QC Samples Quiz

GIFSQ ID: ART-1455

Test your knowledge with this free 20-question QC Samples quiz from Food Laboratory Management, part of the…

What this quiz covers

QC samples are the evidence behind every result a laboratory reports. It starts with the blanks, the most misunderstood tools in the laboratory. A reagent blank checks reagents alone, while a method blank goes through the whole procedure, digestion, extraction, dilution and measurement, so it also captures contamination from glassware, equipment and handling. A method blank detects contamination from reagents, glassware or the procedure, and if it shows analyte signal, the answer is investigation, because with no sample present any signal must have come from the process itself. QC data must be reviewed before sample results are released, because releasing results from an out of control batch risks reporting wrong data to customers.

A matrix spike is a real sample fortified with a known amount of analyte before preparation, and its recovery shows whether the method works in that matrix; poor recovery points to matrix interference or analyte loss during preparation. A laboratory control sample is a clean matrix spiked with analyte, made independently of calibration, verifying that the whole method performs correctly without matrix effects. A surrogate standard is added to every sample before extraction and travels the whole procedure, monitoring extraction efficiency sample by sample. Duplicate analysis means analysing the same sample twice to assess precision, and duplicates that differ greatly signal poor precision, possibly from inhomogeneous samples or unstable measurement.

QC samples are the evidence behind every result

QC samples must be treated exactly like routine samples, so they truly represent batch performance, and they are sometimes prepared blind so analysts cannot treat them differently. QC acceptance limits define objectively when a batch is in control or must be rejected, and QC samples are included at a defined frequency, typically in every analytical batch. Internal QC is the laboratory own batch level controls, while external QC is interlaboratory comparison or proficiency testing. When QC results fail acceptance limits, the batch is stopped, the cause is investigated and corrected, and the batch is reanalysed if needed. Responsibility sits with laboratory management through procedures, with analysts executing and reviewers verifying, and seven consecutive rising points on a control chart signal systematic drift developing, to be investigated before limits are breached.

How to use this quiz

Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the standards and practices a laboratory analyst is expected to know by heart.

QC Samples Quiz
Uncategorized

Choose how you want to play

Sound effects
🔈 Volume
Show explanations
🔊 Auto-read every question
Reading speed

A short note: QC samples in the laboratory

QC samples run alongside test samples to verify control for that batch. They are included at a defined frequency, typically every batch. QC data must be reviewed before release: a batch out of control risks reporting wrong data to customers.

A method blank takes pure reagents through the whole method, detecting contamination from reagents, glassware or the procedure. A reagent blank checks reagents alone. If a method blank shows analyte signal, investigate the process.

A matrix spike adds a known amount of analyte, checking recovery. Poor spike recovery suggests interference or loss is biasing results. An LCS is clean matrix spiked with analyte, independent of calibration, verifying method performance. A surrogate standard is added to every sample to track extraction efficiency through the whole procedure.

A duplicate analysis analyses the same sample twice to assess precision. If duplicates differ greatly, precision is poor, possibly from inhomogeneous samples or unstable measurement. Seven rising points in a row suggest systematic drift.

QC samples are treated exactly like routine samples. They are sometimes prepared blind, so analysts treat them normally. QC acceptance limits define when a batch passes or must be rejected. When QC fails acceptance limits, stop, investigate, correct, and reanalyse if needed.

Internal QC uses the lab's checks; external QC uses interlaboratory comparison or proficiency tests. Laboratory management ensures QC in each batch through procedures, analysts executing, reviewers verifying.

Remember: treat QC samples like routine samples, and never release results from an out-of-control batch.

Questions and answers

Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.

Question 1: What is a reagent blank versus a method blank?

  • A. Reagent blank checks reagents alone; method blank goes through the whole procedure
  • B. One of the two blanks is liquid in practice
  • C. The two blanks are exactly identical in daily practice
  • D. One of the two blanks is solid in practice

Correct answer: Reagent blank checks reagents alone; method blank goes through the whole procedure

A reagent blank contains only the reagents (no sample) and is measured directly, showing the background signal contributed by the reagents alone. A method blank goes further: it is taken through every step of the full procedure : digestion, extraction, dilution and measurement : so it also captures contamination from glassware, equipment and handling. Both are routine QC tools used to monitor the validity of results, an expectation set by ISO/IEC 17025:2017 clause 7.7.

