SOP: Rapid Screening for Listeria Species in Foods and Environmental Samples with a Lateral-Flow Test (MFLP-71)

GIFSQ ID: SOP-1065

Speed is the point of this screen. Two enrichments grow whatever Listeria is present, a heat step releases flagellar antigen, and a lateral-flow strip answers with a blue line: genus-level evidence in 52 hours, ahead of full cultural confirmation. A strip, though, only ever gives a presumptive result, and any positive still goes forward to plating and identification by a cultural method.

One species escapes the test entirely. Antigen targeted here is the flagellar B antigen shared by every Listeria species except L. grayi, so a negative strip never speaks to that species. Sample type also changes the front end: foods go in at 25 g, while sponges and swabs carry their own broth volumes, and compositing rules differ between them.

Purpose

Purpose is the rapid detection of Listeria spp. in foods and environmental samples, to determine compliance with Sections 4 and 7 of the Food and Drugs Act. Presumptive positives are determined in 52 hours and must then be confirmed culturally.

Scope

Validated scope covers raw meat, raw unprocessed seafood, raw frozen vegetables, cheese made from unpasteurized milk, and environmental samples. Use on other foods is possible but requires validation first, following Compendium criteria, with validation data forwarded to the Microbiological Methods Committee so the application can be expanded. Sample units may be analysed individually or composited according to the sampling scheme in the Listeria policy; analysing units separately preserves information about how Listeria is distributed. Work follows ISO/IEC 17025:2005 (or latest version), and any variation on a listed medium formulation, whether a commercial product or made from scratch, needs equivalency shown by the laboratory.

Equipment and Materials

  • Primary enrichment: PALCAM broth
  • Secondary enrichment: UVM 2 broth
  • Listeria Rapid Test kit (Oxoid FT0405), including a positive control of non-viable Listeria monocytogenes (3 vials) supplied with the kit
  • Glass tubes; 1.0 mL pipettes or pipettor and tips
  • Stomacher, blender or equivalent
  • Heat plate or incubator capable of 80 ± 2°C
  • Incubators capable of maintaining 30°C and 35°C
  • Control cultures: Listeria monocytogenes, Listeria innocua or Listeria seeligeri
  • Confirmation media and reagents: Listeria Selective Agar (Oxford); PALCAM Selective Agar, or other selective media used in the cultural methods MFHPB-30 or MFHPB-07; antisera specific to Listeria (optional)

Procedure

  1. Hold sample units refrigerated or frozen before analysis, according to the nature of the product, shelf-stable foods excepted. Thaw frozen samples in a refrigerator, or under time and temperature conditions that prevent microbial growth or death, and analyse units as soon as possible after receipt. Keep a sample-to-broth ratio of 1 part sample material to 9 parts sterile enrichment broth throughout. Incubation tolerances are 24 h ± 2 h and 48 h ± 4 h.
  2. Prepare sterile PALCAM broth and clean the working-area surface with a suitable disinfectant before starting.
  3. Make the analytical unit representative: agitate liquids or free-flowing materials until homogeneous, and for a solid, take portions from several locations within the sample unit. Analytical units may be combined to reduce workload.
  4. Set up environmental samples in PALCAM broth. Add each environmental sponge or large swab to 100 mL of broth, or composite up to 10 sponges or large swabs with 100 mL of broth per sponge. Place smaller swabs, such as cotton tips, in 10 mL portions of broth, or composite up to 10 swabs with 10 mL of broth per swab.
  5. Set up food samples in PALCAM broth. Add 25 g or mL of food to 225 mL of broth in a blender jar or stomacher bag. For alternate analytical unit sizes, keep the 1 part sample to 9 parts broth ratio. Composites may combine analytical units up to 125 g or mL in total (for example, 125 g or mL of food to 1125 mL of broth), and a composite should contain no more than five analytical units. Blend, stomach or vortex environmental and food preparations as required for thorough mixing.
  6. Incubate the primary enrichment mixture and controls for 26 h at 35°C.
  7. Start secondary enrichment. Agitate the PALCAM broth enough to mix, let large food particles settle, then transfer 1 mL into 9 mL of UVM 2. Incubate for 26 to 48 h at 30°C.
  8. Hold the secondary enrichment for possible confirmation. After removing the aliquot for the test device, UVM 2 cultures may optionally be refrigerated for up to 72 hours; this gives the laboratory flexibility, because a refrigerated culture can later be resuspended and plated when a positive result needs confirmation.
  9. Run the test device. Bring the required number of devices to room temperature. Transfer 2 mL of secondary enrichment to a glass test tube and heat on a hot plate or in a water bath at 80°C for 20 minutes. Cool the sample, then, within 1 h of heating, transfer 135 µL to the Sample Window of the device. Incubate the device at room temperature for 20 minutes.
  10. Run both controls with every set of samples. Reconstitute and use the kit’s positive antigen control according to the manufacturer’s instructions. For a negative control, transfer 135 µL of uninoculated UVM 2 broth to a separate device.
  11. Read the result. Look for a blue line in the Control Window and in the Result Window. Any blue line in the Result Window, even a faint one, counts as a presumptive positive, provided a blue line also appears in the Control Window.
  12. Handle the two reading exceptions. Occasionally, excess flagellar antigen gives a strong reaction in the Result Window with no reaction in the Control Window: prepare a 1 in 10 dilution of the heated sample in fresh UVM 2 and retest on a new device. If no line appears in either window, prepare another test from the same extract, provided the extract is not more than 1 hour old.
  13. Confirm presumptive positives culturally. Using the refrigerated secondary enrichment broth, carry out the plating and confirmation steps of a cultural method, MFHPB-07 or MFHPB-30, as appropriate.

