SOP: Rapid PCR Screening for Listeria Species on Environmental Surfaces (MFLP-112)
GIFSQ ID: SOP-1014
Speed comes from pairing one enrichment day with a real-time PCR read. Sponges and swabs from food-area surfaces grow for 24 hours in LPT broth, a small portion is lysed mechanically, and GENE-UP® Listeria spp. 2 (LIS 2) chemistry then looks for Listeria DNA using fluorescence resonance energy transfer, known as FRET, backed by a melt-curve check.
Each reaction carries its own internal amplification control, so a silent sample channel does not automatically mean absence. Software weighs amplification and melting behaviour together before calling a result, and every presumptive positive still goes forward to culture.
Purpose
Purpose is rapid detection of Listeria species on environmental surfaces, supporting compliance work under Sections 4 and 7 of the Food and Drugs Act and other relevant federal regulations. Presumptive findings can be ready after a 24 h enrichment, well ahead of a full cultural workup.
Scope
Scope covers a variety of validated environmental surface samples collected as sponges or swabs. Coverage extends across Listeria spp. with one named exception: Listeria grayi sits outside assay detection, so a negative screen never excludes that species. Collection devices arrive moistened with one of these agents: neutralizing buffer (NE), Dey and Engley neutralizing broth (D/E), buffered peptone water, Letheen broth or HiCap™ neutralizing buffer. Laboratories work under ISO/IEC 17025 competence requirements, and any substituted medium formulation needs proven equivalency before use.
Equipment and Materials
- GENE-UP® Thermocycler System (ref. 4704266): thermocycler, computer, GENE-UP® Routine software, MPS1000™ Mini PCR Plate Spinner (110V), Troemner vortex mixer, GENE-UP® tips (Biotix), GENE-UP® Lysis Tube Holder, GENE-UP® PCR Tube Holder, GENE-UP® Heavy Rack Holder, GENE-UP® Lysis Rack Adaptor, GENE-UP® Lysis Tube Remover and system user manual
- GENE-UP® Listeria spp. 2 (LIS 2) Kit (ref. 423106): ready-to-use freeze-dried pellets holding sample-specific primers and probes plus an internal amplification control, strip tubes and caps, and one pouch containing a control buffer vial of 850 µL
- GENE-UP® Lysis Kit (ref. 414057): two pouches, each holding one rack of 96 lysis tubes
- LPT enrichment broth, dehydrated (500 g, ref. 420984; 5 kg, ref. 420985) or ready to use (6 × 100 mL, ref. 410846; 20 × 10 mL, ref. 410845; 4 × 3 L Dilubag™, ref. 410849)
- Wetting agents: neutralizing buffer (NE), Dey and Engley (D/E) neutralizing broth, buffered peptone water, Letheen broth or HiCap™ neutralizing buffer (World Bioproducts)
- Calibrated adjustable variable-volume pipette, single or multichannel, with compatible sterile filter pipette tips; sterile stomacher bags; incubator holding 35 ± 1.0°C
Procedure
- Keep incoming sample units refrigerated inside the laboratory and begin analysis as soon as possible after receipt. Prepare and sterilize media and reagents by manufacturer directions.
- Bring prepared LPT broth to room temperature, 15-25°C, before adding any sample.
- Handle sponges this way: place each sponge into a sterile stomacher bag and add 100 mL LPT broth per sponge. Compositing may combine up to 10 sponges, still at 100 mL of broth for every sponge, with broth fully covering the sponges. Mix by hand for 1 minute.
- Handle swabs separately: place each swab into a tube holding 10 mL LPT broth. Compositing may combine up to 10 swabs, still at 10 mL for every swab. Rotate swabs by hand for 1 min.
- Incubate prepared sponges and swabs for 24 h ± 2 h at 35 ± 1°C.
- Mix each incubated sample by hand. Build a plate map inside GENE-UP® Routine software and use that map to count out lysis tubes from the lysis kit. Prepare blank, positive and negative controls according to laboratory requirements.
- Seat lysis tubes in the GENE-UP® Lysis Tube Holder, cutting strips apart when fewer than 8 tubes are needed, then clip the holder onto the GENE-UP® Heavy Rack Holder. Pipette 20 µL of enriched sample into each lysis tube, following plate-map positions exactly; lysis buffer shows red.
- Move the lysis tube holder from the heavy rack onto the GENE-UP® Lysis Rack Adaptor. Vortex at 2000-2200 rpm, depending on maximum mixer speed, for 5 minutes. Afterwards return the holder to the Heavy Rack Holder and move straight into PCR setup.