Question 2: Why must QC data be reviewed before sample results are released?

  • A. QC data must be reviewed before sample results are released because of customers like waiting
  • B. Releasing results from an out-of-control batch risks reporting wrong data to customers
  • C. QC data must be reviewed before sample results are released because of review takes time
  • D. QC data must be reviewed before sample results are released because it is tradition

Correct answer: Releasing results from an out-of-control batch risks reporting wrong data to customers

QC data must be reviewed before sample results are released because an out-of-control batch means the analytical system was not performing within its validated limits while those samples were measured. Releasing results from such a batch risks reporting wrong values to customers, with direct food-safety consequences. Review-before-release is the gate that keeps nonconforming work contained, as required by ISO/IEC 17025:2017 clause 7.10.

Question 3: What is a method blank?

  • A. A sample of pure reagents taken through the entire method to detect contamination
  • B. An empty form
  • C. A blank stare
  • D. A missing result

Correct answer: A sample of pure reagents taken through the entire method to detect contamination

A method blank is a portion of analyte-free material or reagents carried through the entire analytical method, from sample preparation to final measurement. Its purpose is to reveal contamination introduced by reagents, glassware or the procedure itself, which would otherwise be mistaken for analyte in the sample. Monitoring validity of results with QC activities such as blanks is an expectation of ISO/IEC 17025:2017 clause 7.7, and a blank that should be clean but is not tells the analyst the process : not the sample : added the signal.

Question 4: In analytical quality control, what does a method blank detect?

  • A. The analyte naturally present in the sample
  • B. The instrument's detection limit
  • C. Contamination introduced by reagents, glassware or the analytical procedure
  • D. The analyst's proficiency

Correct answer: Contamination introduced by reagents, glassware or the analytical procedure

A method blank detects contamination introduced by reagents, glassware or the analytical procedure. Because no sample is present, any analyte signal appearing in the blank must have come from the process itself : dirty glassware, contaminated reagents or carryover. Investigating and eliminating that source is how the laboratory proves the reported result belongs to the sample.

Question 5: What does poor spike recovery suggest?

  • A. Poor spike recoveries suggest the analyst is fast
  • B. Poor spike recoveries suggest the spike was too small to matter
  • C. Poor spike recoveries suggest the instrument is new
  • D. Matrix interference or analyte loss during preparation is biasing results

Correct answer: Matrix interference or analyte loss during preparation is biasing results

Poor spike recovery means the measured amount of a known added analyte is lower or higher than expected, which points to matrix interference suppressing or enhancing the signal, or to analyte lost during preparation steps. It is a bias diagnostic, not a comment on analyst speed or instrument age. Recovery acceptance limits are established during method validation and checked batch by batch, as described in the Eurachem guide on method validation.

Question 6: Why should QC samples be treated exactly like routine samples?

  • A. QC samples should be treated exactly like routine samples because QC samples are special
  • B. QC samples should be treated exactly like routine samples because it is not important
  • C. So they experience the same conditions and truly represent batch performance
  • D. QC samples should be treated exactly like routine samples to save effort

Correct answer: So they experience the same conditions and truly represent batch performance

QC samples must experience exactly the same preparation, reagents, instruments and timing as routine samples so that their results truly reflect how the batch performed. If QC samples were handled more carefully, they would present a flattering picture that the customer samples never enjoyed. Identical treatment is what makes QC representative : a core expectation of ISO/IEC 17025:2017 clause 7.7.

Question 7: What is a duplicate sample analysis?

  • A. Analyzing the same sample twice to assess precision
  • B. Running the instrument twice with no sample
  • C. Copying results
  • D. Testing two different foods

Correct answer: Analyzing the same sample twice to assess precision

Duplicate sample analysis means analyzing the same sample twice to assess precision : the agreement between repeat measurements made under the same conditions. The difference between duplicates is usually expressed as relative percent difference and compared against acceptance limits. Poor duplicate agreement signals inhomogeneous samples or an unstable measurement process.

Question 8: Who is responsible for ensuring QC is performed in each batch?

