Food Technology Loksewa – Calibration

GIFSQ ID: ART-1194

Test your knowledge with this free 20-question Calibration quiz from Food Quality Assurance & Quality Control, part…

What this quiz covers

This Loksewa set is about calibration, the discipline that keeps every measurement honest: calibration compares an instrument’s readings against a traceable reference standard under defined conditions. Traceability is the chain behind that comparison, an unbroken series of comparisons linking the measurement back to national or international standards, each with a stated uncertainty. The set then draws the line between calibration and adjustment: calibration compares and documents the error, while adjustment changes the instrument to reduce it, and a calibration is repeated afterwards to verify it worked.

The middle questions are the checks a laboratory performs: the thermometer ice-point check at 0 C in a properly made ice-water bath, the boiling-point check at 100 C at sea-level pressure, corrected for altitude, the pH meter calibrated with standard buffers at pH 4.00, 7.00 and 10.00, and the balance calibrated with certified test weights. Around those sit the system questions: the calibration interval set by criticality, use and past stability; the certificate that states results, uncertainty, traceability and date; the as-found and as-left data that show the instrument before and after adjustment; working standards for routine calibrations and reference standards preserved to calibrate them; and accreditation to ISO/IEC 17025. The set closes with two ideas: accuracy as closeness to the true value against precision as closeness of repeated readings to each other, and measurement uncertainty as the quantified doubt around a result.

Accuracy is not precision: the pair every calibration question turns on

Accuracy and precision are the two words this set confuses on purpose, so hold them apart with one picture. Accuracy is closeness to the true value: an accurate instrument reads what the standard says it should. Precision is closeness of repeated readings to each other: a precise instrument gives the same reading again and again, even if every reading is wrong by the same amount. A precise instrument can be inaccurate when it carries a bias, which is why calibration matters: it finds the bias and documents it. When an instrument is found out of calibration, never keep using it quietly: quarantine it, and assess whether past measurements and the product they judged are affected.

How to use this quiz

Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the numbers and definitions a food technologist is expected to know by heart.

Food Technology Loksewa – Calibration
Uncategorized

Choose how you want to play

Sound effects
🔈 Volume
Show explanations
🔊 Auto-read every question
Reading speed

A short note: calibration of measuring instruments

Calibration checks an instrument against a traceable reference standard. Traceability means an unbroken chain of comparisons to national or international standards. Calibration differs from adjustment: calibration documents error, while adjustment changes the instrument to reduce error. A calibration certificate should state results, uncertainty, traceability and date. Uncertainty is the doubt about the result, expressed as a range. Accuracy is closeness to the true value; precision is closeness of repeated readings. As-found and as-left data show the instrument's state when received and after adjustment. Calibration records must be kept to demonstrate traceability.

A thermometer's ice-point check is at 0 degrees in an ice-water bath, and the boiling-point check is at 100 degrees at sea level, corrected for altitude. A pH meter is calibrated with standard buffers, commonly pH 4, 7 and 10. A balance is calibrated with certified test weights. A thermocouple for CCP monitoring should be calibrated against a traceable reference across its range of use.

The calibration interval depends on the measurement's criticality, use and past stability. Calibration status should be clearly identified with dated labels showing due dates. Handle, store and protect reference standards to preserve their accuracy. Temperature, humidity and vibration can shift the measurement. Working standards serve routine calibrations; reference standards calibrate them. A laboratory calibrating for others usually holds ISO/IEC 17025 accreditation. If an instrument is out of calibration, quarantine it and assess past measurements and product.

Remember: quarantine an out-of-calibration instrument at once.

Questions and answers

Note: the player shuffles the questions and the answer options on every attempt, so the numbers in the player will not match the numbers in the questions and answers below. Match each answer by its wording.

Question 1: Calibration of a measuring instrument means:

  • A. Adjusting it until it looks right
  • B. Cleaning the instrument
  • C. Comparing its readings against a traceable reference standard under defined conditions
  • D. Repairing broken parts

Correct answer: Comparing its readings against a traceable reference standard under defined conditions

Calibration compares an instrument's readings against a traceable reference standard under defined conditions : it measures the error rather than guessing at it (VIM, JCGM 200:2012; ISO/IEC 17025 §6.4).

