SOP: Rapid PCR Screening for Salmonella in Foods and Surfaces (MFLP-38)
GIFSQ ID: SOP-1039
Speed comes from a single enrichment here. Food or a surface sponge sits overnight in buffered peptone water, a small aliquot is lysed, and real-time PCR then asks whether Salmonella DNA is present, with an internal control in every well watching for failed amplification. Software calls the result, but curve shape still needs a human check before anything is released.
A screen is not a confirmation. Any presumptive positive goes forward into cultural isolation, while a clean negative can be reported once controls and curves behave. Two extraction formats cover throughput: single tubes for small runs and a deepwell plate when batches grow.
Purpose
Purpose is rapid detection of Salmonella spp. in foods and on environmental surfaces for compliance under Sections 4 and 7 of the Food and Drugs Act and other relevant federal regulations. Positive screening results must be confirmed with a cultural method.
Scope
Scope covers all foods and a variety of environmental surfaces, validated as a set. Environmental devices are moistened with Dey-Engley (D/E) neutralizing broth or Letheen broth, and collection plus transport of sponges and swabs follows the environmental sampling procedure in MFLP-41: devices are kept cold, protected from contamination, and analysed as soon as possible after receipt. Work is performed under ISO/IEC 17025 competence requirements, and any substituted medium formulation needs equivalency shown by the laboratory using it.
Equipment and Materials
- iQ-Check Salmonella II PCR Detection kit (Bio-Rad catalog # 3578123): Reagent A lysis reagent, 1 bottle (20 mL) with magnetic stirbar; Reagent B fluorescent probes, 1 tube (0.55 mL); Reagent C amplification mix, 2 tubes (2.2 mL each); Reagent D PCR negative control, 1 tube (0.5 mL); Reagent E PCR positive control, 1 tube (0.25 mL)
- Buffered peptone water (BPW); D/E neutralizing broth; Letheen broth, for moistening environmental devices
- Conical screw-cap sterile tubes, 1.5 mL (tube extraction only); deepwell microplates 12.5 cm × 8.5 cm × 4 cm, catalog # 3594500 (deepwell extraction only); pre-pierced sealing film, catalog # 3600040 or equivalent (deepwell only)
- Low-profile unskirted or skirted PCR plates, 48 or 96 well, white (catalog # MLL4851XTU, MLL9651XTU, HSP9955XTU, HSP9655XTU); low-profile 0.2 mL PCR 8-tube strips without caps, thin-walled, white (catalog # TLS0851XTU); optical flat 8-cap strips, ultraclear (catalog # TCS0803XTU) or sealing film (catalog # MSB1001XTU)
- Powder-free gloves; sterile filter tips for 20 µL, 200 µL and 1000 µL micropipettes; stomacher bag with incorporated filter; micropipettes 20 µL, 200 µL and 1000 µL; positive and negative control cultures
- Bench-top centrifuge, maximum 10,000–12,000 × g, for 1.5 mL tubes (tube extraction only); dry heat block for 1.5 mL tubes capable of 95–100°C (tube extraction only); thermoshaker, Bio-Rad catalog # 3594995 base and 3594996 deepwell block or equivalent (deepwell only)
- CFX Manager Software Industrial Diagnostic Edition (IDE), catalog # 3593893; real-time PCR system, CFX96 (catalog # 3600037) or MiniOpticon; incubator capable of 37 ± 1°C; magnetic stir plate; stomacher, blender or equivalent; vortex apparatus
Procedure
- Hold sample units before analysis refrigerated or frozen according to product nature, except for shelf-stable foods. Thaw frozen units in a refrigerator or under time and temperature conditions that prevent microbial growth or death, and analyse units as soon as possible after receipt.
- Moisten sterile sponges and swabs with D/E neutralizing broth or Letheen broth. Collect and transport environmental samples following MFLP-41 practice.
- Pre-warm BPW to approximately 37°C before use. Make liquids and free-flowing materials homogeneous by agitation; for solids, build the analytical unit from portions taken at several locations so it represents the whole unit.
- Combine 25 g or 25 mL of each food with 225 mL BPW and blend, stomach or vortex as needed for thorough mixing. Place each environmental sponge in 100 mL BPW, or composite up to 10 sponges with 100 mL BPW per sponge; place each swab in 10 mL BPW, or composite up to 10 swabs with 10 mL BPW per swab. Stomach sponges for 2 min on high speed and vortex swabs for 2 min on high speed. Incubate every sample at 37°C ± 1.0°C for 21 ± 1 h.
- Prepare for analysis by switching on the dry heat block or thermoshaker early enough to reach 95–100°C, and switching on the thermal cycler and computer. Choose tube extraction (step 6) or deepwell extraction (step 7).
- Tube extraction: keep lysis reagent stirring on a magnetic stir plate at medium speed. Dispense 100 µL lysis reagent into each 1.5 mL screw-cap conical tube. Add a 100 µL aliquot of enriched sample, avoiding food particles, and mix by pipetting up and down at least 3 times. Heat tubes in the block at 95–100°C for 15 min, vortex at high speed, then centrifuge at 10,000–12,000 × g for 2 min.
- Deepwell extraction: keep lysis reagent stirring at medium speed as above. Dispense 100 µL lysis reagent into each well of a deepwell microplate, add a 100 µL aliquot of enriched sample to each well, and seal with pre-pierced sealing film. Shake in the thermoshaker at 95–100°C and 1,300 rpm for 15–20 min, then remove the plate and cool it in a refrigerator for 10 min.