  • A. The equipment vendor is responsible for ensuring QC is performed in each batch
  • B. Nobody is responsible for ensuring QC is performed in each batch
  • C. The customer is responsible for ensuring QC is performed in each batch
  • D. Laboratory management through procedures, with analysts executing and reviewers verifying

Correct answer: Laboratory management through procedures, with analysts executing and reviewers verifying

Responsibility for QC is shared across defined roles: laboratory management must establish QC procedures and frequencies in the management system, analysts execute the QC within each batch, and reviewers verify the QC data before results are released. ISO/IEC 17025:2017 clauses 5 and 6.2 place this accountability on documented roles with demonstrated competence. No outside party : equipment vendor or customer : carries it.

Question 9: If a method blank shows analyte signal, what does it indicate?

  • A. The method is perfect
  • B. Contamination somewhere in the analytical process that must be investigated
  • C. The analyst is skilled
  • D. Excellent sensitivity

Correct answer: Contamination somewhere in the analytical process that must be investigated

Analyte signal in a method blank means contamination entered somewhere in the analytical process : reagents, glassware, the laboratory environment or carryover : and it must be investigated before the batch is accepted. The laboratory should identify the source, eliminate it and reanalyze; reporting sample results against a contaminated blank would bias every result in the batch. Blank acceptance criteria are set in the method's QC requirements.

Question 10: Why are QC acceptance limits established?

  • A. To define objectively when a batch is in control or must be rejected
  • B. QC acceptance limits are established to save time in practice
  • C. QC acceptance limits are established to punish analysts in practice
  • D. QC acceptance limits are established to reduce paperwork in practice

Correct answer: To define objectively when a batch is in control or must be rejected

QC acceptance limits give an objective, documented boundary between a batch that is in control and one that must be rejected or investigated. They are typically derived from method validation data, historical performance or published method criteria : never from convenience. Without limits, 'in control' would be a matter of opinion rather than evidence.

Question 11: How often should QC samples be included?

  • A. Once a year is how often should QC samples be included
  • B. At a defined frequency, typically in every analytical batch
  • C. Never is how often should QC samples be included
  • D. Only when auditors visit is how often should QC samples be included

Correct answer: At a defined frequency, typically in every analytical batch

QC samples are included at a defined frequency, typically in every analytical batch, so that each set of reported results is backed by evidence that the process was in control at that time. The frequency is documented in the method or QC procedure. Annual or audit-only QC cannot validate the batches run in between.

Question 12: What is a spiked sample (matrix spike)?

  • A. A spiked sample (matrix spike) is a broken sample
  • B. A spiked sample (matrix spike) is a frozen sample
  • C. A spiked sample (matrix spike) is a sample with hot sauce
  • D. A sample fortified with a known amount of analyte to check recovery

Correct answer: A sample fortified with a known amount of analyte to check recovery

A matrix spike (spiked sample) is a real sample fortified with a known amount of analyte before preparation, and its recovery shows whether the method works in that particular matrix. Percent recovery is compared against acceptance limits to detect matrix effects. It differs from a laboratory control sample, which uses a clean matrix rather than a real sample.

Question 13: What is the purpose of a surrogate standard in organic analysis?

  • A. A compound added to every sample to monitor extraction efficiency through the whole procedure
  • B. The purpose of a surrogate standard in organic analysis is a backup instrument
  • C. The purpose of a surrogate standard in organic analysis is to replace the analyst
  • D. The purpose of a surrogate standard in organic analysis is a fake result

Correct answer: A compound added to every sample to monitor extraction efficiency through the whole procedure

A surrogate standard is a compound chemically similar to the target analytes : but absent from real samples : added to every sample, blank and standard before extraction. Because it travels through the whole procedure, its recovery monitors extraction efficiency and flags losses on a sample-by-sample basis. This practice is standard in organic trace analysis, for example in US EPA methods for semivolatile organics and pesticides.

Question 14: What is internal QC versus external QC?

  • A. Internal QC is always cheaper; external QC always costs much more
  • B. Internal QC checks the instruments; external QC checks the analysts instead
  • C. Internal QC uses the lab's samples; external QC uses interlaboratory comparison or proficiency testing
  • D. Internal QC is fully optional; external QC is always strictly mandatory

Correct answer: Internal QC uses the lab's samples; external QC uses interlaboratory comparison or proficiency testing

Internal QC is the laboratory's own batch-level control system: blanks, duplicates, spikes and control samples run alongside routine work. External QC is independent assessment from outside the laboratory, chiefly interlaboratory comparisons and proficiency testing schemes. ISO/IEC 17025:2017 clause 7.7 expects both : continuous internal monitoring plus regular participation in proficiency testing.