Question 2: Measurement traceability means:

  • A. Following a calibration manual loosely
  • B. An unbroken chain of comparisons linking the measurement to national or international standards
  • C. Tracking staff attendance
  • D. Tracing a product lot

Correct answer: An unbroken chain of comparisons linking the measurement to national or international standards

Metrological traceability is an unbroken chain of comparisons, each with stated uncertainty, linking the measurement back to national or international standards (VIM; ISO/IEC 17025 §6.4.6).

Question 3: Calibration differs from adjustment because:

  • A. Calibration compares and documents error; adjustment changes the instrument to reduce error
  • B. Adjustment needs no standard
  • C. They are identical
  • D. Calibration repairs the instrument

Correct answer: Calibration compares and documents error; adjustment changes the instrument to reduce error

Calibration compares and documents the instrument's error against a standard; adjustment physically changes the instrument to reduce that error : and a calibration is then repeated to verify the adjustment worked (VIM).

Question 4: A thermometer's ice-point check is performed at:

  • A. 0°C in a properly made ice-water bath
  • B. Room temperature
  • C. 100°C
  • D. 37°C

Correct answer: 0°C in a properly made ice-water bath

The ice-point check is performed at 0°C in a properly made ice-water bath : the equilibrium temperature of melting ice and water at atmospheric pressure (standard thermometer verification practice).

Question 5: The boiling-point check of a thermometer is done at 100°C:

  • A. Only with distilled ice
  • B. At sea-level atmospheric pressure (corrected for altitude)
  • C. At any altitude without correction
  • D. Only in winter

Correct answer: At sea-level atmospheric pressure (corrected for altitude)

Water boils at 100°C only at sea-level atmospheric pressure, so the boiling-point check must be corrected for altitude (local barometric pressure) : at altitude the uncorrected reading is systematically low.

Question 6: A pH meter is calibrated using:

  • A. Vinegar and baking soda
  • B. Standard buffer solutions, commonly pH 4.00, 7.00 and 10.00
  • C. Distilled water only
  • D. Any fruit juice

Correct answer: Standard buffer solutions, commonly pH 4.00, 7.00 and 10.00

A pH meter is calibrated with standard buffer solutions of known pH : commonly 4.00, 7.00 and 10.00 : chosen to bracket the expected sample pH (standard pH measurement practice).

Question 7: A weighing balance is calibrated with:

  • A. Any heavy object
  • B. Bags of flour
  • C. Coins
  • D. Certified test weights traceable to national standards

Correct answer: Certified test weights traceable to national standards

A weighing balance is calibrated with certified test weights traceable to national standards : random heavy objects have no known, stable mass (OIML R 111 weight classes; ISO/IEC 17025).

Question 8: The calibration interval (frequency) for an instrument should be based on:

  • A. The colour of the instrument
  • B. Staff preference
  • C. A fixed ten-year period for all instruments
  • D. How critical the measurement is, how often it is used, and its past stability

Correct answer: How critical the measurement is, how often it is used, and its past stability

The calibration interval should be set from how critical the measurement is, how often the instrument is used, and its past stability history : not from habit or a one-size-fits-all period (ILAC G24; ISO 10012).

Question 9: A calibration certificate should state:

  • A. The operator's opinion of the factory
  • B. Only the instrument's brand
  • C. The results, measurement uncertainty, traceability, and date
  • D. The purchase price

Correct answer: The results, measurement uncertainty, traceability, and date

A calibration certificate must state the results, the measurement uncertainty, the traceability chain and the date : without those, the certificate cannot support decisions (ISO/IEC 17025 §7.8.4).

Question 10: Measurement uncertainty describes:

  • A. A mistake made by the operator
  • B. The age of the instrument
  • C. The cost of calibration
  • D. The doubt about the measurement result, expressed as a range around the value

Correct answer: The doubt about the measurement result, expressed as a range around the value

Measurement uncertainty is the doubt about a measurement result, expressed as a range around the value within which the true value is believed to lie (GUM, JCGM 100:2008). It is quantified doubt, not an operator's mistake.

Question 11: Accuracy differs from precision in that accuracy is:

  • A. The same as precision
  • B. Only about speed
  • C. Closeness to the true value, while precision is closeness of repeated readings to each other
  • D. Unimportant in QC

Correct answer: Closeness to the true value, while precision is closeness of repeated readings to each other

Accuracy is closeness to the true value, while precision is the closeness of repeated readings to each other : a precise instrument can still be inaccurate if it carries a bias (ISO 5725).

