SOP: Thermal Transitions of Packaging Polymers by Differential Scanning Calorimetry (DSC) (ISO 11357 / ASTM D3418)

GIFSQ ID: SOP-0987

Ask a packaging polymer how it will behave in a freezer, a retort or a hot warehouse, and its thermal transitions answer first. Glass transition marks where an amorphous phase stops being glassy and brittle and starts moving like rubber; melting marks where crystalline order gives way; crystallisation shows how readily that order rebuilds. Differential scanning calorimetry, DSC, reads all of these from one small specimen by following heat flow while temperature is programmed up and down.

Used well, a DSC run confirms a material is what it claims to be, fingerprints an unknown, and quantifies how much crystalline phase a film or moulding carries. Used carelessly, with an uncalibrated cell or a programme nobody recorded, it produces elegant curves that mean very little.

Purpose

Aim is to determine thermal transitions and related energies of packaging polymers and bioplastics by DSC: glass transition temperature (Tg), melting temperature (Tm), crystallisation behaviour, enthalpies of fusion and crystallisation (including apparent heat of fusion, ΔHm), and degree of crystallinity derived from heat of fusion. Where a study calls for them, specific heat, oxidation induction time, reaction energy and temperature, cure energy (as in epoxy adhesives) and denaturation temperatures of biopolymers can also be obtained.

Scope

Coverage extends to plastic packaging materials and bioplastics that can be cut or sampled into a DSC pan, including films, mouldings, pellets, powders and bioplastic specimens such as polyhydroxyalkanoate (PHA) samples and soy protein isolate (SPI) or polylactic acid (PLA) films of the kind reported in the source. Both major instrument designs fall in scope: heat-flux instruments, where sample and reference sit on a disk heated by a common furnace, and power-compensation instruments, where separate enclosures around sample and reference are controlled individually. A fully worked heating, cooling and reheating programme from the source is stated in full.

Equipment and Materials

  • Differential scanning calorimeter (heat-flux or power-compensation design) with a small cell, sample and reference positions, sensitive thermocouple or resistance-thermometer sensing, and software for heat-flow recording and peak integration; the worked example in the source used a TA-Q100 instrument (TA Instruments, USA)
  • Cooling system capable of reaching sub-ambient start temperatures required by a programme; the source describes operation from −120°C to 725°C with a cooling system fitted, with an inert atmosphere required above 600°C
  • Sample pans chosen to avoid any reaction between specimen and pan: aluminium (Al), copper (Cu), gold (Au), platinum (Pt), alumina or graphite, with matching lids and a crimping or sealing press as the pan type requires
  • Empty reference pan, matched to the sample pan type, kept empty in the same cell throughout a run
  • Indium standard for calibration, as used in the source worked example
  • Microbalance suitable for weighing milligram specimens accurately
  • Purge gas supply: nitrogen, delivered at a controlled flow; the worked example used nitrogen at 50 mL/min
  • Cutting tools for preparing small, flat specimens that sit in good contact with the pan base
  • Reference value for the theoretical (100% crystalline) heat of fusion of the polymer, from literature, needed only where degree of crystallinity is to be calculated

