SOP: Identifying Vibrio parahaemolyticus Colonies by Real-Time PCR (MFLP-102)
GIFSQ ID: SOP-1003
Confirming a presumptive Vibrio parahaemolyticus colony by biochemistry takes a battery of tests. A multiplex real-time PCR collapses that work into about 2 hours by asking three genetic questions at once: is the species marker tlh present, and are either of the virulence-associated hemolysin genes, tdh or trh, carried alongside it? Fluorogenic probes report each target as amplification proceeds, and a built-in internal amplification control (IAC) shows whether a negative reaction actually worked.
Starting material is a colony, not a food homogenate. Presumptive colonies come off validated selective or differential agar first, and PCR then names them and characterises their hemolysin profile in a single run.
Purpose
Purpose is rapid identification of presumptive V. parahaemolyticus colonies isolated by standard culture on validated plating media, replacing the usual downstream battery of biochemical assays, with simultaneous reporting of tdh and trh status.
Scope
Scope covers individual presumptive colonies from thiosulfate-citrate-bile salts-sucrose (TCBS) agar or CHROMagar Vibrio, and colonies on tryptic soy agar supplemented with 1.5% sodium chloride (TSA-NaCl) provided those colonies were first isolated from TCBS agar or CHROMagar Vibrio. Application includes compliance work under food legislation. Identification rests on three targets amplified together — tlh (thermolabile hemolysin gene, specific to V. parahaemolyticus), tdh (thermostable direct hemolysin gene) and trh (tdh-related hemolysin gene) — plus the IAC. No substitutions are permitted for the specified Platinum Taq DNA polymerase.
Equipment and Materials
- Thermal cycler, Roche LightCycler 96 or equivalent; biohazard hood and/or PCR workstation with UV lights; block heater capable of shaking and holding 1.5 mL microcentrifuge tubes (Eppendorf Thermomixer or equivalent)
- Adjustable micropipettors covering 1.0 µL to 1000 µL with aerosol barrier filter tips; vortex mixer; microcentrifuge for 1.5 mL tubes; mini-centrifuge for strips of 8 × 0.2 mL PCR tubes or a centrifuge with 96-well plate adaptors; freezer at −20°C; refrigerator at 2°C to 8°C; ice and bucket; chiller blocks for 0.2 mL and 1.5 mL tubes
- Disposable gloves; 1.5 mL microcentrifuge tubes and racks; thin-walled 0.2 mL PCR tubes with optical caps or film, in strip or 96-well plate format
- Sterile nuclease-free water (ultra-pure or molecular grade); dNTP mix, PCR grade, 10 mM; Platinum Taq DNA polymerase (ThermoFisher catalogue 10966-034) with its supplied 10X PCR buffer minus MgCl2 and 50 mM MgCl2
- Primers and probes as listed in the Procedure; IAC template DNA; EcoRI restriction endonuclease; 10 mM Tris-HCl (pH 8.0); for IAC quantification, a small-volume spectrophotometer (ThermoFisher NanoDrop-1000 or equivalent), a heating block or water bath at 37°C and 65°C, and electrophoresis equipment
- Controls: positive control V. parahaemolyticus carrying tlh, tdh and trh (e.g. ATCC 49529); negative control another Vibrio spp. (non-V. parahaemolyticus); plus a no-template water control in each run
- All reagents are molecular-grade and nuclease-free, prepared in a dedicated clean work area; work is to be done in line with ISO/IEC 17025 requirements, and variants of listed formulations need an equivalency demonstration
Procedure
- Lyse colonies. Pick one isolated colony with morphology typical of V. parahaemolyticus from TCBS agar, CHROMagar Vibrio or TSA-NaCl and inoculate 100 µL of sterile nuclease-free water; vortex to resuspend. Set up the same way a positive control colony, a negative control colony, and a no-template control consisting of an un-inoculated tube of 100 µL sterile nuclease-free water.
- Heat each tube at 95°C for 5 minutes with constant shaking at approximately 400 rpm. Snap cool in an ice bath for 10 minutes and store refrigerated at 2°C to 8°C until PCR. If PCR cannot be completed within 24 hours, store lysates at −20°C. Centrifuge tubes to bring down condensate before use.
- Use these primers (5′ → 3′), obtained as de-salted oligonucleotides without HPLC purification: TDH-F TCC CTT TTC CTG CCC CC; TDH-R CGC TGC CAT TGT ATA GTC TTT ATC; TRH-F TTG CTT TCA GTT TGC TAT TGG CT; TRH-R TGT TTA CCG TCA TAT AGG CGC TT; TLH-F ACT CAA CAC AAG AAG AGA TCG ACA A; TLH-R GAT GAG CGG TTG ATG TCC AA; IAC-F CTA ACC TTC GTG ATG AGC AAT CG; IAC-R GAT CAG CTA CGT GAG GTC CTA C.
