SOP: Rapid Gene-Based Screening for E. coli O157:H7 in Foods with the Assurance GDS System (MFLP-16)
GIFSQ ID: SOP-1020
Shiga-toxigenic Escherichia coli O157:H7 demands a fast screen that does not confuse look-alikes. Assurance GDS builds selectivity into its primers and probes, aimed at a highly conserved target sequence, so O157 organisms lacking H7 or NM status and unrelated organisms carrying O157 antigen stay silent where antibody assays might react.
Concentration does heavy lifting before amplification begins. A magnetic capture device pulls target cells out of a short enrichment, away from competing flora and food debris, and that same concentration step later feeds cultural confirmation plates when a screen turns positive.
Purpose
Purpose is detection of E. coli O157:H7 and toxigenic E. coli O157:NM (non-motile) in foods, supporting compliance work under Sections 4 and 7 of the Food and Drugs Act and other relevant federal regulations.
Scope
Scope covers the validated category Raw Meat Products plus fruit juices, leafy greens and sprout process water. Use on other foods assumes fitness for purpose only after validation against Compendium criteria, with data forwarded to the Microbiological Methods Committee so validated commodities can be added. Sample handling preserves viability: keep units refrigerated or frozen by product nature, except shelf-stable foods; thaw frozen units in a refrigerator or under conditions preventing growth or death; analyse as soon as possible after receipt. Build each analytical unit representatively, agitating liquids or free-flowing materials to homogeneity and taking solid portions from several locations.
Equipment and Materials
- Assurance GDS for E. coli O157:H7 Tq kit, catalogue no. 71007-100, 100 tests (BioControl Systems): Amplification Tubes Tq with lyophilized detection reagents and an internal positive control, 25 strips of 4 tubes; O157 Sample Concentration Reagent, 1 vial; Resuspension Buffer Tq, 1 vial; Wash Solution, 1 bottle
- BioControl mEHEC enrichment medium, catalogue no. 65012
- Incubator holding 42°C; stomacher, blender or equivalent; adjustable micropipettes and a repeat pipette
- Assurance GDS Rotor-Gene system with computer and software; Assurance GDS sample concentration device (PickPen™); Assurance GDS vortex mixer; gel cooling block
- Sample wells and sample wells base; resuspension plates; pipette tips and PickPen tips; adhesive film strips
- Culture confirmation media and procedure per MFHPB-10 for plating concentrated samples
Procedure
- Clean the working area surface with disinfectant, and change gloves before handling reagents at both the concentration stage and the assay stage.
- Prepare mEHEC by the fresh route: pre-warm sterile deionized water, 225–1500 mL per sample, at 42°C overnight; on the day of use aseptically add BioControl mEHEC at 7.1 g per 225 mL water, keeping that ratio at larger volumes (47.4 g per 1500 mL), mix to dissolve, use within 6 h, and pre-warm finished medium to 40 ± 2°C before adding sample. Alternatively prepare in advance and autoclave: 31.6 g medium per litre deionized water, stir to dissolve, dispense, autoclave at 121°C for 15 minutes, and still pre-warm to 40 ± 2°C before use.
- Enrich a 25–64 g portion of raw meat product, fruit juice, leafy greens or sprout process water in pre-warmed mEHEC at 1 portion sample to 9 portions medium, for example 25 g in 225 mL. Stomach, vortex or homogenize as needed for thorough mixing. Incubate at 42°C for 6.5–18 h; sprout irrigation (process) water needs a minimum enrichment of 8 h.
- Enrich a larger 65–375 g raw meat portion at 1 portion sample to 4 portions mEHEC, for example 375 g in 1500 mL. Incubate at 42°C for 8–24 h.
- Set up concentration: vortex O157 Concentration Reagent, transfer 20 µL into each required well on the sample wells base, and cover wells with adhesive strips. Transfer 1 mL Wash Solution into additional sample wells, one per sample, and cover those too. Add 45 µL Resuspension Buffer Tq to resuspension plate wells and cover with adhesive strips.
- Peel the film from one strip of Concentration Reagent wells only. Add 1.0 mL incubated sample to each well, avoiding food particles and using a fresh pipette tip per sample, then reseal that strip with a new adhesive film before opening another strip. Return enrichment broths to the 42°C incubator immediately so incubation continues for confirmation; broths may incubate up to 18 h for 25–64 g samples or 24 h for 65–375 g samples, then refrigerate them for confirmation use.
- Place sealed wells holding Concentration Reagent plus sample on the vortex mixer at 900 RPM for 5–15 minutes, adjusting speed so liquid never touches the adhesive film.