- Build the PCR mix by combining amplification solution (Reagent C) with fluorescent probes (Reagent B) in the proportions given in the kit package insert chart. Dispense 45 µL of mix into each well of a 48- or 96-well PCR plate following the plate setup, pipetting carefully to avoid bubbles. Add 5 µL of each sample extract or control to its assigned well. Run one kit positive control (Reagent E) and one kit negative control (Reagent D) with each assay, alongside the laboratory’s standard broth and enrichment controls; kit controls are ready-to-use solutions.
- Seal the plate with optical caps or sealing film, load it into the thermal cycler reaction module and close the lid. Verify plate setup and protocol selection, start the run from the software, and save data to the chosen location when prompted; starting the save starts the PCR protocol. A report is generated at the end of the run.
- Understand what the instrument measures: the kit probe for the Salmonella target carries carboxyfluorescein (FAM), so fluorescence rises only when that conserved target sequence is amplified; a synthetic internal control in every reaction is amplified in parallel and read through a second fluorophore (HEX), so a well with no target signal but a valid internal control signal can be trusted as a true negative rather than a failed reaction.
- Verify amplification curves before releasing results. A valid curve shows a flat baseline, then a rapid exponential rise in fluorescence, then a plateau. Any other shape means the PCR is repeated from step 8. Curves that stay atypical after that repeat make the sample presumptive positive, handled under step 12.
- Confirm presumptive positives without delay: from the original enrichment broth, start cultural method MFHPB-20 at the selective enrichment stage, inoculating tetrathionate brilliant green broth and Rappaport-Vassiliadis soya peptone broth, and continue through plating and confirmation there.
- Prepare the three media from scratch only when commercial product is unavailable, exactly as follows. BPW: peptone 10.0 g, sodium chloride 5.0 g, disodium phosphate 3.5 g, potassium dihydrogen phosphate 1.5 g in 1.0 L distilled water, pH 7.2 ± 0.2, autoclaved 15 min at 121°C. Letheen broth: beef extract 5.0 g, proteose peptone No. 3 10.0 g, polysorbate 80 5.0 g, lecithin 0.7 g, sodium chloride 5.0 g in 1.0 L distilled water, pH 7.0 ± 0.2, autoclaved 15 min at 121°C. D/E neutralizing broth: tryptone 5.0 g, yeast extract 2.5 g, dextrose 10.0 g, sodium thioglycolate 1.0 g, sodium thiosulfate 6.0 g, sodium bisulfite 2.5 g, polysorbate 80 5.0 g, lecithin (soybean) 7.0 g, bromocresol purple 0.02 g in 1.0 L distilled water, pH 7.6 ± 0.2, autoclaved 15 min at 121°C.
Recording Results
Check kit controls before reading any sample: an assay whose positive or negative control falls outside its expected pattern in the CFX Manager IDE summary is invalid, and the PCR is repeated from the mix-preparation step. Read samples against that control check together with curve shape: a sample with a typical Salmonella (FAM) curve is presumptive positive, a sample with no target signal and a valid internal control (HEX) signal is negative, and a sample with neither signal is inhibited. Inhibition on a first run triggers a repeat from the PCR mix step; a sample that inhibits more than once, or that gives atypical curves more than once, is treated as presumptive positive and confirmed culturally. Report valid negatives as Salmonella Not Detected, and record enrichment time, extraction format, control outcomes, curve assessment and confirmation outcome for every unit.
Quality Control
Each assay carries the kit positive and negative controls plus the laboratory’s standard broth and enrichment controls, including positive and negative control cultures carried through enrichment. Incubator discipline matters because enrichment is the only growth step: temperatures at or below 37°C may vary by ± 1.0°C, while any higher growth temperature in a linked cultural confirmation must hold within 0.5°C. Keep lysis reagent homogeneous on the stir plate while dispensing, since settled reagent gives uneven lysis between wells.
Troubleshooting
- Curves lack the flat-then-exponential-then-plateau shape: repeat the PCR from mix preparation; persistent atypical curves convert the sample to presumptive positive for cultural confirmation.
- Controls outside their expected pattern: the run is invalid regardless of sample appearance; repeat the PCR from mix preparation.
- Inhibition pattern, no target and no internal control signal: repeat from mix preparation once; repeated inhibition is handled as presumptive positive, not as negative.
- Food particles carried into lysis: re-extract a fresh aliquot of enrichment, sampling the liquid phase only.
- Bubbles in PCR wells: dispense mix slowly and check wells before sealing, since bubbles distort fluorescence readings.
Interferences and Limitations
PCR detects target DNA after enrichment; it does not by itself prove viable Salmonella, which is why positives require cultural confirmation. Food components can inhibit amplification, and the internal control exists to expose that failure rather than let it pass as a negative. Compositing sponges or swabs dilutes a single positive device across the pool, so a pooled negative speaks for the pool as collected, not for each surface separately. Results outside the validated foods and surface types need separate validation before regulatory use.
Safety Notes
- Treat enrichments as potentially carrying Salmonella; use standard enteric precautions, gloves and biohazard disposal throughout.
- Handle the 95–100°C heat block, thermoshaker and freshly heated tubes and plates with thermal protection, and allow deepwell plates their refrigerated cooling step before handling extracts.
- Use powder-free gloves and sterile filter tips as specified, changing gloves between enrichment handling and PCR setup as general laboratory practice to protect the assay from contamination.
Source Method
Adapted in original wording from Health Canada method MFLP-38, Detection of Salmonella spp. in Foods and Environmental Surfaces Using iQ-Check Salmonella II PCR Detection Kit, February 2023, Compendium of Analytical Methods, Volume 3. Kit control and sample interpretation tables in the source are presented as instrument-software summaries; control validity and inhibition handling above follow the decision rules stated in that method’s text.