SOP: Identifying Vibrio parahaemolyticus by Multiplex PCR for R72H, tdh and trh (MFLP-23)

GIFSQ ID: SOP-1026

Not every Vibrio parahaemolyticus colony carries the same risk. Strains bearing the thermostable direct hemolysin gene (tdh) or its relative trh are the ones tied to illness, so identification has to answer two questions at once: is this really V. parahaemolyticus, and which hemolysin genes does it hold? A single multiplex PCR does both, reading the R72H species marker alongside tdh and trh, and it names presumptive colonies within about 4 hours.

Band size on a gel is the whole verdict, which makes the companion figure part of the method: a typical gel shows how the three products line up against a 50 bp ladder, and any lane pattern outside the expected table is sent back for repetition and biochemical confirmation rather than interpreted creatively.

Purpose

Purpose is the rapid, specific identification and characterization of Vibrio parahaemolyticus strains in foods, seawater, and environmental and clinical samples, working from presumptive positive colonies produced by the MFLP-37 detection method or equivalent methodologies. Used as the identification step inside a quantitative Most Probable Number (MPN) method, the procedure also generates quantitative results for the strains.

Scope

Scope covers presumptive colonies on selective/differential agar routinely used for this pathogen, such as TCBS. Primers are highly specific for V. parahaemolyticus and, under the stated reaction conditions, do not amplify DNA from any other Vibrio species or from non-Vibrio organisms. In most instances the PCR replaces the usual battery of downstream biochemical assays, saving time, labour and cost.

Equipment and Materials

  • Thermal cycler (Stratagene Robocycler or Applied Biosystems GeneAmp 9700) or equivalent; microwave oven or hot plate; submarine gel casting tray and buffer reservoir, power pack and mini-combs or equivalent
  • Shortwave UV transilluminator for stained gels; photodocumentation or digital imaging system
  • Adjustable micropipettors covering 0.5–10 µL, 10–100 µL, 20–200 µL and 200–1000 µL, with aerosol barrier filter tips (required)
  • Block heater capable of shaking and holding 1.5 mL microcentrifuge tubes (VWR Thermomixer) or equivalent; vortex mixer; microcentrifuge; strip-fuge for PCR tubes
  • Microcentrifuge tubes 1.5 mL; thin-walled PCR strip tubes and/or individual tubes with caps, 0.2 mL; DNA molecular weight marker (50 bp DNA ladder recommended)
  • Reagents: agarose; ultrapure nuclease-free, pyrogen-free sterile water; orthophosphoric acid (85%, 1.679 g/mL); Orange G or equivalent loading buffer; glycerol; 100 mM dATP, dCTP, dGTP and dTTP; ethidium bromide 10 mg/mL; EDTA; Taq DNA polymerase 5 units/µL; 50 mM MgCl2; 10X PCR buffer minus MgCl2; Tris[hydroxymethyl]aminomethane; Tris hydrochloride; de-ionized water for electrophoresis
  • Primers, commercial de-salted oligonucleotides (no HPLC purification needed): TDH-L 5′ GTAAAGGTCTCTGACTTTTGGAC 3′ and TDH-R 5′ TGGAATAGAACCTTCATCTTCACC 3′, product 270 bp; TRH-L 5′ TTGGCTTCGATATTTTCAGTATCT 3′ and TRH-R 5′ CATAACAAACATATGCCCATTTCC 3′, product 486 bp; VP33 5′ TGCGAATTCGATAGGGTGTTAACC 3′ and VP32 5′ CGAATCCTTGAACATACGCAGC 3′ for the R72H target, product 387 or 320 bp
  • Controls: positive control, a well-characterized V. parahaemolyticus strain such as ATCC 49529; negative controls, non-Vp material, plus a no-template water control