- Remove and discard film from one strip after vortexing, and peel film from the matching Wash Solution strip. Load tips onto the PickPen, extend its magnetic tips into the first sample row, and stir gently for 30 seconds, moving continuously from surface to well bottom. Tap tips lightly against well sides to shed excess droplets.
- Move the PickPen into the matching Wash Solution wells and swirl gently for 5–10 seconds without releasing particles into the wash. Transfer to the matching resuspension plate row, submerge tips, retract the magnets, and tap gently so particles release into Resuspension Buffer Tq. Cover with an adhesive strip, then repeat the peel, capture, wash and release steps for every remaining strip with fresh tips per strip, and cover the finished resuspension plate with adhesive film strips.
- Freeze the gel cooling block before first use at −25 to −15°C for at least 6 hours; it changes to its ready colour when frozen, returns to the freezer between uses, and may need up to 2 h to be ready again. Complete Rotor-Gene set-up and data entry before moving samples. Place Amplification Tubes Tq from their foil pouch into the frozen block, reseal the pouch, and open tubes.
- Transfer 30 µL from each resuspension well into its Amplification Tube Tq, press each lid firmly closed, and inspect every tube visually for a secure seal. Invert tubes and shake with a snapping motion to mix, load them into the Rotor-Gene in sequential order starting at position 1, and start the cycle by the Assurance GDS user manual. Amplification and detection take about 75 minutes, primers and probes generate a target-specific fluorescence only where the conserved sequence is present, and a separate procedural-control fluorescence reads in every tube. Start the Rotor-Gene within 20 minutes of adding samples, and never open an Amplification Tube once the PCR process has started.
Recording Results
Read outcomes from the Rotor-Gene results table at run end. Positive means presumptive positive for E. coli O157:H7; Negative means negative for E. coli O157:H7; No Amp means amplification did not occur. For a No Amp outcome, repeat the whole procedure starting from concentration set-up (fresh gloves, Concentration Reagent wells onward); if No Amp repeats, contact BioControl Systems Technical Service at (800) 245 0113. Dispose of all run waste under biohazardous waste protocols, and record enrichment size branch, enrichment time and the table outcome for each sample.
Quality Control
Every Amplification Tube carries its own internal positive (procedural) control read on a separate fluorescence channel, so a tube without control signal cannot support a negative call and surfaces as No Amp instead. Gel block readiness is judged by its colour change after freezing at −25 to −15°C. Incubator accuracy follows the general rule: settings at or below 37°C allow ± 1.0°C, while higher settings such as 42°C work here must hold within 0.5°C unless a method states otherwise, because excess heat can kill target organisms.
Troubleshooting
- No Amp on a sample: repeat from concentration set-up with fresh gloves and reagents; a repeat No Amp goes to the manufacturer’s technical service rather than further in-house repeats.
- Liquid touching adhesive film during vortexing: lower the speed from 900 RPM until the film stays dry, keeping mixing time inside 5–15 minutes.
- Particles released into Wash Solution: swirl more gently for 5–10 seconds and keep magnet tips extended until the resuspension well.
- Delayed Rotor-Gene start: any run not started within 20 minutes of sample addition to Amplification Tubes is compromised; prepare fresh tubes from retained resuspension material.
Interferences and Limitations
Presumptive positives are never final: cultural confirmation decides whether E. coli O157:H7 or O157:NM is truly present. Confirmation starts from a fresh 1 mL portion of mEHEC incubated at least 8 h (25–64 g samples) or 10 h (65–375 g samples), repeating concentration exactly as in the main procedure except that 35 µL Wash Buffer replaces Resuspension Buffer Tq in the resuspension plate, then plating the resulting volume across confirmation media per MFHPB-10. Meanwhile the broth keeps incubating to a full 18 h or 24 h by size branch. If no confirmed colonies appear from the early broth, repeat cultural confirmation from the full-time broth, and only a negative confirmation from that full 18 h or 24 h enrichment lets a presumptive positive be reported negative.
Safety Notes
- Treat enrichments as potentially carrying Shiga-toxigenic E. coli; follow biohazardous waste disposal protocols for all assay waste, as the method directs.
- Never open Amplification Tubes after amplification starts, protecting both staff and the laboratory from amplified DNA.
- Handle hot autoclaved media, disinfectants and concentrated reagents with gloves and eye protection, as generic laboratory practice.
Source Method
Adapted in original wording from Health Canada method MFLP-16, Detection of Escherichia coli O157:H7 in foods: Assurance GDS for E. coli O157:H7 Tq Gene Detection System, December 2015, Compendium of Analytical Methods, Volume 3.