SOP: Detecting Listeria monocytogenes Rapidly by ADIAFOOD Real-Time PCR (MFLP-31)
GIFSQ ID: SOP-1033
Speed matters with Listeria monocytogenes, yet a growth step still comes first. Two enrichments build numbers up, then a real-time PCR read-out with molecular beacon probes flags presumptive positives within about 3 hours of enrichment ending, far sooner than screening plates alone would manage.
Specificity carries this design. Primers and beacons in kit tubes target L. monocytogenes alone, so fluorescence rises only when target DNA amplifies. Every positive remains presumptive until culture confirms it.
Purpose
Purpose is rapid presumptive detection of L. monocytogenes in validated foods, supporting compliance work under Sections 4 and 7 of the Food and Drugs Act. Cultural confirmation is mandatory for any positive result.
Scope
Scope covers foods validated for this system: raw fish and crustaceans, raw and raw processed fruits and vegetables, heat processed dairy products, dry dairy products, and shelf stable acid foods. Environmental samples fall outside this draft. ADIAFOOD held AOAC-RI Performance Tested Method status from 2004, Certificate No. 040402.
Sample handling follows cultural Listeria practice, stated here in full: keep units refrigerated or frozen according to product nature, apart from shelf-stable foods; thaw frozen units in a refrigerator or under conditions that prevent both growth and death; analyse promptly after receipt; build a representative analytical unit by mixing liquids or free-flowing material until homogeneous, or by taking solid portions from several locations.
Equipment and Materials
- ADIAFOOD kit: extraction buffer EX-1, lyophilised extraction reagent EX-2, detection buffer DT-1, lyophilised detection reagent DT-2, detection microplates or strip tubes with pre-dispensed PCR reagents, control strip tubes with detection strip kits, PCR optical-grade caps
- ADIAFOOD extraction kit with extraction plates, domed caps and seals; strip holders; PCR capping tool
- AES-validated real-time PCR thermocycler: 96-well (or 48-well) microplate, low and high profile, or 96 (48) × 0.2 mL tube capacity; excitation peak 485 to 520 nm; emission detection peak 500 to 600 nm
- Conventional thermocycler for DNA extraction (optional); centrifuge with rotor and microplate carriers; vortex equipped for microplates; PCR enclosure (optional)
- Two multi-channel pipettors and one single-channel pipettor; aerosol barrier tips; disposable multi-channel pipettor basins; sterile 2 mL microcentrifuge screw cap tubes; powder-free gloves
- Listeria enrichment broth (LEB); Palcam agar slants; 0.1% peptone water; stomacher filter bags; disposable serological pipettes; denatured alcohol and bleach
- Sterile, DNase-free water, pipettes, tips and any other material touching samples or PCR reagents, autoclaved beforehand where applicable to remove DNase and other contamination
Procedure
- Pre-enrich each analytical unit in LEB for 24 h at 30°C, using kit and cultural-method enrichment proportions for the food under test.
- Transfer 200 μL of pre-enrichment broth onto a Palcam agar slant. Incubate at 35°C for 18 to 24 hours.
- Wash growth from slant surface with 2 mL sterile peptone water, resuspending cells into a suspension for DNA extraction. Refrigerate remaining suspension, because confirmation work later depends on it.
- Reconstitute extraction reagent: dispense EX-1 from its squeeze bottle into EX-2 vial, recap, and mix by inversion until lyophilised reagent dissolves completely. Pour reconstituted EX-2 into a multi-channel pipettor basin.
- Aliquot 90 μL reconstituted EX-2 into each well of extraction microplate. Add 10 μL cell suspension per well, placed according to a written sample assignment layout.
- Seal extraction wells with domed cap strips provided. Place plate into thermocycler and run DNA extraction program; cycling lyses bacterial cells.
- Remove plate once program finishes; organisms are now lysed and inactivated. Centrifuge extraction microplate for 5 minutes at 1,800 × g. Supernatant now holds extracted DNA ready for amplification.
- Reconstitute detection reagent in same fashion: DT-1 into DT-2 vial, recap, mix by inversion until fully dissolved, then pour into a fresh pipettor basin.
- Take a detection microplate or strips from pouch and set strips on a holder where applicable. Aliquot 15 μL reconstituted DT-2 into each detection well.
- Transfer 10 μL extracted DNA from extraction plate into corresponding detection wells. Seal detection plate or strips with optical-grade caps provided.
- Seal extraction microplate with provided seals and store at 4°C until analysis is complete.
- Invert detection plate or strips vigorously 4 times so well contents mix thoroughly. Tap gently to clear air bubbles from well bottoms, vortex for 1 minute at maximum speed, then centrifuge for 1 minute at 1,800 × g so PCR mixture, including sample DNA, collects at well bottoms.
- Secure detection plate or strips firmly in PCR instrument with well A1 at upper left corner. Start detection program and run to completion.
- Remove and discard detection plate or strips after program ends. Keep refrigerated cell suspension until confirmation decisions are settled.
Recording Results
Read fluorescence against negative controls. A positive test shows fluorescence above baseline produced by negative controls within 40 cycles of amplification; baseline cut-off value is determined and set by system itself. A negative test normally shows no visible fluorescence. Record result per sample together with enrichment times, slant incubation, extraction and detection run identifiers, and control behaviour. Report positives as presumptive only, pending culture.
Quality Control
Kit controls run with each batch, and negative controls set baseline for every run. Fluorescence above baseline cut-off in a negative control invalidates test results; repeat analysis, taking precautions to remove possible error sources. Work should meet ISO/IEC 17025 competence requirements, and supervisor oversight applies throughout. Layout discipline matters: a written well map prevents sample-to-result mix-ups across extraction and detection plates.
Troubleshooting
- Negative control fluoresces: invalidate run, decontaminate work area and pipettors, use fresh aerosol barrier tips and reagents, then repeat from retained material.
- Bubbles or uneven well volumes: tap plate, vortex and centrifuge exactly as written before starting detection; low or misplaced liquid distorts fluorescence reads.
- Reagent not dissolved: recap and invert again until no lyophilised material remains visible; partly dissolved EX-2 or DT-2 weakens extraction or detection.
- Weak growth on Palcam slant: check LEB incubation reached full 24 h at 30°C and slant received its full 18 to 24 hours at 35°C before washing.
Interferences and Limitations
Validation reaches only food groups named under Scope; other matrices need their own validation before routine use. DNase or stray DNA contamination can create false fluorescence, which is why sterile DNase-free consumables are non-negotiable here. A PCR signal shows target DNA was present in enriched suspension; viability and identity still rest on culture, so a molecular positive never stands alone as a final result.
Safety Notes
- Treat enrichments and slants as potentially carrying L. monocytogenes; wear powder-free gloves and use standard pathogen handling and biohazard disposal.
- Use denatured alcohol and bleach for decontamination as laboratory practice directs, keeping bleach away from PCR reagents and extracted DNA.
- Handle hot thermocycler blocks and sealed plates with care, and discard used detection plates as contaminated consumables.
Source Method
Adapted in original wording from Health Canada method MFLP-31, Identification of Listeria monocytogenes using the Adiafood Rapid Pathogen Detection System, a Real-Time PCR Technique, October 2011, Compendium of Analytical Methods, Volume 3.