Food Technology Loksewa – NBSM Calibration

GIFSQ ID: ART-1274

Test your knowledge with this free 20-question Calibration quiz from NBSM / Naptoul — part of the Food…

What this quiz covers

This set is calibration: what it is, how the standards hierarchy works under Nepal Standard Measurement and Weight Act, 2025, and how a laboratory keeps its measurements trustworthy. Calibration is the operation that establishes, under specified conditions, the relation between quantity values with their measurement uncertainties and the corresponding indications of the instrument: the VIM definition. It determines and documents the error; it does not repair the instrument, adjust it to zero, or certify the operator.

The middle questions cover the hierarchy of standards. Under the Act, the main standard is authenticated and stamped, and checked against the national standard at prescribed periodic intervals with the date of check certified. The secondary standard is prepared to test the accuracy of the working standard, and the working standard is checked against the national prototype, except that special sets of working standards for precious metals and jewels are checked against the main standard. A secondary or working standard that has not been checked and stamped within the prescribed time limit has no legal status and shall not be used for the purposes of the Act.

Adjustment changes the instrument; calibration only measures it

The distinction that matters most in this set is adjustment against calibration: adjustment physically changes the instrument to reduce its error, whereas calibration only determines and documents the error. That is why as-found data matter so much: they record the instrument performance before any adjustment, showing how it was performing in service, and they are the evidence for judging past measurements. When an instrument is found out of tolerance at recalibration, the correct action is to quarantine it and assess the validity of measurements made since its last acceptable calibration.

How to use this quiz

Take the quiz cold first, then read the explanations below the player. Each answer gives the reason behind the key, with the definition and the laboratory practice a food technologist is expected to know by heart.

Food Technology Loksewa – NBSM Calibration
Uncategorized

Choose how you want to play

Sound effects
🔈 Volume
Show explanations
🔊 Auto-read every question
Reading speed

Answer explanations

Worked through the quiz? Here is why each correct answer is correct: the concept behind it, in plain terms.

