SOP: Detecting Salmonella in Foods with the 3M Molecular Detection System (MFLP-100)
GIFSQ ID: SOP-1001
Molecular screening shortens the wait for a Salmonella answer without abandoning culture. Enrichment grows viable cells first, loop-mediated isothermal amplification (LAMP) then copies a target sequence at a single temperature, and a bioluminescent readout reports that copying as it happens. Presumptive positives appear in real time, negatives only at the end of the run, and every positive still goes forward to cultural confirmation.
Only two transfer steps separate enrichment from result: enriched sample into lysis solution, then lysate into reagent tubes. Care at those transfers, especially around cross-contamination and protein residue, decides whether speed stays reliable.
Purpose
Purpose is rapid detection of Salmonella spp. in foods for compliance work, using the 3M Molecular Detection Assay 2: Salmonella test kit on the 3M Molecular Detection Instrument after enrichment in buffered peptone water.
Scope
Scope covers foods generally, validated for all foods except chocolate based products, spices, powdered dairy products and whole nuts; do not apply this screen to those excluded matrices. A positive is presumptive and must be confirmed culturally. Keep sample units refrigerated or frozen according to product nature, except for shelf-stable foods. Thaw frozen units in a refrigerator, or under time and temperature conditions that prevent microbial growth or death, and analyse as soon as possible after receipt. For a representative unit, agitate liquids or free-flowing materials until homogeneous, and for solids take portions from several locations within the unit.
Equipment and Materials
- 3M Molecular Detection Instrument and Accessory Kit (MDS100): Speed Loader Tray, Chill Block Insert, Chill Block Tray, Cap/Decap Tool, Lysis, Cap/Decap Tool, Reagent, and empty lysis and reagent trays
- 3M Molecular Detection Assay 2: Salmonella spp. test kit (MDA2SAL96): Lysis Solution (LS) tubes (96, as 12 strips of 8), Salmonella spp. Reagent tubes (96, as 12 strips of 8), extra caps (96), Reagent Control (RC) tubes (16), and Quick Start Guide
- Buffered peptone water (BPW), prepared and sterilised following manufacturer instructions; 3M Molecular Detection Heat Block Insert (MDSHBIN); 3M Molecular Detection Matrix Control (MDMC96), optional
- Powder-free gloves; sterile barrier filter tips adaptable to 20 µL micropipettes; sterile stomacher bags; single-channel 20 µL micropipette; 8-channel 20 µL micropipette; computer workstation
- Incubators settable at 37 ± 1°C and 41.5 ± 0.5°C; dry block heating unit able to fit the Heat Block Insert (15 cm × 9.5 cm) and reach 100°C; stomacher, blender or equivalent
- Work is to be done in line with ISO/IEC 17025 requirements for testing laboratories, and any variation in a listed medium formulation needs an equivalency demonstration by the laboratory using it
Procedure
- Ready the bench and instruments: have sterile BPW prepared and acclimatised to room temperature, clean the working surface with a suitable disinfectant, switch on the block heater with the Heat Block Insert in place so it can reach 100°C before analysis, and log in to the 3M Molecular Detection Software with the instrument turned on.
- Add 9 volumes of room temperature BPW to each food sample, usually 225 mL for a 25 g unit, in a stomacher bag. Up to 15 analytical units may be composited into one test sample, for example 375 g or mL, while keeping the 1:10 dilution ratio. Blend, stomach or vortex as required for thorough mixing.
- Incubate samples for 18–26 h at 37 ± 1°C. Incubate kefir instead for 18–26 h at 41.5 ± 0.5°C.
- Set up lysis. One 3M Matrix Control per matrix type is suggested so matrix effects on the assay can be seen. Let LS tubes warm on the bench for 2 h and invert capped tubes to mix. One LS tube is needed for each sample plus a negative control. Cut LS strips to the number needed and place tubes in an empty rack.
- Remove enrichment from the incubator and gently agitate. Decap only one LS strip at a time, using the Cap/Decap Tool: Lysis, setting the tool with caps aside on a clean surface, and use a new pipette tip for every transfer. Transfer 20 µL of enriched sample into its LS tube, repeating until every sample is placed. Finally transfer 20 µL of sterile BPW into an LS tube as the negative control (NC), unless the laboratory is running its own internal controls.
- Verify the Heat Block Insert is at 100 ± 1°C. Place the uncovered rack of LS tubes in the insert and heat for 15 ± 1 min; the lysis solution should turn from pink (cool) to yellow (hot). Move the rack to the Chill Block Insert for 5 ± 1 min, during which the solution should return from yellow to pink, then remove the rack from the chill block.
