GC Quiz
GIFSQ ID: ART-1444
Test your knowledge with this free 20-question GC quiz from Food Analysis & Analytical Chemistry, part of…
What this quiz covers
Gas chromatography is the workhorse for volatile compounds in food analysis, and this quiz covers it from injection to detector. Separation rests on differences in volatility and partitioning between a gaseous mobile phase and a stationary phase, and the retention time of a compound, the time from injection to the peak maximum, is characteristic under fixed conditions. Common carrier gases are helium, hydrogen and nitrogen, and temperature programming, raising the oven temperature during the run, elutes compounds with a wide boiling point range in a reasonable time with sharp peaks. Modern capillary columns give much higher efficiency than packed columns, and overlapping peaks are resolved with a longer column, a slower ramp or a different stationary phase. A polar stationary phase such as wax retains polar compounds more strongly, and column bleed, slow degradation of stationary phase at high temperature, raises the baseline.
The injection and detection choices decide what the chromatogram can say. In split injection the split ratio vents most of the sample so only a small fraction enters the column, while splitless injection is preferred for trace-level analytes. An internal standard is added to correct for injection variability and sample losses, improving accuracy, and the amount of an analyte is normally proportional to its peak area or height. The detectors each have their own strengths: the flame ionization detector responds to most organic compounds that burn in the hydrogen-air flame, the electron capture detector is especially sensitive to halogenated compounds such as organochlorine pesticides, and the thermal conductivity detector is considered universal because it detects any compound whose thermal conductivity differs from the carrier gas. Samples must be free of water and non-volatile residues, which damage the column, extinguish the FID flame and contaminate the inlet, and polar compounds are derivatized, by silylation or methylation, to make them volatile and stable.
The column does the separating, the detector does the seeing
That split explains the two most common applications in this set: solid-phase microextraction coupled to GC concentrates volatile aroma compounds from the headspace onto a fiber without solvents for flavor analysis, headspace GC measures volatile aroma compounds and residual solvents without injecting the food matrix at all, and fatty acid profiling of edible oils starts by converting the oil to fatty acid methyl esters, FAME, so the fatty acids become volatile enough to run.
How to use this quiz
Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the principles, detectors and practices a food analyst is expected to know by heart.
Choose how you want to play
A short note: gas chromatography
Gas chromatography separates food components by volatility and partitioning. Carrier gases: helium, hydrogen and nitrogen. Retention time runs from injection to peak top under fixed conditions. In GC, an analyte's amount tracks its peak area.
Samples must contain no water and no non-volatile residues, which damage the column and extinguish the FID flame. Derivatization makes polar, non-volatile compounds volatile and thermally stable. Fatty acid profiling starts by turning oil into fatty acid methyl esters, FAME. An internal standard at known concentration corrects for injection variability and sample losses.
A wax phase holds polar compounds tighter than a non-polar phase does. Temperature programming elutes wide boiling ranges in reasonable time with sharp peaks. Overlapping peaks resolve with a longer column, a gentler ramp or a different phase. Capillary columns beat packed columns on efficiency and resolution. Column bleed is the stationary phase breaking down at high temperature, raising the baseline.
The flame ionization detector responds to most organics burning in the hydrogen-air flame. The electron capture detector is especially sensitive to halogenated compounds like organochlorine pesticides. A thermal conductivity detector is universal: it senses any compound unlike the carrier gas in thermal conductivity. Split injection vents most of the sample, letting only a small fraction enter the column; splitless injection suits trace-level analytes. SPME concentrates volatile aromas from the headspace onto a fibre without solvents, and headspace GC suits volatile aromas and residual solvents with no matrix injected.
Practical takeaway: keep water out, use an internal standard, program the temperature.
Questions and answers
Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.
Question 1: An internal standard in quantitative GC analysis is added to:
Correct answer: Correct for injection variability and sample losses, improving accuracy
An internal standard : a compound added at known concentration before analysis : experiences the same injection variability and sample losses as the analyte, so the analyte-to-standard ratio corrects for both. Chromatography references (e.g., Grob's GC texts) present this as the route to accurate quantification. External standards alone cannot compensate for injection-to-injection variation.
Question 2: Solid-phase microextraction (SPME) coupled to GC is used in flavor analysis to:
Correct answer: Concentrate volatile aroma compounds from the headspace onto a fiber without solvents
SPME exposes a coated fiber to the sample headspace, where volatile aroma compounds partition onto the coating and are then thermally desorbed into the GC : with no solvent at all. Pawliszyn's 1997 monograph introduced the technique. It concentrates trace flavor volatiles that direct injection would miss.
Question 3: The retention time of a compound in GC is:
Correct answer: The time from injection to the peak maximum, characteristic under fixed conditions
Retention time is measured from injection to the peak maximum, and under fixed column, temperature, and flow conditions it is characteristic of a compound. IUPAC's chromatography nomenclature defines it this way. It is a qualitative identifier; the peak's area carries the quantitative information.
Question 4: Why must GC samples be free of water and non-volatile residues?
Correct answer: They damage the column, extinguish the FID flame and contaminate the inlet
Water damages many stationary phases and extinguishes the FID's hydrogen flame, while non-volatile residues foul the inlet liner and the column head. Clean, volatile samples protect both column and detector. GC practice guides list sample cleanliness as the first defense against these failures.
Question 5: The electron capture detector (ECD) is especially sensitive to:
Correct answer: Halogenated compounds such as organochlorine pesticides
The electron capture detector responds to compounds that capture thermal electrons : most sensitively halogenated molecules such as organochlorine pesticides. Its nickel-63 source makes it the classic detector for chlorinated residues. Alcohols, sugars, hydrocarbons, and water give little or no ECD response.
