SOP: Identifying Listeria monocytogenes Isolates by PCR of the hlyA Gene (MFLP-78)
GIFSQ ID: SOP-1064
Colonies do the travelling in this method; the PCR itself starts from growth already isolated on selective agar. A touch of a suspect colony is lysed in buffer and detergent, and that crude lysate, no DNA purification step, goes straight into a reaction built around two primers for the listeriolysin gene, hlyA. One band of a defined size on an agarose gel then answers the identification question, usually within 4 hours for the PCR stage.
Specificity carries the whole design. Primers chosen from the published hlyA sequence amplify L. monocytogenes DNA and, per the source, do not amplify DNA from any other Listeria species or non-Listeria organism, so a correct-size product is a species call rather than a genus hint.
Purpose
Purpose is the rapid identification of Listeria monocytogenes, to determine compliance with Sections 4 and 7 of the Food and Drugs Act. In most instances the procedure can replace the usual battery of biochemical screening and confirmation tests, saving time, labour and cost, although, as the Procedure states, within MFHPB-30 and MFHPB-07 this PCR step is optional and cannot be used as a replacement for their biochemical confirmation steps.
Scope
Validation covers isolates from foods, and environmental and clinical samples, obtained using Health Canada methods MFHPB-30, MFHPB-07 and MFLP-74. Work begins at the point where presumptive positive colonies are visible on selective plating media; sample enrichment, isolation and plating themselves follow those methods. Typical colonies are as described in MFHPB-30, and plates named in the source include PAL, OX and/or MOX selective agars.
Equipment and Materials
- Thermocycler; microwave oven or hot plate; submarine gel casting tray and buffer reservoir, power pack and an appropriate mini-comb; shortwave UV light table (transilluminator); photodocumentation system (optional), including Polaroid camera (hand-held or fixed), hood and Polaroid 667 film
- Adjustable micropipettors covering 0.5–10 µL, 10–100 µL and 200–1000 µL, with filtered pipet tips; block heater with a block accommodating 1.5 mL microfuge tubes; vortex mixer
- Microfuge tubes of 1.5 mL capacity and of 0.2 mL (200 µL) capacity, or equivalent; DNA molecular size marker suitable for differentiating PCR products within the expected range for L. monocytogenes
- Chemicals and reagents: agarose; boric acid; bromophenol blue; dATP; dCTP; dGTP; dTTP; ethidium bromide, SYBR Safe, SYBR Green or equivalent; EDTA; Taq DNA polymerase; MgCl2; KCl; sucrose; PCR buffer 10X; Triton X-100; Tris[hydroxymethyl]aminomethane (for example Trizma base); mineral oil (optional, depends on method used)
- Primers for a 730 bp hlyA fragment: primer A, 5′-CATTAGTGGAAAGATGGAATG-3′, and primer B, 5′-GTATCCTCCAGAGTGATCGA-3′. Primers need not be of high purity; they can be synthesised on a suitable oligonucleotide synthesis system to manufacturer instructions, or synthesis can be contracted out to a university or biotechnology company
- Positive control: a colony or broth culture (Trypticase Soy Broth) of a L. monocytogenes laboratory strain
Procedure
- Prepare and enrich food samples under MFHPB-30, MFHPB-07 or MFLP-74 for isolation of L. monocytogenes. At the stage where presumptive positive colonies are observed on plating media, sample a number of colonies, in addition to the mandatory biochemical tests, and subject each individual colony to the PCR below. Process a positive control, a colony or broth culture (Trypticase Soy Broth) of a L. monocytogenes laboratory strain, along with each set of samples. Prepare all reagents in a laminar flow cabinet that has never been exposed to L. monocytogenes or PCR products, and autoclave all buffers, distilled water, pipets, pipet tips and other materials contacting samples or PCR reagents before use, to remove DNases and other contaminants.