- Use these probes (5′ → 3′): TDH probe TGA CAT CCT ACA TGA CTG TG, 5′ FAM and 3′ MGBNFQ; TRH probe AGA AAT ACA ACA ATC AAA ACT GA, 5′ VIC and 3′ MGBNFQ; TLH probe CGC TCG CGT TCA CGA AAC CGT, 5′ TxRED and 3′ BHQ2; IAC probe AGC TAG TCG ATG CAC TCC AGT CCT CCT, 5′ Cy5 and 3′ IAbRQSp. FAM is 6-carboxyfluorescein, TxRED is Texas Red, Cy5 is Cyanine-5, MGBNFQ is minor groove binding non-fluorescent quencher, BHQ2 is Black Hole Quencher 2, and IAbRQSp is Iowa Black RQ Dark Quencher.
- Prepare IAC DNA. Purchased IAC DNA arrives lyophilised in a pIDTSMART-KAN plasmid with an EcoRI site, total size 2.3 kb, or an equivalent vector. Rehydrate with 10 mM Tris-HCl (pH 8.0) to approximately 50 to 100 ng/µL, linearise with EcoRI, and dilute as quantified below; alternatively propagate the IAC in-house in competent E. coli by transformation with a commercial kit such as One Shot TOP10 Electrocomp E. coli or an acceptable alternative. Its sequence (GenBank FJ357008.1, 5′ → 3′) is: TAC AGC ACC CTA GCT TGG TAG AAT CGA TCA GCT ACG TGA GGT CCT ACG ACG ATC GCC AAG CAT GCC CTA GCT AAG ATG CAT CGA TTG CTC ATC ACG TAC GTT AGG TCG ACT AGG AGG ACT GGA GTG CAT CGA CTA GCT AAG ATG GTT CGA TTG CTC ATC ACG AAG GTT AGG TCG ACT ACG AAC GAG TCG TAC GTT AGG TT.
- Linearise plasmid for the IAC stock by combining 40 µL purified plasmid DNA, 12 µL restriction endonuclease buffer supplied with the enzyme, 60 µL RNase/DNase-free water and 10 µL EcoRI (10 U/µL) in a microcentrifuge tube, following the enzyme manufacturer’s current instructions. Incubate 4 hours at 37°C, then inactivate the enzyme for 20 minutes at 65°C. Verify by electrophoresis against undigested plasmid, diluting both (e.g. 1 in 10) in DNase-free water if desired: the EcoRI fragment is 2.3 kbp (2.1 kbp plasmid plus 200 bp IAC), a single band should appear for linearised plasmid, and undigested plasmid should run as a larger band because of supercoiling.
- Quantify linearised plasmid with a small-volume spectrophotometer and calculate copy number per µL as (concentration in g/µL ÷ molar mass of the IAC plasmid in g/mol) × 6.023 × 1023, taking molar mass from the plasmid supplier. Worked example from the source: 8.1 × 10−9 g/µL divided by 1.3 × 106 g/mol, times 6.023 × 1023, gives 3.75 × 109 copies per µL. Dilute digested plasmid in 10 mM Tris-HCl (pH 8.0) to a stock of not less than 650 fg/µL, aliquot in small volumes (e.g. 20 µL) to limit freeze-thaws, and store at −20°C. Dilute the stock in ultra-pure nuclease-free sterile water immediately before master mix preparation to reach 20 fg/µL.
- Make master mix fresh and use within 6 hours; keep all stocks at −20°C until use. Per single reaction (23 µL master mix plus 2 µL template): ultra-pure nuclease-free sterile water 4.72 µL; 10X PCR buffer 2.5 µL (1X final); 50 mM MgCl2 2.5 µL (5 mM); 10 mM dNTPs 1.0 µL (400 µM); 5 µM TRH-F 1.0 µL (200 nM); 5 µM TRH-R 1.0 µL (200 nM); 5 µM TDH-F 1.0 µL (200 nM); 5 µM TDH-R 1.0 µL (200 nM); 5 µM TLH-F 0.375 µL (75 nM); 5 µM TLH-R 0.375 µL (75 nM); 5 µM IAC-F 2.0 µL (400 nM); 5 µM IAC-R 2.0 µL (400 nM); 5 µM TRH probe 0.375 µL (75 nM); 5 µM TDH probe 0.375 µL (75 nM); 5 µM TLH probe 0.75 µL (150 nM); 5 µM IAC probe 0.75 µL (150 nM); IAC DNA at 9375 copies/µL (20 fg/µL) 1.0 µL (375 copies/µL, 620 aM final); Platinum Taq DNA polymerase (8 U/µL) 0.28 µL (2.2 U per reaction). Total 23.0 µL. For a batch, multiply each volume by the number of samples including controls and by 1.1 to allow for pipetting loss. Adjust IAC final concentration if needed to achieve a Cq of approximately 25.
