SOP: Monochrome Staining of Bacteria
GIFSQ — Global Institute for Food Safety & Quality
Principle
Certain dyes bind to bacterial cell walls and raise the contrast of the cells under the microscope. In monochrome (simple) staining a single dye is applied to a fixed smear, so the shape and size of the organisms — the morphology — can be observed and the nature of the staining noted.
Apparatus required
- Glass slides, 3 × 1 in.
- Wire loop (sterilizable in a flame)
- Bunsen burner or spirit lamp
- Filter paper
- Microscope with oil immersion lens
Reagents
a. Ziehl’s carbol fuchsin: solution A — mix 0.3 g of basic fuchsin with 10 ml of ethyl alcohol (95%); solution B — mix 5 g of phenol with 95 ml of distilled water. Mix solutions A and B.
b. Loeffler’s methylene blue: dissolve 0.3 g of methylene blue in 30 ml of ethyl alcohol.
c. Safranin: prepare a 2.5% solution of safranin O in 10 ml of ethyl alcohol (95%) and 100 ml of distilled water.
Procedure
Clean the glass slides thoroughly and wipe them with a dry cloth — on a clean grease-free slide, a drop of water spreads as a thin film
↓
Heat a wire loop to redness in the flame, let it cool, and transfer a loopful of the bacterial suspension from the broth to the centre of the slide (if the preparation is from a solid medium, touch the sterile loop lightly on the colony and transfer it to a drop of water placed on the slide); reflame the loop and cool it
↓
Spread the drop into a thin film on the slide and let it air dry
↓
Fix the dry smear by passing the slide rapidly through the low flame of a Bunsen burner or spirit lamp — apply only enough heat that the slide can be tolerated on the back of one’s palm
↓
Cover the smear with several drops of any one of the dyes: allow 15–20 sec in the case of carbol fuchsin, or 3–5 min in the case of methylene blue
↓
Wash carefully with water, press dry with filter paper, and examine under the microscope using the oil immersion lens; note the shape of the cells and the nature of the staining
Precautions
- The slide must be grease-free.
- Let the smear air dry completely before fixing; apply only enough heat that the slide can be tolerated on the back of one’s palm.
- Wash the dye off carefully with water.
References
Ranganna, S., Handbook of Analysis and Quality Control for Fruit and Vegetable Products, 2nd edition.