- Make the PCR reaction mixture ahead where convenient. Per single reaction tube, combine 32.5 µL sterile distilled H2O, 5.0 µL 10X PCR buffer, 4.0 µL of 25 mM MgCl2, 0.5 µL of 100 µM primer A, 0.5 µL of 100 µM primer B, and 0.5 µL each of 20 mM dNTPs (dCTP, dATP, dTTP, dGTP), for a final volume of 44.5 µL. In practice a large batch equivalent to 100 reactions may be prepared and 44.5 µL aliquots dispensed into 0.2 mL microfuge tubes well ahead of time and stored at −20°C until use, stable for up to one year; all stock solutions are likewise stored at −20°C until use. Commercial 10X PCR buffer, generally supplied with the Taq solution, usually consists of 500 mM KCl, 100 mM Tris-HCl (pH 9.0) and 1% (w/v) Triton X-100.
- Lyse the cells. Pick a small portion of a suspect colony, lightly touching the surface with a loop or needle is sufficient, and suspend it in 50 µL of 1X PCR buffer in a sterile 1.5 mL microfuge tube. For a negative control, leave one tube of 50 µL of 1X PCR buffer devoid of cells.
- Add 50 µL of 2% (w/v) Triton X-100 to each tube holding cell suspension or negative control, and vortex. Incubate at 100°C for 10 minutes, cool to room temperature, and centrifuge the tube at 2000 rpm for a few seconds.
- Set up each reaction by adding 5 µL of cell lysate or negative control to 44.5 µL of PCR reaction mixture in a microfuge tube suitable for thermocycler use, and vortex. Where the thermocycler has no heated lid, add 2 drops (100 µL) of sterile mineral oil to each tube and place 2 drops of mineral oil in each well of the thermocycler.
- Place tubes into the thermocycler and start the cycling program. Different makes of thermal cycler may perform differently, so cycling parameters may need optimising for each model. Enzyme units required may vary with the Taq used: follow manufacturer instructions. Where a hot start Taq polymerase is used, follow the manufacturer protocol in place of the manual hot start procedure.
- Run Version a, the manual hot start program: hold at 80°C for 10 min; denature for 5 min at 94°C; add 0.5 units of Taq; denature for 3 min at 94°C; then run 30 cycles of denaturation for 1 min at 94°C, annealing for 1 min at 55°C and extension for 2 min at 72°C; finish with a final extension of 2 min at 72°C.
- Run Version b, the hot start Taq program, as the alternative branch: denature for 10 min at 94°C; add 0.5 units of Taq; denature for 3 min at 94°C; then run 30 cycles of denaturation for 1 min at 94°C, annealing for 1 min at 55°C and extension for 2 min at 72°C; finish with a final extension of 2 min at 72°C. Version b may need optimisation, especially of annealing temperatures, to reflect conditions obtained with Version a. With Version b a 0.2 mL tube is used, and a typical reaction volume is 22.5 or 45 µL of PCR reaction mixture with 2.5 or 5 µL, respectively, of template.
- Handle Taq addition carefully during either hot start. Where mineral oil is in use, insert the tip carefully under the mineral oil layer, return the tube to the machine immediately after adding the enzyme, and use a separate pipet tip for each addition of Taq.
- After PCR completes, remove tubes and analyse by agarose gel electrophoresis. If analysis cannot happen immediately, tubes can be stored at −20°C.
- Prepare a 1.2% (w/v) agarose gel in 0.5X TBE (Tris-Borate-EDTA) buffer. Dissolve the agarose by stirring on a hot plate or by microwaving for 1–2 minutes on high power, ensuring it is completely dissolved, a clear liquid with no particles in suspension, and make up volume with 0.5X TBE buffer if evaporation has reduced it. Cool the agarose to 50°C, pour it into a gel tray, add a well-forming comb, and allow the gel to set for about 30 min.
- Prepare samples for electrophoresis in clean microfuge tubes by mixing 5 µL of tracking dye with 10 µL of PCR sample. Where oil was used, take the sample from under the oil layer and wipe the pipet tip with a tissue after taking the sample.
- Load and run the gel. Once set, remove the comb and place the tray with gel in the electrophoresis apparatus. Fill the reservoir with 0.5X TBE buffer, covering the gel to a depth of 5 mm. Gently pipet 10–15 µL of each prepared sample into the wells of the submerged gel, and pipet 10–15 µL of a suitable DNA molecular size marker into an empty well. Connect the power supply with the cathode (−, black) at the top, near the sample wells, and the anode (+, red) at the bottom end of the gel. Apply 80–100 volts (approximately 30 mA) to each gel and run for about 1 hour, or until the tracking dye has spread about two-thirds the length of the gel.