Question 6: Derivatization (for example silylation or methylation) before GC is done to:
Correct answer: Make polar, non-volatile compounds volatile and thermally stable
Silylation and methylation replace polar hydrogens with non-polar groups, making polar or non-volatile analytes volatile and thermally stable enough to survive the GC inlet and column. This is standard practice for acids, alcohols, and sugars. Without derivatization, such compounds adsorb, tail, or decompose.
Question 7: A polar stationary phase (such as wax) in GC will:
Correct answer: Retain polar compounds more strongly than a non-polar phase does
'Like dissolves like' governs GC retention: a polar (wax) stationary phase interacts more strongly with polar analytes, holding them longer than a non-polar phase would. Non-polar compounds, conversely, elute relatively faster from the polar phase. Phase selection follows this polarity-matching principle.
Question 8: In GC quantification, the amount of an analyte is normally proportional to:
Correct answer: Its peak area (or height)
In GC quantification, the amount of analyte is proportional to its peak area (peak height serves for narrow, symmetrical peaks). The detector's integrated response, calibrated against standards, converts area to mass or concentration. Retention time identifies; area quantifies.
Question 9: Temperature programming in GC (raising the oven temperature during the run) is used to:
Correct answer: Elute compounds with a wide boiling-point range in a reasonable time with sharp peaks
Temperature programming starts the oven cool to resolve volatile compounds, then ramps the temperature so high-boiling compounds elute in a reasonable time as sharp peaks. A single isothermal temperature cannot serve both ends of a wide boiling range. McNair and Miller's GC text presents programming as the standard solution.
Question 10: The flame ionization detector (FID) responds to:
Correct answer: Most organic compounds that burn in the hydrogen-air flame
The flame ionization detector burns eluting compounds in a hydrogen-air flame, and the ions formed from C-H bonds generate the signal : so it responds to most organic compounds. It does not respond to water, CO2, or permanent inorganic gases. This carbon-selective, wide-linear-range behavior makes FID the workhorse food GC detector.
Question 11: Gas chromatography separates food sample components based on:
Correct answer: Differences in volatility and partitioning between a gaseous mobile phase and a stationary phase
GC separates components by their volatility and by how they partition between the gaseous mobile phase and the stationary phase. Compounds that spend more time in the stationary phase elute later. IUPAC's definition of chromatography rests on this differential distribution.
Question 12: Common carrier gases for gas chromatography include:
Correct answer: Helium, hydrogen and nitrogen
Helium, hydrogen, and nitrogen are the standard GC carrier gases: helium for inertness, hydrogen for speed and efficiency, nitrogen where economy matters. Oxygen, chlorine, and fluorine are reactive and have no place as carriers. Gas selection balances efficiency, safety, and detector compatibility.
Question 13: For fatty acid profiling of edible oils by GC, the oil is first:
Correct answer: Converted to fatty acid methyl esters (FAME)
Edible oils are converted to fatty acid methyl esters (FAME) : usually by base- or acid-catalyzed methylation : because intact triglycerides are too heavy and polar for GC. The volatile methyl esters separate cleanly by chain length and unsaturation. AOCS Method Ce 1-62 and AOAC 996.06 are the reference FAME procedures.
Question 14: If two peaks overlap in a GC chromatogram, the analyst can improve resolution by:
Correct answer: Using a longer column, slower temperature ramp or a different stationary phase
Overlapping peaks are resolved by increasing the column's separating power: a longer column, a slower temperature ramp, or a stationary phase with different selectivity. Larger injections only worsen overlap by broadening peaks. Resolution fundamentals (Grob, McNair and Miller) all point to these three levers.
Question 15: In split injection for GC, the split ratio is used to:
Correct answer: Vent most of the injected sample so only a small fraction enters the column, avoiding overload
In split injection the split ratio sets how the injected vapor divides: most is vented through the split line and only a small fraction enters the column, preventing overload from concentrated samples. Trace analysis uses splitless injection instead. The ratio is the analyst's control over column loading.
Question 16: Splitless injection is preferred over split injection when:
Correct answer: Trace-level analytes need maximum sensitivity
Splitless injection transfers nearly the entire injected sample to the column, maximizing sensitivity for trace-level analytes. The trade-off is slower transfer and potential solvent effects, acceptable when every picogram counts. Split injection's venting would discard the signal trace work needs.
Question 17: Compared with packed columns, modern capillary GC columns provide:
Correct answer: Much higher separation efficiency and resolution
Modern wall-coated open-tubular (capillary) columns deliver far more theoretical plates per meter than packed columns, giving much higher separation efficiency and resolution. This efficiency revolutionized GC in the 1980s. Packed columns survive only in niche applications.
Question 18: Column bleed in GC refers to:
Correct answer: Slow degradation of the stationary phase at high temperature, raising the baseline
Column bleed is the slow thermal degradation of the stationary phase at high temperature, releasing fragments that raise the detector baseline. It is a column-aging phenomenon, not a gas leak or syringe problem. Low-bleed phases and staying below the phase's temperature limit minimize it.
Question 19: Headspace GC is particularly useful in food analysis for:
Correct answer: Volatile aroma compounds and residual solvents without injecting the food matrix
Headspace GC samples only the vapor above the food, so volatile aroma compounds and residual solvents are measured without injecting the food matrix itself. The matrix never reaches the inlet or column. This is the standard route for flavor volatiles and packaging-solvent residues in foods.
Question 20: A thermal conductivity detector (TCD) is considered universal because it:
Correct answer: Detects any compound whose thermal conductivity differs from the carrier gas
The thermal conductivity detector is universal because it responds to any compound whose thermal conductivity differs from the carrier gas : no flame, ionization, or derivatization required. It detects organics and inorganics alike, though with lower sensitivity than FID. Carrier-gas choice (often helium) sets the contrast.