SOP: Anisidine Value Determination in Edible Oils

GIFSQ ID: SOP-0932

Oxidation in edible oils happens in two acts. Peroxide value catches the first one, while hydroperoxides are still building up. But peroxides are unstable. They break down into aldehydes, mainly 2-alkenals and 2,4-dienals, which are what actually make oil taste stale and painty. The p-anisidine value measures those secondary oxidation products, so it tells you how far oxidation has really gone, not just where it started.

Purpose

This procedure describes the spectrophotometric determination of the p-anisidine value (p-AV) of edible oils and fats as an index of secondary oxidation, following the principle of AOCS Official Method Cd 18-90 and ISO 6885.

Scope

Applies to refined and crude vegetable oils, marine oils, and used frying oils. Not suitable for strongly colored oils (such as crude palm oil) without additional blank correction, and not for oils containing added antioxidants that react with p-anisidine.

Equipment and Materials

  • UV-Vis spectrophotometer capable of measurement at 350 nm, with 10 mm quartz cuvettes
  • Analytical balance, readability 0.1 mg
  • Volumetric flasks, 25 mL; graduated pipettes
  • Glass-stoppered test tubes
  • p-Anisidine reagent: 0.25 g p-anisidine dissolved in 100 mL glacial acetic acid (prepare fresh weekly, store in amber bottle)
  • Glacial acetic acid, analytical grade
  • Isooctane (2,2,4-trimethylpentane), spectrophotometric grade

Procedure

  1. Weigh the oil sample into a 25 mL volumetric flask and dissolve in isooctane to the mark. Use 0.5 g for heavily oxidized oils, up to 4.0 g for fresh refined oils.
  2. Pipette 5 mL of the sample solution into each of two glass-stoppered test tubes, marked A (test) and B (blank).
  3. Add 1.0 mL of p-anisidine reagent to tube A. Add 1.0 mL of glacial acetic acid to tube B.
  4. Stopper both tubes, shake vigorously, and let stand for exactly 10 minutes in the dark at room temperature.
  5. Measure the absorbance of both solutions at 350 nm against isooctane within 10 minutes of the reaction time ending.
  6. Record the absorbance of the test solution (As) and the blank (Ab).

Calculation and Reporting

Calculate the p-anisidine value as: p-AV = 25 times (1.2 times As minus Ab) / m, where m is the mass of oil in the 25 mL flask, in grams. Report to one decimal place. Fresh refined vegetable oils typically show p-AV below 10. Values above 20 indicate significant secondary oxidation, and used frying oils often exceed 30. For a complete oxidation picture, report p-AV alongside peroxide value; the Totox value (2 times PV plus p-AV) combines both. The typical values are kept as printed and must be confirmed by the technical reviewer (see the flag above).

Safety

Glacial acetic acid is corrosive and its vapors irritate eyes and airways. Work in a fume hood, wear nitrile gloves and safety glasses. p-Anisidine is a suspected carcinogen; avoid skin contact and inhalation of the powder when preparing reagent. Isooctane is highly flammable; keep away from ignition sources.

Quality Control

  • Prepare the p-anisidine reagent fresh weekly and store it in an amber bottle. A blank absorbance above 0.2 means the reagent has oxidized; discard it and prepare fresh.
  • Read promptly. The chromophore is stable for about 10 minutes after the reaction period, and yellow color fading before reading invalidates the result.
  • Repeat with a smaller sample weight when absorbance is above 2.0; the sample is too concentrated.
  • Warm the flask gently to 40 C and swirl until clear before pipetting when the oil has not dissolved completely; poor repeatability usually comes from incomplete dissolution.
  • Record the blank absorbance with every batch and do not report sample results when the blank or the reagent check fails.

Troubleshooting

  • High blank absorbance (above 0.2): the p-anisidine reagent has oxidized. Discard and prepare fresh.
  • Poor repeatability: the oil did not dissolve completely. Warm the flask gently to 40 C and swirl until clear before pipetting.
  • Absorbance above 2.0: the sample is too concentrated. Repeat with a smaller sample weight.
  • Yellow color fading before reading: read promptly; the chromophore is stable for about 10 minutes after the reaction period.

Records

Record the sample weight, reagent preparation date, test and blank absorbances, calculated p-AV, peroxide value where reported alongside, and the analyst name and date for every batch.

References

AOCS Official Method Cd 18-90 (p-anisidine value), as named on the live page. ISO 6885 (animal and vegetable fats and oils, determination of anisidine value), as named on the live page. Owner or technical reviewer to confirm the editions in use.