  1. In metrology, calibration is the operation that
    Correct answer: establishes, under specified conditions, the relation between quantity values with measurement uncertainties and the corresponding indications of the instrument. Calibration is the operation that, under specified conditions, establishes the relation between quantity values (with measurement uncertainties) and the corresponding instrument indications: the VIM definition. It determines and documents the error; it does not repair a broken instrument, adjust it to zero, or certify the operator.
  2. Under Nepal’s Standard Measurement and Weight Act, 2025, the secondary standard is prepared
    Correct answer: in order to test the accuracy of the working standard. Under the Standard Measurement and Weight Act, 2025, the secondary standard is prepared in order to test the accuracy of the working standard: it sits one echelon above the working standard in the hierarchy, not as a replacement for the main standard or a sale item for traders.
  3. Under the Act, the main standard is
    Correct answer: authenticated and stamped, and checked against the national standard at prescribed periodic intervals with the date of check certified. The main standard is authenticated and stamped, and checked against the national standard at prescribed periodic intervals with the date of check certified, per the Act standard-hierarchy provisions. It is not kept secret and never checked, used directly in retail shops, or checked only once at manufacture.
  4. Under the Act, the working standard is checked against the national prototype, except that special sets of working standards for precious metals and jewels are checked against the
    Correct answer: main standard. Under the Act, the working standard is checked against the national prototype, except that special sets of working standards for precious metals and jewels are checked against the main standard, the higher-accuracy exception for the most valuable trade. Note: this is the Act own exception as worded, so learn it as the stated rule for precious metals and jewels rather than as the general hierarchy.
  5. A secondary or working standard that has not been checked and stamped within the prescribed time limit
    Correct answer: has no legal status and shall not be used for the purposes of the Act. A secondary or working standard that has not been checked and stamped within the prescribed time limit has no legal status and shall not be used for the purposes of the Act: the stamping is what gives the standard its legal standing. It does not become the main standard or pass into public sale.
  6. The interval between calibrations of a measuring instrument should be decided on the basis of
    Correct answer: its drift history, frequency and conditions of use, required accuracy and manufacturer’s guidance. The interval between calibrations should be decided from the instrument drift history, frequency and conditions of use, required accuracy and the manufacturer guidance: a risk-based interval, not a fixed ten years for every instrument, and certainly not its colour or the operator date of birth.
  7. A proper calibration certificate must include
    Correct answer: the measurement results together with their measurement uncertainty and the traceability chain. A proper calibration certificate must include the measurement results together with their measurement uncertainty and the traceability chain: the three elements that make a calibration usable for decisions. Opinions, advertisements and purchase invoices are not part of a certificate.
  8. When a measuring instrument is found to be out of tolerance at recalibration, the correct action is to
    Correct answer: quarantine it and assess the validity of measurements made since its last acceptable calibration. When an instrument is found out of tolerance at recalibration, the correct action is to quarantine it and assess the validity of measurements made since its last acceptable calibration: past results may be compromised, and ignoring the finding or adjusting it secretly without records would hide that risk.
  9. Reference standards in a calibration laboratory should be
    Correct answer: reserved for calibrating working standards and protected from wear and damage. Reference standards in a calibration laboratory should be reserved for calibrating working standards and protected from wear and damage: using them for routine measurements would wear them down and destroy the lab top accuracy. They are not lent freely or stored in the open on the bench.
  10. Calibration laboratories that offer services to outside customers commonly demonstrate their competence through accreditation to
    Correct answer: ISO/IEC 17025. Accreditation to ISO/IEC 17025 (General requirements for the competence of testing and calibration laboratories) is the recognized way calibration laboratories demonstrate technical competence to outside customers. ISO 14001, ISO 22000 and ISO 50001 cover environment, food safety and energy, not laboratory competence.
  11. A check standard is used in a calibration laboratory to
    Correct answer: monitor the stability of the measurement process over time. A check standard is a stable artefact measured routinely alongside normal calibrations to monitor the stability of the measurement process over time: a control-chart device, not a replacement for the national standard and not a way to impress auditors.
  12. In an interlaboratory comparison (proficiency testing) for calibration,
    Correct answer: the same travelling artefact is measured by the participating laboratories and the results are compared. In an interlaboratory comparison (proficiency testing) for calibration, the same travelling artefact is measured by all participating laboratories and the results are compared: only with a common artefact can differences between laboratories be seen. Different artefacts, secret results and a single laboratory would defeat the purpose.
  13. The test weights used to calibrate a balance obtain their own accuracy through
    Correct answer: calibration against higher-echelon standards, maintaining an unbroken traceability chain to the national standard. Test weights obtain their accuracy by calibration against higher-echelon standards, maintaining an unbroken traceability chain to the national standard. A manufacturer word alone cannot establish traceability, and visual inspection or weighing them on the same balance proves nothing.
  14. Recalibration of an instrument becomes necessary
    Correct answer: after repair, adjustment, overload or shock, or when the calibration interval expires. Recalibration becomes necessary after repair, adjustment, overload or shock, or when the calibration interval expires: any event that can change the instrument metrological behaviour restarts the clock. A brand-new instrument still needs its first calibration, and the manufacturer request alone is not the trigger.
  15. Environmental conditions such as temperature, humidity and vibration must be controlled during calibration because they
    Correct answer: can influence both the instrument and the artefact being calibrated. Temperature, humidity and vibration can influence both the instrument and the artefact being calibrated (thermal expansion, electrical drift, mechanical disturbance), so the environment is controlled as a metrological factor, not only for operator comfort or fire regulations.
  16. The calibration status of each instrument in service should be
    Correct answer: clearly identified, for example by a label showing the calibration date and the next due date. The calibration status of each instrument in service should be clearly identified, typically a label showing the calibration date and the next due date, so no one uses an out-of-date instrument. Memory, guesswork or no record at all would let an expired instrument stay in use.
  17. Adjustment of a measuring instrument differs from calibration in that adjustment
    Correct answer: physically changes the instrument to reduce its error, whereas calibration only determines and documents the error. Adjustment physically changes the instrument (turning a screw, entering a correction factor) to reduce its error, whereas calibration only determines and documents the error without changing the instrument. They are distinct operations, and adjustment is not done only at the time of manufacture.
  18. As-found data recorded during calibration represent
    Correct answer: the instrument’s performance before any adjustment, showing how it was performing in service. As-found data record the instrument performance before any adjustment: how it was actually performing in service since its last calibration. They are the evidence for judging the validity of measurements made in the intervening period, not factory defaults or the next due date.
  19. Calibration records are retained primarily to
    Correct answer: provide evidence of traceability and support the validity of decisions made using past measurements. Calibration records are retained primarily to provide evidence of traceability and to support the validity of decisions made using past measurements, for example, product acceptance decisions. They are quality evidence, not decoration or idle paperwork.
  20. A working standard differs from a reference (master) standard in that the working standard is
    Correct answer: used for routine calibrations while the reference standard is preserved for calibrating the working standard. A working standard differs from a reference (master) standard in that the working standard is used for routine calibrations while the reference standard is preserved for calibrating the working standard: the hierarchy keeps the best accuracy at the top and spends the lower accuracy on daily work.