Question 12: If an instrument is found out of calibration, the correct response is to:

  • A. Quarantine it, and assess whether past measurements and product are affected
  • B. Blame the supplier
  • C. Keep using it quietly
  • D. Throw away all records

Correct answer: Quarantine it, and assess whether past measurements and product are affected

An out-of-calibration instrument must be quarantined so it cannot be used, and past measurements and affected product must be assessed : every decision made with it is now suspect (ISO 10012; ISO 9001:2015 §7.1.5.2).

Question 13: Calibration status of instruments should be:

  • A. Clearly identified, e.g. with dated labels showing due dates
  • B. Kept secret
  • C. Remembered by one person
  • D. Marked only when broken

Correct answer: Clearly identified, e.g. with dated labels showing due dates

Calibration status must be clearly identified : typically with dated labels showing the next due date : so nobody unknowingly uses an overdue instrument (ISO 10012 §7.2).

Question 14: Reference standards used for calibration must be:

  • A. Shared with other departments freely
  • B. Left on the bench
  • C. Handled, stored and protected to preserve their accuracy
  • D. Used as everyday working tools

Correct answer: Handled, stored and protected to preserve their accuracy

Reference standards used for calibration must be handled, stored and protected to preserve their accuracy, and never used as everyday working tools : their degradation would corrupt every calibration beneath them (ISO/IEC 17025 §6.4.9).

Question 15: Environmental conditions during calibration matter because:

  • A. They have no effect
  • B. They make the room comfortable
  • C. Temperature, humidity and vibration can affect the measurement
  • D. They are only for staff health

Correct answer: Temperature, humidity and vibration can affect the measurement

Temperature, humidity and vibration can shift a measurement, so calibration must be performed under controlled and recorded environmental conditions (ISO/IEC 17025 §6.3).

Question 16: "As-found" and "as-left" data on a calibration record show:

  • A. The factory address
  • B. The instrument's condition when received and after any adjustment
  • C. Staff attendance
  • D. The purchase history

Correct answer: The instrument's condition when received and after any adjustment

"As-found" data show the instrument's condition when received; "as-left" data show its condition after any adjustment : together they reveal whether it was drifting and what the adjustment fixed (standard calibration record practice).

Question 17: A laboratory performing calibrations for others is commonly accredited to:

  • A. ISO 9001 only
  • B. A trade association letter
  • C. No standard is needed
  • D. ISO/IEC 17025

Correct answer: ISO/IEC 17025

A laboratory performing calibrations for others is commonly accredited to ISO/IEC 17025 : the international standard for the competence of testing and calibration laboratories (ISO/IEC 17025:2017).

Question 18: Working standards differ from reference (master) standards in that working standards are:

  • A. More accurate than reference standards
  • B. Disposable
  • C. Never calibrated
  • D. Used for routine calibrations, while reference standards are preserved for calibrating the working standards

Correct answer: Used for routine calibrations, while reference standards are preserved for calibrating the working standards

Working standards do the routine calibrations, while reference (master) standards are preserved and used only to calibrate the working standards : protecting the top of the traceability chain (ISO 10012; VIM).

Question 19: Calibration records must be retained to:

  • A. Demonstrate traceability and support decisions made with past measurements
  • B. Fill archives
  • C. Impress auditors only
  • D. Avoid buying new instruments

Correct answer: Demonstrate traceability and support decisions made with past measurements

Calibration records must be retained to demonstrate traceability and to support decisions already made with past measurements : they are the audit trail of every measurement (ISO/IEC 17025 §7.5).

Question 20: A thermocouple used for CCP monitoring should be calibrated:

  • A. Once at purchase and never again
  • B. Against a traceable reference at temperatures covering its range of use
  • C. Only if it looks damaged
  • D. By guessing

Correct answer: Against a traceable reference at temperatures covering its range of use

A thermocouple used for CCP monitoring must be calibrated against a traceable reference at temperatures covering its range of use : monitoring a critical limit with an unproven sensor is a HACCP failure (Codex CXC 1-1969; NACMCF HACCP principles).