SOP: Lactose Determination by Polarimetry
GIFSQ ID: SOP-1075
Lactose twists light. As plane-polarized light passes through a lactose solution, the plane of polarization rotates, and the angle of that rotation is directly proportional to how much lactose is present. Polarimetry turns this optical quirk into a fast, reagent-light assay that dairy labs have trusted for generations, especially for milk powder and whey where lactose dominates the solids.
Purpose
This procedure describes the determination of lactose in milk and dairy products by protein precipitation with Carrez reagents followed by polarimetric measurement, based on AOAC Official Method 896.01 principles.
Scope
Applies to liquid milk, milk powder, whey, and whey powder. Not suitable for fermented products where lactose has been partially converted to lactic acid, or for products containing other optically active sugars at significant levels.
Equipment & Materials
- Polarimeter with sodium D-line source (589 nm), readability 0.01°
- Polarimeter tube, 200 mm, with end caps
- Analytical balance, readability 0.1 mg
- Volumetric flasks, 100 mL
- Carrez solution I: 15 g potassium ferrocyanide per 100 mL water
- Carrez solution II: 30 g zinc sulfate per 100 mL water
- Water bath or temperature control at 20°C
Procedure
- Weigh the sample into a 100 mL volumetric flask: about 10 g of liquid milk or 2 g of milk powder. Record to 0.1 mg.
- Add approximately 50 mL of distilled water and swirl to dissolve or disperse.
- Add 5 mL of Carrez solution I, mix, then add 5 mL of Carrez solution II and mix again. Proteins and fat precipitate.
- Dilute to the 100 mL mark with water, mix thoroughly, and let stand 10 minutes.
- Filter through dry filter paper, discarding the first 10 mL of filtrate.
- Rinse the polarimeter tube with a little filtrate, then fill it completely, avoiding air bubbles. Cap the tube.
- Equilibrate the tube at 20°C for 10 minutes. Lactose mutarotates, so the reading must be taken on the equilibrated solution.
- Measure the optical rotation (P) in degrees, taking the mean of at least three readings.
Recording Results
Calculate lactose content as:
Lactose % = (P × V × 100) / ([α]D20 × l × m)
where P is the observed rotation in degrees, V is the flask volume in mL, [α]D20 is the specific rotation of lactose (+52.3°), l is the tube length in decimeters, and m is the sample mass in grams. Report to 0.1 percent. Typical values: cow milk 4.6 to 5.0 percent, whey powder 70 to 75 percent.
Troubleshooting
- Cloudy filtrate: the Carrez precipitation was incomplete. Repeat with fresh Carrez solutions and ensure the 10-minute standing time.
- Drifting readings: the solution has not reached 20°C or mutarotation equilibrium. Extend equilibration time.
- Air bubbles in the tube: refill carefully; even a small bubble in the light path ruins the reading.
- Unexpectedly low results: check for lactose crystallization in stored powder samples; redissolve fully with gentle warming before analysis.
Safety Notes
Carrez reagents contain cyanide-complexed and zinc salts. Wear nitrile gloves and avoid ingestion. The polarimeter uses an intense light source; never look directly into the beam path. Dispose of zinc-containing waste per local regulations.