- Pipette 23 µL master mix into 0.2 mL thin-walled PCR tubes or a 96-well plate, add 2 µL cell lysate or control, and pipette up and down once to mix. Cap tightly, using a capping tool if needed, or apply optical film to a plate. Vortex gently, spin briefly to collect liquid at the bottom, and keep tubes on ice or chilled until loading into the pre-heated thermal cycler.
- Run this program: initial denaturation 1 minute at 95°C, then 45 cycles of denaturation 5 seconds at 95°C and annealing and extension 45 seconds at 59°C. Different cycler brands or models may perform differently, so optimise cycling parameters if a model other than the specified one is used and demonstrate equivalence.
Recording Results
Read each reaction against a minimum endpoint fluorescence (EPF) of 0.25 for tlh, tdh, trh and IAC on the Roche LightCycler 96 platform; the threshold exists to reduce background noise. A laboratory using another platform must show results equivalent to the LightCycler 96 with EPF set at 0.25 for each target. Examine every amplification plot for baseline, exponential, linear and plateau phases to confirm proper amplification. Interpret the pattern as follows: tlh positive alone, with IAC positive or negative, is V. parahaemolyticus; tlh plus tdh is tdh-positive V. parahaemolyticus; tlh plus trh is trh-positive V. parahaemolyticus; tlh, tdh and trh together is tdh-positive trh-positive V. parahaemolyticus; all three targets negative with IAC positive is non-V. parahaemolyticus; and all targets including IAC negative is an invalid result, i.e. reaction failure.
Quality Control
Process a positive control colony, a negative control colony of another Vibrio species, and a no-template water control alongside every batch of colonies, through lysis and PCR alike. Judge negative validity through the IAC: a negative result counts as valid only if the IAC Cq sits within two cycles of the IAC in the negative control, at a Cq of approximately 25. Master mix age is a control point, prepare and use within 6 hours, stocks held at −20°C, as is reagent identity for the polymerase, for which no substitution is permitted. Verify IAC plasmid linearisation by electrophoresis before quantification whenever a new stock is made.
Troubleshooting
- All targets and IAC negative: record an invalid result from reaction failure; check master mix age, polymerase identity, cycling program and template addition before repeating.
- IAC absent or delayed in a negative sample: treat as evidence of PCR inhibition, results are not valid; the IAC Cq must fall within two cycles of the negative control IAC.
- IAC signal missing in a strongly positive sample: expected behaviour at high target load, because IAC is present at low concentration so it does not interfere with tlh, tdh or trh; judge the sample by its target pattern.
- Weak or noisy fluorescence: confirm the 0.25 EPF threshold is applied per target, inspect plot phases, and on a non-LightCycler platform re-check demonstrated equivalence at that threshold.
- Multiple bands after EcoRI digestion: linearisation failed or was incomplete; repeat digestion following the enzyme manufacturer’s instructions and re-verify against undigested plasmid.
Interferences and Limitations
PCR inhibition is the central limitation, and the IAC exists to expose it: absence or delay of the IAC reaction means a negative cannot be trusted. Identification applies only to colonies already isolated on validated selective or differential media, with the extra rule that TSA-NaCl colonies must trace back to TCBS or CHROMagar Vibrio isolation. Platform differences matter as well, an alternative cycler needs optimised parameters and equivalence to the LightCycler 96 at EPF 0.25. The assay identifies colonies and their hemolysin genes; it does not enumerate organisms in a food or prove toxin production.
Safety Notes
- Handle V. parahaemolyticus cultures and lysates in a biohazard hood with disposable gloves and standard precautions for a foodborne pathogen, and dispose of cultures and PCR waste as biohazardous material.
- Do PCR set-up in a dedicated clean area or PCR workstation, using aerosol barrier tips and nuclease-free consumables, to protect both staff workflow and reaction integrity.
- Take care with the 95°C shaking heater, ice baths and UV lights in the PCR workstation as ordinary laboratory practice requires.
Source Method
Adapted in original wording from Health Canada method MFLP-102, Identification of Vibrio parahaemolyticus Colonies by Real-Time Polymerase Chain Reaction, July 2017, Compendium of Analytical Methods, Volume 3.