SOP: Detecting Listeria monocytogenes in Foods with the 3M Molecular Detection System Version 2 (MFLP-111)
GIFSQ ID: SOP-1012
Loop-mediated isothermal amplification (LAMP) drives this screen. Multiple primers recognise distinct regions of a Listeria monocytogenes gene, a strand-displacing Bst polymerase copies target continuously at one temperature, and pyrophosphate released during copying feeds a bioluminescent reaction read in real time — more than 109 copies can arise in as little as 15 minutes.
Enrichment sets the timetable and branch points. Raw dairy alone needs secondary enrichment, raw meat and some seafood enrich longer and more dilutely, and transfer volumes differ between those paths.
Purpose
Purpose is rapid detection of Listeria monocytogenes in foods after 24 hours of enrichment in most categories, for checking compliance with Sections 4 and 7 of the Food and Drugs Act and other relevant federal requirements. Presumptive positives report in real time; negatives display after the assay completes, within 60–75 minutes.
Scope
Validated categories are specific: raw meat products; fruit and vegetable-based products; the food type “other” of ready-to-eat (RTE) meat and poultry; the food types “raw” and “frozen” of dairy products; and the food types “raw fish and crustaceans” and “frozen” of fish and seafood products. A commodity outside that validated application must first be properly validated against Compendium criteria, with validation data forwarded to the Microbiological Methods Committee so the application can be expanded. Work under ISO/IEC 17025 competence requirements, and prove equivalency before substituting any medium formulation.
Equipment and Materials
- 3M Molecular Detection Assay Listeria monocytogenes test kit, MDA2LMO96: Lysis Solution (LS) tubes, 96 (12 strips of 8); Listeria monocytogenes Reagent tubes, 96 (12 strips of 8); extra caps, 96; Reagent Control (RC), 16 tubes; Quick Start Guide
- MDS100 Instrument and Accessory Kit: Speed Loader Tray, Chill Block Insert and Tray, Cap/Decap Tool—Lysis, Cap/Decap Tool—Reagent, empty lysis and reagent tray
- 3M Molecular Detection Heat Block Insert, MDSHBIN; computer workstation; single-channel 20 µL micropipette and 8-channel 20 µL micropipette with sterile barrier filter tips adaptable to 20 µL
- Dry block heating unit fitting the insert (15 cm × 9.5 cm) and reaching 100°C; incubator(s) settable at 37 ± 1°C; stomacher, blender or equivalent; sterile stomacher bags; powder-free gloves
- Demi-Fraser Broth containing ferric ammonium citrate (FAC), formula in the procedure; Fraser broth containing FAC for secondary enrichment of raw dairy
Procedure
- Hold sample units refrigerated or frozen according to product nature until analysis, except for shelf-stable foods. Thaw frozen units in a refrigerator, or under conditions that prevent both growth and death, and analyse as soon as possible after receipt.
- Have sterile Demi-Fraser broth with FAC prepared and at room temperature. Disinfect the work surface, switch on the block heater with the Heat Block Insert in place to reach 100°C before analysis, and log in to the software with the instrument on.
- Prepare Demi-Fraser with FAC where it is made in-house: basal medium per litre of distilled water is pancreatic digest of casein 5.0 g, peptic digest of animal tissue 5.0 g, beef extract 5.0 g, yeast extract 5.0 g, sodium chloride 20.0 g, mono-potassium phosphate 1.35 g, disodium phosphate anhydrous 9.6 g, esculin 1.0 g, lithium chloride 3.0 g, nalidixic acid 0.01 g and acriflavine HCl 0.0125 g, pH 7.2 ± 0.2. Suspend in 1.0 L distilled water, heat to boiling to dissolve, autoclave at 121°C for 15 min and cool to room temperature. Supplement is FAC 0.5 g in 10.0 mL distilled water, filter sterilized; aseptically add 10.0 mL FAC solution to the cooled basal medium, mix and dispense. Store refrigerated in darkness; acriflavine can photo-oxidize into compounds inhibitory to Listeria spp.
- Make a representative unit: mix liquids and free-flowing materials until homogeneous; for solids, take portions from several locations. Add aseptically 25 g or mL of food to the enrichment volume for its category below. Composites may combine units up to 125 g or mL, for example 125 g in 1125 mL or 2375 mL of broth. For alternate unit sizes keep 1 part sample to 9 parts broth, or 1 to 19 for foods on the 1 in 20 dilution.
- Enrich at 37 ± 1°C by category, blending, stomaching or vortexing first for thorough mixing. Fruit and vegetable-based products, frozen dairy products, and “other” RTE meat and poultry: Demi-Fraser with FAC, 225 mL (1 in 10), primary enrichment 24–26 h, no secondary enrichment, sample analysis volume 20 µL.