Procedure

  1. Calibrate before a measurement series. Run the indium standard under the laboratory’s calibration routine for the instrument and confirm temperature and enthalpy response are acceptable before any test specimen is committed. Record calibration identity and date with the run file.
  2. Select a specimen representative of the material lot. General working in the source uses specimens of 0.5 to 100 mg; the worked example below used 2 to 3 mg per specimen. Small specimens favour sharp, undistorted transitions, so stay at the lower end unless signal strength forces otherwise.
  3. Cut the specimen so it lies flat in the pan base, giving good thermal contact. Weigh it accurately and note mass to the precision the enthalpy calculation needs, since every enthalpy result is normalised by this mass.
  4. Load the specimen into a pan made of a material that cannot react with it: aluminium, copper, gold, platinum, alumina or graphite are the choices the source lists. Seal or crimp as the pan design requires, place the pan on the sample position in the cell, and set an empty pan of the same type on the reference position.
  5. Establish the specified atmosphere in the cell. For the worked programme, purge with nitrogen at 50 mL/min and keep that flow steady for the whole run. Where any part of a programme exceeds 600°C, an inert atmosphere is required, without exception.
  6. Programme temperature as the study design requires. The source worked example, for PHA-type specimens on a TA-Q100 calibrated with indium, ran as follows: heat from −80°C to 180°C at a heating rate of 10°C/min under nitrogen at 50 mL/min; hold at 180°C for 1 min; quench (cool) the molten specimen to −80°C at a cooling rate of 80°C/min; then heat a second time from −80°C to 180°C at 10°C/min. Adopt this programme only for materials and questions it suits, and record every segment, rate, hold and gas flow actually used.
  7. Record heat flow continuously through every segment. Output appears as heat flow, usually expressed in milliwatts, in W/g when normalised by specimen mass, or in W/g·°C in specific-heat units, with endothermic events showing heat absorption and exothermic events showing heat evolution.
  8. Evaluate thermal properties from the second heating run where a heat, hold, quench and reheat programme has been used, as the source example did. A first heating carries the specimen’s prior processing and storage history; controlled cooling followed by reheating gives the comparable, history-erased picture. Where a study deliberately examines as-received behaviour, evaluate the first heating as well and label each curve unambiguously.
  9. Read Tg, Tm and ΔHm from the evaluated curve following the recording rules below. Note crystallisation events on cooling or on reheating (cold crystallisation) wherever they appear, since they change how a melting enthalpy may be interpreted.
  10. Run a fresh specimen rather than reheating a degraded one whenever a curve looks suspect. Note any sign of decomposition, pan reaction or mass loss in the run record instead of quietly discarding a run.

Recording Results

Report for each specimen: material and lot identity, specimen mass, pan material, instrument type (heat-flux or power-compensation), calibration standard used (indium in the source example), purge gas and flow rate, the complete temperature programme with every rate and hold, and which heating run was evaluated. Temperature measurement repeatability quoted in the source is ±0.1°C; do not report transition temperatures to a finer precision than the instrument and calibration support.

Determine Tg from the step change in heat flow (baseline shift) in the glass transition region, Tm from the melting endotherm, and enthalpies by integrating the corresponding peaks, reporting ΔHm as apparent heat of fusion following the source example. For illustration, SPI film showed Tg at 117.7°C and PLA film at 29.5°C in the source: treat such values as material-specific findings, never as acceptance limits for other materials.

Calculate degree of crystallinity from DSC by normalising the measured heat of fusion against the theoretical heat of fusion of the fully crystalline polymer:

Degree of crystallinity (%) = (Measured heat of fusion / Theoretical heat of fusion) × 100

Authentic 100% crystalline specimens are rare, so literature values are often used for the theoretical heat of fusion: name the value and its source on the report whenever this calculation is made. Where a specimen cold-crystallises during heating, take care that the melting enthalpy used reflects crystalline material present before that heating event, and state how crystallisation enthalpy was handled. Beyond transitions themselves, DSC data can support statements on general thermal stability, reaction kinetics, exothermal energy of cure and degree and rate of cure, and a qualitative judgement on whether a polymer blend is miscible or immiscible, provided the transition temperatures of its components are sufficiently well separated: report such interpretations as interpretations, separate from measured values.

Quality Control

  • Verify calibration with indium at the frequency laboratory practice sets, and whenever the cell, sensor or purge arrangement has been disturbed. A failed check stops testing until resolved.
  • Run an empty-pan baseline under the same programme when weak transitions are expected, so cell asymmetry is not mistaken for specimen behaviour.
  • Replicate a determination on a second, independently prepared specimen when a result will drive a material acceptance or rejection decision.
  • Check pan and specimen after a run for leakage, reaction or charring. Any of these invalidates enthalpy values from that run.
  • Keep programme, gas flow and specimen mass range consistent within a comparison set. Tg and peak shape shift with heating rate and thermal history, so curves from unlike programmes must not be compared as equals.