Question 15: What should happen when QC results fail acceptance limits?

  • A. QC results fail acceptance limits should report the samples anyway
  • B. QC results fail acceptance limits should delete the QC data
  • C. QC results fail acceptance limits should blame the customer
  • D. Stop, investigate the cause, correct it, and reanalyze the batch if needed

Correct answer: Stop, investigate the cause, correct it, and reanalyze the batch if needed

When QC results fail acceptance limits, the batch is stopped: the analyst investigates the cause, takes corrective action, and reanalyzes the samples if the failure invalidates their results. Sample results must not be reported from a failed batch. ISO/IEC 17025:2017 clause 7.10 requires nonconforming work to be controlled and its impact on previous results assessed.

Question 16: What does it mean if duplicate results differ greatly?

  • A. It means if duplicate results differ greatly great accuracy
  • B. It means if duplicate results differ greatly the samples are identical
  • C. It means if duplicate results differ greatly the analyst is fast
  • D. Poor precision, possibly from inhomogeneous samples or unstable measurement

Correct answer: Poor precision, possibly from inhomogeneous samples or unstable measurement

Duplicates that differ greatly indicate poor precision : the measurement system is not reproducing itself. Common causes are inhomogeneous samples, a frequent issue with foods, or an unstable instrument or procedure. Large duplicate differences trigger investigation, and the batch may need reanalysis after better homogenization or method checks.

Question 17: Why are QC samples sometimes prepared blind (unlabeled as QC)?

  • A. To prevent analysts treating them differently from routine samples
  • B. It saves labels
  • C. QC samples are sometimes prepared blind (unlabeled as QC) to confuse staff
  • D. Blind samples are cheaper

Correct answer: To prevent analysts treating them differently from routine samples

Blind QC samples are prepared so that analysts cannot identify them as QC, preventing any conscious or unconscious special treatment. If analysts knew which samples were QC, they might run them more carefully, defeating the purpose of representative monitoring. Blinding keeps the QC honest and the batch assessment genuine.

Question 18: What is a laboratory control sample (LCS)?

  • A. A dirty beaker taken from the laboratory washing sink area daily
  • B. A sample provided personally by the laboratory manager this morning
  • C. Clean matrix spiked with analyte, made independently of calibration, verifying method performance
  • D. A customer complaint about the recent test results

Correct answer: Clean matrix spiked with analyte, made independently of calibration, verifying method performance

A laboratory control sample is a clean matrix spiked with a known amount of analyte, prepared independently of the calibration standards. It verifies that the whole method : preparation plus measurement : performs correctly in the absence of matrix effects. LCS recovery is a primary batch acceptance criterion in EPA and AOAC-based quality control programs.

Question 19: What does a trend of seven consecutive points rising on a control chart suggest?

  • A. A trend of seven consecutive points rising on a control chart suggests the chart is broken
  • B. A systematic drift developing that should be investigated before limits are breached
  • C. A trend of seven consecutive points rising on a control chart suggests excellent precision
  • D. A trend of seven consecutive points rising on a control chart suggests normal random variation

Correct answer: A systematic drift developing that should be investigated before limits are breached

Seven consecutive points rising on a control chart is a classic non-random pattern: even though no point has crossed a control limit, the steady drift signals a systematic change developing, such as instrument drift, degrading reagents or a shifting process. Control-chart trend rules : the Nelson rules published in the Journal of Quality Technology in 1984 : treat such runs as assignable-cause signals. The correct response is investigation now, before limits are breached, rather than waiting for a failure.

Question 20: What is the purpose of analyzing QC samples alongside test samples?

  • A. The purpose of analyzing QC samples alongside test samples is to train new staff only
  • B. To verify the analytical process is in control for that batch
  • C. The purpose of analyzing QC samples alongside test samples is to use up reagents
  • D. The purpose of analyzing QC samples alongside test samples is to fill the autosampler

Correct answer: To verify the analytical process is in control for that batch

QC samples are analyzed alongside test samples to verify that the analytical process was in control for that specific batch. They provide the documented evidence behind every reported result. Without batch-level QC, results rest on assumption rather than proof : which is why ISO/IEC 17025:2017 clause 7.7 makes monitoring the validity of results a requirement, not an option.