- Set up amplification. Cut Reagent tube strips to the number needed: one Reagent tube per sample and for the NC, placed in an empty rack, plus one RC tube. Transfer sample lysates into Reagent tubes first, then the NC, and hydrate the RC tube last. Decap one Reagent strip at a time with the Cap/Decap Tool: Reagent and discard those caps, using a new tip for each transfer.
- For each transfer, draw 20 µL of lysate from the upper portion of fluid in the LS tube, avoiding carry-over of protein residue from the bottom, which may interfere with amplification. Dispense at an angle so pellets are not disturbed, and mix by gently pipetting up and down 5 times. A multi-channel pipette can speed this step. Cap filled Reagent tubes with the provided extra caps and press firmly with the rounded side of the Cap/Decap Tool: Reagent in a back-and-forth motion to seal.
- Complete the controls in the reagent rack: transfer 20 µL of NC lysate into a Reagent tube as the negative control, with the same angled dispense and 5 gentle mixes, and transfer 20 µL of NC lysate into the RC tube to serve as the positive control, mixed the same way.
- Load capped tubes into a clean, decontaminated Speed Loader Tray, close and latch its lid, review and confirm the configured run in the software, click Start and select the instrument. When the instrument lid opens automatically, place the Speed Loader Tray inside and close the lid to start. Results arrive within 60 min, although positives may be detected sooner.
- Confirm every presumptive positive culturally, starting from the retained primary BPW enrichment. Plate the enrichment directly onto selective agars used for cultural Salmonella isolation, and at the same time transfer an aliquot into the selective enrichments used in that cultural method. If direct plating confirms the result, the selective enrichment work can be stopped. If direct plating is negative, continue with enrichment incubation, plating, isolation and confirmation through to completion of the cultural method.
Recording Results
Record software calls exactly as given. An algorithm interprets the light output curve from nucleic acid amplification; results are analysed automatically and colour-coded. A Positive or Negative call rests on analysis of several unique curve parameters. Negative samples do not read zero, because the system and amplification reagents carry a background relative light unit (RLU) signal. Presumptive positives are reported in real time, while Negative and Inspect results display only after the run finishes. Record sample identity, enrichment time and temperature (including the kefir branch where used), matrix control result where run, NC and RC behaviour, and each presumptive positive’s disposition into cultural confirmation. An unusual light output labelled Inspect must be repeated beginning at the lysis step; an Inspect result that persists on repeat goes to cultural confirmation rather than being reported as negative.
Quality Control
Run the NC and the RC in every assay as described, samples first, NC next and RC hydrated last. Where a growth temperature of 37°C or below is specified the tolerance is ± 1.0°C, while higher growth temperatures must be held within 0.5°C because of potential lethality to the organism being isolated. Watch the lysis colour changes, pink to yellow on heating, yellow back to pink on chilling, as a check that heating and cooling happened. One matrix control per matrix type is suggested, since food components can affect amplification.
Troubleshooting
- Inspect result on a sample: repeat the assay for that sample starting at the lysis step; if Inspect persists, proceed with cultural confirmation.
- Amplification appears inhibited: check that lysate was taken from the upper fluid layer and that no protein residue from the bottom of the LS tube was carried over; review the matrix control for that food type.
- Lysis solution stays pink during heating: verify the Heat Block Insert has reached 100 ± 1°C before loading and that tubes sat uncovered for the full 15 ± 1 min.
- Suspected cross-contamination: audit strip handling: one strip decapped at a time and a fresh tip for every transfer, for both LS and Reagent tubes.
- Positive screen, negative direct plates: do not close the sample; continue selective enrichment, plating, isolation and confirmation to completion.
Interferences and Limitations
Excluded matrices, chocolate based products, spices, powdered dairy products and whole nuts, fall outside validation. Protein residue from the bottom of a lysis tube can interfere with amplification, and matrix components can affect assay behaviour, hence the suggested matrix control per matrix type. A molecular result remains presumptive: only cultural confirmation settles a positive, and a negative depends on enrichment having grown viable cells within 18–26 h.
Safety Notes
- Treat enrichments as potentially carrying Salmonella; work with powder-free gloves and standard enteric precautions, and disinfect the work surface before analysis as directed.
- Take care around the 100°C heat block and hot lysis tubes; allow the chill step to finish before handling tubes for transfer.
- Dispose of lysis and reagent tubes, tips and enrichments as biohazardous laboratory waste following local rules, as generic laboratory practice requires.
Source Method
Adapted in original wording from Health Canada method MFLP-100, Detection of Salmonella spp. in Foods Using the 3M Molecular Detection System Test Kit Version 2, May 2020, Compendium of Analytical Methods, Volume 3.