- Raw meat products, plus frozen fish and seafood and raw fish and crustaceans: Demi-Fraser with FAC, 475 mL (1 in 20), primary enrichment 28–32 h, no secondary enrichment, sample analysis volume 20 µL.
- Raw dairy products: Demi-Fraser with FAC, 225 mL (1 in 10), primary enrichment 20–24 h, then secondary enrichment by transferring 0.1 mL into 10 mL Fraser broth with FAC for 20–24 h, sample analysis volume 10 µL taken from the secondary enrichment.
- Allow LS tubes to warm to room temperature; one per sample plus a negative control (NC). Cut strips to the number needed in an empty rack. Remove enrichment from the incubator and gently agitate.
- Invert capped LS tubes to mix. Decap one strip at a time with the Cap/Decap Tool—Lysis and use a new tip per transfer to avoid cross-contamination; set the tool with caps aside on a clean surface, retaining caps in order if tubes may need storing.
- Transfer enriched sample into LS tubes: 20 µL per sample generally, or 10 µL from the secondary enrichment for raw dairy products. Repeat until every sample and control occupies its corresponding LS tube.
- Verify the Heat Block Insert is at 100 ± 1°C. Heat the rack of LS tubes in the insert for 15 ± 1 min; solution should change from pink (cool) to yellow (hot). Cool the uncovered rack in the Chill Block Insert for 5 ± 1 min, until solution returns to pink, then remove it.
- Select Reagent tubes, one per sample plus the NC, cutting strips as needed, in an empty rack. Decap one strip at a time with the Cap/Decap Tool—Reagent, discard those caps, and use a new tip for each transfer.
- Transfer 20 µL of lysate from the upper fluid in each LS tube into its Reagent tube, at an angle to avoid disturbing pellets, mixing by pipetting up and down 5 times. Carryover of protein residue from the bottom of LS tubes may interfere with amplification, so take care pipetting. A multi-channel pipette can speed this step.
- Cover Reagent tubes with the provided extra caps, pressing tightly with the rounded side of the Cap/Decap Tool—Reagent in a back-and-forth motion. Load capped tubes into a clean, decontaminated Speed Loader Tray; close and latch its lid.
- Review and confirm the run in the software, click Start and select the instrument; its lid opens automatically. Place the tray in the instrument and close the lid. Results arrive within 60–75 min, positives possibly sooner.
Recording Results
Software analyses the light output curve automatically and colour-codes each result, deciding “Positive” or “Negative” from several curve parameters. Negatives do not give zero, because system and reagents carry background relative light unit (RLU) signal. Unusual output is labelled “Inspect”: repeat from the lysis step, and if Inspect persists, confirm culturally. Record branch, enrichment times, transfer volume, lots, controls and every outcome.
Confirmation is cultural, in substance: for foods screened after primary enrichment only, plate the presumptive positive Demi-Fraser with FAC broth (24 h to 32 h as applicable) directly onto selective agars and continue isolation and confirmation culturally. For raw dairy, plate the presumptive positive Fraser with FAC secondary enrichment the same way.
Quality Control
Include the negative control in every run and Reagent Control tubes as the kit directs. Incubation at 37 ± 1°C and lysis at 100 ± 1°C are validity conditions, verified before samples are committed. As generic lab practice, verify micropipette delivery at 10 µL and 20 µL, because the raw dairy branch depends on a 10 µL transfer.
Troubleshooting
- Repeated “Inspect” result: repeat once from lysis; if Inspect persists, stop repeating and confirm culturally.
- Lysis solution fails to turn yellow, or fails to return to pink: check block insert temperature at 100 ± 1°C, heating time of 15 ± 1 min and chill time of 5 ± 1 min before proceeding.
- Suspected inhibition: check for protein residue carryover from the bottom of LS tubes and confirm lysate was taken from the upper fluid at an angle without disturbing pellets.
Interferences and Limitations
Validation is category-specific, so foods outside the listed types are unvalidated until proven otherwise. Protein residue, pellet disturbance and cross-contamination from decapping several strips at once threaten amplification. Background RLU means a low non-zero reading is expected for negatives and must not be reinterpreted as a weak positive. Presumptive positives and persistent Inspect results need cultural confirmation.
Safety Notes
- Treat all enrichments as potentially carrying L. monocytogenes; wear powder-free gloves and use biohazard disposal for lysis and reagent tubes.
- Take care around the 100°C block and hot LS tubes; handle lithium chloride and acriflavine media with gloves and eye protection.
- Never open reagent tubes after amplification; dispose of them capped to contain amplified DNA.
Source Method
Adapted in original wording from Health Canada method MFLP-111, Detection of Listeria monocytogenes in Foods Using the 3M Molecular Detection System Test Kit Version 2, April 2021, Compendium of Analytical Methods, Volume 3.