technologies (2005).

Question 10: Ion-pair chromatography is used in HPLC to:

  • A. Retain charged analytes on reversed-phase columns via pairing reagents
  • B. Ion-pair chromatography is used in HPLC to separate proteins by size
  • C. Ion-pair chromatography is used in HPLC to measure pH
  • D. Ion-pair chromatography is used in HPLC to detect metals

Correct answer: Retain charged analytes on reversed-phase columns via pairing reagents

Ion-pair chromatography adds a pairing reagent to the mobile phase that associates with charged analytes, giving them the neutral character needed to be retained on reversed-phase columns. The technique is explained in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 11: For HPLC analysis of sugars, which detector is most commonly used?

  • A. UV detector at 254 nm
  • B. Flame ionization detector
  • C. Electrochemical detector only
  • D. Refractive index (RI) detector

Correct answer: Refractive index (RI) detector

Sugars have no useful UV chromophore, so the refractive index detector : which responds to any change in the eluent's refractive index : became the classic HPLC detector for them. HPLC-RI determination of sugars follows AOAC 982.14.

Question 12: A guard column is installed before the analytical HPLC column to:

  • A. A guard column is installed to increase the separation efficiency
  • B. A guard column is installed to heat the mobile phase
  • C. A guard column is installed to detect the analytes early
  • D. Trap particulates and strongly retained contaminants, protecting the main column

Correct answer: Trap particulates and strongly retained contaminants, protecting the main column

The guard column sits ahead of the analytical column and sacrifices itself: it traps particulates and strongly retained contaminants before they can foul the expensive main column. Guard-column protection is recommended in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 13: Normal-phase HPLC (polar stationary phase, non-polar mobile phase) is best suited to:

  • A. Normal-phase HPLC (polar stationary phase, non-polar mobile phase) is best suited to highly water-soluble sugars
  • B. Separation of isomers and non-polar compounds such as fat-soluble vitamins and carotenoids
  • C. Normal-phase HPLC (polar stationary phase, non-polar mobile phase) is best suited to inorganic anions
  • D. Normal-phase HPLC (polar stationary phase, non-polar mobile phase) is best suited to proteins

Correct answer: Separation of isomers and non-polar compounds such as fat-soluble vitamins and carotenoids

Normal-phase HPLC, with its polar stationary phase and non-polar mobile phase, excels at separating isomers and non-polar compounds such as fat-soluble vitamins and carotenoids. Mode selection guidance is given in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 14: Smaller stationary-phase particle sizes in HPLC columns generally give:

  • A. Higher efficiency and resolution, but higher backpressure
  • B. Lower backpressure and poorer resolution
  • C. No change in performance
  • D. Longer column lifetimes only

Correct answer: Higher efficiency and resolution, but higher backpressure

Smaller stationary-phase particles shorten diffusion paths and raise efficiency and resolution : at the cost of substantially higher backpressure. The particle-size/efficiency relationship follows van Deemter theory, summarized in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 15: Water-soluble vitamins such as ascorbic acid and B-vitamins in foods are commonly analyzed by:

  • A. GC-FID without derivatization
  • B. Reversed-phase HPLC with UV or fluorescence detection
  • C. Flame photometry
  • D. Titrimetry only

Correct answer: Reversed-phase HPLC with UV or fluorescence detection

Water-soluble vitamins such as ascorbic acid and the B-vitamins are routinely measured by reversed-phase HPLC with UV or fluorescence detection, which pairs the right separation mode with sensitive detection. Vitamin HPLC methods are compiled in Nielsen, Food Analysis (5th ed., 2017).

Question 16: Fluorescence detection in HPLC is especially valuable for:

  • A. Fluorescence detection in HPLC is especially valuable for all inorganic ions
  • B. Fluorescence detection in HPLC is especially valuable for saturated fats
  • C. Compounds with native fluorescence or fluorescent derivatives, such as aflatoxins and some vitamins
  • D. Fluorescence detection in HPLC is especially valuable for water indeed

Correct answer: Compounds with native fluorescence or fluorescent derivatives, such as aflatoxins and some vitamins

Fluorescence detection is prized for compounds that fluoresce natively or can be derivatized to do so : aflatoxins and several vitamins among them : because it is both sensitive and selective. Aflatoxin determination by HPLC with fluorescence detection follows AOAC 999.07 (1999).

Question 17: Caffeine in beverages is routinely quantified by reversed-phase HPLC with:

  • A. UV detection near 272 nm
  • B. RI detection
  • C. Flame ionization
  • D. Conductivity detection

Correct answer: UV detection near 272 nm

Caffeine absorbs strongly in the UV, so reversed-phase HPLC with UV detection near 272 nm is the standard way to quantify it in beverages. Caffeine determination by HPLC is specified in ISO 20481:2008 (coffee and coffee products).

Question 18: Peak tailing in HPLC chromatograms is commonly caused by:

  • A. Secondary interactions with unreacted silanol groups or column overloading
  • B. Too high a flow rate passing through the column
  • C. Using a guard column fitted to the system
  • D. Degassing the mobile phase thoroughly before use

Correct answer: Secondary interactions with unreacted silanol groups or column overloading

Peak tailing usually comes from secondary interactions between analytes and unreacted silanol groups on the silica, or from simply overloading the column. These causes are catalogued in Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 19: Why are HPLC samples and mobile phases filtered (typically 0.45 µm) and degassed?

  • A. HPLC samples are and mobile phases filtered (typically 0.45 µm) and degassed to increase the flow rate
  • B. HPLC samples are and mobile phases filtered (typically 0.45 µm) and degassed to change the pH
  • C. To prevent particles from clogging the column and bubbles from disturbing the detector
  • D. HPLC samples are and mobile phases filtered (typically 0.45 µm) and degassed to concentrate the analytes

Correct answer: To prevent particles from clogging the column and bubbles from disturbing the detector

Filtering (typically 0.45 µm) removes particles that would clog the column and fittings, while degassing keeps dissolved air from forming bubbles that disturb the detector. Both are standard system-care steps per Snyder, Kirkland & Glajch (2nd ed., 1997).

Question 20: System suitability testing before an HPLC run verifies that:

  • A. System suitability testing before an HPLC run verifies that the laboratory is clean
  • B. System suitability testing before an HPLC run verifies that the samples are sterile
  • C. System suitability testing before an HPLC run verifies that the analyst is trained
  • D. The chromatographic system meets resolution, precision and peak-shape criteria for the method

Correct answer: The chromatographic system meets resolution, precision and peak-shape criteria for the method

System suitability testing confirms the chromatographic system still meets the method's criteria : resolution, precision, peak shape : before any sample results are trusted. Suitability requirements are defined in USP <621> (Chromatography).