- Stain and view the gel. Remove it from the tray and place it in ethidium bromide (EtBr) solution (10 µg/mL) for 15 min. Lift the gel out with a gel scoop, rinse briefly with tap water, and visualise DNA bands under ultraviolet light (shortwave) on a transilluminator. Photograph gels, using Polaroid 667 film or other suitable methods, to facilitate analysis and for record keeping.
Recording Results
Judge every lane against the marker run on the same gel. Amplicon generated from L. monocytogenes hlyA sequences by this method is a double-stranded DNA fragment of 730 bp, so a positive test yields a 730 bp fragment appearing as an intense band on an EtBr-stained agarose gel. A negative test normally produces no visible bands. Bands that do not correspond to the 730 bp amplicon, non-specific amplification products, an extremely rare occurrence, are considered negative. Record colony source and plate, control outcomes, cycling version used, and band size for each isolate tested.
Quality Control
Every set carries a positive control of a L. monocytogenes laboratory strain and a negative control tube of 1X PCR buffer without cells, carried through lysis and amplification alongside the samples. A DNA molecular size marker runs on every gel so the 730 bp call rests on migration comparison, not estimate. Reaction mixture aliquots and stock solutions are held at −20°C, with aliquots stable for up to one year. Contamination discipline is part of control here: reagents are prepared in a cabinet never exposed to L. monocytogenes or PCR products, and everything contacting samples or reagents is autoclaved first to remove DNases and other contaminants.
Troubleshooting
- 730 bp band in the negative control: contamination is present in Triton X-100, mineral oil, the PCR reaction mixture, or bacteria in aerosol form. Discard those reagents, review every point where aerosol contamination may have occurred, and re-analyse the samples with new reagents.
- Bands at the wrong size: read them as negative, non-specific amplification products do not count, however intense they appear.
- No band from a suspect colony: check that only a light touch of colony was taken, that lysis ran the full 10 minutes at 100°C, and that cycling parameters suit the thermocycler model in use; models differ and may need optimisation.
- Weak or absent signal after switching to hot start Taq: optimise the Version b protocol, especially annealing temperature, against conditions obtained with the standard Version a protocol, and follow the enzyme manufacturer instructions for unit requirements.
- Cloudy, uneven gel: redissolve the agarose completely, clear liquid, no particles, and restore any volume lost to evaporation with 0.5X TBE before pouring.
Interferences and Limitations
Identification applies only to colonies already isolated as presumptive positives; this procedure does not enrich, isolate or enumerate the organism from the original food. Within MFHPB-30 and MFHPB-07 the PCR is an optional step and cannot replace their biochemical confirmation steps, even though the source notes it can in most instances replace the usual battery of biochemical screening and confirmation tests. Aerosol-borne bacteria or amplified product are the standing interference: one contaminated reagent or aerosol event invalidates a run through its negative control. Mineral oil handling also affects recovery, sample for the gel must be drawn from under the oil layer with the tip wiped afterwards.
Safety Notes
- Ethidium bromide is a mutagen: use extreme caution, wear double gloves when handling it, and always wear gloves and protective clothing, changing gloves often, especially when handling gels or electrophoresis buffer or whenever contamination with EtBr is possible.
- Discard EtBr-contaminated waste in a designated receptacle for disposal by hazardous waste contractors, and decontaminate gel buffer in the correct filters before disposal down the drain.
- Ethidium bromide can be replaced by SYBR Safe, SYBR Green or equivalent; follow manufacturer recommendations for handling and disposal of the substitute.
- UV light can cause eye damage: wear safety goggles when using the transilluminator.
- Take care with the 100°C lysis step, hot agarose, and microwave or hot plate heating of gels.
Source Method
Adapted in original wording from Health Canada method MFLP-78, Identification of Listeria monocytogenes from foods and environmental samples by the Polymerase Chain Reaction (PCR), February 2011, Compendium of Analytical Methods, Volume 3. This revised method replaced MFLP-78 dated January 2006.