- Count PCR tubes from the LIS 2 kit against the plate map and seat them in the GENE-UP® PCR Tube Holder, cutting strips apart when fewer than 8 are needed. Remove transportation caps, tap strips on the bench so pellets settle at tube bottoms, then open caps carefully and visually confirm a freeze-dried pellet sits in every tube.
- Transfer 10 µL of lysed sample into its mapped PCR tube using a 10 µL Biotix filter tip on a single or multichannel pipette; lysed samples show red. For a negative control, pipette 10 µL of control buffer instead of lysate. Never agitate lysate before aspirating, because solid material must remain at the tube bottom, and inspect every tip for absence of beads and bubbles and for correct volume.
- Cap each strip using the GENE-UP® Lysis Tube Remover tool, cutting strips apart when fewer than 8 caps are needed. Set the PCR tube holder into the plate centrifuge, balance the centrifuge, and spin for 10 seconds.
- Start the run by following on-screen prompts on the GENE-UP® instrument and view results there. During cycling, paired FRET probes hybridize to an internal sequence of amplified target: one probe carries fluorescein at its 3′ end and its partner carries LC Red 640 at its 5′ end, so fluorescence arises only when both probes sit close together on template DNA. Once cycling ends, products are melted, and software combines the real-time amplification curve with the melt peak before calling any result.
Recording Results
Interpretation arrives automatically once a run finishes, judged from sample signal at 640 nm alongside internal amplification control signal at 705 nm. Positive sample signal means positive, whether the control channel reads positive or negative. Negative sample signal with a positive control signal means negative, and report such outcomes as Listeria spp. Not Detected. Negative signals in both channels mean invalid. Record plate-map position, sample type and compositing, enrichment time and temperature, control outcomes and the software call for every unit.
Treat every invalid result as a presumptive positive rather than a failure to discard. Confirm presumptive positives culturally from retained enrichment broth: plate and confirm by a cultural Listeria method, using Oxford agar plus at least one chromogenic agar, because isolating Listeria monocytogenes or other Listeria spp. from enrichment broth requires a chromogenic agar alongside Oxford agar.
Quality Control
Blank, positive and negative controls accompany each run, prepared as laboratory requirements specify, with a control-buffer negative carried through PCR setup. Laboratory supervisors must ensure analysts work within ISO/IEC 17025 competence requirements. Temperature discipline matters: growth steps at or below 37°C allow ± 1.0°C, so incubation runs at 35 ± 1.0°C, while any higher recommended temperature must hold within 0.5°C because excess heat can kill target organisms. As generic lab practice, check pellet presence, tip volume and plate-map alignment before starting a run, since software can only interpret tubes loaded in their mapped positions.
Troubleshooting
- Both channels negative (invalid): do not report a negative; treat the sample as presumptive positive and confirm culturally from enrichment broth as above.
- Missing pellet in a PCR tube: replace that tube before adding lysate; a tube without its freeze-dried pellet holds no primers, probes or internal control.
- Beads or bubbles seen in a tip: discard the transfer and repeat with a fresh sterile filter tip, confirming correct volume visually before dispensing.
- Cloudy or disturbed lysate: remember lysate must never be agitated before aspiration; draw from liquid while solids stay settled at the tube bottom.
- Sponges partly dry after broth addition: verify 100 mL per sponge was dispensed, including across composites, so broth completely covers every sponge.
Interferences and Limitations
Detection excludes Listeria grayi, the standing gap in species coverage. Results remain presumptive until culture confirms them, because PCR signals DNA presence through amplification and melt behaviour rather than demonstrating an isolate. Compositing caps at 10 sponges or 10 swabs, and enrichment beyond 24 h ± 2 h or outside 35 ± 1°C falls outside validated conditions. Sample types beyond the validated variety of environmental surfaces carry no performance claim under this procedure.
Safety Notes
- Treat every enrichment as potentially carrying Listeria spp., including L. monocytogenes; wear gloves and use biohazard disposal for bags, tubes and tips as generic lab practice.
- Keep PCR tubes capped after the run and dispose of them unopened, as generic practice to contain amplified DNA.
- Balance the plate centrifuge before its 10-second spin and keep hands clear of the vortex mixer during the 5-minute lysis run.
- Handle neutralizing and wetting agents by their safety data sheets, and clean spills of enrichment broth as potentially infectious material.
Source Method
Adapted in original wording from Health Canada method MFLP-112, Detection of Listeria Species in a Variety of Environmental Surface Samples Using GENE-UP® Listeria spp. 2 (LIS 2) Method, October 2021, Compendium of Analytical Methods, Volume 3.