SOP: Water Vapour Transmission Rate by Infrared Sensor (ASTM F1249)

GIFSQ ID: SOP-0994

Moisture crossing a packaging film can stale a crisp product, cake a powder or feed mould, and gravimetric cup testing answers that risk slowly, over days of weighing. Infrared sensing answers it directly. Film separates a chamber kept saturated with water vapour from a dry nitrogen stream, and an infrared detector tuned to the wavelength water absorbs watches the nitrogen for whatever vapour has crossed.

Book 1 calls this route faster and more accurate than gravimetric measurement, which is why it became the standard instrumental method for film water vapour transmission. Accuracy still depends on holding a genuine humidity gradient across the specimen: vapour must truly face a saturated side and a dry side, or the number produced belongs to some smaller, unknown gradient.

Purpose

Purpose is to measure the water vapour transmission rate of plastic film and sheeting with a modulated infrared sensor under ASTM F1249, expressed as mass of water vapour per unit area per unit time.

Scope

Scope covers plastic films and sheeting sealed between a wet chamber and a dry chamber in an isostatic arrangement, meaning equal static pressure on both faces with a difference in water vapour partial pressure providing the driving force. Book 2 gives the useful range of ASTM F1249 as 0.002 to 6.45 g/100 in.2/day. Existing gravimetric methods in this library, the dish, cycle and whole-package procedures, remain separate methods; values from a cup test and from this sensor method should be reported under their own method names. A specification-sheet example in Book 1 pairs this standard with a moisture vapour transmission rate measured at 38°C and 100% relative humidity, which illustrates the kind of condition a result must be reported with.

Equipment and Materials

  • Two-chamber permeation cell in which the specimen is sealed between a wet chamber and a dry chamber, of the isostatic type used in the PERMATRAN-W instrument line named in Book 2
  • Modulated infrared sensor able to detect water vapour at 6.3 µm, with an optical filter that blocks light of all other wavelengths
  • Sponge soaked in high-performance liquid chromatography (HPLC) grade water, fitted in the wet chamber to keep its air saturated with water vapour
  • Dry nitrogen carrier gas circulating through the dry chamber and on to the sensor
  • Film or sheeting specimens, flat and undamaged, with thickness recorded where results will be compared across gauges
  • Temperature control for the cell, since transmission rates belong to the temperature at which they were measured

Procedure

  1. Prepare the wet chamber. Fit the sponge soaked in HPLC grade water so the air in that chamber stays saturated with water vapour throughout the run. A sponge allowed to dry out quietly lowers the driving force and the result with it.
  2. Mount the specimen. Seal the film between the wet and dry chambers so vapour can reach the dry side only by crossing the specimen.
  3. Start the dry carrier. Circulate nitrogen through the dry chamber and onward to the infrared sensor. Static pressure should stay the same on both sides of the film; flow here is a partial-pressure measurement, not a pressure-driven one.
  4. Shine the measuring wavelength through the carrier stream. Direct infrared light at 6.3 µm, the wavelength at which water vapour absorbs infrared radiation, through its optical filter and then through the nitrogen and transmitted-vapour mixture on its way to the detector.
  5. Read the detector response. Amount of infrared light reaching the detector corresponds to the water vapour concentration in the carrier stream, and that concentration corresponds to the water vapour transmission rate of the specimen.
  6. Hold conditions until the reading is steady. Continue the run until the sensor response settles at its working level for the conditions set, then take the transmission rate from that steady response. Record temperature and the humidity arrangement used alongside every result.

Recording Results

Report water vapour transmission rate as mass per unit area per unit time, in g/m2/day or g/100 in.2/day, with test temperature, humidity conditions, specimen identity and thickness. Underlying the unit is the general transmission relationship used throughout the source chapter: amount transmitted Q divided by area A and time t. For comparison, the gravimetric form of that relationship in the source is written WVTR = slope / area, giving g H2O/(m2 day) from a weight-versus-time plot; the infrared instrument arrives at the same quantity from vapour concentration in the carrier gas instead of from weighings.

Quality Control

Gradient integrity is the control point that matters most. Book 2 explains that for high-transmission materials the desired gradient, in effect 100% relative humidity driving against a dry side, can collapse in practice: vapour leaves through the sample faster than the reservoir below can replace it, leaving an unknown humidity at the sample surface, while moisture gathering on the dry side lifts that side above its intended dry state. Laboratories should therefore satisfy themselves that the wet chamber stayed saturated and the dry side stayed dry for the material under test, and should treat results for very permeable specimens with extra suspicion. Duplicate runs, a check that the sponge remained wet, and reporting the full test conditions are the practical safeguards, applied as general laboratory practice.

Troubleshooting

  • Suspiciously low result on a very permeable film: suspect gradient collapse as described under Quality Control; the specimen may have seen far less than the intended humidity difference, and Book 2 shows cup and sensor results falling away dramatically at high rates for exactly this reason.
  • Reading drifts during a run: check the wet chamber first; a drying sponge or falling saturation lowers the driving force progressively.
  • Result disagrees with a cup-test value: compare ranges and conditions before doubting either method. Book 2 gives the cup method, ASTM E96, a useful range of 0.65 to 3.23 g/100 in.2/day against 0.002 to 6.45 g/100 in.2/day for this sensor method, and Book 1 puts the practical lower limit of gravimetric work near 1 g/m2/day with measurement times of typically 2 to 14 days.
  • Porous specimen gives an odd result: decide whether flow through holes, rather than permeation through the material, is being measured. Porosity responds to a static pressure difference and is expressed in units such as Gurley seconds or Darcies; permeability is movement under equal static pressure with a partial-pressure difference. Confusing the two invalidates the comparison.

Interferences and Limitations

Very high transmission rates strain this method just as they strain the cup test, because neither arrangement can hold a full humidity gradient against a specimen that lets vapour through almost freely; Book 2 describes newer apparatus built for rates from 500 to 100,000 g/m2/day, a different class of measurement from the standard range quoted in the Scope. Humidity-sensitive polymers add a second caution: water content changes diffusion behaviour in materials such as ethylene vinyl alcohol and polyamide, so a result belongs to its stated humidity arrangement. Temperature likewise travels with the result under the Arrhenius-type behaviour described for permeation generally. Finally, a specimen with pinholes or porosity measures flow paths, not molecular permeation, and should be described as such.

Safety Notes

  • Handle nitrogen supplies and pressurised gas lines under cylinder and safety-data-sheet rules, and vent exhaust gas safely, as general laboratory practice.
  • Take care with heated cells and with electrical infrared source equipment, following the instrument maker’s instructions, as general laboratory practice.
  • Use HPLC grade water as specified for the wet chamber; substitutes carrying dissolved contaminants have no place in a measurement built on vapour purity, as general laboratory practice.

Source and Standard

Drafted from Chapter 5 of Food Packaging Materials: Testing & Quality Assurance (Singh, Wani and Langowski, eds., CRC Press, 2017), section 5.3.2 on infrared sensing, and from Chapter 11 of Food Packaging: Testing Methods and Applications (Risch, ed., ACS Symposium Series 753, 2000), which gives the working range of the method and the humidity-gradient problem in high-transmission testing. Governing standard: ASTM F1249, Standard Test Method for Water Vapor Transmission Rate Through Plastic Film and Sheeting Using a Modulated Infrared Sensor.