Recording Results

Record each device by its two windows: Control Window line present or absent, and Result Window line present, faint or absent, together with any dilution-and-retest outcome. A presumptive positive requires a blue line in both windows. Report a completed cultural confirmation at species level where identification supports it. Two reporting rules govern wording. First, where Listeria spp. have not been detected, L. monocytogenes is also understood not to have been detected, because no member of the genus was found. Second, that logic does not run the other way for environmental samples: where Listeria spp. are detected in an environmental sample, results cannot be used to state that L. monocytogenes is absent, no matter how much colony screening or confirmation follows; report every species of Listeria identified in environmental samples.

Quality Control

A positive antigen control from the kit and a negative control of uninoculated UVM 2 run alongside the samples on every occasion; live control cultures of L. monocytogenes, L. innocua or L. seeligeri support the cultural side of the work. Media are commercial preparations made and sterilized to the manufacturer’s instructions, and equivalency must be shown before any formulation variation is used. Verify incubator temperatures: 35°C may be held at 35 ± 1.0°C, and 30°C and 25°C likewise allow ± 1.0°C, whereas higher temperatures such as 43°C or 45.5°C must stay within 0.5°C because excess heat can kill the organisms being isolated.

Troubleshooting

  • Strong Result Window line but no Control Window line: excess flagellar antigen is the likely cause; dilute the heated sample 1 in 10 in fresh UVM 2 and retest on a new device.
  • No line in either window: repeat the test from the same heated extract while it is still within 1 hour of heating; an older extract must not be reused.
  • Faint Result Window line: do not discount it; any visible blue line in the Result Window is a presumptive positive when the Control Window line is present.
  • Device read late or sample applied late: respect both timing limits, application within 1 h of the 80°C heat step and reading after 20 minutes at room temperature; a run outside those limits is not a valid result.
  • Positive screen needing confirmation days later: plate from the refrigerated UVM 2 culture, which the method allows to be held refrigerated for up to 72 hours after the test aliquot is taken.

Interferences and Limitations

Genus coverage has one hole: the flagellar B antigen targeted by the device is common to all Listeria species except L. grayi, so this screen cannot detect L. grayi. Very high antigen loads can also overwhelm the control line and force the dilution-and-retest step. A strip result is presumptive only; species identity, and any statement about L. monocytogenes in environmental samples, comes from cultural confirmation, not from the device. Approval extends only to the validated commodities listed under Scope; other foods need validation before results can be relied on.

Safety Notes

  • Treat every enrichment as potentially carrying L. monocytogenes, a human pathogen; use standard enteric precautions, contain aerosols during blending and stomaching, and dispose of cultures and used devices as biohazard waste.
  • Take care at the 80°C heating step: hot tubes and hot plates burn, and glass tubes should be handled with heat protection.
  • Disinfect the working area before analysis, as the method requires, and again after any spill of enrichment culture.

Source Method

Adapted in original wording from Health Canada method MFLP-71, Detection of Listeria spp. in Foods and Environmental Samples using the Oxoid Listeria Rapid Test, December 2013, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-71 dated February 2011.