Procedure

  1. Take sample units through enrichment, isolation and plating under MFLP-37 or an equivalent method. Once presumptive positive colonies appear on the selective/differential agar (for example TCBS), sample the number of colonies that method specifies and put each individual colony through the multiplex PCR below. Process a positive control colony of a laboratory V. parahaemolyticus strain with the samples, along with a negative (non-Vp) control and a no-template water control.
  2. Lyse the cells: pick an isolated colony and inoculate 100 µL of sterile nuclease-free water, then vortex to resuspend. Set the controls up the same way, the no-template control being an uninoculated tube of 100 µL of sterile nuclease-free water.
  3. Heat the tube to 95°C for 5 minutes, snap-cool it on wet ice for 10 minutes, and hold the lysate under refrigeration until PCR. When PCR cannot be completed within 24 hours, store lysates at −20°C.
  4. Make the PCR master mix per reaction (23 µL master mix plus 2 µL template): 7.22 µL ultrapure nuclease-free sterile water, 2.50 µL 10X PCR buffer, 1.25 µL 50 mM MgCl2, 0.63 µL 10 mM dNTPs, 2.50 µL each of 5 µM VP-32 and VP-33 primers, 2.00 µL each of 5 µM TRH-L and TRH-R primers, 1.00 µL each of 5 µM TDH-L and TDH-R primers, and 0.40 µL Taq DNA polymerase (5 units/µL). Master mix without Taq can be prepared well ahead and stored at −20°C (stable up to one year), with Taq added immediately before use; stock solutions also store at −20°C, thawed reagents other than Taq are held at 4°C, and aliquoting into small working volumes limits extended refrigeration. For batch preparation, multiply per-reaction volumes by the number of analytes including controls and by 1.1 for a slight excess.
  5. Pipette 23.0 µL of master mix into a 0.2 mL thin-walled PCR tube, add 2.0 µL of cell lysate or negative control, and pipette up and down once to mix. Cap tightly, using a capping tool if needed, then vortex and spin the tubes in a strip-fuge. Keep tubes on ice or at 4°C until they go into the programmed, preheated thermal cycler, then start the program.
  6. Cycle as follows: initial denaturation 5 minutes at 94°C; then 40 cycles of denaturation 30 seconds at 94°C, annealing 45 seconds at 58°C and extension 75 seconds at 68°C; final extension 7 minutes at 68°C. Different cycler models perform differently, so parameters may need optimization on models other than those specified. Amplicons may wait under refrigeration until electrophoresis.
  7. Prepare a 2.0% (w/v) agarose gel in 1X TPE, 1X TBE or 1X TAE buffer, dissolving the agarose on a hot plate with stirring or in a microwave for 1–2 minutes at high power, stopping and starting to prevent excessive boiling. Confirm complete dissolution (a clear liquid with no suspended particles) and restore any evaporative loss gravimetrically with de-ionized water.
  8. Cool the agarose to 60°C, add ethidium bromide to a final 0.2 µg/mL, swirl gently, pour into the gel tray, insert a comb, and let the gel set for about 20 to 30 minutes.
  9. Add 3 µL of 10X Orange G (or equivalent loading buffer) to 25 µL of each PCR sample for an approximate final 1X concentration.
  10. Set the gel in the electrophoresis apparatus and fill the reservoir with the same 1X buffer used to make the gel, containing ethidium bromide at 0.2 µg/mL final, covering the gel to a depth of about 3–5 mm. Pipette 15 µL of sample into each well, and load 7.5 µL of 50 bp DNA ladder (0.1 µg/µL) into wells on either side of the samples; on gels with more than 20 wells, add further ladder lanes in the middle.
  11. Connect the power supply with the cathode (−, black) at the well end and the anode (+, red) at the far end. Run at 5–10 volts per cm of gel length (a 12 × 14 cm gel takes 70–140 volts) long enough to resolve the products; a 12 × 14 cm gel at 75 volts needs approximately 1 hour.
  12. Lift the gel out and visualize bands under shortwave UV on the transilluminator, documenting results with the imaging system for analysis and record keeping.