Troubleshooting

  • No visible glass transition: the step may be weak in a highly crystalline material, or masked by enthalpy relaxation or moisture effects. Try a history-erasing first heat and controlled cool, then read Tg on the second heating; a larger specimen within the stated range can also strengthen the signal.
  • Drifting or noisy baseline: check pan seating and contact, purge flow stability, and cell cleanliness. Rerun an empty-pan baseline before blaming the specimen.
  • Melting peak split or shouldered: mixed crystal populations, a blend, or uneven specimen contact can each produce this. Prepare a flatter specimen, confirm identity of the material, and consider whether two components are genuinely present.
  • Unexpected exotherm before melting: cold crystallisation is common in materials such as PET quenched during processing. Record it as a finding, and handle its enthalpy explicitly in any crystallinity calculation.
  • Results shift between runs of one material: audit programme rates, hold times, cooling rate, specimen mass and pan type first. Thermal history differences explain most shifts of this kind.

Interferences and Limitations

Prior heat treatment history, physical treatments such as biaxial stretching, chemical structure, additives and rate of temperature change all influence crystallisation and the transitions built on it, so a DSC curve describes a specimen with a history, not a timeless property of a polymer name. Blend miscibility judgements fail when component transitions lie too close together to separate. Pan reaction falsifies a run outright, which is why pan choice is part of the method. At high temperature, oxidative degradation intrudes unless the atmosphere is inert, and above 600°C inert conditions are mandatory in the source account. DSC is also strongest alongside companion techniques: the source presents thermal analysis as a family (DSC with thermogravimetric and thermomechanical analysis, plus dynamic mechanical thermal analysis) and recommends combining them for full characterisation.

Safety Notes

  • Handle pans and cell parts with tweezers after any run reaching high temperature, and allow the cell to cool before reaching inside.
  • Respect cryogenic hazards where a cooling system uses liquefied gases to reach sub-ambient starts such as −80°C or lower: insulated gloves and eye protection are routine laboratory practice for Dewar and line handling.
  • Ensure purge and cooling exhausts vent safely, and never seal a specimen known to generate gas on heating into a fully hermetic pan without confirming the pan pressure rating.
  • Cut film and moulding specimens with care, and dispose of sharp offcuts immediately.
  • Follow the instrument maker’s interlock and furnace procedures at all times, particularly for any work above 600°C under inert gas.

Source and Standard

Procedure above is drawn, in original wording, from Singh, Wani and Langowski (eds.), Food Packaging Materials: Testing & Quality Assurance, CRC Press, 2017: Chapter 3, Sections 3.4.1 and 3.4.2.1 for DSC principles, specimen masses, pan materials, temperature range and measurands; Chapter 9, Section 9.5.1 for heat-flux and power-compensation designs and for the worked TA-Q100 programme (indium calibration, 2 to 3 mg specimens, −80°C to 180°C at 10°C/min, nitrogen at 50 mL/min, 1 min hold at 180°C, quench at 80°C/min, second heating evaluated); and Chapter 3, Section 3.5.2.4 for crystallinity from heat of fusion. Standards named in the book for DSC work include ASTM D3418, Standard Test Method for Transition Temperatures and Enthalpies of Fusion and Crystallization of Polymers by Differential Scanning Calorimetry, and ASTM D3417 on enthalpies of fusion and crystallisation, alongside ASTM E793, E794, E928 and E1356. The SOP title follows the task assignment in citing ISO 11357 with ASTM D3418: ISO 11357 is not separately discussed in the source sections, so programme, calibration and reporting detail above rests on the book’s account and the ASTM methods it names, and the current ISO 11357 part applicable to the quantity sought should be consulted before testing to that standard is claimed.