Recording Results

Products are double-stranded fragments of defined size: trh 486 bp, R72H 387 or 320 bp, and tdh 270 bp, sized against the ladder run on the same gel. Interpret each lane strictly by the band pattern. R72H positive alone (387 or 320 bp, no 486, no 270): Vibrio parahaemolyticus. R72H plus 270 bp: tdh-positive V. parahaemolyticus. R72H plus 486 bp: trh-positive V. parahaemolyticus. All three bands: tdh-positive, trh-positive V. parahaemolyticus. No bands at all: non-V. parahaemolyticus. The method’s companion figure (Figure 1, a gel image of a typical Vp multiplex PCR) shows exactly this reading practice: 50 bp ladder lanes flank the samples, lane A carries R72H and tdh bands only (trh negative, R72H positive, tdh positive), lane B is positive for all three targets, and lane C is negative for every target. A negative PCR normally shows no visible bands, and bands at sizes other than 486, 387 or 320, and/or 270 bp are non-specific amplification products, counted as negative however rare they are. Any pattern the table does not cover means the multiplex PCR is repeated and the isolate confirmed with the biochemical tests of MFLP-37 or an equivalent method. Optional biochemical confirmation may use API 20E strips or equivalent rapid kits in place of conventional tube media, with one adjustment: use 2% NaCl as the diluent, because some Vibrio spp. will not grow in commercial strip media suspended in physiological saline (0.85% NaCl); when strips fail to identify, continue with conventional tests under an HPFB method or equivalent.

Quality Control

Positive, negative and no-template controls are lysed and amplified alongside every batch of colonies, in the same 100 µL water format as samples. Ladder placement is also a control: flanking lanes on every gel, plus middle lanes on wide gels, keep size calls honest across the whole gel width. Buffer recipes behind the run are fixed: 10X PCR buffer is 200 mM Tris-HCl (pH 8.4) with 500 mM KCl; 10X TPE is built from 108 g Tris, 15.5 g orthophosphoric acid and 40.0 mL of 0.5 M EDTA (pH 8) per litre, stored at room temperature and diluted 1:10 for use; 10X Orange G is 0.25 g Orange G with 40.0 mL glycerol in 60.0 mL water, sterilized, aliquoted in 1 mL portions, stored at −20°C and held at 4°C once thawed; the ladder working solution mixes 50 µL of 50 bp ladder (1 µg/µL), 50 µL of 10X Orange G and 400 µL of water. All buffers, water, pipettes, tips and other materials touching sample preparation, PCR set-up and post-PCR handling (the electrophoresis step excepted) are autoclaved before use to remove DNases and other contaminants, and reagents are prepared in a dedicated work area.

Troubleshooting

  • Expected-size band in the negative control: contamination is presumed; the reagents are candidate sources and are discarded, samples are re-analysed with new reagents, and every aerosol contamination route is reviewed.
  • No bands from the positive control: check Taq addition timing (it goes in immediately before use), lysate storage beyond 24 hours without freezing, and master mix age against the one-year frozen limit.
  • Extra bands at unexpected sizes: these are non-specific products and the sample reads negative; if the pattern repeats, repeat the PCR and confirm biochemically instead of forcing an interpretation.
  • Smeared or unresolved gel: verify the agarose fully dissolved, the running buffer matches the gel buffer, and voltage stayed inside 5–10 volts per cm.

Interferences and Limitations

Identification rests on three loci only: an isolate is speciated by R72H and risk-typed by tdh and trh, and nothing in this procedure speaks to other virulence markers. Thermal cyclers other than the named models may need cycling optimization before band patterns can be trusted. Aged lysates and reagents held too long at refrigeration temperature lose reliability, since long-term stability of PCR reagents at 4°C is unknown. Results inconsistent with the interpretation table are never resolved by judgement at the bench; repetition plus biochemical confirmation is the prescribed route.

Safety Notes

  • Ethidium bromide is a potent mutagen: wear medical grade non-latex gloves whenever handling it, the stained gel, or the running buffer.
  • UV light can damage eyes: wear UV-blocking safety glasses at the transilluminator.
  • Treat presumptive Vibrio colonies and lysates as infectious until the 95°C lysis step is complete; use aerosol barrier tips for all PCR pipetting to protect both samples and reagents.
  • Hot agarose and microwave-heated buffer are burn and boil-over hazards; orthophosphoric acid used in TPE preparation is corrosive and handled with gloves and eye protection.

Source Method

Adapted in original wording from Health Canada method MFLP-23, Specific Detection of Vibrio parahaemolyticus Strains Using a Multiplex Polymerase Chain Reaction (PCR) Based on the R72H Taxonomic Marker and the Hemolysin Genes tdh and trh, March 2006, Compendium of Analytical Methods, Volume 3, including its companion Figure 1 gel image document.